arXiv:2608.27620cs.GRAug 2026

Manuscript Offprint

arXiv:2608.27620

Visual Cue Interactions in AR-Guided Needle Insertion: A Prostate Biopsy-Inspired Phantom Study

Xinrui Zou, Mingxu Liu, Thomas T. Jones, Braden Millan, Sandeep Gurram, Peter A. Pinto, Raisa Z. Freidlin, Alejandro Martin-Gomez

Xinrui Zou, Mingxu Liu, Thomas T. Jones, Braden Millan, Sandeep Gurram, Peter A. Pinto,

Raisa Z. Freidlin,

Alejandro Martin-Gomez

ID

Figure

Figure 1: System overview and visualization combinations. Left: A participant using HoloLens 2 with a tracked biopsy needle and a gel-embedded pelvis phantom. Right: Overlays rendered from the user's perspective, showing all 2×2×22 \times 2 \times 2 combinations of the three visualization cues: Focus-and-context Window (W), Color-proximity (C), and Trajectory guidance (T). Top row (W on): W, W+C, W+T, W+C+T. Bottom row (W off): None (baseline), C, T, C+T. Additional details are provided in Sec. 3.

Abstract— Despite the apparent simplicity of the motor action involved during percutaneous needle procedures, manipulating the tool's direction becomes challenging when clinicians cannot directly visualize internal anatomy and must rely on ultrasound images, which increase cognitive demand. Augmented reality (AR) offers the promise to assist with these tasks by providing pertinent visual information in the clinician's field of view. However, simply providing visual information that ignores meaningful visual cues can complicate depth perception and spatial understanding. In this work, we introduce and evaluate three visualization techniques for needle alignment developed during design sessions with medical experts: a localized focus-and-context window, a color-based proximity encoding, and an explicit trajectory overlay. These techniques were evaluated in a user study (n=26) including clinical experts (n=7) using a prostate-biopsy-inspired phantom. Results from this study suggest that cue effects depended on the surrounding cue configuration and user expertise. For novices, explicit trajectory overlay improved targeting accuracy and reduced retreat behavior, but its effect on completion time varied across cue configurations, with slower performance when the overlay was presented alone. For experts, the focus-and-context window reduced completion time and retreat events, while color-based proximity overlay improved completion time. Subjectively, color cues were often perceived as helpful even when their effects on accuracy were not consistent. These results suggest that AR guidance strategies for percutaneous interventions should consider user expertise and visual context.

Index Terms—Augmented Reality, Mixed Reality, Visualization Techniques, User Studies, Urology, Perception, Prostate Biopsy, Image-Guided Intervention

1 INTRODUCTION

Percutaneous needle procedures are spatially demanding interventions that require clinicians to navigate small instruments through soft tissue toward deep or narrow targets, often with limited visual or tactile feedback. These procedures are common in clinical practice, ranging from biopsies and drainage to localized injections and tumor ablations. Among these procedures, prostate biopsy provides a representative and clinically relevant context because accurate and efficient tissue sampling requires significant experience. Inaccurate insertion can lead to false negatives or suboptimal sampling, delaying diagnosis and compromising downstream treatment decisions [3]. These clinical risks motivate improved spatial guidance.

• Xinrui Zou is with Johns Hopkins University and National Institute of Biomedical Imaging and Bioengineering, NIH. E-mail: xzou8@jhu.edu

• Mingxu Liu is with Johns Hopkins University.

• Thomas T. Jones and Raisa Freidlin are with National Institute of Biomedical Imaging and Bioengineering, NIH.

• Braden Millan, Sandeep Gurram and Peter A. Pinto are with Urologic Oncology Branch, National Cancer Institute, NIH.

• Alejandro Martin-Gomez is with University of Arkansas. Email: alejandro.martin@uark.edu

Manuscript received xx xxx. 202x; accepted xx xxx. 202x. Date of Publication xx xxx. 202x; date of current version xx xxx. 202x. For information on obtaining reprints of this article, please send e-mail to: reprints@ieee.org. Digital Object Identifier: xx.xxxx/TVCG.202x.xxxxxx

To guide such procedures, clinicians rely on medical imaging. Computed Tomography (CT) and Magnetic Resonance Imaging (MRI) are commonly used for pre-operative planning due to their high spatial resolution and soft tissue contrast, respectively. Ultrasound (US), in contrast, is widely used for intraoperative guidance because it provides real-time imaging at the bedside. Across modalities, images are usually shown on external monitors rather than in direct alignment with the patient's body, forcing clinicians to mentally register 2D slices with the 3D physical anatomy during the intervention [8]. The spatial mismatch between the patient's body and the imaging display introduces several challenges: disrupted hand-eye coordination, increased cognitive load, and reduced spatial awareness, especially when navigating toward deep, small, or shifting targets.

Augmented reality (AR) has emerged as a promising approach to mitigate these challenges by superimposing medical imaging content, such as anatomical structures, tool trajectories, or internal targets, directly onto the patient's body. Prior studies have shown that AR can improve targeting accuracy and reduce task completion times during the performance of percutaneous needle procedures [13,23]. This technology offers valuable information for spatial guidance. However, its effectiveness depends not only on what but also on how the information is presented. A central challenge in achieving intuitive and accurate spatial understanding lies in designing overlays that convey depth and spatial relationships clearly, without introducing visual clutter or ambiguity. For example, overlays lacking strong perceptual cues, such as occlusion or shading, can appear disconnected from anatomy, reducing the operator's ability to judge distances and positions accurately [10].

To support spatial understanding in AR-guided interventions, researchers have explored a range of visualization strategies that differ in both their visual structure and the types of depth cues they provide.

Focus-and-context techniques enable the visualization of internal anatomical structures through virtual cutaways or synthetic windows [9]. By leveraging occlusion, the strongest depth cue known, these methods reveal anatomical targets by obscuring the surrounding anatomy. At the same time, they provide minimal but valuable information regarding the occluding structures around the targets, supporting enhanced spatial judgment. Because of these characteristics, focus-and-context approaches help users interpret layered anatomical relationships during tool navigation.

Another family of techniques relies on illustrative encoding, where artificial visual mappings serve to convey depth or proximity. A classic example of these approaches is chromostereopsis, a rendering technique that assigns warmer colors to near objects and cooler tones to distant ones, creating a perceptual gradient that helps distinguish spatial layers [16]. Unlike occlusion or shading, these cues require user interpretation and are intentionally designed to enhance depth understanding without modifying object geometry.

Explicit trajectory-based guidance has often been used in needle-based procedures to improve tool alignment. This includes visualizing the optimal insertion path as a color-coded trajectory, indicating entry points, or dynamically updating indicators to reflect deviation from the ideal route [21]. Such methods offer direct feedback during interaction by encoding how well the user's actions align with the intended task.

Collectively, these strategies address complementary aspects of spatial understanding, yet existing works often evaluate them in isolation. Combining multiple cues is not guaranteed to yield additive benefits because overlays may compete for attention, increase visual clutter, or cause unfavorable speed–accuracy trade-offs.

To address the limited understanding of how AR visual cues interact across user expertise levels, we conducted the same controlled factorial protocol in two cohorts: novices (Study I) and clinical experts (Study II). This design allowed us to analyze overall performance trends across all participants and test whether cue effects varied by expertise. We also conducted within-cohort analyses to characterize expertise-specific effects. We systematically evaluated three representative techniques: focus-and-context, color-based illustrative, and explicit trajectory guidance. These techniques were implemented in a unified AR platform to support controlled execution and assessment of needle-alignment tasks, with prostate biopsy serving as a representative procedural scenario. Accordingly, our evaluation focuses on task performance and user perception in a phantom-based alignment task, rather than clinical efficacy. By evaluating accuracy, task duration, and workload, our contributions are threefold: (1) a unified implementation of three representative AR visualization strategies for needle-alignment guidance; (2) a factorial, multi-metric evaluation of their main and interaction effects, including pooled and expertise-specific analyses; and (3) empirical evidence suggesting that cue benefits are expertise-dependent and not uniformly additive, yielding design considerations for AR-guided needle interventions.