arXiv:2611.12686cs.CVNov 2026

Manuscript Offprint

arXiv:2611.12686

Differentiable Inverse Rendering Recovers Intrinsic Properties from Intravascular Optical Coherence Tomography

Ajay Manicka, Karim Kadry, Sebastian Ruiz-Lopera, Maryam Alsharqi, Sanjana Arji, Samuel Ruiperez-Campillo, Nestor Uribe-Patarroyo, Brett E. Bouma, Farhad Nezami, Elazer R. Edelman

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Abstract

Medical images encode intrinsic tissue properties that indicate disease progression, yet these properties remain entangled with the system-dependent physics of image formation. Differentiable inverse rendering asks whether they are recoverable by inverting the imaging system's forward physics rather than through post-hoc correction. We address this in optical coherence tomography (OCT), focusing on the optical attenuation coefficient, a per-pixel marker of pathology and cellular composition. Conventional recovery relies on closed-form inversions limited to simplified light-tissue models, or learned methods coupled to specific imaging configurations. We instead recover per-pixel attenuation from a single OCT image by gradient descent through a differentiable physics-based forward model. We verified the method against analytical solutions, validated it on five in vitro phantoms, and applied it without modification to 29,269 clinical intravascular OCT images from 128 patients. The inversion needs no training data or segmentations and is portable across platforms and retrospective archives. Replacing the forward model or recovering another optical parameter leaves the optimization procedure unchanged. We demonstrate per-pixel recovery is tractable because gradient information scales to this high-dimensional problem, where gradient-free methods become intractable. Our work supports quantitative tissue characterization from OCT images and provides a framework for extension across imaging modalities and organ systems.

Keywords: Inverse Rendering, Medical Imaging, Differentiable Physics, Tissue Characterization, Visual Computing

Introduction

The interrogation of biological matter through diverse forms of energy constitutes the foundation of modern medical diagnostic practice. Whether utilizing magnetic gradients in magnetic resonance imaging [1], high-frequency sound waves in ultrasound [2], ionizing radiation in computed tomography [3], or backscattered light in optical coherence tomography (OCT) [4], modern imaging modalities share a common operative principle. They apply energy to tissue and infer anatomy from how the energy is differentially reflected, scattered, or absorbed by the underlying material [5]. Historically, the advancement of these technologies has focused on hardware innovations to improve spatial resolution and contrast, providing clinicians with high-fidelity anatomical visualizations of organ systems and diseased tissue [6, 7]. However, while these anatomical maps are essential, they are not direct measurements of tissue composition: each image is a projection of the underlying intrinsic properties through the physics of the imaging system, entangling composition with modality-specific acquisition artifacts. In clinical applications, intrinsic properties such as stiffness, conductivity, and pH dictate disease progression and physiological stability [8–10], yet remain inaccessible from the projected image alone. This paper inverts that projection. Although the imaging chain is a complex, modality-specific forward operator whose inverse is ill-posed, we demonstrate intrinsic tissue composition can be recovered directly from clinical imaging through differentiable inverse rendering, with OCT as our case study.

The technical foundation for this inversion comes from parallel development in computer vision. Physics-based forward models have been refined in medical imaging [6], yet they are almost exclusively used in alternate directions. They simulate, correct, or render images rather than recover tissue properties from them. The computer vision community is closing an analogous gap through differentiable rendering [11, 12], which treats the image-formation process as a differentiable function and uses gradient-based optimization to recover scene parameters from observed images [13]. The principle that gradient descent can invert any differentiable forward model to recover its input extends naturally beyond natural images and has begun to enter medical imaging in modality-specific applications such as x-ray and ultrasound [14–17].

In biomedical optical imaging, the optical attenuation coefficient (OAC) is a particularly informative quantitative tissue property, characterizing how rapidly the signal is attenuated with depth. Attenuation serves as a robust discriminator of pathology across both acoustic [18–21] and optical [22–28] regimes, and is sensitive to fundamental cellular mechanisms, including necrosis and apoptosis [29], making it a functional proxy for tissue viability beyond the shape and size of anatomical features captured by standard imaging. While IV-OCT provides high anatomical resolution [30, 31], the image entangles the OAC with the effects of complex wave propagation and system-specific acquisition parameters, making it difficult to extract intrinsic composition from image intensity [32, 33].

Other biomedical optical modalities recover different optical properties. Photoacoustic and diffuse optical methods derive contrast from optical absorption, where quantitative recovery is itself an ill-posed problem because the internal fluence is unknown [34, 35], and intravascular photoacoustic catheters remain preclinical [36]. OCT attenuation is instead dominated by scattering, the mechanism carrying lipid, calcium, and cellular density information in plaque [37, 38], and is well established as a quantitative marker across many tissue types [39–43]. Because IV-OCT is already standard of care during percutaneous coronary intervention [44, 45], inverting its image formation extracts an additional quantitative channel from images that are already acquired. OCT is a case study here as the framework requires only that the forward operator be differentiable with respect to the property of interest.

Existing approaches to OCT attenuation recovery fall into three categories [46]. Curve-fitting (CF) methods extract a single attenuation coefficient by fitting signal decay over a manually segmented region [47], which is robust for uniform tissues but unable to capture intra-tissue heterogeneity. Depth-resolved (DR) methods compute per-pixel attenuation from the ratio of local intensity to remaining energy [48, 49], but the closed-form formulation constrains them to single-scattering assumptions and makes the resulting maps sensitive to noise sources like speckle and local signal artifacts. Recent work in this area addresses the noise sensitivity issue using despeckling to improve local region homogeneity [50]. Learning-based methods can incorporate richer forward physics through synthetic training data [51], but the learned amortized inverse is fixed after training and coupled to the simulated distribution. The stationary iteration method estimates depth-dependent attenuation and backscattering without explicit layer segmentation [52], and optimization also estimates these parameters via an implicit neural representation of optical properties through an OCT imaging model using multi-angle measurements [53]. Our method instead optimizes any differentiable forward model by gradient descent from a single B-scan.

Our work targets a central question in differentiable inverse rendering of whether intrinsic tissue properties can be inferred directly from medical images by framing property recovery as the inverse of the imaging system's forward physics. To examine this in OCT, we built an automated and generalizable physics-based inverse rendering framework that extracts intrinsic tissue properties directly from standard clinical OCT images without the need for segmentations. The framework addresses all three limitations in a single formulation by delivering per-pixel attenuation while operating on an explicit forward model that can incorporate arbitrary physics and optimizing each image independently. Because the forward model is modular, incorporating additional scattering physics, such as the Extended Huygens-Fresnel model [54], or recovering further optical parameters requires only modifying the forward model, not the inversion procedure.

By inverting the path-integrated Beer-Lambert (B-L) model [55], we recovered depth-resolved, pixel-level OAC maps. We verified our method on five optically homogeneous physical phantoms against the conventional CF method and applied it to 29,269 clinical coronary IV-OCT frames [56–59] (over a 128-patient intravascular lithotripsy cohort) to demonstrate spatially resolved attenuation recovery and faithful image rendering across varying vessel morphologies. IV-OCT routinely guides percutaneous coronary intervention. Images are acquired with a catheter-based scanner that illuminates the arterial wall with a near-infrared laser and reconstructs depth-resolved cross-sections from the backscattered light. Current clinical practice relies on viewing anatomical features such as plaque shape and size, while treatment decisions depend on tissue composition. These results support the potential of per-pixel OAC maps to enable quantitative plaque characterization and guide interventional decision-making [29, 60].

The differentiable inverse rendering approach generalizes to any imaging modality whose forward physics can be expressed as a differentiable operator. Changing the forward physics can change the properties that are recovered. We can also control the optimization loop, loss function, and regularization framework to guide optimization. Our work demonstrates this paradigm in OCT, joining differentiable forward models already demonstrated for other modalities across physics-based medical imaging [61, 62]. We make the following contributions:

  • A gradient-based differentiable inverse rendering framework for per-pixel optical attenuation recovery from OCT images without training data or prior segmentation.

  • Quantitative validation against curve-fitting references across 5,120 B-scan inversions on 5 physical liquid intralipid phantoms each imaged at two focal positions, with systematic characterization of optimizer reproducibility, noise robustness, and parameter sensitivity in agar phantoms.

  • Retrospective application to 128 clinical IV-OCT coronary pullbacks of atherosclerotic tissue demonstrating spatially resolved attenuation and backscattering recovery.

Differentiable inverse rendering posits that intrinsic property recovery from images can be cast as a tractable inverse problem driven by gradient information. Our work contributes evidence for this paradigm in biomedical imaging, with OCT as the use case. We introduce a differentiable inverse rendering framework for OCT with validation across phantoms and demonstration in clinical acquisitions. Gradient information is vital because it captures the intrinsic relationships between tissue regions and guides the optimizer toward a physically consistent solution. The inversion is portable across different imaging systems and applicable to retrospective clinical archives, addressing the domain-shift barrier in clinical AI deployment. Gradient descent through an explicit forward model decouples the complexity of the physics from the availability of an inverse, whereas closed-form estimators require physics simple enough to invert analytically and learned estimators are bound to their training distribution, making the forward model a replaceable component and the approach transferable beyond OCT.