arXiv:2608.30143cs.CVAug 2026

Manuscript Offprint

arXiv:2608.30143

Repeatability Characterisation and Error Budget of a Consumer Structured-Light Scanner for 3-D Wound Geometry:A Rigid-Phantom Study

Pushkal Kumar, Aadit Aggarwal, Karlen Aleksanyan

Pushkal Kumar, Aadit Aggarwal, and Karlen Aleksanyan

Abstract—We characterise the measurement error of a free-hand consumer structured-light scanner used to derive three-dimensional geometric wound descriptors. Rigid wound-care phantoms cannot change, so every difference between repeat scans of one site is measurement error; all repeats come from a single scanner unit, nine sites and 23 scans, so this characterises one instrument. The 95 percent repeatability limit for reconstructed surface area is a factor of 4.8, with a confidence interval from 3.0 to 7.0, so rescanning an unchanged site can shift the reading from a 79 percent decrease to a 377 percent increase. Forty-four of 45 descriptors fall below an intraclass correlation of 0.50, and none of 248 descriptor and pipeline combinations reaches 0.75. An error budget formed by holding the analysis region fixed, leaving sensor and reconstruction untouched, bounds the share of variance attributable to how much surface the operator captured at 75 percent for surface area, 91 percent for hull area and 95 percent for bounding-box diagonal; only hull volume is majority instrumental, at 48 percent, so a better sensor would buy little. No wound is delineated anywhere in the chain, so the comparison against the four-week area reduction used clinically to predict healing, a ratio near 2.1, is a lower bound on the noise an unsegmented pipeline must overcome, not a measurement of wound-area reproducibility. Standardising the analysis region cuts the limit to 2.13, meeting that ratio rather than clearing it. Statistical outlier removal imposes a measured systematic area deficit near 11 percent.

Index Terms—Biomedical measurement, error analysis, measurement uncertainty, structured light, surface reconstruction.

I. INTRODUCTION

CHRONIC wounds are a large and growing burden. Recent Medicare analyses put the affected population at 10.5 million beneficiaries, with annual expenditure in the tens of billions of dollars [1], [2], [3]. Because treatment decisions hinge on whether a wound is closing, measurement is not

incidental to wound care; it is the mechanism by which care is steered.

The measurement in routine use is poor. Ruler-based length ×\times width overestimates area by roughly 40% relative to digital planimetry [4], and its inter-rater intraclass correlation has been reported as low as 0.34 against 0.77 for a smartphone planimetry application [5]. Margin identification, rather than the tracing itself, dominates the error [6], and systematic review has found most instrument evaluations methodologically weak [7].

This has motivated a large body of work on three-dimensional capture. Structured light, stereophotogrammetry, and more recently consumer smartphone depth sensors have all been applied to wounds [8], [9], [10], [11], and the derived quantities have expanded well beyond area into volume, depth distribution, surface curvature, and roughness indices treated as digital biomarkers of healing.

The evidence for that expansion is thinner than its adoption suggests. Area measurement by 3-D systems is reliably good, with intraclass correlations above 0.95 across several studies. Depth and volume are a different matter: commercial 3-D cameras have been reported to underestimate volume against water displacement by 23 to 58% [12], clinical depth agreement can fall to R=0.49R = 0.49 [13], and successive systematic reviews conclude that 3-D wound technologies remain under-evidenced rather than established [14], [15], [16].

A descriptor is useful for monitoring only if a clinically meaningful change in it is larger than the noise with which it is measured. That comparison is rarely made, in part because separating measurement error from biological change in patients requires the biology to hold still, which it does not. Our design sidesteps the problem: we scan rigid synthetic wound models. Their geometry cannot change between captures, so the entire observed spread across repeat scans is attributable to the acquisition-to-descriptor chain.

This paper makes four contributions.

Manuscript submitted September 1, 2026. (Corresponding author: Pushkal Kumar.)

  1. A repeatability characterisation of 45 geometric wound descriptors computed from consumer structured-light scans of rigid phantoms, with a variance decomposition separating between-site morphological spread from within-site measurement error.

This work involved no human subjects and no animal subjects. All measurements were made on commercially manufactured synthetic wound-care training models, so institutional review board approval and informed consent were not applicable.

P. Kumar, A. Aggarwal, and K. Aleksanyan are independent researchers in the USA (e-mail: pushkalku@gmail.com; aadit.aggarwal@gmail.com; karlen.aleksanyan2007@gmail.com).

The authors declare no conflict of interest. This work received no funding. The authors have no financial or personal relationship with VATA, Inc., Shining3D, or Revopoint, whose products are evaluated here.

  1. Per-descriptor noise floors – repeatability coefficients and minimum detectable changes – benchmarked di-

Analysis code and the derived descriptor tables that reproduce every number in this paper are available from the authors.

rectly against the area-reduction thresholds used in wound care.

  1. A curvature estimator that is resolution-independent over a measured window, motivated by the observation that repeat scans of one site differ by up to a factor of 15 in vertex density, which makes per-vertex discrete curvature incomparable between captures.

  2. An error budget apportioning the measurement variance between capture extent and the instrument. Capture extent accounts for 75% of the variance in surface area and 95% in bounding-box diagonal, so a better sensor would buy little; standardising the analysis region recovers most of it in software, cutting the surface-area repeatability limit from 4.6 to 2.1 while raising intraclass correlation from 0.02 to 0.20.

The headline result is cautionary, and we think a useful one: it is specific, it is quantitative, and it supplies the noise floors that a positive claim in this area has to clear. A cautionary headline is not the whole of it, though. The same analysis identifies which error source dominates and shows that one inexpensive change to the processing pipeline removes most of it.

References (27)