OUCH: Objective, Uniform, Controlled Haptic Stimulation for Automated Neurological Assessment

Sherry Choi, Sitao Sunny Li, Joseph Romero, Lillian Chin, Ashwin G. Ramayya, Allison M. Okamura, Ava Chen

Sherry Choi1,2,†^{1,2,\dagger}, Sitao Sunny Li1,†^{1,\dagger}, Joseph Romero3^{3}, Lillian Chin3^{3}, Ashwin G. Ramayya4^{4}, Allison M. Okamura1^{1}, and Ava Chen1^{1}

Abstract—Motor responses evoked by noxious stimulation are used to assess neurological function in hospitalized patients with disorders of consciousness. In standard bedside assessments, these stimuli are applied manually, resulting in substantial variation in force, duration, frequency, and location. Here, we present OUCH (Objective, Uniform, Controlled Haptic stimulation), a wrist-worn device that uses pneumatic actuation to deliver noxious mechanical stimuli. OUCH combines stacked pneumatic pouch actuators, a rigid indenter, and a compliant spacer designed to buckle at a set load, so that the indenter contacts the skin only during actuation. We characterize the device using blocked force tests, clinician evaluation, finite element analysis, and a deformable pressure testbed, and benchmark it against a commercially available handheld pressure algometer. At 200 kPa pouch inflation pressure, our indenter concentrated its load into an estimated contact pressure of 18,500 kPa, over 25 times that of the algometer tip, while requiring less total force. On the deformable testbed, it localized deformation near the surface, while deformation at depth remained comparable to the algometer. As a result, in preliminary perceptual testing, our indenter elicited discomfort at approximately 89% lower applied force than the algometer, demonstrating the feasibility of evoking discomfort with a compact, wearable soft actuator. These results show that combining pneumatic actuation with designed contact geometry enables controlled, repeatable noxious stimulation at the wrist, providing a foundation for objective neurological assessment.

I. INTRODUCTION

Motor response provides one of the few observable windows into neurological function in patients with disorders of consciousness, who are often unable to communicate or follow commands [1]. Current standard bedside assessments for this population, such as the Glasgow Coma Scale (GCS) and the Coma Recovery Scale-Revised (CRS-R), evaluate motor response by a clinician manually delivering a noxious localized pinch or pressure stimulus, then subjectively categorizing the evoked movement [2], [3]. More reflexive responses indicate lower neurological function, and more purposeful movements indicate higher function [4], [5]. However, the high variance in force, duration, frequency, and location of manually applied stimuli across examiners

Figure

Fig. 1: OUCH wearable device and operating principle. (a) Side view and (b) front view of OUCH on the wrist. (c) Schematic of the device at rest and during neurological assessment, where pneumatic pouch actuator expansion deforms the compliant spacer and drives the indenter into contact with the skin while the IMU tracks motion.

and sessions [3], [6], as well as the subjective interpretation of the evoked response, present a challenge for monitoring changes in neurological state [2], [7], [8].

Several approaches have been developed to deliver more controlled mechanical stimuli. Handheld pressure algometers, such as the AlgoMed (Medoc Ltd.), display the applied force and are commonly used to measure pain thresholds. However, these devices require an examiner to manually position the device and apply force against the body [9]. Many compact pneumatic wearables have been developed to apply localized pressures to the skin [10], although these are typically designed for innocuous touch and generate sensations well below pain thresholds. Polianskis et al. measured pressure-pain responses with a computer-controlled pneumatic tourniquet cuff, but their device produced a deep squeeze stimulus rather than the localized pain used to evoke movement in clinical neurological assessment [11]. Other human studies using controlled indentation and impact experiments have shown that small contacts can evoke pain at low forces [12], and that the onset of perceived pain occurs at lower contact pressures than blunt tissue injury [13], [14] or skin penetration [15], suggesting that noxious stimulation need not be injurious. This presents an opportunity to deliver

This work was supported in part by the Stanford Coulter Translational Research Program, the Stanford Wu Tsai Institute Neuroscience:Translate Program, and the Rice University Owl Edge Summer Experience Fund.

†^{\dagger} These authors contributed equally.

1^{1}Department of Mechanical Engineering, Stanford University, Stanford, CA, USA. {sunnyst1, aokamura, avachen}@stanford.edu

2^{2}Department of Mechanical Engineering, Rice University, Houston, TX, USA. sc233@rice.edu

3^{3}Department of Electrical and Computer Engineering, University of Texas at Austin, TX, USA.

4^{4}Department of Neurosurgery, Stanford University, Stanford, CA, USA.

a safe, localized stimulus with a wearable device in a controlled and uniform manner, eliciting a response to the noxious stimulus without causing tissue injury.

In this work, we present OUCH (Objective, Uniform, Controlled Haptic stimulation), a wrist-worn pneumatic device for controlled noxious stimulation (Fig. 1). OUCH combines two widely studied mechanisms in soft robotics—textile pouch actuators and tunable-buckling silicone structures—to adapt pneumatic wearables for noxious stimulation. First, we introduce the design of each component of the device with the goal of delivering localized pain to the body only when the device is inflated. Second, we experimentally quantify how inflation pressure relates to force, contact pressure, and deformation produced by our indenter design on rigid and deformable surfaces, and to perceived discomfort thresholds. Finally, we compare the mechanical response of our indenter geometry against that of the indenter geometry of a clinical handheld algometer. Together, these experiments characterize the force and contact pressure that a soft wearable can generate to evoke a pain response at the wrist, revealing that localizing the load is more effective than increasing force.