arXiv:2609.08770cs.ROSep 2026

Manuscript Offprint

arXiv:2609.08770

A Controlled Comparison of Manual and Teleoperated Intraocular Instrument Motion for an Input Device

Korab Hoxha, Mirza Imamovic, Angelo Henriques, M. Ali Nasseri

Abstract

Input devices for robotic microsurgery are frequently described as preserving the surgeon's trained technique, but the claim is rarely measured. We compared manual and teleoperated intraocular instrument motion with the trocar constraint, the instrument, the eye model and the tracking source common to both conditions, so that the control interface was the only factor varied. Prior comparisons cannot hold the instrument fixed, because a robotic instrument is not the tool used manually. Sixteen participants performed a navigation task on a commercial ophthalmic simulator by hand and through a three-degree-of-freedom input device commanding a five-joint robot. Task outcome was equal but at ceiling: every participant acquired all five targets under both interfaces with no retinal or lens injury. Execution differed on every measure. Teleoperated trials took three times as long at a quarter of the median speed, covered less than half the angular working range, and were broken into 3.5 times as many separate movements. Completion time and movement fragmentation improved substantially across four trials of practice and had not plateaued; the measures set by the configured rate ceiling and joint limit changed the least. Finger activity doubled and pinch variability tripled, so reducing instrument degrees of freedom redistributed manual effort rather than reducing it. The interface preserves the outcome and reshapes the execution.

Keywords: Human–robot interaction, leader–follower systems, motion analysis, remote center of motion, surgical robotics, surgical simulation, teleoperation, tremor, vitreoretinal surgery.

1 Introduction

Vitreoretinal surgery treats disorders of the retina and vitreous through instruments introduced into the eye via trocars in the sclera. The relevant anatomy is small and fragile: the internal limiting membrane is 2–3 µm thick, retinal veins are of order 100 µm in diameter, and the retina does not regenerate, so inadvertent contact can cause permanent visual loss. Subretinal injection, an emerging route for gene and cell therapy, requires placing a needle tip within tens of micrometers of the retinal surface and holding it there.

Three features of the procedure shape any robotic assistance. The trocar imposes a remote center of motion (RCM) at the sclera, so once inside the eye the instrument retains four degrees of freedom: three rotations about the entry point and one translation along the instrument axis (Fig. 1) [2]. Two of the rotations aim the instrument tip across the retina; the third is a roll about the instrument axis, which matters for tools with a directional tip but not for the positioning task studied here. The eye is a closed, fluid-filled cavity, so lateral force at the entry port stresses the sclera and can displace the globe. And the surgeon works under a microscope with limited depth cues, at magnifications where hand tremor is visible in the field.

This work has been submitted to IEEE Access for possible publication. Copyright is held by IEEE. Personal use of this material is permitted; permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting or republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works. This version may no longer be accessible without notice.

Figure

Figure 1: Cross-section of an eye with an instrument inserted through a trocar. The degrees of freedom (θ,ϕ,ψ,z)(\theta, \phi, \psi, z) of the tool are given by the arrows

That last point sets the precision limit. Structures of interest are tens of micrometers across, while physiological hand tremor during vitreoretinal surgery spans 100–200 μ\mum peak-to-peak [1]. Robotic assistance has been pursued on the premise that scaling motion and filtering tremor can extend the surgeon's reach beyond that limit.

The effectiveness of such a system depends on the interface through which the surgeon commands it. A robot that can position an instrument to a few micrometers is only as useful as the operator's ability to specify where those micrometers should be. This has motivated a class of application-specific input devices for ophthalmic surgery, and a recurring design argument: that an input device should preserve the movements the surgeon has already spent years learning, rather than require new ones.

That argument is widely made and rarely tested. Input devices are characterized by specification: degrees of freedom (DOF), workspace, peak force, update rate. They are evaluated by task outcome: completion time, error counts, subjective workload. A device may be described as intuitive, or as preserving natural surgical technique, without either property being measured. What would make the claim testable is a direct comparison of the motion a surgeon produces through the interface against the motion the same surgeon produces by hand, on the same task.

Such comparisons exist for laparoscopic and general surgery, where the dynamics of the input device have been shown to alter operator movement measurably [11–15]. In every case the instrument at the tissue differs between conditions: a wristed robotic instrument is not the tool the surgeon holds manually. Where the manual condition is open technique rather than laparoscopic, the kinematic constraint differs as well. In intraocular surgery neither confound is necessary. The trocar imposes a remote center of motion (RCM) in both conditions, and a robot can hold the same unmodified instrument the surgeon would hold, tracked by the same system. The comparison can therefore be made while holding the physical instrument, trocar constraint, eye model, and measurement source constant, allowing the effects of the teleoperation pathway and its control architecture to be examined without simultaneously changing the surgical tool or environment.

This paper reports such a comparison for an input device we call EMIRAS (Ergonomic Manipulator for Intraocular Robotic-Assisted Surgery), built explicitly on the manual-similarity premise. EMIRAS is held in a pen grasp with the wrist rested, pivots about a fulcrum below the fingers, and multiplexes three sensed inputs onto the five joints of a robot, with aiming mapped position-to-position in a frame aligned to the microscope view. Sixteen participants (three vitreoretinal surgeons, six residents, seven novices) performed a navigation task on a commercial ophthalmic simulator, manually and through the device, with the same instrument, the same eye model, the same trocar port and the same tracking source in both conditions.

The purpose of this comparison is not simply to determine whether the teleoperated interface can complete the same task as manual operation, but to determine whether it elicits similar movement behavior. We therefore examine task success together with movement speed, path length, angular workspace, movement fragmentation, mode-engagement behavior, practice effects, and finger-level kinematics. This allows us to identify which characteristics of manual microsurgical movement are retained and which are transformed by the teleoperation architecture.

The contributions of this work are:

  1. A complete system description of a three-DOF input device controlling a five-DOF intraocular robot through phase-specific mode multiplexing, with all control parameters, sensing resolutions, timing and workspace limits reported as configured or measured rather than as specified.

  2. A modality-agnostic basis for comparing intraocular instrument motion. The RCM is recovered per trial from the instrument trajectory by least-squares line intersection, the standard pivot-calibration formulation, and motion is decomposed into insertion depth and aiming angles about it. The method is not new; applying it per trial as a common frame in which manual and teleoperated motion become directly comparable, despite a depth-dependent displacement scaling, is what this contributes.

  3. A controlled comparison in which the physical instrument, eye model, trocar constraint, task, and measurement chain are common to both conditions, allowing the effects of the teleoperation pathway to be examined without confounding them with changes in the surgical tool or environment.

  4. Quantitative characterization of effects the literature does not report: clutch duty cycle and engagement statistics for a surgical input device, tremor amplitude projected in a common physical unit from the handle through the robot joints to the instrument tip, and the learning trajectory of a novel interface against a familiar one. Finger-level tremor is reported separately and in relative terms only, since it does not convert to the same unit.

Section 2 reviews input devices for microsurgery and the existing evidence on manual versus teleoperated movement. Section 3 describes the system. Section 4 details the experimental design and the analysis. Section 5 reports the results, Section 6 discusses their implications for interface design, and Section 7 concludes.

References (42)