arXiv:2609.10047cs.HCSep 2026

Manuscript Offprint

arXiv:2609.10047

Streaming P300 Acquisition and Statistical Signal Validation Across Five EEG Platforms: A Hardware-Agnostic BrainFlow/LSL Pipeline

Isabella Guan, Rui Liu, Fusheng Wang

Isabella Guan

Lake Washington High School Kirkland, WA, USA isabellaguan2009@gmail.com

Rui Liu

Dept. of Computer Science Stony Brook University Stony Brook, NY, USA ruiliu1@cs.stonybrook.edu

Fusheng Wang

Dept. of Computer Science & Biomedical Informatics Stony Brook University Stony Brook, NY, USA fusheng.wang@stonybrook.edu

Abstract—P300 spellers offer people with severe motor impairment, such as ALS, an effective communication channel and remain one of the most established surgery-free alternatives to intracortical interfaces. Advanced language models have made spellers faster and more robust, yet the hardware beneath them is under-studied. We present a hardware-agnostic, real-time P300 acquisition pipeline built on BrainFlow and Lab Streaming Layer (LSL) that runs unchanged across consumer- and research-grade EEG headsets, with permutation tests of signal separability. Using a standard 6 x 6 row/column paradigm, we piloted five configurations: a custom dry system, a custom wet/gel system, Emotiv Flex, Emotiv EPOC X, and Muse 2. The custom systems and EPOC X showed weak or inconsistent signal separability, Muse 2 had the highest acquisition reliability despite limited centro-parietal coverage, and Flex showed the most promising signal. In 20 further Flex sessions varying subject, timing, and phrase length (131 target characters), a peak-amplitude permutation test and a cross-validated xDAWN decoder both detected a significant target response under two channel-exclusion policies, with decoder AUC reaching about 0.72 after 15 repetitions. Character accuracy depended heavily on evaluation methodology: in-sample majority voting reached 94.7%, whereas character-held-out accuracy was 31.3% with evidence accumulated across repetitions, about three times that of held-out majority voting. These analyses indicate that Flex captured a detectable, if still weak, P300 under the studied conditions, while broader participant-level validation and improved decoding remain necessary.

Index Terms—brain-computer interface, P300 speller, permutation test, statistical signal validation, cross-validation, multi-platform EEG comparison, BrainFlow, Lab Streaming Layer, EEG

I. INTRODUCTION

The P300 speller enables communication without muscle movement by detecting a positive event-related potential elicited around 300 ms after a subject attends to a flashing target in a row/column grid [1], making it a promising augmentative channel for individuals with severe motor impairment, such as Amyotrophic lateral sclerosis (ALS), brainstem stroke, for whom eye movement may not provide reliable access.

Intracortical BCIs achieve high communication rates [2] but require neurosurgery unsuitable for many candidates. Non-invasive, scalp-EEG approaches such as the P300 speller remain the most established surgery-free alternative, with a

growing literature on making them faster and more robust: dynamic stopping [3], P300/eye-tracking hybrids [4], and LLM-assisted next-character prediction [5], reviewed in [6].

What is comparatively under-studied is the hardware layer beneath these paradigm-level improvements: which consumer- and research-grade headsets can actually deliver a usable P300 signal, and how acquisition reliability and target-related responses vary across recording configurations and sessions. Most demonstrations validate a single, fixed headset, even though consumer-grade EEG hardware varies widely in fidelity relative to medical-grade systems [7], and same-paradigm comparisons across multiple headsets through a shared pipeline remain uncommon – leaving budget-limited labs unable to tell a paradigm's limitations from a given headset's.

This paper describes a hardware-agnostic streaming acquisition pipeline and a permutation-test framework for evaluating P300 signal separability across platforms, with per-session quality checks identifying acquisition failures before platform-level aggregation. Based on the pilot comparison (Section IV-A), we conducted a larger Flex study with variation across sessions and subjects (Section IV-C). We also compare three character-decoding strategies on the same 20-session pool to examine differences between in-sample and held-out accuracy and the effect of combining evidence across repetitions (Section IV-E).

The paper is guided by three research questions:

  • What signal quality and acquisition reliability are observed for each headset under a common pilot protocol?
  • Once a promising platform is identified from a small pilot, does its signal remain detectable across additional subjects, sessions, and phrase lengths?
  • How much does the resulting character-decoding accuracy depend on evaluation methodology and on how repetition-level evidence is combined?