arXiv:2608.26000eess.SPAug 2026Cited by 1 paper on PaperVantage

Manuscript Offprint

arXiv:2608.26000

CardioFusion-AI: Robust ECG--PPG Fusion for Multimodal Physiological Monitoring Under Signal Degradation

Navaneetha Krishnan Kamalakannan, Janakiraman Kamalakannan

Navaneeth Krishnan K and Janakiraman K

Abstract—Wearable electrocardiogram (ECG) and photoplethysmogram (PPG) sensors are complementary but individually fragile: motion artifact, poor contact, and sensor dropout routinely degrade one or both signals. Fusion strategies that assume both modalities are equally trustworthy can become less reliable than a single clean modality once degradation sets in, yet the comparative robustness of different ECG–PPG fusion mechanisms under graded degradation and complete modality loss remains poorly characterized. We present CardioFusion-AI, a framework whose signal-processing front end (R-peak and systolic-peak detection, an Orphanidou-type signal-quality index, and beat-by-beat pulse transit time estimation) is validated on 53 real intensive-care recordings (848 windows, heart-rate mean absolute error 1.61 bpm ECG / 2.78 bpm PPG) and a real annotated fetal ECG database (R-peak F1 0.89–0.98). Building on this validated front end, we conduct a controlled, physiologically-grounded synthetic degradation study comparing eight ECG–PPG fusion strategies – unimodal baselines, three fixed-weight fusion variants, attention fusion, and two sample-adaptive gated fusion mechanisms (one conditioned only on learned embeddings, one additionally conditioned on the validated signal-quality index) – across six degradation regimes spanning graded corruption and complete modality loss, over five independent training seeds. Attention fusion achieved the lowest descriptive overall error (1.66±0.43 bpm). Both adaptive gates correctly reallocated weight toward the healthy modality under complete modality loss but showed near-zero correlation between gate weight and signal quality when both modalities were present but graded-degraded (r ≈ 0.10–0.24), and signal-quality conditioning produced a large, specific improvement under missing-PPG conditions (1.56±0.59 bpm, near the 1.48 bpm unimodal ceiling). With only five training seeds, no pairwise comparison survives Holm-corrected significance testing; we report effect sizes and confidence intervals accordingly. These results indicate that modality availability and modality quality are functionally distinct problems for adaptive fusion.

Index Terms—Adaptive fusion, electrocardiography, multimodal biosignal fusion, photoplethysmography, physiological monitoring, robust wearable sensing, signal quality assessment, signal-quality-aware fusion.

I. INTRODUCTION

Manuscript submitted for review.

N. Krishnan K. is with the Department of Electronics and Communication Engineering, Saveetha School of Engineering (e-mail: knavaneeth385@gmail.com) and J. K is with the Government Kilpauk Medical College and Hospital (e-mail: kjanaki2408@gmail.com).

WEARABLE physiological monitoring increasingly relies on multiple complementary biosignals rather than any single sensor. Electrocardiography (ECG) and photoplethysmography (PPG) are the most common pairing in consumer and clinical wearables: ECG provides a direct electrical readout of cardiac timing, while PPG provides an optically derived pulse waveform from which heart rate, pulse timing, and – in combination with ECG – pulse transit time (PTT) can be estimated [1]. Because the two modalities are captured through different physical transduction mechanisms, they are also affected by different degradation processes, which is a large part of why fusing them is attractive: a corruption that disables one modality frequently leaves the other intact.

In practice, however, both modalities degrade routinely and often simultaneously. Motion artifact corrupts PPG far more readily than ECG; poor electrode contact or perspiration corrupts ECG independently of PPG; and either sensor can lose contact entirely for extended periods in ambulatory use. A fusion method that implicitly assumes both inputs are equally trustworthy can therefore become less reliable than a single well-behaved modality once one input deteriorates – naive averaging of a clean signal with a corrupted one degrades the estimate relative to simply discarding the corrupted input.

This motivates signal-quality-aware fusion: mechanisms that adjust how much each modality contributes based on how trustworthy it currently is. Several families of fusion strategy have been proposed for multimodal biosignal and, more broadly, multimodal machine learning problems, including early/input-level concatenation, feature-level fusion via a learned combination MLP, late/decision-level fusion with fixed or learned per-modality weights, cross-modal attention, and sample-adaptive gating in which a small network learns to reweight modalities on a per-input basis [2]. Adaptive gating is conceptually the most attractive of these for degraded wearable sensing, since it promises to reweight modalities exactly when and where degradation occurs, rather than applying a single global weighting scheme.

Existing evaluations of these fusion mechanisms are, however, typically conducted under clean or narrowly controlled conditions, or evaluate a single proposed mechanism against a small number of baselines. The comparative behavior of a broad set of fusion mechanisms – including whether an adaptive, signal-quality-conditioned gate actually learns to track graded signal degradation, as opposed to only the binary presence or absence of a modality – has not, to our knowledge,

IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS

been systematically characterized under a controlled degradation protocol with matched architectures, matched training procedures, and an explicit accounting for the statistical power of the comparison.

In this work we address that gap directly. We build CardioFusion-AI, a two-part evaluation: (i) a signal-processing front end – peak detection, an Orphanidou-type signal-quality index (SQI) [3], and beat-by-beat PTT estimation – validated against real, independently sourced physiological data; and (ii) a controlled, physiologically-grounded synthetic degradation study in which eight ECG–PPG fusion architectures, sharing an identical encoder backbone and training protocol, are compared across six degradation regimes spanning graded single- and dual-modality corruption and complete modality loss. We deliberately frame this work in terms of physiological monitoring robustness rather than disease diagnosis or clinical risk stratification: the evidence we present concerns the reliability of physiological estimation under signal degradation, not diagnostic or prognostic performance, and should not be read as either.

The central hypothesis motivating this study is:

Signal-quality-aware and multimodal fusion mechanisms can improve the robustness of physiological estimates under ECG and/or PPG signal degradation, compared with fixed or unimodal estimation strategies; and sample-adaptive fusion guided explicitly by signal-quality information will show greater robustness to modality-specific degradation than fusion strategies without such conditioning.

As we show, the data support the first clause but only partially support the second, in a way that is itself informative about the mechanism.

The contributions of this paper are threefold:

  1. A signal-quality-validated ECG–PPG fusion framework and controlled degradation-and-modality-loss evaluation protocol for physiological estimation, built on physiologically-grounded synthetic signal generation with independently controllable, graded per-modality corruption.

  2. A systematic, matched-protocol comparison of eight unimodal and multimodal fusion strategies – including fixed-weight, attention-based, and implicit- and explicit-signal-quality-conditioned adaptive gating – across six degradation regimes and five independent training replicates, with pre-registered, multiplicity-corrected statistical analysis.

  3. A mechanistic analysis of learned modality allocation showing that adaptive gates reliably detect complete modality absence but exhibit near-zero correlation with graded signal quality when both modalities are present, and that explicit signal-quality conditioning improves this behavior selectively rather than uniformly.