Low Dose CT Reconstruction via Edge-preserving Total Variation Regularization

Z. Tian, X. Jia, Kehong Yuan, T. Pan, Steve B. Jiang

Zhen Tian 1,3^{1,3}, Xun Jia 2,3^{2,3}, Kehong Yuan 1^{1}, Tinsu Pan 4^{4}, Steve B. Jiang 2,3^{2,3}

5

1^{1}Department of Biomedical Engineering, Graduate School at Shenzhen, Tsinghua University, Shenzhen, Guangdong 518055, China

2^{2}Department of Radiation Oncology, University of California San Diego, La Jolla, CA 92093, USA

10

3^{3}Center for Advanced Radiotherapy Technologies, University of California San Diego, La Jolla, CA 92093, USA

4^{4}Department of Imaging Physics, University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA

15

20

Email: sbjiang@ucsd.edu

25

30

High radiation dose in CT scans increases a lifetime risk of cancer and has become a major clinical concern. Recently, iterative reconstruction algorithms with Total Variation (TV) regularization have been developed to reconstruct CT images from highly undersampled data acquired at low mAs levels in order to reduce the imaging dose. Nonetheless, the low contrast structures tend to be smoothed out by the TV regularization, posing a great challenge for the TV method. To solve this problem, in this work we develop an iterative CT reconstruction algorithm with edge-preserving TV regularization to reconstruct CT images from highly undersampled data obtained at low mAs levels. The CT image is reconstructed by minimizing an energy consisting of an edge-preserving TV norm and a data fidelity term posed by the x-ray projections. The edge-preserving TV term is proposed to preferentially perform smoothing only on non-edge part of the image in order to better preserve the edges, which is realized by introducing a penalty weight to the original total variation norm.

†^{\dagger} Zhen Tian and Xun Jia have contributed equally to this work and should be considered co-first authors.

During the reconstruction process, the pixels at edges would be gradually identified and given small penalty weight. Our iterative algorithm is implemented on GPU to improve its speed. We test our reconstruction algorithm on a digital NCAT phantom, a physical chest phantom, and a Catphan phantom. Reconstruction results from a conventional FBP algorithm and a TV regularization method without edge preserving penalty are also presented for comparison purpose. The experimental results illustrate that both TV-based algorithm and our edge-preserving TV algorithm outperform the conventional FBP algorithm in suppressing the streaking artifacts and image noise under the low dose context. Our edge-preserving algorithm is superior to the TV-based algorithm in that it can preserve more information of low contrast structures and therefore maintain acceptable spatial resolution.

1. Introduction

X-ray computed tomography (CT) has been extensively used in clinics to provide patient volumetric images for a number of purposes nowadays. However, by its nature, CT scans expose a high x-ray radiation dose to the patient which may result in a non-negligible lifetime risk of cancer (Hall and Brenner, 2008; de Gonzalez et al., 2009; Smith-Bindman et al., 2009). This fact has become a major concern for the clinical applications of CT scans, particularly for pediatric patients, who are more sensitive to radiation and have a longer life expectancy than adults (Brenner et al., 2001; Brody et al., 2007; Chodick et al., 2007). Therefore, it is highly desirable to reduce CT imaging dose while maintaining clinically acceptable image quality.

A simple way to reduce the x-ray dose is to lower mAs levels in CT data acquisition protocols. Nonetheless, this approach will result in an insufficient number of x-ray photons detected at imager and hence elevate the quantum noise level on the sinogram. As a consequence, the quality of the CT images reconstructed from a conventional filtered backprojection (FBP) algorithm (Deans, 1983) will be degraded by the noise-contaminated sinogram data. Another way to reduce imaging dose is to decrease the number of x-ray projections acquired by, e.g., operating the x-ray generator in a high-frequency pulsed model in future with hardware modification. Yet, this will cause serious streaking artifacts in the reconstructed CT images, as the FBP algorithms require that the number of projections should satisfy the Shannon sampling theorem (Jerri, 1977).

Recently, compressed sensing algorithms (Donoho, 2006) have been applied to the CT reconstruction problem. In particular, Total Variation (TV) methods (Rudin et al., 1992) have presented their tremendous power in CT reconstruction with only a few x-ray projections (Sidky et al., 2006; Song et al., 2007; Chen et al., 2008; Sidky and Pan, 2008). In such approaches, an energy function of a TV form is minimized subject to a data fidelity condition posed by the x-ray projections. Since this energy term corresponds to image gradient, minimizing it will effectively remove those high spatial gradient parts such as noise and streaking artifacts in the reconstructed CT images. One disadvantage of this TV approach is its tendency to uniformly penalize the image gradient irrespective of the underlying image structures. As a result, edges, especially those of low contrast regions, are sometimes over smoothed, leading to the smoothed edge to some extent and loss of low contrast information. To resolve this issue, an edge guided compressive sensing reconstruction algorithm has recently been proposed in solving an MRI image reconstruction problem (Guo and Yin, 2010). In such an approach, edges are detected during the reconstruction process and are purposely excluded from being smoothed by the TV norm, by which the reconstruction can be speed up. Though this method can be easily generalized to the CT reconstruction problems, when it comes to the low dose case where insufficient or/and noisy x-ray projections are used in the reconstruction, a high quality edge detection algorithm is required to avoid incorrect detection of those fake edges caused by image noise and streaking artifacts. The sophisticated edge detection algorithms usually pose a high computation burden in the reconstructions. Moreover, they may require fine tuning of some case dependent parameters, making it hard to control their efficacy.

Another disadvantage of the TV-based reconstruction methods are the associated time consuming computation process due to their iterative nature. Generally speaking, the prolonged computational time makes the iterative CT reconstruction approaches prohibitive in many routine clinical applications. Recently, high-performance graphics processing units (GPUs) have been reported to speed up heavy duty computational tasks in medical physics, such as CBCT reconstruction (Xu and Mueller, 2005, 2007; Li et al., 2007; Yan et al., 2008; Jia et al., 2010b), deformable image registration (Sharp et al., 2007; Samant et al., 2008; Gu et al., 2010), dose calculation (Gu et al., 2009; Hissoiny et al., 2009; Jia et al., 2010a) and treatment plan optimization (Men et al., 2009; Men et al., 2010). In principle, high computation efficiency can be expected utilizing GPU in our CT reconstruction problem.

In this work, we generalize the TV-based CT reconstruction algorithm to an edge-preserving TV (EPTV) regularization form to reconstruct CT images under undersampling and/or low mAs situation. In particular, an EPTV norm is designed by introducing a penalty weight to the original TV norm, which enables the algorithm to automatically locate the sharp discontinuities of image intensity and adjust the weight adaptively to adopt the progressively recovered edge information during the reconstruction process. This regularization term automatically ensures that less smoothing is performed on edges to better preserve the edges, as will be seen in below. Our reconstruction algorithm is implemented on GPU to speed up the computation process.