arXiv:2609.08992cs.LGSep 2026

Manuscript Offprint

arXiv:2609.08992

Physics-Informed Deep Learning for False Ventricular Tachycardia Alarm Reduction in the ICU

Athanasios Papastathopoulos-Katsaros, Alexandra Stavrianidi, Zhandong Liu

Abstract

False ventricular tachycardia (VT) alarms are a leading contributor to alarm fatigue in intensive care units. We propose a deep learning framework combining a 1D SE-ResNet with ICU-realistic data augmentations and a physics-informed auxiliary reconstruction task based on the three-element Windkessel hemodynamic model, implemented as a differentiable forward simulation. By requiring the network's latent representation to produce physiologically plausible arterial pressure waveforms, artifact-driven ECG patterns are penalized while true VT remains coherent across modalities. Evaluated on the VTaC benchmark under a strict real-time protocol (10 s pre-alarm window), our method achieves a Challenge Score of 85.08 ± 1.65, a 5-point improvement over prior state-of-the-art. Ablation studies confirm that the physics-informed objective is the primary performance driver, providing gains in accuracy, ~2× label efficiency, and more localized and clinically meaningful ECG segments.

1. Introduction

Ventricular tachycardia (VT) is a life-threatening arrhythmia characterized by anomalous ventricular beats exceeding 100 bpm [1]. Because prolonged VT can rapidly lead to sudden cardiac death, ICU monitors are tuned for high sensitivity, making VT alarms among the most prone to false positives [2, 3]. This contributes to alarm fatigue, a critical patient-safety concern [4, 5]. Reducing false VT alarms is a critical problem at the intersection of machine learning and healthcare.

A key difficulty in detecting false alarms is that sensor detachment, patient movement, and electrical interference produce ECG artifacts that closely resemble true arrhythmias, confounding data-driven classifiers. The VTaC benchmark [6] provides over 5,000 multi-institutional ICU recordings with ECG, photoplethysmography (PLETH),

and arterial blood pressure (ABP), enabling evaluation. Prior real-time methods, including supervised CNNs, contrastive models [7], and cross-modal VAEs [8], achieve Challenge Scores up to 80.08 [9]. Foundation models [10] and retrospective approaches [11] report higher AUCs, but the former rely on large external pretraining corpora and the latter use post-alarm information, while still remaining vulnerable to severe sensor noise and previously unseen artifact patterns.

We address this gap by embedding physiological structure into the learning process. Our architecture jointly optimizes classification with cross-modal physics-informed reconstruction of ABP (via a Windkessel simulation) and of PLETH via a data-driven decoder. This forces the network to ensure that any ECG pattern classified as VT produces a plausible hemodynamic response, thereby penalizing artifact-driven predictions. Our method achieves a Challenge Score of 85.08 ± 1.65, a ~5-point improvement over prior state-of-the-art, operating strictly within a 10 s real-time window without external pretraining data.

References (18)