MiBOT: A head-worn robot that modulates cardiovascular responses through human-like soft massage

Alice Mylaeus, Stephanie Vogt, Berken Utku Demirel, Marcel Gort, Mirko Meboldt, Manuel Meier, Christian Holz

Alice Mylaeus¹*, Stephanie Vogt²*, Berken Utku Demirel¹, Marcel Gort², Mirko Meboldt², Manuel Meier¹, and Christian Holz¹

Abstract—Massage therapy is helpful for the rehabilitation of various diseases, such as headaches caused by migraines and stress. Existing robotic systems have focused on massage therapy on the torso and limbs, but performing massage motions through suitable actuation on a person's head has been a challenge. In this paper, we present MiBOT, a head-worn massage robot that actuates two soft tactors to produce touch motions mimicking human massage. A key design principle behind MiBOT is its silent actuation, which we achieve through pneumatic artificial muscles in conjunction with a controller loop to respond to contact pressure. We evaluated the effectiveness of MiBOT in a controlled study and assessed subjects' blood pressure and heart rate levels while applying MiBOT. We found that our mechanical system generated positive and conclusive quantitative outcomes that are similar to the human-administered massage, decreasing participants' mean systolic and diastolic blood pressure by 2.8 mmHg and 1.7 mmHg, respectively, as well as calming their heart rate by 8–10% on average.

I. INTRODUCTION

According to the Global Burden of Disease (GBD) study, headache disorders, which include migraine and tension-type headaches, are among the most prevalent disorders of mankind and disabling conditions worldwide [1]. The World Health Organization (WHO) estimates that the prevalence among adults of current headache disorder (symptomatic at least once within the last year) is about 50%. Studies have estimated that half to three-quarters of adults aged 18–65 years in the world have had headaches in the last years and, among those individuals, 30% or more have reported migraine. Headache on 15 or more days every month affects 1.7–4% of the world's adult population [2].

Some studies found that massage may be as effective as medication in the preventative management of headaches [3]. This is particularly true of those caused by migraines, where the activation of the trigeminal nerve releases calcitonin gene-related peptide (CGRP) from the trigeminovascular system [4] and causes the release of pain-producing inflammatory substances around the nerves and blood vessels of the head [5]. Massage therapy is a manipulation of the soft tissue in humans by causing movement or applying pressure to the body. It improves blood circulation within the body, thus promoting healing in the patient [6]–[8]. Moreover, chronic excessive use of medication to treat headaches can cause even more frequent headache episodes which, again, increases the need

for medication, leading to persistent headaches in the end-stage [2]. This makes massage therapy the preferable solution to break this cycle. Massages are also an important part of healthy lifestyles to relieve stress and relax the body [9], [10] while reducing chronic pain, anxiety, and depression and improving the quality of sleep and life [11], [12].

At present, most types of massage actions require massage therapists. However, such manual massages require considerable resources, most notably labor cost and time [13]–[15]. Furthermore, it is challenging for therapists to continuously exert suitable levels of force for a long time without fatigue. This decreases the number of treated patients and the effect of therapies. Thus, developing massage robots with human-like touch has become an important research topic in robotics to produce automated, repeatable massages that could become a standard part of therapy [6], [9], [16], [17]. Here, a self-actuated robotic system would be capable of operating without fatigue and have the potential to consistently treat a large number of patients [6].

To date, several studies designed a mechanical therapy unit and verified the feasibility of developing an intelligent massage robot [18]. For example, Jones et al.'s developed robotic system focused on the lumbar and thoracic portions of a patients' back, generating proper trajectories and controlling motions such that the robot produced optimal massage movements with proper force [6]. Luo et al. leveraged electromyographic (EMG) signals as part of a multi-finger robot hand control to create human-like massage on a

Figure

Fig. 1. MiBOT is a head-massage robot that integrates two symmetrically actuated massaging units, each mounting pneumatic artificial McKibben muscles to actuate a soft tactor on either side of the face. MiBOT dynamically adjusts the pressure applied to the wearer's head through a sensor-driven control loop to dynamically conform to various head shapes and topological features, while performing massage motions reminiscent of human touch.

  • These authors contributed equally to this work.

¹Department of Computer Science, ETH Zürich, Switzerland. ²Department of Mechanical Engineering, ETH Zürich, Switzerland. This work was not supported by any organization.

Figure

Fig. 2. The compliant tactor and the pulleys for providing guidance and force transfer. MiBOT integrates compliant mechanisms (a) and pulleys (b) to transform unidirectional motion into circular massage motions that approximates human-like touch.

Figure

Fig. 3. (left) An actuation block with the stepper motor, 3:1 gear reduction, and a rack to compress the mounted syringes. (right) A sound-insulated box houses five actuation blocks; three drive two syringes for circular massage motion and two drive one syringe each for tactor pressure adaptation.

person's shoulder [19], analyzing EMG signals to distinguish painful and comfortable feelings. Depending on a person's feelings and exerted massage force, their system obtains the appropriate range of input for the robotic hand. Peng et al. imitated human palm and finger motions to produce actions of massage including "pushing, picking up, and kneading" [20]. Luo et al. proposed an anthropomorphic dual-arm robot that utilizes techniques through the impedance control of both cartesian space and joint space [11]. They evaluated massages using participants' electroencephalogram (EEG) signals during therapy.

Previous studies on human-like massage have focused on the back, shoulders, and arms (e.g., [6], [11], [19]–[24]). They may not apply to headaches or specifically migraine and tension-related headaches, as they are caused by inflammatory substances around the head's blood vessels [2]. Previous massage robots that targeted the temple region have not been evaluated for the robot's effect on the person's cardiovascular system to measure the sympathetic response and, thus, the quantified outcome. Since massage therapy is effective in reducing blood pressure (BP) and heart rate (HR) [25], we argue that, optimally, a massage robot would produce similar effects compared with human-administered massage. In other words, for a massage robot to be considered successful and beneficial, it should be able to reduce blood pressure and heart rate to a comparable extent as a human massage.

In this paper, we introduce MiBOT, a novel wearable massage robot that produces human-like massage sensations on a person's temple regions. MiBOT, where the overall design is shown in Fig. 1, operates at a non-disturbing noise level and adapts to complex target topologies as well as a wide range of different head shapes. We evaluate the effect of the presented robot system on nine participants by monitoring their heart rate (HR) and blood pressure values while comparing the same quantitative metrics with a commercially available head massage. Overall, our results indicate that the massage delivered by the designed robot can modulate the cardiovascular system while showing a significant relaxation effect on participants.

We make the following contributions in this paper:

  • MiBOT, a novel human-like massage robot system that administers soft tactor actuation on the temple region of participants to alleviate headaches,
  • a feedback control loop scheme that allows massage motions to follow a defined trajectory with proper force despite naturally varying head topologies, and
  • an experimental evaluation of MiBOT's efficacy and effect on cardiovascular responses with nine participants. Our results show that our robotic system decreased participants' heart rates and blood pressure levels. To the best of our knowledge, we are the first to evaluate the effect of a wearable massage robot targeting the temple region on the human cardiovascular system and quantitatively demonstrate its effect.