Toward Trustworthy Robot-Assisted Sliding Palpation for Shallow Vessel Localisation with a Calibrated Digital Twin

Piotr Blaszyk, Wen Fan, Kaizhong Deng, Daniel Elson, Dandan Zhang

Piotr Blaszyk ID^{ID}, Wen Fan ID^{ID}, Kaizhong Deng ID^{ID}, Daniel Elson ID^{ID}, and Dandan Zhang ID^{ID}

Imperial College London, UK

Abstract. Reliable localisation of shallow subsurface vessels matters for safe robot-assisted venous access and vessel-aware manipulation, but collecting diverse tactile data on physical hardware is costly, slow, and can degrade soft vision-based tactile sensors. We present a robot-assisted sliding-palpation framework in which a calibrated digital twin generates labelled tactile sequences, reducing dependence on real-world data collection. The twin models sensor–vessel contact, is calibrated against real palpation trajectories by Bayesian-optimisation-based domain adaptation, and is randomised over sliding direction and contact conditions. A spatio-temporal graph neural network trained on simulated marker trajectories performs per-node vessel classification and yields a human-verifiable top-view localisation map through 2D-to-3D-to-2D geometric projection. Using three datasets (Sim, Silicone, and Meat, the latter a raw-meat phantom with vessel models at nominal depths of 0-30 mm) we evaluate four [train]→[test] models: Sim→Sim, Sim→Silicone, Sim→Meat, and Meat→Silicone. The calibrated twin, validated on four canonical interactions, yields a simulated-to-real marker-alignment MAE of 0.50 mm at the point of deepest contact. After reprojection onto a 1 mm top-view grid, with each model's decision threshold set to favour few false alarms over high sensitivity, predicted vessel pixels lie on average 1.05-5.49 mm from the nearest true vessel pixel across the four models (1.05-1.31 mm for all but Sim→Meat, whose larger domain shift marks the current limit of simulation transfer). These results demonstrate progress toward trustworthy tactile palpation through calibrated simulation, interpretable localisation, and transparent reporting of cross-domain degradation. Code, model weights, and data are publicly available on GitHub and Zenodo. 1^{1}

Keywords: tactile palpation · medical robotics · graph neural networks

1^{1} We give Zenodo DOIs here, as they are permanent. Simulation and learning code [10]: https://doi.org/10.5281/zenodo.21958186. Robot control code [9]: https://doi.org/10.5281/zenodo.21958190. Datasets and model weights [8]: https://doi.org/10.5281/zenodo.21958107.

P. Blaszyk et al.

1 Introduction

Working principle of the ViTacTip hybrid tactile sensor

(a) Sensing cross-section applied contact

(b) Internal camera image

Figure

(c) Hybrid decoding — one camera, two channels

Marker channel (TacTip-like)

Marker motion -- contact force, shear force & incipient slip

Vision channel (GelSight-like)

Direct image -- fine surface texture & local 3-D geometry

Both signals are extracted simultaneously from a single internal camera frame.

Fig. 1: The working principle of the ViTacTip sensor used in this work. This figure was inspired by the work of Fan et al. [15].

Shallow blood vessel detection is an important capability in robot-assisted minimally invasive surgery (RAMIS), where it helps prevent accidental vessel injury during manipulation and tissue interaction [33]. It is also essential for clinical procedures such as catheterisation for intravenous injections [24], blood sampling, and interventional treatments including cardiac stenting [30]. In oncology applications, the spatial distribution of vasculature surrounding a tumour is a key factor when determining electrode placement for irreversible electroporation [23].

Vessel detection is supported by several imaging modalities. Magnetic resonance imaging (MRI) and computed tomography (CT) can map vessels at high resolution and depth but are expensive and, for CT, expose the patient to radiation. Ultrasound (US) is the clinical gold standard for intra-operative detection of shallow and deep vessels [22]; near-infrared imaging (NIR) is common in phlebotomy for superficial vessels [31]; and optical coherence elastography (OCE) is an emerging approach with finer resolution than either [19]. All rely on specialised imaging equipment.

Their cost is a key limitation: NIR and US probes cost a few thousand dollars (e.g., 1,800[3]and1,800 [3] and4,800 [4]), and commercial ophthalmic optical coherence tomography (OCT) probes were reported at 40,000–40,000–150,000 in 2021 [37]. They also need specialised maintenance, infrastructure, and trained operators, which hinders deployment in resource-constrained settings, particularly in low- and

A

Simulated Digital Twin

Real Finetune

Figure

B

Real scenario #2: Meat phantom

Real Scenario #1: Silicone Phantom

Vein Detection

Fig. 2: Vessel-detection pipeline. Domain adaptation calibrates the sensor Young's modulus and sensor–vessel contact stiffness, while domain randomisation varies the sliding direction and sensor–vessel contact parameters. Training data are generated using the DiffTactile-based digital twin, and an ST-GNN performs per-node binary classification on the marker graph (magenta: vessel; cyan: no vessel). Robot kinematics are then used to project these predictions into a top-view vessel map (white: predicted vessel). Real ViTacTip trajectories collected on the silicone vascular phantom (Real Scenario #1) and the raw-meat phantom (Real Scenario #2) provide labelled tactile videos for evaluating the ST-GNN.

middle-income countries [38], motivating lower-cost approaches to shallow vessel detection.

Vision-based tactile sensors (VBTSs) provide a compact and comparatively low-cost approach to high-resolution tactile perception. By imaging the deformation of a compliant sensing surface during contact, VBTSs can capture rich information about local geometry and contact mechanics [5,25,40]. Fig. 1 illustrates the operating principle of one such sensor, the ViTacTip [15]. A key practical advantage of VBTSs is their affordability. Commercial devices such as the GelSight Mini and DIGIT cost approximately 500and500 and350, respectively [1,2], while laboratory-built designs can reduce fabrication costs further. For example, MagicTac reports a minimum manufacturing cost of only £4.76 [16]. In comparison, point-of-care ultrasound systems typically cost several thousand pounds, and conventional cart-based systems can cost tens of thousands of pounds [32]. Ultrasound remains an established modality for vascular access and offers direct imaging of subsurface anatomy at substantially greater depths [26]. However, for the localisation of shallow subsurface vessels, the substantially lower hardware cost, compact form factor, and ease of robotic integration of VBTSs motivate their investigation as a complementary sensing modality.

P. Blaszyk et al.

Beyond their cost advantage, VBTSs provide a high degree of sensing versatility. The rich visual information generated by contact-induced deformation can be used to infer a range of tactile quantities, including surface geometry, contact location, force, and slip [17]. Recent developments further demonstrate the potential to integrate multiple sensing modalities within a single compact robotic end-effector. For example, MagicGripper combines tactile, visual, and proximity sensing to support diverse contact-rich manipulation tasks [14]. This multifunctionality enables a common sensing platform to support different perception and manipulation objectives through task-specific processing and inference, without requiring dedicated hardware for each function. Such flexibility is particularly valuable for compact and resource-constrained robotic systems and further motivates the use of VBTSs for robot-assisted palpation.

In the intended clinical workflow, an end-effector-mounted VBTS performs a brief tactile pre-scan of the skin to localise superficial vessels before any invasive action. The resulting vessel map can support automated venous access tasks, such as intravenous cannulation and blood sampling, or vessel-aware manipulation during RAMIS. By presenting the localisation result as a human-verifiable top-view map, the framework retains clinician oversight and enables confirmation of vessel location before needle insertion or instrument interaction.

Vessel detection from VBTS deformation patterns typically needs learning-based models, and hence a large and diverse dataset spanning the contact conditions met during palpation. Collecting it on physical hardware is slow and labour-intensive, and the soft elastomer and delicate optics can degrade under repeated interaction. Digital twins (DTs) address this by simulating tactile interactions at scale without physical wear [13], generating common and rare contact scenarios with automatic ground-truth labels and greater diversity, which can improve generalisation over limited real datasets [29]. Two complementary strategies mitigate the sim-to-real gap: domain adaptation, which aligns DT parameters with their real-world counterparts, and domain randomisation, which adds variability to approximate real conditions [41].

To the best of our knowledge, this study is the first to combine a VBTS with sliding palpation and a DT for shallow vessel detection, and asks how accurately and reliably shallow vessels can be localised from VBTS deformation observed while sliding across them. The main contributions of this work are as follows:

– We formulate shallow-vessel localisation from continuous robot-executed sliding palpation using a vision-based tactile sensor, enabling spatially continuous sensing along the scanning trajectory.

– We develop and calibrate a digital twin of sensor–vessel interaction and use domain-randomised simulation to train a spatio-temporal graph neural network (ST-GNN) for per-node vessel segmentation without silicone-phantom training data.

– We evaluate sim-to-real transfer across two physical experimental domains: a silicone vascular phantom and a raw-meat phantom containing tubular vessel surrogates at nominal depths of 0–30 mm, including explicit analy-

Table 1: Comparison of related work: part 1

ProjectTargetSensorPalpationApproachBPTrainTest
MiniTac [27]TVBTSTMn/aRP, EV
DIGIT Pinki [12]TVBTSTMn/aRP, EV
Bewley et al. [7]TVBTSTEEn/aRP
Beasley et al. [6]BVCapacitiveSEEYn/aP, IV
Hampson et al. [20]BVTactile arrayTEEYn/aP
3D CNN [11]BVNIRnoneEENRIV
OursBVVBTSSEENDTP, M

Abbreviations: target (T = tumour, BV = blood vessel), palpation (T = tapping, S = sliding), approach (M = modular, EE = end-to-end), BP = use blood pulsation data (N = no, Y = yes), train set (R = real-world data, DT = digital twin), test set (P = synthetic phantom, EV = ex-vivo human tissue, IV = in-vivo human tissue, M = raw meat phantom)

Table 2: Comparison of related work: part 2

ProjectModelAccuracy [%]Error [mm]
MiniTac [27]MLP, SVM100n/a
DIGIT Pinki [12]RN18, VMAE98, 100n/a
Bewley et al. [7]LR, GKDE95, 701
Beasley et al. [6]MMD, THR, LRn/a1.3
Hampson et al. [20]MM, GCn/a0.2, 0.6, 6.0
3D CNN [11]3D CNN87n/a
Oursa\text{Ours}^{\text{a}}ST-GNNn/a1.1, 1.2, 5.5, 1.3

Abbreviations: RN18 = ResNet-18; VMAE = VideoMAE; LR = linear regression; GKDE = Gaussian kernel density estimation; MMD = min-max detection; THR = thresholding; MM = mathematical model; GC = Gaussian curve; ST-GNN = spatio-temporal graph neural network.

a^{a} Mean L2 error of the top-view vessel map (Table 4), for Sim→Sim, Sim→Silicone, Sim→Meat, Meat→Silicone in that order.

sis of performance degradation under the larger domain shift, seed-to-seed variability, and a temporal-window ablation.

– We transform tactile predictions using robot kinematics into a human-verifiable top-view vessel map, evaluated per pixel on all three datasets, that can provide spatial input for subsequent robot planning.