Simulating Vein Compressibility for Ultrasound-Guided Femoral Access: Speed, Accuracy, and Phantom Validation

Amit Manicka, Laurent Coulon, Darwin Ang, Michael Loushin, Timothy M. Kowalewski

Amit Manickaa,∗^{a,*}, Laurent Coulon, PhD,b^{b} Darwin Ang MD, PhD, MPHb,bb^{b,bb}, Michael Loushin, MD,c^{c} and Timothy M. Kowalewski, PhDa,∗^{a,*}

a^{a}Department of Mechanical Engineering, University of Minnesota, Minneapolis, Minnesota 55455, United States of America

b^{b}Lightside Surgical, Bellevue, Washington 98005, United States of America

bb^{bb}Healthcare Corporation of America, HCA HealthONE Trauma Quality Improvement Program Univ., Ocala, Florida

c^{c}Visible Heart Laboratory, Department of Anesthesiology, University of Minnesota Medical School, Minneapolis, Minnesota 55455, United States of America

In ultrasound-guided femoral access, probe pressure deforms the tissue and changes the vascular anatomy a robot observes. A simulator must reproduce these mechanics fast enough for robotic interaction, yet the computational effort needed for a clinically meaningful signal like vein collapse before artery is unclear. We test the feasibility of a physics-based simulation in NVIDIA's Isaac Sim for Healthcare, with Stable Neo-Hookean tissue solved by Vertex Block Descent on one GPU. A simulation of a low-fidelity cylindrical benchtop phantom with known geometry and mechanical properties pressed by a robot predicted its artery within 5 percentage points at every depth. It matched the vein within 3 points through 11 mm of a 25.4 mm press, and followed the collapse behavior of the low-fidelity physical phantom. Across element size, time step and solver iterations, wall clock per frame scaled linearly with solver work, but accuracy did not improve with more work. The chosen setting cut the solver's time per frame factor of 976, from 29,700 ms to 30.4 ms, within 1.03 points of the finest reference setting. The physics solver alone took 30.4 ms per frame, and with ultrasound and graphics rendering took to full stack took 56.4 ms per frame. Finally, seeding the digital simulation with realistic literature-based mechanical properties, the leg model compressed the vein far more than the artery at every setting, reproducing the expected clinical signal (vein collapse before artery) use to distinguish them. In total, this suggests that the NVIDIA Isaac Sim platform used may be suitable to augment training for reinforcement learning or online simulation for autonomous femoral access with ultrasound.

Keywords: Robotic ultrasound; femoral vascular access; deformable tissue simulation; finite element method; sim-to-real validation.

1. Introduction

Simulation is now a standard environment for training and evaluating medical robots,1−3^{1-3} but the mechanics of current robot-learning frameworks4,5^{4,5} remain strongest for rigid bodies,6,7^{6,7} and probe-control policies for robotic ultrasound are typically trained on static image volumes or rigid scenes.1,2,8,9^{1,2,8,9} Deformable solvers exist, but their calibration against real tissue is rarely reported,7,10,11^{7,10,11} so a simulator can look anatomically correct while behaving mechanically inaccurately.

This matters because probe pressure, needle insertion and vessel compression all change the anatomy the robot observes,1,12^{1,12} and the ultrasound image measures the deformed anatomy rather than a fixed model.13,14^{13,14} The chain runs from a robot action to a contact force, to the deformed

tissue, to the image that drives the next action (Fig. S1). Sim-to-real transfer depends more on observations that stay consistent with probe motion than on single-frame realism.15^{15}

We target fully automated access to both the common femoral vein and artery for emergency medicine and resuscitation. In these settings, aortic balloon occlusion needs arterial access,16^{16} extracorporeal life support can need both vessels,17^{17} and an absent pulse makes ultrasound replace palpation.18,19^{18,19} Probe pressure collapses the thin-walled vein while the artery holds its lumen,20^{20} and clinical guidance uses this compressibility to tell them apart.21,22^{21,22} Although color Doppler can also separate them,18^{18} we do not assume it is available in emergency use, and thus rely on the B-mode image. Most robotic ultrasound systems regulate contact force to suppress probe-induced deforma-

2 A. Manicka et al.

tion. 13,23−25^{13,23-25} Only one has used compression to classify the vessels. 26^{26} Compression also has a cost: a compressed vein is a smaller target, so clinicians press lightly during puncture. 27^{27} We envision a brief, deeper press that labels the vessels, then a lighter contact that keeps the vein open for the needle. However, both need a model that predicts vein and artery lumen area at each probe depth. In the Seldinger technique, the needle, dilator and catheter each compress the vein again, and every later step inherits the puncture site chosen under the probe. 28^{28} This work therefore starts with probe contact.

A useful simulator must reproduce this behavior fast enough for a robot to interact with it, and for the thousands of repetitions that learning demands. 2,4^{2,4} We report cost as wall-clock milliseconds per frame, where a frame is 16.7 ms of simulated time, one image at the 60 Hz rate of an ultrasound display. Speed is not the only concern: a numerically accurate solution can still rest on material parameters that do not represent real tissue. 6,11^{6,11} This work therefore asks how much computational effort is required to reproduce a clinical signal faithfully enough for a robot to interact with? We test whether Isaac Sim is feasible for this task. We build an interactive Stable Neo-Hookean model 29^{29} of the femoral region, solved on the GPU with Vertex Block Descent, 30^{30} whose deformation drives the rendered ultrasound image. We first compare a simulation of a simple benchtop phantom of known materials with robot presses on it. We then map how element size, time step and iterations trade accuracy against speed, first on a tissue slab, then on the leg. Finally, we give the leg model literature-based tissue properties and test whether it separates the femoral vein from the artery. Supplementary material (Sections S1–S3, Figs. S1–S7, Tables S1–S9), videos and the simulation code are available online. a^{a}