Spectral Features Dominate BCG Respiratory-Event Detection: A Large-Scale Patient-Independent Comparison of Feature Groups in Sleep Apnea Patients

Israel Campero Jurado, Zoe Bousraou, Lara Benning, Sara Padilla Neira, Alexander Breuss, Robert Riener, Esther Irene Schwarz, Elisabeth Wilhelm

Abstract

Unobtrusive ballistocardiographic (BCG) sensing is a promising modality for long-term sleep-apnea monitoring, yet it remains unclear which signal features are most discriminative for respiratory-event detection. We present a literature-guided, patient-independent comparison of ten BCG feature groups using a 512-sensor capacitive pressure mat recorded simultaneously with respiratory polygraphy in 155 patients (52 female, 103 male) undergoing in-hospital evaluation for obstructive sleep apnea. Features were extracted from six spatially distinct signal channels, yielding a 191-dimensional feature vector spanning general statistical, time-domain, frequency-domain, wavelet, frame-energy, and nonlinear complexity descriptors.

Under strict leave-one-patient-out cross-validation for binary classification of respiratory-event windows versus event-free reference windows, Random Forest and Histogram Gradient Boosting achieved AUC-ROC of 0.967 and 0.969 and AUC-PR of 0.977 and 0.979, respectively. Feature-importance analysis revealed that frequency-domain features dominate discrimination: breathing-band power in the 0.1–0.4 Hz range accounted for 30.3% of to-

*Corresponding author

Email address: i.c.j.campero.jurado@rug.nl (Israel Campero Jurado) tal discriminative information across all spatial channels, and Fast Fourier Transform (FFT) spectral-shape descriptors of the adaptively preprocessed channel contributed a further 15.1%. Area Under the Curve (AUC) and curve-length features provided the main complementary time-domain evidence (21.5%), whereas wavelet-derived and nonlinear features contributed smaller secondary effects (10.4% combined across 59 features). Frequency-domain and time-domain features together accounted for 67% of total discriminative information, demonstrating that a compact, interpretable subset of the full feature library achieves clinically relevant performance under patient-independent validation and providing an empirical basis for feature selection in future BCG systems.

Keywords: Obstructive sleep apnea, BCG, Unobtrusive sensing, Feature extraction, Machine learning, Polygraphy synchronisation, Signal processing

1. Introduction

Ballistocardiographic (BCG) sensing using force or pressure sensors integrated into or placed beneath the mattress, in the bed frame, or at the foot of the bed is an attractive modality for long-term sleep monitoring because it captures respiratory effort, posture, and gross body movement without requiring sensors to be attached to the body, and is therefore comfortable enough for repeated overnight use [1, 2, 3]. Bed-based BCG has been investigated for sleep-disordered breathing (SDB) detection in a number of studies, and it has been shown that the BCG signal carries information sufficient to distinguish respiratory events from normal-breathing segments [4, 5, 6, 7, 8, 9]. However, these studies differ substantially in sensing hardware, feature definitions, prediction targets, and validation protocols (see Table 1). As a direct consequence, it remains unclear which feature groups contribute most to the discrimination of respiratory-event windows from normal-breathing windows in BCG-based detection. Prior work has used general statistical descriptors [4, 5, 7, 8, 9], time-domain waveform features [4, 6, 9], frequency-domain band power [4, 5, 7], FFT spectral-shape descriptors [10, 11], wavelet-derived features [6, 5, 9], and nonlinear complexity measures [6, 5, 10], but none has compared all of these under a single, patient-independent experimental protocol.

Establishing a feature ranking matters for two practical reasons. First, it identifies which BCG signal properties are indicative of apneic or hypop-neic events, grounding the classification in signal physiology rather than in benchmark results alone. Second, it provides an empirical basis for deciding which feature groups are worth computing in future real-time or resource-constrained systems and which can be omitted without meaningful performance loss. The goal of the present work is therefore to answer the following research questions under strict leave-one-patient-out (LOPO) cross-validation in a cohort of 155 patients:

  1. Which feature categories carry the most discriminative information for BCG respiratory-event detection?

  2. Does adaptive preprocessing (high-pass filtering, quality-score-based sensor selection, and polarity-aligned averaging) amplify feature discriminability compared with raw spatial aggregates?

  3. Which feature categories provide complementary discriminative information beyond the dominant tier?

We extracted 191 features organised into ten groups, informed by a structured review of prior BCG, electrocardiogram (ECG), and acoustic SDB studies, and applied them to six signal channels derived from a 512-sensor capacitive BCG recorded synchronously with respiratory polygraphy. Random Forest feature importance (mean decrease in impurity) provides the group ranking. We address these questions through a systematic study in which all ten feature groups are applied to six signal channels derived from a 512-sensor capacitive BCG recorded synchronously with respiratory polygraphy in 155 patients, and evaluated under strict leave-one-patient-out cross-validation. Their discriminative contributions are quantified using Random Forest mean decrease in impurity, providing a standardised, patient-independent feature-group ranking for BCG respiratory-event detection.

References (22)