Contrast-Free Autonomous Navigation of Untethered Endovascular Microrobots Using Single-Plane Fluoroscopy

Husnu Halid Alabay, Tuan-Anh Le, Ping Wang, Hakan Ceylan

ABSTRACT

Reliable three-dimensional (3D) navigation of magnetically actuated untethered microrobots remains a major barrier to clinical translation. X-ray fluoroscopy is the standard real-time imaging modality for endovascular procedures, but single-plane fluoroscopy provides only a two-dimensional (2D) projection, eliminating depth information and complicating autonomous navigation. Recovering this information through biplane imaging or repeated contrast-enhanced angiography increases procedural complexity, radiation exposure, or contrast burden. Here, we introduce VISTA (Virtual Integration for Spatial Tracking and Autonomy), a digital twin framework enabling contrast-free autonomous navigation under single-plane fluoroscopy. VISTA reconstructs vascular anatomy as a 3D digital twin, discretizes the vessel centerline into navigation milestones, and assigns the detected 2D robot position to the nearest projected milestone. Consecutive milestones define the local vessel orientation used to generate magnetic actuation commands, converting single-plane fluoroscopic observations into topology-constrained navigation states without requiring contrast injection during navigation. VISTA is demonstrated across anatomically distinct vascular phantoms under continuous flow and within the inferior vena cava of a live rat in vivo. Compared with conventional fluoroscopic human-in-the-loop control, VISTA reduced navigation time by up to 62%, corrective actuation commands by up to 98%, and radiation exposure by up to 57%. These results establish VISTA as a digital twin-guided framework for contrast-free autonomous navigation of untethered endovascular microrobots using widely available single-plane fluoroscopy.

Keywords: Digital twin; Fluoroscopic imaging; Magnetic actuation; Autonomous navigation; Millirobot; Endovascular robotics

INTRODUCTION

Untethered endovascular microrobots offer a fundamentally different approach to minimally invasive intervention. Their small size, wireless actuation, and passive compliance enable access to anatomically tortuous vasculature, including regions with fragile or inflamed vessel walls, where conventional tethered devices cannot reach without mechanical trauma [1-8]. These properties position such systems to expand endovascular therapy toward more precise, atraumatic treatment across a broader range of cardiovascular and neurovascular conditions [9, 10]. Realizing this potential, however, depends critically on the ability to localize, track, and navigate untethered robots reliably under clinically feasible conditions.

Magnetic actuation has been widely adopted for remote control of endovascular microrobots, enabling torque- and force-based manipulation without physical tethers [11-13]. Effective magnetic navigation requires continuous estimation of the robot's 3D position and orientation in order to update the external magnetic field in real time. In laboratory settings, stereo optical imaging triangulates robot pose effectively [14]. In vivo, optical line-of-sight is physically inaccessible, as biological tissue opacity eliminates this option entirely, necessitating imaging modalities compatible with clinical endovascular workflows.

X-ray fluoroscopy is the clinical standard for real-time image guidance during endovascular procedures, including angiography, balloon angioplasty, stenting, and embolization [15-17]. Fluoroscopy readily visualizes radiopaque interventional devices but does not inherently delineate the surrounding vessel lumen. Vascular anatomy is therefore conventionally visualized through contrast-enhanced angiography, with additional contrast-enhanced angiographic acquisitions performed as needed during device navigation to re-establish vascular context [18]. Although this approach is well established for catheter-based interventions, repeated contrast administration increases the cumulative procedural contrast burden, making contrast-sparing strategies particularly desirable in patients at increased risk of contrast-associated renal complications [19, 20].

This limitation becomes more consequential for untethered microrobots. Conventional catheters remain mechanically tethered and can be directly advanced, retracted, or repositioned by the operator. Once released into the circulation, an untethered robot lacks this direct mechanical control and must instead be localized and actuated remotely. Reliable navigation therefore requires knowledge of both the robot location and the surrounding vascular trajectory. Repeated contrast-enhanced angiography could intermittently recover this vascular context, but only transiently, requiring additional contrast administration whenever the vessel trajectory must be re-visualized. A navigation framework that retains vascular information throughout the procedure could instead provide persistent anatomical context without repeatedly administering contrast. Errors in localization or magnetic-field orientation may otherwise result in stalling, unintended downstream migration, or incorrect branch selection.

A second limitation is inherent to the imaging geometry itself. Single-plane fluoroscopy collapses 3D vascular anatomy onto a 2D projection, eliminating depth and obscuring local vessel orientation and branching context [21, 22]. A microrobot may therefore be clearly visible fluoroscopically while its position along the 3D vessel trajectory remains ambiguous (Fig. 1A). In complex geometries, multiple distinct 3D locations can produce similar or overlapping 2D projections, directly complicating closed-loop magnetic navigation [23]. Biplane fluoroscopy can partially recover depth by providing simultaneous projections from two imaging orientations, but introduces additional radiation exposure, calibration requirements, and potential workspace conflicts with external magnetic actuation hardware [24-27]. Moreover, biplane systems are not universally available, limiting their scalability across clinical environments [9]. An alternative approach is therefore needed that extracts sufficient spatial information for autonomous navigation from widely available single-plane fluoroscopy without requiring repeated contrast administration.

These limitations also increase the demands placed on manual magnetic control. Operators must interpret 2D fluoroscopic projections, infer motion relative to the 3D vascular trajectory, manage contrast administration, and simultaneously adjust the external magnetic actuator under dynamic flow [28]. Because untethered magnetic actuation lacks direct mechanical coupling to the device, errors in field orientation or timing can result in off-target motion, migration, or deformation. Operator reaction time and fatigue may therefore limit the consistency of purely manual fluoroscopy-guided magnetic navigation in complex vascular geometries [29, 30].

Prior approaches to endovascular microrobot navigation have incorporated flow-aware planning and model-based steering [31], but experimental validation has largely been restricted to planar geometries, optically accessible environments, or non-clinical imaging modalities. Fluoroscopy-guided studies in vascular phantoms have similarly not addressed the depth ambiguity introduced by single-plane projection imaging in anatomically realistic 3D vascular networks with overlapping vessel trajectories [32]. Consequently, autonomous navigation using single-plane fluoroscopy remains limited by the absence of persistent 3D vascular context.

We previously introduced an X-ray-guided magnetic manipulation platform combining fluoroscopic imaging with a virtual workspace and robot avatar for human-in-the-loop 2D steering [33]. That platform achieved low-latency fluoroscopic control but did not resolve depth ambiguity or support autonomous navigation through 3D vascular geometries. In related work, we developed EndoBot, a physiologically adaptive soft magnetic microrobot for atraumatic image-guided endovascular therapy [34]. Together, these developments motivated a navigation framework that could combine real-time fluoroscopic robot detection with pre-existing 3D vascular information to enable autonomous magnetic control.

Here, we present Virtual Integration for Spatial Tracking and Autonomy (VISTA), a digital twin framework for contrast-free autonomous navigation of untethered endovascular microrobots using single-plane fluoroscopy. VISTA represents the known 3D vascular centerline as a sequence of discrete navigation milestones and associates each 2D fluoroscopic robot detection with the corresponding state along this predefined vascular trajectory. Consecutive milestones provide the local vessel orientation used to configure the rotating magnetic field for closed-loop actuation (Fig. 1B). Because the vascular topology is encoded within the digital twin, fluoroscopy is used to track the radiopaque robot rather than repeatedly visualize the vessel lumen, eliminating the need for contrast administration during navigation. Rather than attempting unconstrained reconstruction of the robot's 3D pose from a single projection, VISTA leverages known vascular topology to recover the spatial information required for milestone-to-milestone autonomous control.

We validate VISTA using our helical soft microrobot, EndoBot, across diverse anatomically distinct vascular phantoms under continuous flow and within a live rat inferior vena cava model in vivo, demonstrating autonomous closed-loop navigation with reduced operator intervention, lower cognitive burden, shorter navigation times, lower radiation exposure, and fewer corrective magnetic actuation commands compared with conventional fluoroscopic human-in-the-loop control. By transforming single-plane fluoroscopy from a passive visualization modality into an active input for closed-loop navigation, VISTA establishes a scalable digital twin-guided framework for clinically deployable untethered endovascular microrobots.