Measurement-Aware Multimodal LLM for No-Reference Low-Dose CT Quality Assessment

Kerem Basim, Tolga Demiroglu, Mehmet Ozan Unal, Metin Ertas, Isa Yildirim

Kerem Basim1,∗^{1,*}, Tolga Demiroglu1,∗^{1,*}, Mehmet Ozan Unal1^{1}, Metin Ertas2^{2}, Isa Yildirim1^{1}

1^{1}Dept. of Electronics and Communication Eng., Istanbul Technical University, Istanbul, Turkey

2^{2}Dept. of Electrical and Electronics Eng., Istanbul University, Istanbul, Turkey

{basim21, unalmehmet, iyildirim}@itu.edu.tr, tolgadem7@gmail.com, ertas@istanbul.edu.tr

∗^{*}Equal contribution

Abstract—No-reference image quality assessment (IQA) of low-dose CT predicts the score that radiologists would assign to a slice without a full-dose reference, as needed to compare dose protocols and denoisers. Trained CT IQA networks agree well with radiologists but output only a score, whereas multimodal large language models (MLLMs) can explain a score but perform poorly on CT when only prompted. We propose a measurement-aware MLLM, built on Qwen3.8-27B, that rates a slice from its image together with six image-domain measurements of noise, sharpness, high-frequency content and streaks, given to the language model as percentiles in text. On the LDCT-IQA 2023 test set, it reaches an Overall score of 2.735, on par with the retrained challenge winner (2.735) and MedSigLIP-FiLM (2.713), and substantially above its base MLLM used as a zero-shot agent. Shuffling the measurements at test time reduces the score by 0.048, mostly through noise, but the same measurements also lead the model to assign higher scores to over-smoothed denoised images.

Index Terms—Low-dose CT, image quality assessment, multimodal large language models, fine-tuning, interpretability

I. INTRODUCTION

Lowering the dose of a CT scan reduces radiation risk but increases noise and streak artifacts, and the deep learning denoisers that remove them can over-smooth fine anatomy. Comparing dose protocols and denoisers therefore requires image quality assessment (IQA) that agrees with radiologists, and manual scoring by radiologists is impractical at scale [1]. Since no full-dose reference exists in clinical practice, the assessment must be no-reference, that is, a model predicts the mean opinion score (MOS) of radiologists from the slice alone. The LDCT-IQA 2023 challenge [1] established this task as a benchmark of abdominal CT slices scored by radiologists from 0 (worst) to 4 (best).

On this benchmark, trained networks achieve high correlation with the MOS. The challenge was won by an ensemble of convolutional and transformer classifiers [2], later work adapted medical foundation models such as MedSigLIP-FiLM [3], and a recent method reports a higher score [4]. Such scores, however, depend on the evaluation protocol. Rounding predictions to the 0.2 MOS grid changes Kendall's τb\tau_{b}, and choosing a model configuration by its test-set score inflates the result. Moreover, these networks output only a score, whereas multimodal large language models (MLLMs) can also explain a score in words. Prompted general MLLMs [5] and a vision-language model paired with ChatGPT [6] assess CT quality in text, but on LDCT-IQA prompting remains substantially below trained networks (Table I). A medical MLLM fine-tuned for CT quality [7] also remains below them. For natural images and videos, MLLMs are either fine-tuned to output a quality score [8], [9] or given IQA tools that they call and reason over in an agent loop [10]–[13].

For CT, the natural counterpart of such tools are the physical measurements with which image quality is conventionally characterized, such as noise magnitude and edge sharpness [14], [15], and an MLLM can read them as text. However, it remains unclear whether an MLLM fine-tuned with these measurements reaches the accuracy of trained CT IQA networks, and whether it actually uses them. We propose a measurement-aware MLLM that rates a slice from the image together with six measurements in four groups (noise, sharpness, high-frequency content and streaks), given as percentiles in text (Fig. 1b). We compare this model with its base MLLM, Qwen3.8-27B, used as a zero-shot agent over precomputed measurements and visual views (Fig. 1a). Our contributions are (i) a measurement-aware MLLM that is on par with the strongest trained CT IQA networks on LDCT-IQA 2023 (Table I); (ii) ablation and test-time shuffling experiments showing that the model relies on the measurement values, mainly on noise, and that the same inputs lead it to assign higher scores to over-smoothed denoised images (Tables II and III, Fig. 2); and (iii) an evaluation under a fixed protocol, with the effect of rounding to the 0.2 MOS grid and configurations selected on the test set reported separately.