arXiv:2609.07857cs.ROSep 2026

Manuscript Offprint

arXiv:2609.07857

Scene Graph-Driven Haptic Feedback for Safety Enhancement in Robotic Ophthalmic Surgery via Physically Simulated iOCT

Danial Arbabi, Korab Hoxha, Angelo Henriques, Mirza Imamovic, M. Ali Nasseri

Danial Arbabi¹, Korab Hoxha¹, Angelo Henriques¹, Mirza Imamovic¹,²,³, and M. Ali Nasseri¹,²,⁴

Abstract— Robotic ophthalmic surgery offers high precision but introduces a "sensory gap" by decoupling the surgeon from their instrument, resulting in a loss of tactile feedback. This paper presents a novel haptic feedback system for subretinal injection tasks leveraging Scene Graphs (SG). The system bridges the sensory gap by analyzing a physically simulated intraoperative Optical Coherence Tomography (iOCT) feed to construct a real-time surgical SG. The SG serves as a semantic abstraction layer for the surgical scene, which is then utilized by a deterministic, rule-based engine to generate state-dependent haptic feedback on a robotic input device. The system was evaluated in a user study (N=16) using an anthropomorphic head phantom and a custom-built surgical robot. Results demonstrate that the SG-driven haptic feedback improved surgical precision, reducing needle alignment error by 14% (p=0.044p = 0.044) and improving System Usability Scale (SUS) scores by 8% (p=0.015p = 0.015), while maintaining comparable task completion times. A needle trajectory analysis revealed the emergence of a safer "Align-then-Approach" strategy, in which our haptic negative reinforcement prompted users to fine-tune the tool's trajectory before approaching the retinal target. This work suggests that SGs can effectively serve as the direct computational foundation for real-time, safety-enhancing context-aware haptic feedback in robotic microsurgery.

I. INTRODUCTION

Subretinal injection is a surgical procedure used to treat pathologies such as age-related macular degeneration (AMD) [1]. The procedure requires sub-millimeter precision [2], often exceeding the physiological limits of human surgeons due to hand tremor. Thus, robotic systems have been introduced [3], [4] to filter and scale motion, enhancing the precision and improving safety.

However, current robotic systems introduce a sensory gap [5], effectively blocking direct tactile feedback from the instrument. Furthermore, interaction forces in subretinal injections drop below the threshold of human perception,

Figure

Fig. 1: Schematic of subretinal injection. The needle targets the subretinal space between the neurosensory retina and the RPE [9]. Concept adapted from [10].

shifting reliance entirely to visual estimation and possibly increasing cognitive load [6].

To address this, strategies using haptic feedback [7] have been proposed. While force sensors, used to measure forces acting on the instruments and replicate them to an input device, often require complex hardware changes, using Visual-to-Haptic feedback offers an alternative. This approach allows computer vision to infer geometry and forces via existing infrastructure, such as iOCT.

In this work, we propose a novel framework that leverages SGs as the direct foundation for a real-time, context-aware feedback system, connecting visual perception and haptic feedback. Unlike raw sensor data and distance metrics, SGs provide a structured representation of a surgical scene [8], which we then combine with a rule-based engine to infer context-aware vibrotactile cues for the surgeon. This system aims to synthesize kinesthetic forces from semantic vibrotactile cues.

This paper makes the following contributions:

  1. A perception pipeline that builds a real-time SG from a simulated iOCT by having the side view from a physical eye phantom.

  2. A rule-based inference engine mapping SG relations to specific haptic feedback patterns.

The results of this work were validated in a user study (N=16) on an existing surgical robotic platform.