When can multi-fragment 2D/3D registration be trusted? Calibrated certification and uncertainty-driven view planning for pelvic fracture fluoroscopy

Bowen Xiang, Michael I. Miga

Abstract

Registering a segmented CT to fluoroscopy can locate every fracture fragment but can guide surgery only when trusted. After each shot, our closed loop registers all fragments, bounds their errors with a calibrated sandwich covariance, flags implausible or ambiguous ones, and stops or plans the next view. On simulated fluoroscopy of 60 held-out fracture CTs with realistic image uncertainty, 95% bounds covered 83% of unflagged fragments and 91% of those within 5 mm, but from the second view on 14% failed by over 5 mm, a third with bounds of at most 2 mm. A fragment's shift when a view was added separated tightly bounded failures best (exploratory; AUC 0.94, bound 0.81). Requiring that the newest view moved no fragment beyond its bound, preferred post hoc, lowered false stops at 2 mm from 23% to 6% overall, with 345 instead of 362 correct ones but more trajectories uncertified; delaying the stop did not, and other signals at equal acceptance left 10–15% false. With a segmentation error shared by all views (median Dice 0.89), 21% of this rule's stops were false against 10% without it (15 cases; exploratory). Minimax planning certified sooner than the clinical protocol without the check (4.20 versus 4.72 shots, Holm p<0.001p < 0.001), but no criterion reached the pre-specified endpoint, a true 2 mm error, significantly sooner, even in the true basin. On real fluoroscopy of five intact pelves no 2 mm certificate was false. Trust takes evidence across views, not only a calibrated local bound.

Keywords: 2D/3D registration, view planning, uncertainty calibration, Fisher information, fluoroscopy, pelvic fracture

1. Introduction

Percutaneous fixation of pelvic fractures is guided by fluoroscopy: the surgeon reduces the displaced fragments and places screws in narrow bony corridors, checking them in inlet, outlet and lateral views, which with conventional imaging has taken a median of more than a hundred shots per screw (Alemdaroglu et al., 2014), with the corresponding dose (Prost et al., 2024). Registering the pre-operative CT, segmented into its fragments, to the intra-operative images shows where every fragment is in three dimensions (Han et al., 2021; Grupp et al., 2020b). Such a registration can guide the next step only if the system also knows when its result is accurate, and which view to acquire when it is not.

Both are questions about uncertainty, and both are harder for several fragments than for one bone. One image barely constrains a fragment's depth, and small fragments overlapped by large ones are weakly constrained in any view, so whether a further view resolves a fragment depends on the fracture. A decision to stop ("every fragment is within 2 mm") needs a per-fragment error bound that holds in practice, not a residual or a similarity score. A bound derived from the registration's local fit, however, in-

herits its blind spot: intensity-based registration can converge to a wrong local optimum, a wrong basin, where the local fit can look as well determined as at the truth.

We study when multi-fragment 2D/3D registration can be trusted, in a closed loop that after each shot registers all fragments to all acquired views, computes the covariance of the fragment poses from rendered image derivatives, calibrates it against held-out errors and certifies the case once the 95% bound on the worst fragment's error is at most a threshold; otherwise it acquires the candidate view that minimises the worst fragment's predicted error (Fig. 1; method in Fig. 2). Flags for implausible and ambiguous fragments and a check of stability under the next view add evidence from beyond the local fit. We ask three questions (Fig. 1d): are the calibrated bounds reliable; how does the certificate fail, and which evidence available at run time reveals it; and does choosing the views help? Our contributions are:

  • a certificate for intensity-based multi-fragment 2D/3D registration: per-fragment bounds from the sandwich covariance of a tempered maximum-a-posteriori (MAP) estimator with a gradient-correlation loss, computed from rendered derivatives and calibrated by one factor, combined with failure flags and a sequential stability check, with ablations that separate what makes the bounds valid from what makes the stop decision reliable;

*Corresponding author. Email address: bowen.xiang@vanderbilt.edu (Bowen Xiang)

(a) Stop, or acquire another view?

after every C-arm shot, all fragments are

registered to all acquired views

Figure

(b) Confidently wrong bounds

objective along one fragment's pose (schematic): the bound reflects the curvature at the estimate

Figure

S1: 14% of fragment registrations failed by >5 mm, 32% of these with a bound of at most 2 mm

(c) Evidence across views

the same objective after one more view: the truth stays, a wrong basin rarely does

Figure

S1: requiring stability cut false stops at 2 mm from 23% to 6%; an error shared by all views shifts every view's objective alike and can pass the check

(d) The loop studied, and three questions

Figure

Q1 Are the calibrated bounds reliable?

coverage, and how often registrations fail grossly

Q2 How does the certificate fail, and what reveals it?

wrong basins, run-time evidence, stop rules

Q3 Does choosing the views help?

shots to a certificate and to a true 2 mm error

Figure 1: When can a multi-fragment registration be trusted? (a) After every C-arm shot, the first anterior–posterior (AP), all fragments are registered to all acquired views (PENGWIN case 031, five moving fragments; outlines: the estimated fragments), and the system must stop or acquire another view. (b) A calibrated local bound reflects the curvature of the objective at the estimate, which can be as sharp in a wrong basin as at the truth (schematic): in the main study (S1), 14% of fragment registrations from the second view on failed by more than 5 mm, 32% of them with a bound of at most 2 mm. (c) All views agree at the true pose, but a wrong basin rarely survives a further view unchanged, so an estimate that moves beyond its bound is suspect: requiring stability lowered the false stops at 2 mm from 23% to 6% of first certificates (S1, all policies; preferred post hoc). A model error shared by all views shifts every view's objective alike and can pass this check (setting S5). (d) The loop studied and its three questions: Q1 is answered in Section 5.1, Q2 in Sections 5.2–5.4, Q3 in Section 5.6. Fig. 2 details the method. The curves in (b) and (c) and the diagram in (d) are schematic; panel (a) and the numbers come from the study's data.

an analysis of how the certificate fails: registrations into wrong basins, many with confident bounds, and their causes, separated by oracle ablations into optimisation failures and errors of the estimator; how well the signals available at run time detect them; and how stop rules built on these signals compare, with controls for stopping later;

  • sequential view planning on the calibrated posterior, with a worst-fragment (minimax) criterion as one instance, compared with A-, D- and E-optimal design, expected information gain with full and diagonal Fisher information, an image-gradient heuristic, the clinical protocol, maximal angular separation and random views;

an evaluation fixed before the test data were used, on 60 held-out fracture cases with per-view image uncertainty matched to real fluoroscopy, with a tracked and an untracked C-arm and at lower noise, and on real fluoroscopy of cadaver specimens; and, added later with its readings fixed in advance, a stress test with a segmentation error shared by all views.

The calibrated bounds covered 83% of unflagged fragment registrations and 91% of those within 5 mm of the truth, against a nominal 95%. From the second view on

14% of fragment registrations failed by more than 5 mm, 32% of them with a bound of at most 2 mm. Among such tightly bounded fragments, a fragment's shift when a view was added separated failures best (AUC 0.94, the bound itself 0.81), also when every error above 2 mm counted as a failure. Over all view policies, 23% of first certificates were false under the bound and flags alone and 6% with the stability check (planner: 11/38 and 1/29), at the cost of more trajectories without a certificate; delaying the stop did not reduce false certificates, requiring two consecutive certificates left 10% with fewer correct ones, and rules thresholding the other signals at the same acceptance left 10–15% false (exploratory). When the registration model's fragment boundaries were off by about 2 mm in every view, the stable rule's false share was 21% against 10% on the same 15 cases (exploratory; within sampling variation at these counts), in the direction expected of an error that need not move the estimate when a view is added. Planning certified sooner than the clinical protocol under the bound and flags alone, and on real images of intact pelves it certified sooner than random views on every specimen, without a false certificate. After correction for multiple comparisons, however, no criterion reached the pre-specified primary endpoint, a true worst-fragment error of 2 mm, significantly sooner than the pro-tocol, even when registrations were kept in the true basin. We report these negative results with the positive ones: gross failures, not wide posteriors, limited both accuracy and certification, and trusting a multi-fragment registration took evidence across views, not only a calibrated local bound; model errors that all views share may weaken both.