arXiv:2609.01018physics.med-phSep 2026

Manuscript Offprint

arXiv:2609.01018

PyDoseRT Proton: A GPU Pencil-Beam Engine with a Convolutional Residual-Correction Network for Fast Proton Dose Calculation

Lukas Zimmermann, Hermann Fuchs, Attila Simkó, Gerd Heilemann

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Lukas Zimmermann1,2^{1,2}, Hermann Fuchs1,2^{1,2}, Attila Simkő3^{3}, and Gerd Heilemann1,2^{1,2}

1^{1} Department of Radiation Oncology, Medical University of Vienna, Vienna, Austria

2^{2} Christian Doppler Laboratory for Image and Knowledge Driven Precision Radiation Oncology, Medical University of Vienna, Vienna, Austria

3^{3} Department of Diagnostics and Intervention, Umeå University, Umeå, Sweden

Abstract. Architecture category. Hybrid method: a physics-based analytical pencil-beam (PB) dose engine followed by a 3-D convolutional residual-correction network (RepVGG-U-Net).

We addressed the DoseRAD2026 proton dose-prediction task with PyDoseRT Proton, a GPU-accelerated engine implemented in PyTorch and augmented by a learned residual toward Monte Carlo (MC) accuracy. A double-Gaussian PB kernel was calibrated to GATE/Geant4 integrated depth doses in water in two stages: a classical per-energy curve fit, then a gradient-based fit of the full 3-D dose through the PyTorch physics engine as it retains a differentiable execution path for gradient-based optimization of dose-dependent objectives. The engine computes each beamlet on a beam's-eye-view (BEV) lattice with variance-preserving Gaussian splitting, an analytic nuclear halo, and a Fermi-Eyges heterogeneity term, then rotates the result into the patient frame. Additionally, a compact residual U-Net predicts an additive correction in BEV space. It is conditioned on voxelwise material-label embeddings, a discrete energy embedding and spot size. The same model was used for all anatomical sites (thoracic and abdominal). It was trained with a patient-space L1L_1 objective emphasizing the scored high-dose region and multi-scale BEV deep supervision. The submitted CT configuration obtained preliminary-test beamlet MAE 0.0066, image-z IDD distance 0.0025, plan MAE 0.0049, 98.30% gamma pass rate (1%/1 mm), and DVH error 0.460.

Keywords: Proton therapy · Dose calculation · Pencil-beam engine · Residual learning · Monte Carlo.

1 Introduction

Intensity-modulated proton therapy delivers highly conformal dose but is sensitive to range uncertainty and anatomical changes, which motivates dose engines that are both fast enough for online adaptation and accurate enough to replace analytical pencil-beam (PB) algorithms near tissue heterogeneities. Monte Carlo

L. Zimmermann et al.

(MC) transport is the accuracy reference but is too slow for iterative planning; analytical PB algorithms [4] are fast but degrade in lung and at bone/air interfaces. Recent work showed that a differentiable dose engine enables gradient-based plan optimization end to end [2].

We extend this idea to protons for the DoseRAD2026 challenge. Our method couples a proton PB engine, calibrated to MC in water through an autograd-enabled copy of its forward model, with a compact convolutional network that learns an additive residual toward MC. The zero-initialized residual anchors optimization at the analytical solution while retaining an explicit physics baseline instead of predicting dose directly from anatomy.