arXiv:2608.28714eess.IVAug 2026

Manuscript Offprint

arXiv:2608.28714

Evaluating the Safety of Deep Learning-Based Brain MRI Reconstruction

Dat Tat Mai, Thai Viet Pham, Thu Nguyen Thi Dang, James Jin Kang

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Dat Tat Mai, Thai Viet Pham, Thu Nguyen Thi Dang, and James Jin Kang

Abstract—Objective: Deep learning accelerates brain MRI four- to tenfold, but learned models can erase a lesion or invent tissue, and pixel-averaged scores such as PSNR and SSIM miss both. We review whether evaluation practice could detect this blind spot. Methods: Following PRISMA 2020 we searched seven databases without date limit, included 263 studies (1995–2026), appraised them with QUADAS-2 and matched instruments, and synthesised narratively. Categories came from titles, abstracts and controlled vocabulary; every prevalence figure is a floor. Duplicate screening achieved substantial agreement (Fleiss' κ=0.877\kappa = 0.877); appraisal agreement was substantial (0.788; 0.390 where observable); extraction is unaudited. Results: Only 18 of 263 studies (6.8%) record a fidelity metric and a reader assessment on the same data, leaving the field's central proxy unmeasured. Reader studies mostly measure reader-versus-reader agreement, which is weak: fastMRI 2020 concordance was 0.457 and 0.386 (Kendall's W), improving only where SSIM had diverged. Erasing a 100 mm3^3 lacunar infarct moves global PSNR about 0.03 dB under the stated error model. The corpus grew fivefold; reader assessment fell from 32% to 18%, recovering to 21%. Among the four largest families the most hallucination-associated (generative, 39%) is among the least reader-evaluated (11.3%); self-supervised is higher still (47%) with no readers. One study in twenty both releases code and runs readers; none evaluated a model observer; no named dataset covers acute stroke or haemorrhage. Conclusions: On these floors, evaluation practice cannot certify diagnostic safety. We derive five requirements safety-oriented evaluation must meet.

Index Terms—Deep learning, brain MRI reconstruction, reconstruction safety, hallucination, image quality assessment, systematic review.

I. INTRODUCTION

Brain MRI is widely used because it can clearly distinguish different types of brain tissue and reveal subtle abnormalities that are difficult to detect with other imaging modalities. However, this image quality comes at the cost of long acquisition times. Unlike conventional imaging methods that capture an image more directly, MRI builds an image from a sequence of raw signal measurements (samples of spatial frequencies, known as k-space) collected over time. This prolonged acquisition reduces patient throughput and increases sensitivity to patient motion, which can introduce artifacts and degrade image quality [1]. To address these limitations, acceleration techniques such as parallel imaging, including Sensitivity Encoding (SENSE) [2], and compressed sensing [3] were developed to reduce the amount of data that must be collected. More recently, deep learning has enabled further acceleration by reconstructing high-quality images from incomplete MRI measurements, through cascaded convolutional networks [4], model-based unrolled networks [5, 6], generative adversarial networks [7], and score-based diffusion models [8], at accelerations from fourfold to, in exploratory work, an order of magnitude [9].

The reconstructed images look convincing, and that is the problem. Small changes in the input can produce large errors in the output [10, 11]. Four failure types recur across the studies we reviewed: strong priors smooth away faint structure; motion causes ghosting the model never saw in training [12]; a change of scanner or sampling degrades results in ways that in-distribution testing misses [13]; and generative models can invent plausible tissue [14]. The scores used to certify these methods, peak signal-to-noise ratio (PSNR) and structural similarity (SSIM) [15], agree only partially with radiologist quality judgement even on conventional image degradations [16]; and SNR, though reproducible across observers, tracks no-reference quality models only moderately [17], so an image can score well while the error that matters goes unrecorded. We call this the metric blind spot: standard image-quality metrics cannot catch these small but clinically serious errors, because a lesion of a few cubic millimetres barely moves an average taken over millions of voxels.

Parts of the field have responded, training quality metrics on radiologist rankings [18], attaching uncertainty maps to reconstructions [19], or running reader studies that test an accelerated protocol against a reference one directly [9, 20]. Causal and counterfactual methods can ask the most direct question, whether a lesion survives a controlled change of acquisition, but have mostly been applied to classification rather than reconstruction [21, 22]. These responses share a root cause. Acceleration works because a learned prior fills in information the measurement does not contain, and that same filled-in information can hide a real finding or create a false

Dat Tat Mai and James Jin Kang are with the School of Science, Engineering & Technology, RMIT University Vietnam, Ho Chi Minh City, Vietnam (e-mail: dat.mai2@rmit.edu.vn; james.kang@rmit.edu.vn).

Thai Viet Pham is with the School of Computing Technologies, RMIT University, Melbourne, VIC 3000, Australia (e-mail: s4229249@student.rmit.edu.au).

Thu Nguyen Thi Dang is with the School of Health and Biomedical Science, RMIT University, Melbourne, VIC 3000, Australia (e-mail: S4205224@student.rmit.edu.au).

Corresponding author: Dat Tat Mai.

one, while the field certifies the result with averaged pixel scores that cannot see either event.

To our knowledge no systematic review has assembled the evidence on this problem. Existing surveys cover architectures and benchmark performance [23, 24]; none asks how often conventional metrics miss clinically important errors, why they miss them, or what evaluation would have to look like to catch them. We address that through three questions:

RQ1 How reliably do conventional fidelity metrics track radiologist diagnostic ratings in deep learning-based accelerated brain MRI, and how wide is the evaluation reporting gap in the literature?

RQ2 What underlying mathematical mechanisms cause standard metrics to overlook localized pathology erasure and generative hallucinations in deep learning-based reconstruction?

RQ3 What can each evaluation paradigm in current use establish about prior-induced diagnostic error in deep learning-based MRI reconstruction, and what requirements for safety-oriented evaluation follow from those limits?

We contribute: (C1) a synthesis of what 263 primary studies report about agreement between pixel-wise metrics and radiologist judgement; (C2) four mathematical mechanisms that explain the divergence, with an exact identity bounding the PSNR cost of erasing a lesion; (C3) an analysis of where the evaluation gap comes from, showing that it co-occurs with a division of evaluation practice between research communities and with a benchmark infrastructure lacking acute pathology; and (C4) five requirements that safety-oriented evaluation must meet. The auditing framework these requirements motivate is developed and evaluated in a companion methods paper, not here.

Three boundaries should be stated at the outset, because they determine what the findings can and cannot be asked to support. First, this is a review of evaluation practice, not of reconstruction performance: we count what studies measured, not how well their methods performed, and nothing here estimates how often a deployed reconstruction actually erases a lesion. Second, the mechanisms in RQ2 are mathematical properties of the metrics, derived under a stated error model; they establish that a focal error can pass unrecorded, not that it commonly does. Third, prevalence counts rest on labels extracted from what studies report, so every one of them is a lower bound on practice rather than an estimate of it, and we treat the direction of that bias explicitly in Section V-B. The contribution is a diagnosis of what the field's current instruments are capable of detecting, which is logically prior to, and considerably cheaper to establish than, a measurement of how often they fail.

References (100)

Couldn't load all references. Reload the page to try again.