arXiv:2611.07087WAVES: Wearable Active Vibroacoustic Elasticity Sensing for Low-Cost and Non-Invasive Muscle Stiffness Monitoring
Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.
Abstract
Muscle stiffness is a physiologically meaningful biomarker associated with a wide range of musculoskeletal and neurological conditions. However, existing muscle stiffness assessment methods, such as shear wave elastography (SWE), manual palpation, and handheld mechanical probes, remain impractical for continuous, everyday monitoring due to bulky hardware, operator dependence, and limited self-use. Meanwhile, wearable muscle-sensing modalities such as Electromyography (EMG) and Mechanomyography (MMG) primarily capture muscle activation rather than passive mechanical properties of tissue at rest, leaving a critical gap in portable monitoring of resting muscle stiffness.
In this work, we present WAVES, a wearable system for continuous monitoring of muscle stiffness via active vibroacoustic sensing. WAVES injects controlled vibroacoustic excitation into soft tissue and estimates muscle stiffness by measuring shear-wave propagation velocity through Time-Difference-of-Arrival (TDoA) analysis across spatially distributed wearable sensors. To achieve accurate TDoA estimation with low-cost wearable hardware, we develop a physiology- and quality-guided vibroacoustic sensing framework that effectively screens and refines TDoA for accurate stiffness estimation. We evaluate WAVES through characterization experiments using ballistic jelly and end-to-end on-body evaluation with human participants. Results demonstrate that WAVES can robustly estimate muscle stiffness across multiple users and different superficial muscles while enabling lightweight, wearable deployment for everyday use. WAVES is able to achieve a muscle stiffness estimation error of 13.1% or less across multiple users and multiple muscle groups.
ACM Reference Format:
Shubham Rohal, Dong Yoon Lee, and Shijia Pan. 2026. WAVES: Wearable Active Vibroacoustic Elasticity Sensing for Low-Cost and Non-Invasive Muscle Stiffness Monitoring. In . ACM, New York, NY, USA, 15 pages. https://doi.org/10.1145/nnnnnnn.nnnnnnn
One of the most critical biomarkers for muscle health is the resting muscle stiffness, which reflects a wide range of physiological conditions and musculoskeletal changes [28, 56]. For example, athletes may experience increased passive muscle stiffness during training or fatigue, which elevates the risk of non-contact strain injuries [25, 80]. Office workers with a sedentary lifestyle may develop neck and upper-extremity musculoskeletal disorders due to posture-induced muscle stress [17]. Despite differences in manifestation, these conditions share a common underlying physical phenomenon – changes in the viscoelastic properties of the tissue – reflecting mechanical alterations in the muscle and fascia that often occur independently of voluntary contraction [54]. Although muscle stiffness is not the sole indicator of muscle health, it provides a physiologically meaningful and broadly applicable biomarker that enables quantitative assessment of underlying mechanical tissue conditions beyond instantaneous muscle activation [5].
The industry gold standard for muscle stiffness measurement – shear wave elastography (SWE) [67] – is primarily used in clinical settings due to its high cost and bulky form factor. SWE uses an ultrasound probe to image subcutaneous (in between skin and muscle) layers and assess their stiffness; however, beyond its physical and financial barriers, the method also requires a trained sonographer to operate, making it unsuitable for continuous monitoring. Similarly, manual palpation is a clinical diagnostic method in which physicians palpate patients' muscles to qualitatively assess muscle stiffness [21, 34, 53], leading to subjective and inconsistent measurements. Portable devices, such as Myoton [1], estimate muscle stiffness by applying a brief mechanical impulse to the skin and measuring the resulting damped oscillatory response of the underlying tissue [26, 67]. Crucially, operating such a device effectively requires the second person to manually align the measurement probe perpendicular to the muscle, limiting its self-operating capabilities. Wearable Doppler Elastography (WDE) [33, 41] directly measures tissue motion and mechanical wave propagation using miniaturized ultrasound sensing. However, these systems still rely
Permission to make digital or hard copies of all or part of this work for personal or classroom use is granted without fee provided that copies are not made or distributed for profit or commercial advantage and that copies bear this notice and the full citation on the first page. Copyrights for components of this work owned by others than the author(s) must be honored. Abstracting with credit is permitted. To copy otherwise, or republish, to post on servers or to redistribute to lists, requires prior specific permission and/or a fee. Request permissions from permissions@acm.org.
Conference'17, Washington, DC, USA
© 2026 Copyright held by the owner/author(s). Publication rights licensed to ACM.
ACM ISBN 978-x-xxxx-xxxx-x/YYYY/MM
https://doi.org/10.1145/nnnnnnn.nnnnnnn
Shubham Rohal, Dong Yoon Lee, and Shijia Pan

(b)

Figure 1: WAVES sensing principle illustration.
on ultrasound imaging hardware, stable acoustic coupling, and relatively complex sensing pipelines, which limit their practicality for low-cost, lightweight everyday use.
A broader class of portable muscle-sensing technologies includes wearable approaches such as Electromyography (EMG) [23, 44], and Mechanomyography (MMG) [48, 83]. While these modalities are portable and widely adopted for monitoring muscle activity, they are not designed to measure muscle stiffness at rest. EMG measures the electrical depolarization of muscle fibers, whereas MMG captures the mechanical vibration and deformation generated during voluntary contraction. Although muscle contraction can transiently increase muscle stiffness, contraction-induced stiffness differs fundamentally from the passive mechanical and viscoelastic properties of muscle and fascia tissues at rest. Consequently, these modalities cannot reliably capture persistent changes in resting stiffness, such as fascial densification or chronic muscle tightness, which may occur without significant electrical or mechanical activation signals. This leaves a critical gap in the continuous, portable, and everyday monitoring of resting muscle stiffness.
We present WAVES, a wearable sticker-based system that measures and tracks muscle stiffness via active vibroacoustic sensing. Skin-mounted stickers are already widely used in everyday life for a variety of health and wellness applications – from adhesive bandages for wound protection, to kinesiology (KT) tape for musculoskeletal support, and skin-mounted biosensors such as continuous glucose monitors (CGMs) and electrocardiogram (ECG) patches for continuous physiological monitoring. WAVES adopts the sticker form factor to enable high-quality vibroacoustic elasticity sensing on the body parts, as illustrated in Figure 1 (a). By injecting an acoustic excitation signal into the muscle and measuring its propagation via spatially distributed sensors. The muscles' stiffness directly affects shear-wave propagation speeds [9]. WAVES
calculates the shear wave velocity as the Time Difference of Arrival (TDoA) between sensor pairs and estimates the muscle stiffness accordingly, as depicted in Figure 1 (b). In this way, muscle stiffness can be estimated without relying on purely data-driven models that are often sensitive to inter-person physiological variability.
However, this is non-trivial due to several challenges. (C1) Accurately estimating the Time-Difference-of-Arrival (TDoA) between closely spaced wearable sensors using low-cost vibroacoustic hardware is difficult due to limited spatial separation, low signal-to-noise ratio (SNR), and hardware-induced timing uncertainty. (C2) Vibroacoustic propagation in soft tissue is highly heterogeneous and physiologically dependent. Variations in tissue composition, anatomical structure, posture, and muscle state introduce complex attenuation, dispersion, and propagation dynamics, making stiffness-related propagation features difficult to robustly characterize across users and real-world conditions. (C3) Wearable deployment introduces unintended mechanical coupling and spurious propagation paths through the skin interface, adhesive structure, and body movement, contaminating tissue propagation measurements and reducing the robustness of stiffness estimation in everyday use. To tackle these challenges, we co-design physiology-guided hardware with sensing-quality-aware TDoA estimation to enable reliable measurement of shear-wave propagation under wearable constraints. Specifically, we design a deployment-quality enhancement and validation pipeline to account for constraints in low-cost wearables. We compensate for noisy measurements by screening and refining TDoAs that deviate from physiological and geometric constraints. By adopting this physics-driven approach, we reduce the impact of inter-subject variability and improve robustness across users.
We conduct both characterization (on ballistic jelly) and end-to-end on-body experiments to evaluate the system's performance across different people and multiple muscle groups. WAVES achieves an average stiffness estimation error of less than 13% across multiple participants and muscle groups. This level of accuracy falls within the 20% error margin commonly considered acceptable for non-clinical and everyday physiological monitoring applications [77].
In summary, the contributions of this work include:
We present WAVES, a wearable sticker-based active vibroacoustic sensing system for continuous, unobtrusive monitoring of muscle stiffness in everyday settings.
We introduce a wearable vibroacoustic active sensing framework that models soft-tissue propagation and estimates muscle stiffness through physics-informed shear-wave velocity analysis without relying on purely data-driven models.
We co-design a physiology-guided wearable hardware and a sensing-quality-aware estimation pipeline that mitigates spurious mechanical coupling, motion interference, and low-SNR propagation challenges for robust on-body muscle stiffness estimation.
We evaluate WAVES through experiments on ballistic jelly and human subjects, demonstrating accurate and robust muscle stiffness estimation across diverse sensing conditions.
The rest of the paper is organized as follows. First, we introduce background knowledge on the causes of muscle stiffness and on acoustic wave propagation and reflection in muscles in Section 2. Then, we present the WAVES system design and implementation in
Section 3. Next, we report the experimental results in Section 4 to evaluate the design's effectiveness. Finally, we discuss the future directions in Section 6 and conclude in Section 7.
2 Background
This section introduces the physiological significance of muscle stiffness, the need for continuous monitoring, and the acoustic principle for shear-wave-based stiffness measurement, which motivate and inspire the design of WAVES.
Muscle Stiffness: Causes and Needs. A resting muscle in a healthy state should be soft and flexible, similar to a rubber band. This flexibility allows for healthy blood flow to absorb shock. However, under various physiological and pathological conditions, muscles may lose their ability to fully relax and instead remain partially contracted, entering a state referred to as prolonged stiffness [50]. Unlike the stiffness due to muscle contraction, this stiffness is involuntary and electrically silent. As a result, electromyography (EMG) becomes ineffective at capturing such changes. Here, we list several common causes of elevated resting muscle stiffness.
Prolonged sitting. During prolonged sitting, postural muscles such as the trapezius remain in a low-level, sustained contraction [79]. This continuous activation leads to metabolic accumulation and fascial dehydration/cross-linking [36]. Consequently, the local muscle–fascia complex exhibits elevated passive tension and increased viscoelastic stiffness [57]. Therefore, continuous monitoring of these muscles' stiffness can reveal sustained low-level mechanical loading before pain symptoms appear, enabling early detection and intervention.
Over-training. When muscles are repeatedly loaded beyond recovery capacity, it causes micro-damage, and the damaged fibers trigger inflammatory responses and edema [56]. This will elevate intramuscular pressure and tissue density, while nociceptor-mediated protective reflexes further increase baseline muscle tone [64]. Together, these processes generate passive tension and increase the muscle's effective viscoelastic stiffness even at rest [35]. Therefore, continuous tracking of muscle stiffness can inform persistent mechanical changes after exercise, help identify inadequate recovery, and reduce the risk of strain injury.
Acoustic-based Muscle Stiffness Measurements. Traditional Shear Wave Elastography (SWE) leverages shear wave propagation in tissue to quantify stiffness by measuring wave velocity. An ultrasound transducer generates a localized ultrasound push pulse (acoustic radiation force) to induce shear waves – propagate at a lower velocity – in tissues and to track their propagation using fast plane-wave ultrasound imaging [67]. Ultrasound imaging will generate tissue displacement maps, enabling precise tracking of shear-wave-induced displacement (micrometer scale) and the calculation of shear speed. The shear modulus of the tissue is calculated as , where is the tissue density and is the shear speed [67]. The velocity measurement of the reflected signal can be translated to the Young's modulus () or stiffness of the medium:
where is Poisson's ratio and is the density of the medium. For incompressible media, the Poisson's ratio is , i.e., [52]. In this work, we leverage the established relationship
between shear-wave velocity and tissue stiffness to enable wearable monitoring of muscle stiffness.
3 WAVES System Design
The design of WAVES is inspired by the shear wave elastography (SWE) – the gold standard for muscle stiffness measurement – where muscle stiffness directly affects the propagation dynamics of mechanically induced shear waves in soft tissues [9]. Stiff tissues generally exhibit higher shear-wave propagation velocities due to their increased resistance to deformation [9]. Based on this principle, WAVES actively injects controlled vibroacoustic excitation into the muscle and estimates the muscle stiffness by analyzing the propagation characteristics of the resulting mechanical waves, i.e., their velocity. In contrast to bulky and expensive ultrasound imaging systems, WAVES achieves a lightweight wearable form factor through physiology-guided active vibroacoustic sensing with compact wearable hardware (Section 3.1). However, measuring shear wave velocity in muscle with low-cost actuators and sensors is challenging for a few reasons – heterogeneous tissue propagation, attenuation, and dispersion of vibroacoustic signals, limited spatial separation between wearable sensors, and motion-induced interference during on-body deployment. WAVES leverages physics-informed filtering and propagation-aware signal analysis to robustly estimate shear-wave velocity (Section 3.2), and applies quality-aware screening and refinement strategies to mitigate sensing uncertainty and physiological variability (Section 3.3). Figure 2 depicts the overview with key modules in WAVES.
3.1 Physiology-Guided Active Vibroacoustic Sensing Hardware Design
WAVES adopts a sticker-based active vibroacoustic sensing design that consists of one bone-conduction transducer and multiple inertial measurement units (IMUs) to measure shear wave propagation speed on the target muscle. The transducer and the IMUs are placed in a linear array on the muscle with carefully derived distances between them, as shown in Figure 4.
3.1.1 Shear Wave-Guided Active Vibroacoustic Sensing Design. Since the sensing principle of WAVES is based on the propagation properties of mechanically induced shear waves, we first need to understand the shear wave generation and propagation in viscoelastic soft tissues.
Generation Model. Generating shear waves in soft tissue requires localized out-of-plane mechanical excitation [8]. Biological soft tissues are often modeled as viscoelastic, nearly incompressible materials. When localized out-of-plane mechanical excitation is applied to the surface, the resulting displacement field is governed by the Navier-Cauchy equation [42]:
where denotes the displacement vector field, is the density of material, and are the Lame parameters describing compressional and shear behavior, respectively.
Propagation Model. Because biological soft tissues are nearly incompressible, , causing volumetric deformation to be strongly
Shubham Rohal, Dong Yoon Lee, and Shijia Pan

Figure 2: The system overview of WAVES, consisting of three main modules. First, we adopt physiology-guided wearable hardware to enable high-quality data acquisition. The acquired signals are then used to estimate shear-wave velocity through Time Difference of Arrival (TDoA) analysis. Finally, physiology- and consistency-guided screening and refinement are applied to improve TDoA accuracy for robust stiffness estimation.

(a) High quality shear wave ()

(b) Medium quality shear wave ()

(c) Low quality shear wave ()
Figure 3: Quantifying deployment quality with Hodogram analysis of wavelet-filtered vibroacoustic signals at different accelerometers shown in Figure 1. The metric , calculated as the ratio between the ranges of the and axes, quantifies shear-wave quality relative to other vibration noise. Low values indicate deployments with poor signal quality that require physical adjustment.
constrained while transverse shear propagation becomes the dominant measurable mechanical response [8]. Therefore, shear-wave propagation provides a more informative and measurable indicator for stiffness estimation in soft tissue [10]. The displacement field generated by the localized out-of-plane mechanical excitation is often modeled using Helmholtz decomposition, which can be separated into compressional (P-wave) and shear (S-wave) components:
where represents longitudinal compressional waves and represents transverse shear waves. This indicates that the P-wave particle motion is parallel to the propagation direction, whereas the S-wave particle motion is perpendicular to propagation [20]. The transverse displacement associated with S-wave propagation produces measurable surface-normal vibration that can be captured by skin-mounted IMU sensors.
Active Sensing Design. The penetration depth and propagation behavior of the generated shear waves are strongly frequency dependent [76]. Higher-frequency excitation produces shorter wavelengths and improved temporal localization but experiences stronger attenuation and shallower penetration depth. In contrast, lower-frequency excitation propagates deeper into tissue with reduced attenuation, enabling mechanically induced waves to interact with deeper muscle structures [76]. Therefore, the actuation frequency in WAVES must balance penetration depth, propagation stability, and temporal resolution for reliable wearable stiffness sensing.
We select actuation transducers that can effectively provide controllable mechanical excitation normal to the skin surface, making commonly available rotational vibromotors, such as eccentric rotating mass (ERM) actuators [82], unsuitable for reliable shear-wave generation. Bone-conduction transducers, on the other hand, are designed to generate localized out-of-plane vibroacoustic excitation

Figure 4: WAVES form factor and mounting mechanism.
through direct mechanical coupling with solid media. They use an internal vibrating mass driven by an electromagnetic actuator to generate controllable normal displacement at the contact surface [38]. This enables efficient mechanical coupling of vibration energy into soft tissue and produces repeatable shear-wave propagation suitable for stiffness sensing.
To generate temporally localized and distinguishable propagation events, WAVES uses short-duration actuation pulses instead of continuous excitation. Short pulses reduce temporal overlap between successive propagation responses and improve the separability of wavefront arrivals across sensing units, which is critical for reliable TDoA estimation. However, unlike ultrasound-based SWE systems that generate near-impulsive excitation through acoustic radiation force, the pulse duration achievable by WAVES is constrained by the dynamics of the bone-conduction transducer and the mechanical loading imposed by the soft tissue interface. Specifically, the actuator requires sufficient excitation duration to overcome inertial response and establish stable mechanical coupling before generating a measurable shear-wave event. Therefore, the pulse duration in WAVES is determined by jointly considering the hardware response characteristics and the expected propagation delay range between sensing units. We select the minimum duration sufficient to generate stable, repeatable excitation at the target frequency.
3.1.2 Physiology- and Quality- Guided Sensor Deployment Strategy
The hardware design must jointly account for the expected shear-wave velocity range in soft tissue, anatomically feasible sensor placement on target muscles, and the sampling-rate requirements to resolve small TDoA differences between closely spaced sensors.
The propagation velocity of shear waves in biological soft tissue is substantially lower than compressional waves and typically falls within a limited range - 2 to 8 m/s - determined by tissue stiffness and physiological conditions [10]. Because WAVES estimates muscle stiffness through propagation delay measurements, the spatial separation between sensing units directly affects the measurable TDoA range. Smaller sensor spacing would reduce propagation delay differences between sensors and increase the required sampling rate for accurate measurement. Excessive spacing increases propagation attenuation and exposure to heterogeneous tissue dynamics,

(a) Raw signal pairs.
(b) Filtered signal.

(c) CWT filtered signal.

(d) Peak detection.
Figure 5: Signal examples. (a) and (b) depict the shear wave actuation signals captured by two sensors at distances of 3.5cm and 7cm from the actuator. (c) is the continuous wavelet transform (CWT) for the signal depicted in (a). (d) shows the peak detection on the signals in (c).
resulting in higher noise impact on the sensor signal. Therefore, the hardware configuration – sensor layout and sampling rate – needs to balance TDoA resolution and sensing reliability under wearable deployment constraints.
On the other hand, sensor placement is also constrained by anatomical and physiological factors. Because shear-wave propagation in muscle is anisotropic, propagation behavior depends strongly on the direction relative to muscle fibers. Muscle fibers form aligned mechanical structures, so along-fiber propagation encounters fewer structural boundaries and less scattering than cross-fiber propagation. As a result, longitudinal propagation more consistently reflects the intrinsic mechanical behavior of the target muscle fibers. To address these constraints, WAVES adopts a compact skin-mounted sensing geometry that places sensing units along physiologically meaningful propagation paths while minimizing rigid structural coupling between components.
To systematically quantify sensing quality and assess the effectiveness of the shear-wave propagation captured at a distance d, WAVES leverages hodogram analysis on the wavelet-filtered vibroacoustic signals. A hodogram visualizes the multi-axis particle motion trajectory of the sensed vibration signal over time, enabling characterization of propagation coherence, directional consistency, and structural interference patterns [63]. Because shear-wave propagation induces structured transverse motion, high-quality propagation signals typically produce concentrated and geometrically
Shubham Rohal, Dong Yoon Lee, and Shijia Pan
consistent hodogram trajectories, whereas noisy or distorted measurements exhibit scattered and irregular motion patterns.
Figure 3 illustrates three representative hodograms generated from signals captured at different sensing distances. At the shortest propagation distance, as shown in 3 (a), the received signals exhibit coherent directional motion with concentrated hodogram trajectories, indicating strong mechanical coupling and stable shear-wave propagation. As the propagation distance increases, attenuation, dispersion, and heterogeneous tissue interactions progressively distort the propagation waveform, resulting in broader and less structured hodogram patterns. At excessive sensing distances, the measured signals become dominated by noise and spurious propagation artifacts, producing highly scattered hodogram distributions with weak directional consistency.
We use Hodogram-based signal-quality assessment to identify reliable sensing configurations and to reject low-confidence propagation measurements. The mechanically induced shear wave generates a dominant transverse displacement normal to the skin surface. As a result, valid propagation events should exhibit motion geometry dominated by the z-axis, which can be quantified via Hodogram axis range ratio . Measurements with a z-to-x ratio below 1.25 are rejected because they typically indicate weak shear-wave coupling, excessive structural vibration, or spurious propagation artifacts.
3.1.3 Spurious Propagation Path Mitigating Skin-Sensor Interface. A key challenge in wearable vibroacoustic sensing is that the wearable device itself may introduce unintended mechanical propagation paths in addition to the target tissue propagation path. Because mechanically induced waves can propagate through rigid structural components, loose skin coupling, and wearable substrates, the sensing signals may, in this case, contain structural vibrations unrelated to intrinsic tissue mechanics. Such spurious propagation paths distort propagation-delay estimation and reduce the reliability of stiffness sensing. On the other hand, many existing wearable sensing systems adopt garment-based form factors that prioritize flexibility and freedom of movement. However, these loosely coupled interfaces introduce unstable skin contact and additional structural vibration paths that are undesirable in propagation-based muscle sensing, where stable, localized mechanical coupling is critical for reliable wave transmission and measurement.
To address this challenge, WAVES adopts a compact, adhesive, skin-mounted interface that minimizes the transmission of structural vibrations outside the target tissue path while maintaining stable mechanical coupling between the sensing components and the skin surface. Figure 4 depicts the skin-mounted interface on a person's thigh, where the IMU array is aligned with the muscle fiber direction and attached to the skin via the adhesive patches [40]. The adhesive patch material ensures stable mechanical coupling between the skin and the sensing components for efficient shear-wave generation and transmission, while mechanically isolating adjacent components to reduce vibration leakage and cross-component interference.
3.2 Shear-Wave Signal Processing for Time Difference of Arrival (TDoA) Estimation
The raw signals captured by the IMU array are then processed to estimate the shear-wave velocity. However, directly estimating shear-wave velocity from wearable IMU measurements is challenging due to low SNR, propagation attenuation, waveform dispersion, posture-induced interference, and structural vibration artifacts introduced during on-body deployment.
To address these challenges, WAVES adopts a physics-informed signal processing pipeline that explicitly incorporates the propagation characteristics of shear waves in viscoelastic soft tissue. Instead of relying on purely data-driven feature extraction, the system leverages propagation-aware filtering, wavelet-domain temporal analysis, and physiological signal quality constraints to robustly estimate Time Difference of Arrival (TDoA) between sensing units. The resulting TDoA measurements are then used to derive shear-wave propagation velocity for downstream stiffness estimation.
3.2.1 Reduce Posture-Induced Artifacts. Although WAVES is designed to estimate resting muscle stiffness under passive muscle conditions, posture changes may still introduce motion artifacts and alter the propagation characteristics of the captured vibroacoustic signals. In order to reduce such posture-induced motion artifacts, WAVES applies zero-phase filtering [74]. In this way, the temporal alignment of the propagation wavefront is preserved without phase distortion that leads to TDoA estimation errors. Figure 5 depicts the example signals, where the raw signal and filtered signal are presented in (a) and (b), respectively.
3.2.2 Reduce Propagation Artifacts. Although the actuation signal is designed to be narrowband, imperfections in actuator response and mechanical coupling introduce frequency leakage, causing different spectral components to propagate with different attenuation and phase characteristics in soft tissue. To effectively isolate the target propagation band, WAVES applies Continuous Wavelet Transform (CWT)-based filtering with the corresponding scale. CWT jointly preserves time and frequency information and has been proven effective for impulsive and dispersive vibroacoustic signal TDoA estimation [45]. WAVES adopts the first-order Gaussian Wavelet , because it is commonly used for transient onset detection [55]. The wavelet scale corresponding to the excitation frequency is extracted for further analysis. Figure 5 demonstrates the CWT filtered signal.
3.2.3 Peak-based TDoA Calculation. One of the well-known challenges for TDoA estimation is the noisy onset detection. To acquire a smooth and stable characterization of the propagation model, we first apply the Hilbert Transform [70] to extract the envelope of the extracted scale, denoted as . An adaptive peak threshold based on the statistical distribution of the envelope signal – mean and standard deviation – is then derived as . The first detected peak satisfying is treated as the propagation onset. This adaptive thresholding strategy improves robustness to signal-amplitude variation due to propagation and inter-subject physiological variability.

(a) Physiological Constraint Screening (PCS). (b) Cross-Sensor Consistency Screening (CCS). (c) Screened trials after PCS and CCS.
Figure 6: Quality-aware TDoA estimation screening.
3.3 Quality-Aware TDoA Screening and Refinement
Although propagation-aware signal processing can estimate TDoA with mitigated propagation and physiological variability, there may still be uncertainty factors that affect the measurement. To further refine the estimation for higher accuracy, we leverage the physical relationships among multiple TDoA measurements from different sensor pairs to quantify estimation quality. The intuition is as illustrated in Figure 6, where the x-axis is the TDoA measured between sensor pair 0-1 and the y-axis is the TDoA measured between sensor pair 0-2. We can derive the relationship between and as below
where , , are distances between the actuator and sensors as depicted in Figure 4. Therefore, the ideal estimation values should fall on the line that satisfies this relationship, which is marked by the red line in Figure 6 (a). The blue, yellow, and brown solid lines in Figure 6 (a) marks the TDoA calculated from the ground truth value of the muscle stiffness at 330 N/m, 410 N/m, and 510 N/m, respectively, where . The intersection of these lines and the TDoA ratio line in red indicates the ideal measurements under corresponding conditions. However, due to various physical factors, such as sensor noise, tissue heterogeneity, propagation attenuation, and residual structural vibration artifacts, the TDoA estimation may be erroneous. WAVES conducts the estimation assessment based on the following criteria to determine whether the estimation is reliable or not.
3.3.1 Physiological Constraint Screening (PCS): Individual TDoA Estimation Quality Assessment and Refinement. First, the estimated propagation velocity must fall within the expected shear-wave velocity range in biological muscle tissues. When the TDoA measurement-based velocity estimation falls outside of the physiologically plausible propagation velocity range, an erroneous peak detection is likely to occur. Since active sensing via bone-conduction transducers uses short-duration actuation pulses (on the 1/10-second
timescale), while resting muscle stiffness changes occur at a much coarser timescale (hours to weeks), the system can discard these erroneous measurements and acquire new ones. Therefore, for individual TDoA measurement, we conduct the physiological-based filtering as shown in Figure 6 (b). The green lines are the TDoA ranges calculated using the known parameter and the range of shear-wave propagation speeds in soft tissues. Individual estimates that fail this criterion are rejected because they typically indicate poor coupling, excessive structural vibration, or spurious propagation paths.
3.3.2 Cross-Sensor Consistency Screening (CCS) and Refinement (CCR). Although the remaining measurements satisfy individual quality constraints, estimation uncertainty may still arise from waveform distortion and onset-detection errors. Fortunately, WAVES employs multiple sensors, yielding multiple pairs of TDoA measurements. Because all sensor pairs observe the same propagation event, their TDoA measurements must satisfy the geometric relationship described in Equation 4. As aforementioned, the red line in Figure 6 represents the ideal relationship between TDoA measurements from different sensor pairs, where the ratio of the measured TDoAs matches the ratio predicted by the physical relationship of the sensor placements. To account for noise and propagation artifacts, we consider measurements whose TDoA ratios fall within a bounded region around the ideal ratio as reliable candidates for stiffness estimation. The bounded region is marked by the red dashed lines in Figure 6(b), and Figure 6(c) shows the measurements that satisfy the proposed propagation-consistency criterion. WAVES further projects the measurements onto the ideal ratio relationship, refining the estimates by enforcing physical consistency among TDoA measurements.
3.3.3 Statistical Aggregation across Repeated Measurements. The propagation consistency analysis helps identify reliable measurements. However, there may still be measurement variability due to various factors, such as transient changes in coupling, viscoelastic damping, and measurement noise. Because the time scale for muscle stiffness changes much more coarsely than that of individual sensing events, multiple measurements can be viewed as
Shubham Rohal, Dong Yoon Lee, and Shijia Pan
repeated observations of the same physiological state. Therefore, WAVES aggregates repeated measurements to further reduce the uncertainty of the estimation. WAVES employs a Gaussian Mixture Model (GMM) for clustering to aggregate multiple TDoA estimates by identifying the dominant cluster and its centroid as the aggregated TDoA value. This aggregated TDoA value is then converted into shear-wave velocity using the known sensor placement geometry. This repeated-measurement aggregation effectively reduces estimation variance, thereby improving robustness to transient physiological and sensing artifacts.
4 System Evaluation
We evaluate WAVES through three sets of experiments: (1) Sensing capability characterization on multiple tissue simulants prepared using ballistic gel to control stiffness level and ranges; (2) On-body data collection with multiple users across different muscle groups (thigh, trapezius, and forearm); (3) Continuous monitoring experiment on a user during a workout session.
4.1 Prototype Implementation
We customize printable circuit boards (PCBs) for reliable, comfortable data collection. The sensing platform is built around the ESP32-S3 Mini N8 microcontroller unit (MCU) [2], which performs actuator control, signal acquisition, and wireless communication. The platform is designed to interface with two actuators and six IMU sensors, enabling data collection across various muscle groups. For example, the symmetric design allows simultaneous data collection on both sides of the trapezius muscles. A GD-01 bone conduction speaker [31] with a MAX98357A amplifier [3] is adopted as the shear wave generating actuator. BMI270 triaxial Inertial Measurement Units (IMUs) [11] are used for capturing shear wave signals. The system is driven by a 5V Lithium-Ion battery and supports wireless communication via the User Datagram Protocol (UDP).
Given the expected shear-wave velocity range in skeletal muscle (2–8 m/s) [29] and the typical length of target muscles (10–20 cm) [61], we design the sensor spacing to provide sufficient propagation distance for reliable TDoA estimation while maintaining compatibility with wearable deployment on common muscle groups. The distances between the bone-conduction transducer and three IMU sensors are cm, cm, and cm, respectively. Based on prior literature, shear waves at frequencies between 200 and 300 Hz [69] is ideal for vibroacoustic muscle sensing. The bone-conduction transducer (BCT) generates excitation impulses at frequencies ranging from 200 Hz to 350 Hz in 10 Hz increments. The signal frequency determines how deeply the shear wave will travel, allowing WAVES to effectively capture stiffness. In each trial, the BCT produces 16 pulses, each a 45-millisecond impulse, as the excitation signal. Excitation signals are separated by 200 ms intervals to ensure one actuation to completely dissipate and do not overlap with succeeding signals.
4.2 Experiment Design
We conduct a series of experiments on both tissue simulants and human subjects. This section describes the data collection and ground-truth collection used throughout the evaluation, as well as the evaluation metrics.
Table 1: Human-subject data collection summary.
| Experiment | Participant ID | Measurements # |
|---|---|---|
| Thigh | 1-5 (5M) | 2000 |
| Trapezius | 2,6-9 (2M+3F) | 2000 |
| Forearm | 1,10-13 (5M) | 2000 |
| Cross-day (forearm) | 11 | 1700 |
| Deployment quality | 3,5,6,11,12,14 | 500 |
| Total | - | 8300 |
Data Collection Procedures. We use the following data collection procedure for all experiments in the subsequent sections. First, the system will be installed on the surface of the target soft tissue. Figure 7 shows the different deployments of WAVES system across tissue simulant gels and different human muscles.
For experiments with tissue simulant gels, we use tissue simulants of three different stiffnesses. Each gel block has a length, width, and height of 220 mm, 120 mm, and 25 mm, respectively. The three tissue simulants samples (Sim Gel 1,2,3) are fabricated with the ground truth stiffness measured of 393 N/m (or 31.5 kPa), 593.4 N/m, and 713 N/m, which simulate stiffness to that of resting muscle, moderately stiffer muscle, and high stiffness muscle that is prone to injury, respectively. We collected 200+ trials for each simulant gel sample.
All human-subject recruitment, enrollment, and data collection procedures were conducted in accordance with an Institutional Review Board (IRB)-approved protocol. We use a flexible 3D-printed stencil to minimize distance variance across deployments caused by human error, which is demonstrated in Figure 7 (b). WAVES actuator and sensors are deployed in a line aligning with the underlying muscle fibers' direction. To measure resting muscle stiffness, data were collected while participants maintained relaxed standing or seated postures and avoided voluntary contraction of the target muscle. To ensure participant comfort and compliance, participants were allowed to select the muscle groups to be measured. In total, 14 participants were enrolled in the study, including 11 males and 3 females. For each muscle group investigated, measurements were collected from at least 5 participants, as summarized in Table 1.
Ground Truth Collection Procedures. We utilize the Myoton Pro[4] palpation device to collect ground truth stiffness to verify and evaluate WAVES's stiffness estimation capability. Figure 7 (f) demonstrates an example session of the ground truth collection procedure. A single stiffness reading from the Myoton Pro may be subject to measurement variability due to placement angle and pressure. To ensure an accurate stiffness ground truth, we collect 5 or more stiffness readings from the Myoton Pro and calculate the aggregated mean stiffness for each data collection instance.
Evaluation Metrics. We compare the output of WAVES system, which is the stiffness estimated with the aggregated TDoA estimation from multiple measurement events. Because WAVES rejects uncertain measurements based on physical knowledge and signal quality assessment to ensure estimation accuracy, we consider (1) estimation error, and (2) rejection rate, as the major evaluation metrics that reflects the system's performance. We report the error

(a)

(b)
(c)

(d)
(e)
(f)
Figure 7: Experiment settings. (a) shows the system installation on the ballistic gel for controlled characterization of sensing capability. (b) depicts the 3D-printed installation guide tool used during on-body system installation to ensure consistent sensor placement. (c-e) present the data collection deployment for different muscles (thigh, forearm, and trapezius). (f) shows the ground truth collection using Myoton Pro [4].

Figure 8: Stiffness estimation distribution for three simulant ballistic gels with different stiffness levels. Each data point is calculated with one trial of TDoA measurement.
percentage calculated as to reflect practical estimation quality – the estimation error within 30% is considered good and acceptable for non-clinical use. [77].
4.3 Characterization with Tissue Simulant
To systematically characterize the capability of our low-cost active vibroacoustic elasticity sensing approach, we first conduct controlled experiments using ballistic gel made tissue simulants [13, 78]. Ballistic gels have been widely used as soft-tissue simulants because their mechanical properties can approximate those of biological tissues and can be reproducibly controlled across experiments [6, 7, 14]. As a result, they provide a well-controlled environment for evaluating sensing performance before deployment on human subjects. We simulate different tissue stiffness levels using ballistic gel with varying concentrations of gelatin and water.

Figure 9: Stiffness estimation across ballistic gel samples with different stiffness levels. The x-axis depicts the number of TDoA trials aggregated for a single stiffness estimate.
4.3.1 Overall Stiffness Estimation Performance. We first evaluate whether WAVES can accurately distinguish simulant gel samples with different stiffness levels. Figure 8 shows the distribution of predicted stiffness across three different tissue simulant gels. The x-axis shows the different simulant gels, while the y-axis shows the stiffness in N/m. The ground-truth stiffness measurements for Sim Gel 1, 2, and 3 are shown as horizontal blue, yellow, and green lines, with values of 393 N/m, 593.4 N/m, and 713 N/m, respectively. Three forms of analysis are shown for WAVES output. First, each data point represents an estimate from a single active sensing trial. Built upon that, the box plot shows that mean stiffness is close to ground truth stiffness, and the half violin plot illustrates that the estimated stiffness values are concentrated around the mean with a low stiffness estimation outlier. The GMM applied to the distribution, as illustrated by the violin plot, reports estimated stiffness values of 387.81 N/m, 641.44 N/m, and 747.51 N/m, which correspond to estimation errors of 1.42%, 8.46%, and 4.8%, respectively. We observe that the estimated stiffness is more concentrated around the ground truth for Sim Gel 1, while the estimated stiffness for Sim Gel 2 and Sim Gel 3 is more spread out. This indicates that stiffness estimation becomes more error-prone at higher stiffness levels, likely because
Shubham Rohal, Dong Yoon Lee, and Shijia Pan

(a) Rejection rate for different screening and refinement strategies.

(b) Stiffness estimation error across different baseline methods.

(c) Stiffness estimation error comparison of WAVES with CC and GCC-PHAT baselines.
Figure 10: Evaluation of stiffness estimation performance with rejection rate and estimation error across different baselines as well as different stiffness simulant gels.
higher shear-wave velocity produces smaller TDoA values, making the estimation more sensitive to sampling-rate limitations.
We further evaluate the impact of the number of trials on the overall stiffness estimation. WAVES uses the Gaussian Mixture Model(GMM) to find the centroid locus among stiffness estimation from multiple trials of measurements. We randomly select N different events and predict the overall stiffness from the GMM centroid output. We repeated this process with 20 different random seed values, varying N from 10 to 750 in increments of 10 trials per estimation. Figure 9 shows how the number of trials used for estimation impacts the error in stiffness estimation. We observe that the error is highest for N = 10, as the variation among the estimations is higher, as shown by the min-max estimation. The mean of the error is 10%, with a minimum error of 0.2% and the maximum error of 54%. However, this changes rapidly as the number of trials increases. We observe that the mean stiffness estimation stabilizes when the number of trials reaches 50, and the mean estimation error for all three tissue-simulant gels converges to within 10% of the ground-truth stiffness. For the on-body experiment in Section 4.4, we adopt , which translates to approximately 1-2 minutes of data collection.
4.3.2 Ablation Study. We conduct the ablation study to understand the effectiveness of each modularized design and define the following baselines.
Baseline 1 [w/o GMM]: This baseline uses all the in-trial screening and refinement strategies except using GMM to find the centroid locus for cross-trial refinement; instead, the mean of the TDoA measurements is used to estimate stiffness.
Baseline 2 [w/o (GMM, CCS)]: For this baseline, we modify baseline 1 by removing the Cross-Sensor Consistency Screening (CCS) illustrated by dashed red lines in Figure 6 (b).
Baseline 3 [w/o (GMM, CCS, PCS)]: This baseline is a modified version of baseline 2 by removing the Physiological Constraint Screening (PCS) marked by green lines in Figure 6 (a).
Baseline 4 [Raw, w/o (GMM, CCS, PCS, CCR)]: This baseline is a modified version of baseline 3 by removing the Cross-Sensor
Consistency Refinement (CCR); instead, the raw estimation is obtained by combining the means of TDoA estimations from TDoA-12 and TDoA-13.
Figure 10a and 10b show the rejection and estimation error percentages, respectively, across all baselines and our WAVES. WAVES outperforms all the baselines in stiffness estimation with an average error of 4.86%. The next-best method is baseline 1, with an average estimation error of 16.13%. The rejection rate across WAVES and baseline 1 is the same, as they both perform the same screening process. The stiffness estimation errors for baseline 2 and 3 are 26.9% and 21.4%, respectively. This shows that the screening process can effectively identify and reject low-quality data points, thereby improving the overall system's performance. The baseline 4 performs the worst, with an average estimation error of 34.9%, while Sim Gel 2 demonstrates an error up to 70.2%. On the other hand, the rejection trend is along expectation; with an increase in screening strategies, the rejection rate also increases.
4.3.3 TDoA Estimation Comparison. We also evaluated WAVES's performance against commonly used TDoA estimation methods to further cement the effectiveness of WAVES's wavelet-based TDoA estimation and physiology-based filtering. We select Cross Correlation (CC) [12] and Generalized Cross-Correlation with Phase Transform (GCC-PHAT) [37, 60, 73] as two TDoA baselines. For both baselines, we compute the TDoA for all measurement trials using CC and GCC-PHAT, and take the mean of these TDoAs as the cross-trial aggregation output. Figure 10c shows the stiffness estimation error across these three different methods. WAVES outperforms both baselines across all three stiffness levels with approximately 16× and 22× improvement on average. We also observe that the performance of both the CC and GCC-PHAT baselines degrades with increasing stiffness. For example, on Sim Gel 3 with a stiffness of 713.2 N/m, the estimation errors are 71.26% and 95.37% for CC and GCC, respectively. This may be because increasing stiffness leads to higher shear-wave velocities and, therefore, smaller TDoA values. Combined with signal attenuation in the viscoelastic medium,

Figure 11: Overall stiffness estimation across different muscles and tissue simulant gel samples.

Figure 12: Stiffness estimation errors across three muscle groups.
this makes TDoA estimation with correlation-based methods more error-prone.
4.4 On-Body Stiffness Estimation Analysis
WAVES shows highly accurate stiffness estimation performance on the tissue simulant, which has an ideal, consistent density compared to human muscle. To evaluate and validate WAVES's effectiveness in predicting muscle stiffness, we conduct a thorough on-body evaluation as detailed in Section 4.2.
Figure 11 shows the results for muscle stiffness estimation across 13 participants and three different muscle groups, along with the stiffness estimation for tissue simulant gels for comparison. The ground truth is shown along the x-axis, while the stiffness estimation from WAVES is shown along the y-axis. The dark blue line at the center shows the ideal scenario with both predicted stiffness and ground truth being the same value, while the 20% error region is marked along this line, indicating the acceptable error range for non-clinical usage [77]. The stiffness estimates for the back (trapezius), forearm, and thigh muscles are shown in yellow, red, and pink circles, respectively. Each circle represents stiffness estimates for different users. For all participants, the stiffness estimation error remains within the 20% range across all three muscle groups. While the muscle stiffness across different users ranges from 296 N/m to 428 N/m, WAVES consistently tracks inter-subject and inter-muscle variation without performance degradation.

Figure 13: Same subject muscle stiffness estimation error across three days.

Figure 14: Rejection rate comparison at different deployment quality levels, each with three deployment samples.
4.4.1 Cross Muscle Analysis. Because of physiological differences among muscles, such as the presence of subcutaneous fat and proximity to bones, signal propagation across human muscles can vary. To evaluate WAVES's robustness to these physiological differences, we tested WAVES across three different muscle groups, as discussed in Section 4.4. By evaluating data from five participants per muscle group, we account for inter-subject and intra-subject physiological variability. Figure 12 shows the stiffness estimation error among the thigh – vastus lateralis, back – trapezius muscle, and forearm – brachioradialis, respectively. Similar to Figure 11, we can observe that the muscle stiffness estimation error is under a 20% range. The average stiffness estimation error for the thigh muscle is 13.1%, the highest among the three muscles. The average errors for the forearm and back muscles are 7.4% and 5.9%, respectively.
4.4.2 Multiple Day Deployment Variation. We also tested WAVES's stiffness estimation over multiple days to validate its robustness across different deployments of the same muscle on the same user. For this study, we collected data from subject 11's forearm on three different days and compared stiffness estimation errors. Figure 13 shows the result for stiffness estimation on the same user across the same muscle on different days. We observe that overall muscle stiffness varies across users and days. However, the WAVES's stiffness estimation error is 15.71% or lower across all 3 days of deployment. This shows that with consistent deployment, the WAVES can effectively estimate muscle stiffness.
4.4.3 On-body Deployment Quality Comparison. To evaluate deployment quality, WAVES relies on the signal's energy concentration along the z-axis (normal to the surface) and the x-axis (along the surface, perpendicular to the signal propagation) to assess the shear-wave components as discussed in Section 3.1.2. The ratio is used to verify whether the deployment is valid. Figure 14 shows a comparison of rejection rates across six deployments, with high-quality deployments on the left and low-quality on the right.
Shubham Rohal, Dong Yoon Lee, and Shijia Pan
We observe that in these examples, the rejection rate is lower in high-quality deployments than in low-quality deployments – the identified high-quality deployments can acquire effective measurement trials 5× more than low-quality deployments on average.
5 Related Work
This section introduces related work from three aspects, which align with the technical contributions and provide comparisons to demonstrate novelty.
Muscle Stiffness Measurement and Estimation. Muscle stiffness is an important biomarker for assessing musculoskeletal health, rehabilitation progress, and neuromuscular disorders. Existing approaches estimate muscle stiffness through manual palpation [22, 49, 58], handheld mechanical probing devices such palpation devices [4, 18], and ultrasound shear-wave elastography has become a widely adopted method because shear-wave velocity is strongly correlated with tissue stiffness [24, 46, 65]. However, these systems typically require specialized imaging hardware, precise sensor placement, and substantial computational resources, limiting their suitability for continuous and everyday monitoring. Recent estimation approaches also rely on biomechanical models or indirect physiological measurements such as muscle activation signals, which may not directly capture tissue stiffness under varying conditions. In contrast, WAVES explores a low-frequency active sensing approach that estimates muscle stiffness from shear-wave propagation using a lightweight wearable sensing system designed for continuous deployment.
Wearable Systems for Muscle Health Monitoring. Many wearable solutions have been explored for muscle activation and health monitoring. Electromyography (EMG) is among one of most well explored modality for muscle activity and health. EMG senses the electrical changes in muscle that arise from neuron activity and muscle engagement. Change in this signal during muscle activity is studied to estimate movement and muscle fatigue [15, 19, 32, 51, 81]. The other sets of works like Mechanomyography (MMG) uses the low frequency vibrations generated by the muscle fibers to track active muscle contraction [47, 71] and fatigue [48, 68]. The other set of work use electric impedance sensing for muscle activity and motion monitoring [85, 86]. However, all these works only measure active muscle contraction induced changes and does not measure the physical property like muscle stiffness. There are active vibroacoustic based systems [59, 62] that brings the muscle stiffness sensing to wearable form factor. However, these studies only profile relative stiffness changes and do not estimate or predict stiffness.
Physics-Informed and Quality-Aware Sensing Systems. Signals from low-cost Internet of Things (IoT) sensing systems are often noisy, impacted by heterogeneous deployment conditions, and measurement artifacts. As a result, quality-aware sensing and inference strategies have been explored to improve the robustness in sensing applications [39, 43, 84]. For example, AutoQual quantifies sensor data quality as a data-driven model of upstream sensing system deployment factors that impact downstream task performance, based on expert heuristics [84]. SQEE quantifies sensing data quality as the uncertainty of downstream machine learning models, with poor sensing conditions typically increasing prediction uncertainty [39].
EarFusion leverages the designed modality-specific Signal Quality Indices to determine the multimodal fusion strategies in real-time [43]. Earlier work on acoustic target tracking quantifies sensing data quality by comparing measured propagation delays with those predicted by a physical propagation model and uses the resulting quality estimate to suppress unreliable measurements and prioritize high-quality sensing reports [75].
None of these solutions is applicable to the task of shear-wave velocity estimation via TDoA measurements for a few reasons. First, in wearable sensing systems, quality-impact factors and target information are often tightly integrated and difficult to separate, as in ambient sensing solutions. Second, TDoA estimation is fundamentally a signal-processing problem rather than a machine-learning prediction task. As a result, uncertainty estimates derived from downstream predictive models are either unavailable or only become observable after the TDoA has already been estimated. Last but not least, the target shear-wave signal is often mixed with multiple propagation components and interference sources. Therefore, existing quality-aware sensing approaches do not provide this level of propagation-specific assessment. WAVES explores new ways to quantify wave-specific quality for deployment guidance and to refine TDoA measurement estimation by designing wave-specific propagation-based metrics.
6 Discussion
This section discusses potential directions and new design spaces that build on this paper.
Opportunistic Sensing via Haptic Interaction. In this work, we use a bone-conduction transducer to generate shear waves for active sensing. Meanwhile, vibration transducers have been widely used for haptic interaction [30]. Therefore, one future direction is to explore opportunistic sensing through interaction, in which the same actuation signals used to provide haptic feedback are repurposed as excitation sources for sensing tissue or surface responses. This problem is non-trivial because the design for sensing and interaction has different goals that need to be aligned and managed to jointly optimize the effects. This alignment needs to take into account multiple factors, including sensing quality, user perception/fatigue, and interaction utility, requiring the exploration of new design spaces.
Closed-loop Intervention via Actionable Feedback. While this work focuses on measuring muscle stiffness, various forms of closed-loop intervention can be explored in the future. As discussed in Section 2, muscle stiffness is an actionable physiological indicator associated with fatigue, recovery, injury risk, and rehabilitation progress. Therefore, timely intervention leveraging continuous stiffness measurements would open new doors for muscle health management. For example, at a coarse temporal scale (hours to weeks), a long-term stiffness profile can be leveraged to provide personalized recommendations on training, recovery, and rehabilitation plans. At a fine temporal scale (seconds to minutes), haptic interventions (discussed in Section 6) can provide real-time, muscle-specific feedback to guide exercise execution, stretching, and movement correction.
Together, these interventions at different scales enable a closed-loop framework that provides comprehensive assistance for muscle health management.
7 Conclusion
This paper presents WAVES, a wearable vibroacoustic muscle-sensing system for muscle stiffness estimation. We design WAVES, a low-cost vibroacoustic active-sensing wearable hardware that uses a bone-conduction transducer and an IMU array to profile shear-wave propagation in the muscle to estimate resting muscle stiffness. WAVES uses a physiology-guided deployment and signal processing pipeline to isolate the shear-wave propagation for effective collaborative TDoA estimation. WAVES utilizes a physiology- and geometry-guided screening and refinement process to mitigate error. We validate WAVES across tissue-simulant ballistic gels of varied stiffness and test it with multiple users and muscle groups. WAVES achieves an average stiffness estimation maximum error rate of 13.1% across multiple muscle groups. Our results highlight the potential of wearable vibroacoustic sensing as a practical and scalable approach for long-term muscle health monitoring, opening opportunities for applications in muscular injury prevention and rehabilitation.
References
[1] 2023. Myoton - Muscle Tone, Stiffness, Elasticity measurement device. https://www.myoton.com [Online; accessed 5. Jun. 2026].
[2] 2026. ESP32-S3-MINI-1 & ESP32-S3-MINI-1U Datasheet. https://documentation.espressif.com/esp32-s3-mini-1_mini-1u_datasheet_en.html [Online; accessed 17. Feb. 2026].
[3] 2026. MAX98357A. https://www.analog.com/en/products/MAX98357A.html [Online; accessed 17. Feb. 2026].
[4] 2026. Technology - Myoton. https://www.myoton.com/technology [Online; accessed 17. Feb. 2026].
[5] Abdulrahman M. Alfurahai, Ai Lyn Tan, Philip O'Connor, Paul Emery, and Richard J. Wakefield. 2019. The effect of ageing on shear wave elastography muscle stiffness in adults. Aging Clin. Exp. Res. 31, 12 (Dec. 2019), 1755–1763. doi:10.1007/s40520-019-01139-0
[6] Ester Maria Rodrigues de Andrade, Lucas Paixão, Bruno Melo Mendes, and Telma Cristina Ferreira Fonseca. 2025. Monte Carlo modeling and simulation of a new 3D printed phantom for WBC calibration with ballistic gel as a tissue substitute. Appl. Radiat. Isot. 215 (Jan. 2025), 111565. doi:10.1016/j.apradiso.2024.111565
[7] G. J. Appleby-Thomas, D. C. Wood, A. Hameed, J. Painter, V. Le-Seelleur, and B. C. Fitzmaurice. 2016. Investigation of the high-strain rate (shock and ballistic) response of the elastomeric tissue simulant Perma-Gel . Int. J. Impact Eng. 94 (Aug. 2016), 74–82. doi:10.1016/j.ijimpeng.2016.04.003
[8] Kemal Arda, Nazan Ciledag, Elif Aktas, Bilgin Kadri Aribas, Kenan Köse, Kemal Arda, Nazan Ciledag, Elif Aktas, Bilgin Kadri Aribas, and Kenan Köse. 2013. Quantitative Assessment of Normal Soft-Tissue Elasticity Using Shear-Wave Ultrasound Elastography. Am. J. Roentgenol. (Jan. 2013). doi:10.2214/AJR.10.5449
[9] Michel Bernabei, Sabrina SM Lee, Eric J Perreault, and Thomas G Sandercock. 2020. Shear wave velocity is sensitive to changes in muscle stiffness that occur independently from changes in force. Journal of applied physiology 128, 1 (2020), 8–16.
[10] Jonathon Blank, Matthew Blomquist, Lesley Arant, Stephanie Cone, and Joshua Roth. 2022. Characterizing musculoskeletal tissue mechanics based on shear wave propagation - a systematic review of current methods and reported measurements. Ann. Biomed. Eng. 50, 7 (March 2022), 751. doi:10.1007/s10439-022-02935-y
[11] Bosh. [n.d.]. BM270 datasheet document revision 1.5 document release date March 2023. https://www.bosch-sensortec.com/media/boschsensortec/downloads/datasheets/bst-bmi270-ds000.pdf
[12] Paul Bourke. 1996. Cross correlation. Cross Correlation , Auto Correlation—2D Pattern Identification 596 (1996).
[13] John Breeze and Debra J. Carr. 2016. Physical Models: Tissue Simulants. In Blast Injury Science and Engineering. Springer, Cham, Switzerland, 145–153. doi:10.1007/978-3-319-21867-0_11
[14] Julia Caldwell and James J. Mooney. 2019. Analysis of Soft Tissue Materials for Simulation Development. Simulation in Healthcare 14, 5 (Oct. 2019), 312. doi:10.1097/SIH.0000000000000382
[15] Kang-Ming Chang, Shin-Hong Liu, and Xuan-Han Wu. 2012. A Wireless sEMG Recording System and Its Application to Muscle Fatigue Detection. Sensors 12, 1 (Jan. 2012), 489–499. doi:10.3390/s120100489
[16] Sebastien Francois Martin Chastin, Eduardo Ferriolli, Nathan A. Stephens, Ken C. H. Fearnon, and Carolyn Greig. 2012. Relationship between sedentary behaviour, physical activity, muscle quality and body composition in healthy older adults. Age Ageing 41, 1 (Jan. 2012), 111–114. doi:10.1093/ageing/afr075
[17] SHARMILA CHAUDHURI, MEENA GUPTA, VANDANA PHADKE, and JAS-MINE KAUR CHAWLA. 2024. Development and Validation of an Instrument for Upper Cross Syndrome: A Cross-sectional Study. Journal of Clinical & Diagnostic Research 18, 10 (2024).
[18] Li-ling Chuang, Ching-yi Wu, and Keh-chung Lin. 2012. Reliability, validity, and responsiveness of myotonometric measurement of muscle tone, elasticity, and stiffness in patients with stroke. Archives of physical medicine and rehabilitation 93, 3 (2012), 532–540.
[19] D. F. Collins, T. Cameron, D. M. Gillard, and A. Prochazka. 1998. Muscular sense is attenuated when humans move. J. Physiol. 508, 2 (April 1998), 635–643. doi:10.1111/j.1469-7793.1998.00635.x
[20] J. W. M. Dankhaar. 1985. SEPARATION OF P- AND S-WAVES*. Geophys. Prospect. 33, 7 (Oct. 1985), 970–986. doi:10.1111/j.1365-2478.1985.tb00792.x
[21] Melissa J. Davidson, Poul M. F. Nielsen, Andrew J. Taberner, and Jennifer A. Kruger. 2019. Is it time to rethink using digital palpation for assessment of muscle stiffness? Neurol. Urodyn. 39, 1 (Oct. 2019), 279–285. doi:10.1002/nau.24192
[22] Melissa J. Davidson, Poul M. F. Nielsen, Andrew J. Taberner, and Jennifer A. Kruger. 2020. Is it time to rethink using digital palpation for assessment of muscle stiffness? Neurol. Urodyn. 39, 1 (Jan. 2020), 279–285. doi:10.1002/nau.24192
[23] Carlo De Luca. 2006. Electromyography. Encyclopedia of medical devices and instrumentation (2006).
[24] Sarah F. Eby, Pengfei Song, Shigao Chen, Qingshan Chen, James F. Greenleaf, and Kai-Nan An. 2013. Validation of shear wave elastography in skeletal muscle. J. Biomech. 46, 14 (Sept. 2013), 2381–2387. doi:10.1016/j.jbiomech.2013.07.033
[25] Jan Ekstrand, Markus Walden, and Martin Hägglund. 2016. Hamstring injuries have increased by 4% annually in men's professional football, since 2001: a 13-year longitudinal analysis of the UEFA Elite Club injury study. Br. J. Sports Med. 50, 12 (June 2016), 731–737. doi:10.1136/bjsports-2015-095359
[26] Giovanna Ferraioli, Richard G Barr, André Farrokh, Maija Radzina, Xin Wu Cui, Yi Dong, Laurence Rocher, Vito Cantisani, Eleonora Polito, Mirko D'Onofrio, et al. 2022. How to perform shear wave elastography. Part II. Medical ultrasonography 24, 2 (2022), 196–210.
[27] Peter Francis, Mark Lyons, Mathew Piasecki, Jamie Mc Phee, Karen Hind, and Philip Jakeman. 2017. Measurement of muscle health in aging. Biogerontology 18, 6 (Dec. 2017), 901–911. doi:10.1007/s10522-017-9697-5
[28] Richard L Gajdosik. 2001. Passive extensibility of skeletal muscle: review of the literature with clinical implications. Clinical biomechanics 16, 2 (2001), 87–101.
[29] Jean-Luc Gennisson, Thomas Deffieux, Emilie Macé, Gabriel Montaldo, Mathias Fink, and Mickaël Tanter. 2010. Viscoelastic and anisotropic mechanical properties of in vivo muscle tissue assessed by supersonic shear imaging. Ultrasound in medicine & biology 36, 5 (2010), 789–801.
[30] Yu Huang, Kuanming Yao, Jiyu Li, Dengfeng Li, Huiling Jia, Yiming Liu, Chun Ki Yiu, Wooyoung Park, and Xinge Yu. 2022. Recent advances in multi-mode haptic feedback technologies towards wearable interfaces. Materials Today Physics 22 (Jan. 2022), 100602. doi:10.1016/j.mtphys.2021.100602
[31] Adafruit Industries. 2026. Bone Conductor Transducer with Wires - 8 Ohm 1 Watt. https://www.adafruit.com/product/1674#technical-details [Online; accessed 16. Feb. 2026].
[32] P. A. Karthick, Diptasree Maitra Ghosh, and S. Ramakrishnan. 2018. Surface electromyography based muscle fatigue detection using high-resolution time-frequency methods and machine learning algorithms. Comput. Methods Programs Biomed. 154 (Feb. 2018), 45–56. doi:10.1016/j.cmpb.2017.10.024
[33] Jon-Èmile S Kenny, Chelsea E Munding, Joseph K Eibl, Andrew M Eibl, Bradley F Long, Aaron Boyes, Jianhua Yin, Pietro Verrecchia, Matthew Parrotta, Ronald Gatzke, et al. 2021. A novel, hands-free ultrasound patch for continuous monitoring of quantitative Doppler in the carotid artery. Scientific Reports 11, 1 (2021), 7780.
[34] Simple F. Kothari, Mohit Kothari, Raul F. Zambra, Lene Baad-Hansen, and Peter Svensson. 2014. Standardization of Muscle Palpation—Methodological Considerations. Clin. J. Pain 30, 2 (Feb. 2014), 174. doi:10.1097/AJP.0b013e3e31828c893d
[35] Lilian Lacourpaille, Antoine Nordez, Francois Hug, Antoine Couturier, Christian Dibie, and Gael Guilhem. 2014. Time-course effect of exercise-induced muscle damage on localized muscle mechanical properties assessed using elastography. Acta Physiologica 211, 1 (2014), 135–146.
[36] Sven-Erik Larsson, Ann Bengtsson, Lars Bodegård, KG Henriksson, and Jörgen Larsson. 1988. Muscle changes in work-related chronic myalgia. Acta Orthopaedica Scandinavica 59, 5 (1988), 552–556.
[37] Ran Lee, Min-Seok Kang, Bo-Hyun Kim, Kang-Ho Park, Sung Q. Lee, and Hyung-Min Park. 2020. Sound Source Localization Based on GCC-PHAT With Diffuse-ness Mask in Noisy and Reverberant Environments. IEEE Access 8 (Jan. 2020),
Shubham Rohal, Dong Yoon Lee, and Shijia Pan
7373-7382. doi:10.1109/ACCESS.2019.2963768
[38] Seung Hyun Lee, Ki Woong Seong, Kyu-Yup Lee, and Dong Ho Shin. 2020. Optimization and Performance Evaluation of a Transducer for Bone Conduction Implants. IEEE Access 8 (May 2020), 100448–100457. doi:10.1109/ACCESS.2020.2994323
[39] Shuheng Li, Jingbo Shang, Rajesh K Gupta, and Dezhi Hong. 2022. Sqee: A machine perception approach to sensing quality evaluation at the edge by uncertainty quantification. In Proceedings of the 20th ACM Conference on Embedded Networked Sensor Systems. 277–290.
[40] Zhengwei Liao, Ata Golparvar, Mohammad Javad Bathaei, Filipe Arroyo Cardoso, and Clementine M. Boutry. 2026. Wearable and Implantable Devices for Continuous Monitoring of Muscle Physiological Activity: A Review. Adv. Sci. 13, 15 (March 2026), e09934. doi:10.1002/advs.202509934
[41] Hsiao-Chuan Liu, Yushun Zeng, Chen Gong, Xiaoyu Chen, Piotr Kijanka, Junhang Zhang, Yuri Genyk, Hisham Tchelepi, Chonghe Wang, Qifa Zhou, et al. 2024. Wearable bioadhesive ultrasound shear wave elastography. Science Advances 10, 6 (2024), eadk8426.
[42] Li-Wei Liu and Hong-Ki Hong. 2014. A Clifford Algebra Formulation of Navier-cauchy Equation. Procedia Eng. 79 (Jan. 2014), 184–188. doi:10.1016/j.proeng.2014.06.329
[43] Yang Liu and Alessandro Montanari. 2025. EarFusion: Quality-Aware Fusion of In-Ear Audio and Photoplethysmography for Heart Rate Monitoring. In Proceedings of the 3rd ACM International Workshop on Intelligent Acoustic Systems and Applications. 31–36.
[44] Kerry R Mills. 2005. The basics of electromyography. Journal of Neurology, Neurosurgery & Psychiatry 76, suppl 2 (2005), ii32–ii35.
[45] Mostafa Mirshekari, Shijia Pan, Jonathon Fagert, Eve M Schooler, Pei Zhang, and Hae Young Noh. 2018. Occupant localization using footstep-induced structural vibration. Mechanical Systems and Signal Processing 112 (2018), 77–97.
[46] Naokazu Miyamoto, Kosuke Hirata, Hiroaki Kanehisa, and Yasuhide Yoshitake. 2015. Validity of Measurement of Shear Modulus by Ultrasound Shear Wave Elastography in Human Pennate Muscle. PLoS One 10, 4 (April 2015), e0124311. doi:10.1371/journal.pone.0124311
[47] Frank Mokaya, Roland Lucas, Hae Young Noh, and Pei Zhang. 2015. Myovibe: Vibration based wearable muscle activation detection in high mobility exercises. In Proceedings of the 2015 ACM international joint conference on pervasive and ubiquitous computing. 27–38.
[48] Frank Mokaya, Roland Lucas, Hae Young Noh, and Pei Zhang. 2016. Burnout: a wearable system for unobtrusive skeletal muscle fatigue estimation. In 2016 15th ACM/IEEE International Conference on Information Processing in Sensor Networks (IPSN). IEEE, 1–12.
[49] Corrie Myburgh, Anders Holsgaard Larsen, and Jan Hartvigsen. 2008. A Systematic, Critical Review of Manual Palpation for Identifying Myofascial Trigger Points: Evidence and Clinical Significance. Arch. Phys. Med. Rehabil. 89, 6 (June 2008), 1169–1176. doi:10.1016/j.apmr.2007.12.033
[50] Masatoshi Nakamura, Tome Ikezoe, Yohoi Takeno, and Noriaki Ichihashi. 2011. Acute and prolonged effect of static stretching on the passive stiffness of the human gastrocnemius muscle tendon unit in vivo. Journal of Orthopaedic Research 29, 11 (2011), 1759–1763.
[51] Karen L. Newcomer, Timothy D. Jacobson, David A. Gabriel, Dirk R. Larson, Robert H. Brey, and Kai-Nan An. 2002. Muscle activation patterns in subjects with and without low back pain. Arch. Phys. Med. Rehabil. 83, 6 (June 2002), 816–821. doi:10.1053/apmr.2002.32826
[52] Naotaka Nitta, Makoto Yamakawa, Hiroyuki Hachiya, and Tsuyoshi Shiina. 2021. A review of physical and engineering factors potentially affecting shear wave elastography. Journal of Medical Ultrasound 48, 4 (2021), 403–414.
[53] Paul S. Nolet, Hainan Yu, Pierre Côté, Anne-Laure Meyer, Vicki L. Kristman, Deborah Sutton, Kent Murnaghan, and Nadège Lemeunier. 2021. Reliability and validity of manual palpation for the assessment of patients with low back pain: a systematic and critical review. Chiropr. Man. Therap. 29, 1 (Dec. 2021), 33. doi:10.1186/s12998-021-00384-3
[54] Claudio Orizio, Bertrand Diemont, Fabio Esposito, Enrico Alfonsi, Giovanni Parinello, Arrigo Moglia, and Arsenio Veicsteinas. 1999. Surface mechanomyogram reflects the changes in the mechanical properties of muscle at fatigue. Eur. J. Appl. Physiol. Occup. Physiol. 80, 4 (Aug. 1999), 276–284. doi:10.1007/s004210050593
[55] Melani I. Plett. 2007. Transient Detection With Cross Wavelet Transforms and Wavelet Coherence. IEEE Trans. Signal Process. 55, 5 (May 2007), 1605–1611. doi:10.1109/TSP.2006.890874
[56] Uwe Proske and David L Morgan. 2001. Muscle damage from eccentric exercise: mechanism, mechanical signs, adaptation and clinical applications. The Journal of physiology 537, 2 (2001), 333–345.
[57] Peter P Purslow. 2002. The structure and functional significance of variations in the connective tissue within muscle. Comparative Biochemistry and Physiology Part A: Molecular & Integrative Physiology 133, 4 (2002), 947–966.
[58] Alasdair T. L. Rathbone, Liza Grosman-Rimon, and Dinesh A. Kumbhare. 2017. Interrater Agreement of Manual Palpation for Identification of Myofascial Trigger Points: A Systematic Review and Meta-Analysis. Clin. J. Pain 33, 8 (Aug. 2017), 715. doi:10.1097/AJP.0000000000000459
[59] Shubham Rohal, Shreya Shiriram, VP Nguyen, and Shijia Pan. 2022. FinePose: Fine-Grained Postural Muscle Profiling via Haptic Vibration Signals. In Proceedings of the 2022 Workshop on Body-Centric Computing Systems. 19–24.
[60] Daniele Salvati, Carlo Drioli, Gian Luca Foresti, et al. 2021. Time Delay Estimation for Speaker Localization Using CNN-Based Parametrized GCC-PHAT Features.. In Interspeech. 1479–1483.
[61] David Shier, Jackie Butler, and Ricki Lewis. 1996. Human anatomy and physiology. McGraw-Hill.
[62] Shreya Shirram, Shubham Rohal, Zhizhang Hu, Yue Zhang, Phuc Nguyen, and Shijia Pan. 2022. Sedentary Posture Muscle Monitoring via Active Vibratory Sensing. In 2022 21st ACM/IEEE International Conference on Information Processing in Sensor Networks (IPSN). IEEE, 531–532.
[63] Richard D. Slack, Daniel A. Ebrom, John A. McDonald, and Robert H. Tatham. 1993. Thin layers and shear-wave splitting. Geophysics 58, 10 (Oct. 1993), 1468–1480. doi:10.1190/1.1443361
[64] William T Stauber. 2004. Factors involved in strain-induced injury in skeletal muscles and outcomes of prolonged exposures. Journal of Electromyography and Kinesiology 14, 1 (2004), 61–70.
[65] Mikaela L. Stiver, Seyed Ali Mirjalili, and Anne M. R. Agur. 2023. Measuring Shear Wave Velocity in Adult Skeletal Muscle with Ultrasound 2-D Shear Wave Elastography: A Scoping Review. Ultrasound Med. Biol. 49, 6 (June 2023), 1353–1362. doi:10.1016/j.ultramedbio.2023.02.005
[66] Barbara Strasser. 2020. Importance of Assessing Muscular Fitness in Secondary Care. Front. Genet. 11 (Oct. 2020), 583810. doi:10.3389/fgen.2020.583810
[67] Mihra S Taljanovic, Lana H Gimber, Giles W Becker, L Daniel Latt, Andrea S Klauser, David M Melville, Liang Gao, and Russell S Witte. 2017. Shear-wave elastography: basic physics and musculoskeletal applications. Radiographics 37, 3 (2017), 855–870.
[68] M Tarata, A Spaepen, and R Puers. 2001. The accelerometer MMG measurement approach, in monitoring the muscular fatigue. Measurement Science Review 1, 1 (2001), 47–50.
[69] Wakako Tsuchida, Yoshiki Yamakoshi, Shingo Matsuo, Mayu Asakawa, Keita Sugahara, Taizan Fukaya, Eiji Yamanaka, Yuji Asai, Naotaka Nitta, Toshihiko Ooie, et al. 2020. Application of the novel estimation method by shear wave elastography using vibrator to human skeletal muscle. Scientific reports 10, 1 (2020), 22248.
[70] Timothy Ulrich. 2006. Envelope calculation from the Hilbert transform. Los Alamos Nat. Lab., Los Alamos, NM, USA, Tech. Rep (2006).
[71] Raphael Uwamahoro, Kenneth Sundaraj, and Indra Devi Subramaniam. 2021. Assessment of muscle activity using electrical stimulation and mechanomyography: a systematic review. Biomedical engineering online 20, 1 (2021), 1.
[72] Flavio T. Vieira, Yuanjun Cai, M. Cristina Gonzalez, Bret H. Goodpaster, Carla M. Prado, and Andrea M. Haqq. 2025. Poor muscle quality: A hidden and detrimental health condition in obesity. Rev. Endocr. Metab. Disord. 26, 5 (Oct. 2025), 723–744. doi:10.1007/s11154-025-09941-0
[73] Jiadong Wang, Xinyuan Qian, Zihan Pan, Malu Zhang, and Haizhou Li. [n.d.]. GCC-PHAT with Speech-oriented Attention for Robotic Sound Source Localization. In 2021 IEEE International Conference on Robotics and Automation (ICRA). IEEE, 2021–05. doi:10.1109/ICRA48506.2021.9561885
[74] Jianhong Wang, Yongjiang Ye, Xiang Pan, and Xudong Gao. 2015. Parallel-type fractional zero-phase filtering for ECG signal denoising. Biomed. Signal Process. Control 18 (April 2015), 36–41. doi:10.1016/j.bspc.2014.10.012
[75] Qixin Wang, Wei-Peng Chen, Rong Zheng, Kihwal Lee, and Lui Sha. 2003. Acoustic target tracking using tiny wireless sensor devices. In Information Processing in Sensor Networks. Springer, 642–657.
[76] Xiuming Wang, Yue Hu, Jia'an Zhu, Junxue Gao, Si Chen, Fang Liu, Wenxue Li, Yiqun Liu, and Bilig Ariun. 2020. Effect of acquisition depth and precompression from probe and couplant on shear wave elastography in soft tissue: an in vitro and in vivo study. Quantitative Imaging in Medicine and Surgery 10, 3 (March 2020), 754. doi:10.21037/qims.2020.01.15
[77] Konstantin Warneke, Thomas Gronwald, Sebastian Wallot, Alessia Magno, Martin Hillebrecht, and Klaus Wirth. 2025. Discussion on the validity of commonly used reliability indices in sports medicine and exercise science: a critical review with data simulations. European Journal of Applied Physiology 125, 6 (2025), 1511–1526.
[78] Anna Wermer, Joseph Kerwin, Kelsea Welsh, Ricardo Mejia-Alvarez, Michaelann Tartis, and Adam Willis. 2020. Materials Characterization of Cranial Simulants for Blast-Induced Traumatic Brain Injury. Mil. Med. 185, 1 (Jan. 2020), 205–213. doi:10.1093/milmed/usz228
[79] RH Westgaard and CJ De Luca. 1999. Motor unit substitution in long-duration contractions of the human trapezius muscle. Journal of neurophysiology 82, 1 (1999), 501–504.
[80] Erik Witvrouw, Lieven Danneels, Peter Asselman, Thomas D'Have, and Dirk Cambier. 2003. Muscle flexibility as a risk factor for developing muscle injuries in male professional soccer players: a prospective study. The American journal of sports medicine 31, 1 (2003), 41–46.
[81] Richard Woodward, Sandra Shefelbine, and Ravi Vaidyanathan. [n.d.]. Pervasive Motion Tracking and Muscle Activity Monitor. In 2014 IEEE 27th International
Symposium on Computer-Based Medical Systems. IEEE, 27–29. doi:10.1109/CBMS.2014.43
[82] Zhaoqian Xie, Dongjun Bai, Jianli Ma, Chengyu Jing, and Ming Lu. 2026. Characterization of Bilayer Tissue Moduli and Thickness Via Eccentric Rotating Mass Dynamics. Appl. Mech. Rev. 78, 4 (July 2026). doi:10.1115/1.4071065
[83] Hironori Yasukawa and Yasuo Yoshimura. 2015. Introduction of MMG standard method for ship maneuvering predictions. Journal of marine science and technology 20, 1 (2015), 37–52.
[84] Yue Zhang, Zhizhang Hu, Susu Xu, and Shijia Pan. 2021. AutoQual: task-oriented structural vibration sensing quality assessment leveraging co-located mobile sensing context. CCF Transactions on Pervasive Computing and Interaction 3, 4
(2021), 378–396.
[85] Junyi Zhu, Yuxuan Lei, Aashini Shah, Gila Schein, Hamid Ghaednia, Joseph Schwab, Casper Harteveld, and Stefanie Mueller. 2022. MuscleRehab: Improving unsupervised physical rehabilitation by monitoring and visualizing muscle engagement. In Proceedings of the 35th Annual ACM Symposium on User Interface Software and Technology. 1–14.
[86] Junyi Zhu, Jackson C Snowden, Joshua Verdejo, Emily Chen, Paul Zhang, Hamid Ghaednia, Joseph H Schwab, and Stefanie Mueller. 2021. EIT-kit: An electrical impedance tomography toolkit for health and motion sensing. In The 34th Annual ACM Symposium on User Interface Software and Technology. 400–413.