arXiv:2611.06791A Frame-Level View and Image-Quality Annotation Tool for Fetal Echocardiography: Comparison with Spreadsheet Annotation
Mohamed Azzam¹ | Ziyang Xu¹ | Quinlan Couch² | Zoe Keese² | Andrea Cadlo² | Christine Sedze-DeWolf² | Micaiah Uhrmacher² | Madelyn Graff² | Catherine Fox² | Lauren Vatter² | Paige Jeffryes² | Jason Christensen³,⁴,⁵ | Neil Hamill² | Shibiao Wan⁶ | Ling Li³,⁴,⁵ | Jieqiong Wang¹
¹Department of Neurological Sciences, University of Nebraska Medical Center, Omaha, NE 68198, USA | ²Department of Obstetrics and Gynecology, University of Nebraska Medical Center, Omaha, NE 68198, USA | ³Department of Pediatrics, University of Nebraska Medical Center, Omaha, NE 68198, USA | ⁴Criss Heart Center, Children's Nebraska, Omaha, NE 68114, USA | ⁵Child Health Research Institute, Omaha, NE 68198, USA | ⁶Department of Genetics, Cell Biology and Anatomy, University of Nebraska Medical Center, Omaha, NE 68198, USA
Correspondence: Jieqiong Wang (jiwang@unmc.edu)
Keywords: fetal echocardiography | congenital heart disease | image annotation | standard views | image quality | artificial intelligence
ABSTRACT
Objective: To develop the Cardiologist Annotation Tool (CAT) for frame-level cardiac view and image-quality annotation in fetal echocardiography and compare it with spreadsheet-based annotation.
Methods: CAT provides cine playback, frame navigation, and keyboard-based labeling without manual frame-range entry. Experts annotated 25 studies using both methods, and spreadsheet annotations were audited for predefined data-entry errors. In a randomized crossover study, six medical students annotated two normal studies, one with each method, against an expert reference.
Results: Among 1,746 spreadsheet-annotated files, 6.2% contained major or critical data-entry errors, more frequently in congenital heart disease than normal studies (10.9% vs 3.4%). CAT yielded more annotated view segments per clip (3.6 vs 1.5). Rates of labeled and correctly labeled frames per minute were 1.77 and 1.68 times those with spreadsheets, respectively (both P = 0.031), with gains in all six students. No detectable difference in view-label accuracy was observed.
Conclusion: CAT eliminates manual frame-range entry and enables finer-grained, structured annotation of fetal echocardiography, providing a practical approach to preparing datasets for AI research.
Key points
What does this study add?
-
Dedicated annotation software labels view and image quality per frame on DICOM cine, preventing spreadsheet data-entry errors by construction.
-
Major or critical errors affected 6.2% of expert spreadsheet files; a student crossover estimated efficiency and accuracy.
1 | Introduction
Congenital heart disease (CHD) affects approximately 9 per 1,000 live births [1, 2] and is a leading cause of infant morbidity and mortality [3, 4]. Prenatal detection supports delivery planning at centers with neonatal cardiac care [5], but detection rates vary widely. For critical CHD, reported rates range from 13% to 87% across international screening programs [6] and from 12% to 53% across US states among infants undergoing cardiac surgery [7]. Including outflow-tract and three-vessel views alongside the four-chamber view improves detection [3, 8]. Nevertheless, lesions requiring these additional views are less consistently detected, and performance depends on examiner training [7, 8].
Software has been developed to support prenatal diagnosis from ultrasound findings [9], while artificial intelligence (AI) offers potential support for CHD detection from ultrasound images. Models that identify standard planes in freehand fetal ultrasound [10] and detect CHD from cardiac views [11] rely on identifying relevant frames within ultrasound studies. Developing and evaluating these models requires expert-labeled cardiac views, ideally
1 of 12 accompanied by image-quality annotations to identify diagnostically useful frames [11, 12].
However, fetal echocardiography datasets with both frame-level view and quality annotations remain limited (Section 2). Producing these labels is labor-intensive and requires substantial expert time [13, 14]. An efficient workflow should allow fetal cardiologists to focus on image interpretation rather than manually recording frame ranges and labels. In the spreadsheet workflow evaluated here, experts review cine in a separate viewer and type annotations into spreadsheets. Without checks against file length and existing labels, overlapping, missing, or out-of-range entries may go unnoticed.
To address these challenges, we developed the Cardiologist Annotation Tool (CAT) in close collaboration with fetal cardiologists to make efficient use of their limited annotation time. CAT reads DICOM cine directly, preloads neighboring files, and supports keyboard-based labeling of cardiac views and image quality (Figure 1). Annotations are stored as change points, ensuring that each frame receives a single view and quality label while preventing structural data-entry errors by design.
This study presents three contributions:
-
A dedicated fetal echocardiography annotation tool that records cardiac views and image quality frame by frame directly on DICOM cine and common video and image formats, while preventing structural data-entry errors by design. To our knowledge, no published tool combines these features.
-
An expert audit of 25 fetal echocardiography studies (15 normal and 10 CHD) comparing CAT with spreadsheet annotation. Using a predefined error codebook and validated detectors, we quantified data-entry errors, examined their association with CHD, and assessed annotation agreement and granularity.
-
A randomized crossover study involving six medical students to compare annotation throughput and view accuracy against an expert reference. Frame-based throughput was higher with CAT for all six students, without a detectable difference in view accuracy.
2 | Related Work
2.1 | Annotation practices in fetal ultrasound
Published fetal cardiac datasets illustrate different approaches to expert labeling. Arnaout et al. assigned every frame to one of five screening views or a non-target category and used only diagnostically adequate views to train their view classifier [11]. For HeartAssist, one expert classified 65,324 images from 2,985 fetuses into ten cardiac views [15]. Taksøe-Vester et al. used LabelMe to annotate still images of eight cardiac planes and assign structure-level quality scores [16], while Nurmaini et al. outlined cardiac structures in frames extracted from ultrasound videos [17]. Other fetal ultrasound datasets used customized CVAT or in-house interfaces to annotate images [18–20]. These studies demonstrate variation in annotation granularity and how image quality is assessed.
2.2 | Existing annotation tools
General-purpose tools such as LabelMe, VIA, CVAT, and Label Studio support image labeling, video-frame annotation, or temporal segments [21–26]. Their flexibility allows project-specific label schemas, but fetal cardiac views and image-quality grades require configuration.
Ultrasound-specific tools provide more tailored capabilities. Annotation Web supports ultrasound-video annotation, including sequence-level image-quality fields, and has been adapted for regional quality scoring in adult echocardiography [27, 28]. SlicerUltrasound processes DICOM ultrasound cine and can attach classification labels to entire loops [29]. Most directly relevant, an Oxford fetal heart tool labels each video frame with one of three cardiac views and a visibility state. It works with screen-captured video rather than native DICOM cine and does not record a frame-level quality grade [30, 31]. Across these tools, support for native DICOM cine, frame-level fetal cardiac view and image-quality labeling, and prevention of structural data-entry errors remains distributed across different platforms. None of the reviewed tools reported an evaluation against another annotation method. To address these gaps, CAT was designed to combine these features in a portable desktop application for macOS and Windows, without requiring package installation or annotation configuration. We evaluated CAT against traditional spreadsheet annotation in complementary expert and student studies.
2.3 | Automated and weakly supervised annotation
Other approaches reduce the need for manual frame-level annotation. SonoNet derived standard-plane labels from freeze frames saved during routine scanning [10]. The iFIND and PULSE projects used freeze events, on-screen text, and machine settings to recover information from recorded scans [32–34]. In adult echocardiography, stored DICOM view tags have also provided labels [35]. Another study used AI to extract video clips of the standard cardiac views from full fetal scans for remote second review by experts [36]. These approaches can accelerate dataset preparation, but stored labels reflect what sonographers chose to save, and standard-plane detectors trained on normal hearts may miss frames that show structural defects [37]. Expert-reviewed frame-level labels therefore remain valuable for evaluating automated view and quality predictions.
3 | Methods
We describe the tool and then its two evaluations, an audit of expert annotations and a crossover study in medical students (Table 1).
3.1 | Development of the annotation tool
CAT was developed in close collaboration with fetal cardiologists to maximize the efficiency of their limited annotation time. The software was implemented in Python with a Tkinter-based graphical user interface and is distributed as a bundled executable for macOS and Windows that runs locally on the annotator's computer without installing Python or other packages. The initial version was developed in one month roughly, followed by two
2 of 12
(a) The annotation bottleneck

General-purpose annotation tools
Generic label sets: no fetal cardiac view and image-quality schema
Setup often needs a server, Docker or Python installation, or upload to a cloud service
(b) CAT: Cardiologist Annotation Tool

(c) Evaluation: traditional spreadsheet vs CAT
| Data-entry errors | Throughput |
| Agreement between methods | View accuracy |
FIGURE 1 | Study overview. (a) In traditional annotation, frame ranges are typed into a spreadsheet beside an image viewer, which allows structural data-entry errors; general-purpose annotation tools offer generic label sets and need setup. (b) The Cardiologist Annotation Tool (CAT) records the view and image quality of every frame directly on DICOM cine and common video and image formats. (c) CAT was compared with the traditional spreadsheet in an expert audit and a randomized crossover study of medical students. CHD, congenital heart disease; view abbreviations and quality levels as in Figure 2.
TABLE 1 | Design of the expert audit and the student crossover study
| Expert audit | Student crossover study | |
|---|---|---|
| Question | How often spreadsheet labels contain data-entry errors, and how spreadsheet and CAT labels agree | Whether CAT changes annotation throughput and view accuracy |
| Annotators | Four pediatric and fetal cardiologists (mean 17.5 years of experience): two in the traditional pass and three in the CAT pass; one took part in both | Six medical students, trained by an expert on one example scan |
| Cases | 25 studies (15 normal, 10 CHD); 1,746 files annotated with both methods | Two normal scans (27 and 29 valid files), with invalid decoy files to be left unannotated |
| Design | Original DICOM cine, annotated first on the experts' own spreadsheet and 3 to 5 months later with CAT | Randomized crossover on simplified video and image copies: three students used the spreadsheet template first and three used CAT first, with one day or less between sessions |
| Comparison standard | CAT serves as a comparator, and disagreements are unadjudicated | Expert reference, created with CAT and revised twice by experts |
| Annotation time | Not recorded | Recorded by each student in a template spreadsheet; checked against CAT's event log for CAT sessions |
| Outcomes | Data-entry errors; agreement between the methods on the view, image quality and view boundaries | Annotation time; reference segments, labeled frames and correct frames per minute; view accuracy; per-view recall; data-entry errors |
months of testing and iterative refinement based on feedback from fetal cardiologists. During this period, additional features were incorporated to address their annotation needs and improve usability. The finalized version was used consistently in both experimental studies.
3.2 | Software and annotation workflow
CAT opens a study folder, validates its files, and reads DICOM cine files directly, as well as MPEG-4 (MP4) video and image files (Figure 2). While one file is displayed, neighboring files are decoded in background threads (preloading). The annotator
3 of 12 plays or scrubs through the cine and labels frames with single keys; changes are saved automatically, and labels are exported as comma-separated values (CSV) files.
3.3 | Annotation model
Each frame receives one of eight view labels: the abdominal (situs) view (ABDO), apical four-chamber view (A4CH), left ventricular outflow tract (LVOT), right ventricular outflow tract (RVOT), three-vessel view (3VV), three-vessel and trachea view (3VT), non-target (NT) for a frame that shows none of these six target views, and “?” (uncertain) for frames to recheck. No frame was labeled “?” in either study. The target views correspond to the views of the fetal cardiac screening examination [3]; the aortic and ductal arch views are not among them, so such frames were labeled NT. Target-view frames also receive a quality score on a three-level ordinal scale, judged on the visibility of the cardiac structures that define the view: 2, clearly visible without occlusion or noise; 1, some noise or not fully clear, but acceptable; 0, not adequately visible. Quality has no meaning for NT frames and was ignored for them in every analysis.
The annotator records a label only where it changes, and the label holds until the next change point in the same file or to the end of the file, so the segments tile each file without gaps or overlaps. When a file is first displayed, CAT records NT with quality 2 for its first frame.
3.4 | Error prevention by design
Frame ranges typed into a spreadsheet beside a separate viewer can overlap under different views, leave frames uncovered, run beyond the end of a file, or be unreadable. CAT prevents these structural data-entry errors by construction (Table 2): each frame has one label, ranges derive from change points, frame numbers come from the display, labels from fixed keys, and file names from disk.
3.5 | Studies
Expert audit. Pediatric and fetal cardiologists annotated 25 fetal echocardiography studies, selected at random within two groups: 15 normal studies and 10 with CHD. In the traditional pass, two experts reviewed the original DICOM files in an image viewer and typed frame ranges into a spreadsheet of their own design, with one row per file and one column per view and quality level. After an interval of 3 to 5 months, three experts annotated the same files with CAT where one expert took part in both passes. The four experts had 17.5 years of experience on average. In this setting, annotation time was not measured and there was no reference standard to adjudicate disagreements between both methods. The analysis set comprised 1,746 files (247,477 frames) present in both sources.
Student crossover study. A total of six medical students each annotated two normal scans, one with each method, in a randomized two-period crossover: three students were randomly selected for Group A and the remaining three formed Group B. Group A (student1–student3) used the traditional spreadsheet first and Group B (student4–student6) used CAT first. The students were not told which group they were in or whether other students annotated the same scans, and each annotated separately from the others. Beforehand, the students were trained by an expert on one example scan similar to the two study scans. Each scan included invalid decoy files, which the students were instructed to leave unannotated (Table 1). Each student received renamed, shuffled video and image copies of the scans and recorded the start and end of each session in a template spreadsheet; the spreadsheet template for traditional annotation restricted view and quality to the permitted values. The students' labels were scored against the expert reference, which was created with CAT and revised twice by experts.
3.6 | Outcomes
Data-entry errors. Before any error was counted, we froze an error codebook of 21 codes in six families: formatting, unreadable or ambiguous entries, range validity, overlaps, coverage gaps and row identity. Each code has a certainty tier, provable (wrong by the spreadsheet's own logic or by the file length) or requiring the images, and a severity graded by its effect on a training dataset (critical: a wrong or contradictory view label, or labels on the wrong file; major: labels lost or quality off by two levels; Table 2). The main measure was the share of files with a provable finding of major or critical severity.
Agreement between methods. On frames with exactly one traditional view, we computed view agreement, Cohen's and Gwet's first-order agreement coefficient (AC1) [38, 39], and on frames with the same target view in both methods, linearly weighted for quality [40]. Boundary agreement was the share of view-change boundaries of one method that had a boundary of the other method within frames, for the range of tolerances k shown in Figure 4c. Each traditional frame was classified as agreeing with CAT, differing from CAT, unlabeled, or conflicting. Because annotators and method changed together, this represents a method-plus-rater agreement.
Student measures. Outcomes included self-reported annotation time, reference view segments per minute, labeled and correctly labeled frames per minute, view accuracy, per-view recall, and data-entry errors identified using an adapted codebook. Reference segments per minute was calculated using the fixed number of expert-defined segments, independent of how students divided the clips. Correctly labeled frames were those matching the expert reference. View accuracy was the proportion of reference frames correctly labeled, with unlabeled or conflicting frames counted as incorrect. Predefined flags identified unreliable time records, protocol deviations (including annotation of at least half the decoys), and outliers (robust ). Flagged observations were retained in the primary analysis.
3.7 | Statistical analysis
In the expert audit, per-study error rates were assigned Wilson 95% confidence intervals (CIs) [41]. Pooled error rates and agreement measures were assigned CIs using 2,000 study-level bootstrap resamples stratified by CHD status [42]. Error rates in normal and CHD studies were compared using a permutation test (100,000 permutations). A logistic generalized estimating equation (GEE) model [43] assessed associations between major
4 of 12

(b)

FIGURE 2 | The CAT software. (a) Main window during annotation of a fetal echocardiography study where the displayed frame is labeled A4CH, quality 2. Callouts: 1, session bar; 2, cine frame; 3, view keys (top) and quality keys (bottom), keyboard keys in parentheses; 4, file and frame number; 5, mark file invalid; 6, playback, Check Uncertain and Skip (leave the file for later); 7, frame slider and jump to a file or frame; 8, progress through the study; 9, open a study folder and export the annotations. (b) Annotation workflow. Dashed blocks denote background step. Each label is stored as a change point and holds until the next label in the file. Views: abdominal (ABDO), apical four-chamber (A4CH), left (LVOT) and right (RVOT) ventricular outflow tracts, three-vessel (3VV), three-vessel and trachea (3VT), non-target (NT), uncertain (?). Quality: 2 for clearly visible, 1 for acceptable with some noise, and 0 for not adequately visible.
TABLE 2 | Data-entry error classes in traditional spreadsheet annotation and how CAT prevents each by design
| Error class | Severity | Findings in traditional spreadsheets | How CAT prevents it by design | |
|---|---|---|---|---|
| Experts (files affected) | Students | |||
| Conflicting labels on the same frames | major-critical | 49 (35) | 3 | One label per frame: a label holds from its change point to the next |
| Overlap at a shared endpoint | minor | 39 (36) | 6 | Segment ranges are derived from change points, so adjacent segments cannot share a frame |
| Duplicated range, same label | minor | 0 (0) | 1 | Consecutive identical labels are merged |
| Range beyond the file end | major-critical | 15 (14) | 0 | Frame numbers are taken from the displayed frame; the last segment ends at the file's frame count |
| Invalid range (reversed, or starting before the first frame) | major | 0 (0) | 1 | Ranges are computed, not typed |
| Unlabeled frames (gaps) | major | 65 (60) | 5 | Segments tile the whole file |
| File row with no labels | major | 8 (8) | 0 | The first frame is labeled as soon as the file is displayed |
| Unreadable or ambiguous entry | major-critical | 7 (7) | 1 | Labels come from fixed buttons and keys |
| Formatting irregularity | minor | 60 (45) | 10 | Labels are saved in a fixed, structured format |
| Row identity (duplicate or missing file name, omitted file) | major-critical | 0 (0) | 0 | File names are read from disk |
| Any provable major or critical error | major-critical | 138 (109) | 10 | |
or critical data-entry errors and CHD status, log frame count, file type, and file position, accounting for clustering within studies.
In the student study, paired CAT-to-spreadsheet ratios were calculated for annotation time and throughput measures, while paired
5 of 12 differences were used for view accuracy. Ratios were analyzed on the log scale. We used exact two-sided paired permutation tests [44] and Hodges–Lehmann estimates with exact CIs (96.9% for six pairs) [45, 46]. P values were not adjusted for multiplicity. A separate randomized crossover analysis assessed the method effect while accounting for the combined influence of scan and annotation period. Sensitivity analyses included exclusion of the student with unreliable timing, and CAT active time derived from its event logs. All analyses were conducted in Python.
4 | Results
4.1 | Data-entry errors in the traditional spreadsheets
Of the 1,746 files analyzed in both sources, 109 (6.2%, 95% CI 4.4–8.4) carried at least one provable data-entry error of major or critical severity in the traditional spreadsheet (Figure 3a, Table 2). Such errors occurred in 23 of the 25 studies, with per-study rates of up to 25.0% (Figure 3b).
The most consequential class was conflicting labels: 49 entries in 35 files gave 2,602 frames two different labels, and in 15 of these entries an identical frame range had been entered under two views (Figure 3c). A further 15 ranges in 14 files extended beyond the end of their file, labeling 1,511 frames beyond the frame count read by CAT. Gaps left 5,930 frames unlabeled in 60 files, and eight file rows carried no label at all. CAT prevents these errors by design, so its zero error count is not an empirical estimate (Table 2).
Errors were more frequent in CHD studies, affecting 72 of 659 files (10.9%) compared with 37 of 1,087 normal files (3.4%). The mean per-study error rate was 7.0 percentage points higher in CHD studies (permutation ). In the GEE model, CHD status (odds ratio 1.90, 95% CI 1.13–3.18) and longer files (odds ratio 1.93 per unit of log frame count, 95% CI 1.31–2.82) were associated with higher error odds. Differences in case complexity and annotators may also contribute to the observed association with CHD.
4.2 | Agreement between traditional and CAT annotation
On frames with a single spreadsheet-assigned view, agreement between methods was 78.5% (95% CI 75.4–81.7), with and AC1 = 0.76. Agreement was higher in normal than CHD studies (86.3% vs 72.9%) (Figure 4a). Agreement was lowest for frames CAT labeled as RVOT, 3VV, or 3VT, which spreadsheets often assigned to NT or an adjacent view. Across all traditional frames, 74.9% agreed with CAT, 21.2% differed, and 3.9% were unlabeled or carried conflicting labels, making them unsuitable as single training labels (Figure 4b).
CAT recorded more view segments per clip than spreadsheets (3.6 vs 1.5). Within a tolerance of five frames, 52.9% of spreadsheet view-change boundaries matched a CAT boundary, compared with only 13.4% in the reverse direction (Figure 4c). For frames assigned the same target view, image-quality scores agreed on 60.4% (weighted , 95% CI 0.21–0.36).
4.3 | Student annotation efficiency
Median annotation time was 102.5 min with CAT versus 174.5 min with spreadsheets, with five of six students completing annotations faster using CAT (). Labeled frames per minute (ratio 1.77, 96.9% CI 1.33–2.43) and correctly labeled frames per minute (ratio 1.68, 1.25–2.36) were higher with CAT for all six students (both ). These measures included frames covered by CAT's default and block labels. Reference segments per minute were higher for five students (ratio 1.60, ) (Table 3 and Figure 5a).
The randomized crossover analysis yielded similar throughput ratios of 1.63, 1.77, and 1.71 for reference segments, labeled frames, and correctly labeled frames per minute, respectively (all ). Excluding the student with uncertain timing yielded corresponding ratios of 1.60, 1.88, and 1.87. Analyses using CAT active time derived from event logs also favored CAT, although comparable logs were unavailable for spreadsheets.
Student spreadsheets contained 27 data-entry findings, including 10 provable major or critical errors. CAT output contained no corresponding structural errors, as expected from its design (Table 2).
4.4 | Student view accuracy
Median view accuracy was 58.3% with spreadsheets and 59.0% with CAT, with no detectable difference between methods (paired difference +1.0 percentage point, 96.9% CI -5.8 to +15.1, ) (Table 3). Three of six students achieved higher accuracy with CAT (Figure 5a). With both methods, recall was lowest for outflow-tract and vessel views (Figure 5b,c).
5 | Discussion
5.1 | Principal findings
The expert audit found provable major or critical data-entry errors in 6.2% of spreadsheet-annotated files, more frequently in CHD studies. CAT prevents these structural errors by design. CAT annotations also contained more view segments per clip, although this does not establish more accurate boundaries. Agreement on image-quality grades was low, suggesting the need for clearer scoring guidelines and annotator calibration. In medical students, frame-based annotation throughput was higher with CAT, without a detectable difference in view accuracy.
5.2 | Sources of spreadsheet errors
Spreadsheet annotation requires manually transferring frame ranges and labels from a separate cine viewer. Without checks for file length, overlapping ranges, or missing frames, invalid and contradictory entries may go unnoticed. Even with restricted label choices, student spreadsheets contained 10 provable major or critical errors. CAT avoids manual frame-range entry by using fixed labeling keys and deriving ranges from annotation change points, preventing these structural errors (Table 2). However, it cannot prevent clinically incorrect but otherwise valid labels.
6 of 12

(b)
(c)
FIGURE 3 | Data-entry errors in the traditional spreadsheet annotations of the expert audit (1,746 files, 25 studies). CAT stores one label per change point and so prevents these errors by construction. (a) Files with at least one error of each class, with each class's severity range. (b) Files with a major or critical error, per study and per group (file-weighted). (c) Clip E, illustrating an example of view annotation in both methods: two traditional ranges give different views to the same frames; the overlapping range is drawn beneath (conflicting labels). CAT annotations instead prevent this error and indicates more fine-grained annotation. Intervals: 95% confidence intervals (CIs), Wilson per study, otherwise study-level cluster bootstrap. View and quality abbreviations as in Figure 2.
TABLE 3 | Annotation efficiency and view accuracy of the six medical students with traditional annotation and with CAT, against the expert reference
| Measure | Traditional, median (range) | CAT, median (range) | Effect (96.9% CI) | Students favoring CAT | P |
|---|---|---|---|---|---|
| Annotation time, min | 174.5 (87–216) | 102.5 (54–219) | 0.62 (0.39–1.17) | 5/6† | 0.063 |
| Reference segments/min | 2.33 (1.19–2.94) | 3.29 (1.85–7.52) | 1.60 (0.96–3.57) | 5/6 | 0.094 |
| Labeled frames/min | 34.9 (19.0–69.1) | 60.2 (25.3–116.1) | 1.77 (1.33–2.43) | 6/6 | 0.031 |
| Correct frames/min | 19.4 (11.9–44.2) | 36.2 (14.9–67.5) | 1.68 (1.25–2.36) | 6/6 | 0.031 |
| View accuracy, % | 58.3 (36.9–64.0) | 59.0 (42.1–61.5) | +1.0 (−5.8 to +15.1) | 3/6 | 0.719 |
Effect: Hodges–Lehmann estimate of the paired CAT vs traditional effect with 96.9% CI, a ratio for annotation time and the per-minute measures and a difference for view accuracy. P: exact two-sided sign-flip test. †Took less time with CAT.
5.3 | Errors in CHD studies
The higher error rate in CHD studies is concerning because abnormal cases are relatively uncommon in screening datasets and are important for training diagnostic models [47]. This association may partly reflect differences in file length, case complexity, or annotators. Our error detectors identify logical and structural
inconsistencies but cannot detect incorrect labels that remain internally consistent. Switching between a cine viewer and a separate spreadsheet may introduce additional recording errors, although this was not directly assessed. These findings highlight the importance of validating spreadsheet annotations before model training.
7 of 12

(b)
(c)
FIGURE 4 | Agreement between traditional and CAT annotation of the same files in the expert audit (1,746 files, 25 studies). (a) View confusion on frames with exactly one traditional view; each row (CAT view) sums to 100%. <1 denotes nonzero but under 1%. Overall agreement 78.5% (95% CI 75.4–81.7); Cohen's 0.64 (CI 0.60–0.68). (b) Label status of all 247,477 traditional frames, by CAT view (percentages of the row). (c) Share of view-change boundaries of one method with a boundary of the other within frames, for all studies and separately for normal and CHD studies. View abbreviations as in Figure 2.
5.4 | Throughput and accuracy in students
All three throughput measures favored CAT. Labeled and correctly labeled frames per minute were higher for all six students, while reference segments per minute was higher for five of six students but not significantly (). The frame-based measures include automatically assigned default labels, whereas reference segments per minute is less affected by labeling style. The crossover analysis yielded similar effect estimates, although these did not reach statistical significance. View accuracy did not differ detectably between methods. Recall was lowest for outflow-tract and vessel views with both methods (Figure 5b), suggesting that recognizing these views remains challenging regardless of the annotation method.
5.5 | Comparison with existing tools
Existing tools support ultrasound video annotation but differ in their scope and capabilities. The Oxford fetal heart annotation tool supports frame-level cardiac view and visibility labeling in
AVI videos [30]. Annotation Web supports image-quality assessment alongside other annotation tasks in ultrasound videos [27]. CAT combines direct DICOM cine review with frame-level cardiac view and image-quality labeling, using predefined labels and automatic frame-range generation to reduce manual data entry.
5.6 | Limitations
The expert audit lacked an independent reference standard, so disagreements between methods could not be adjudicated. Spreadsheet and CAT annotations were performed in a fixed order, 3–5 months apart, by partly different experts. Consequently, differences may reflect both the annotation method and the annotators. Expert annotation time was not recorded, so potential time savings for cardiologists remain unmeasured.
The student study included only six medical students who were aware of the annotation method and used simplified scans rather than DICOM files. Scan and period coincided within each crossover group, limiting their separation in the paired analysis. Annotation time was self-reported, with event-log verification
8 of 12

(b)
● (a, b) CAT ■ (a, b) traditional ○ (a) outlier □ (b) median, CAT □ (b) median, traditional ● (c) CAT ■ (c) traditional
(c)

FIGURE 5 | Student crossover: six medical students each annotated one normal scan with each method (Group A, student1–student3, traditional first; Group B, student4–student6, CAT first), scored against the expert reference. (a) Paired change per student: reference segments, labeled frames and correct frames per minute, as CAT ÷ traditional; view accuracy, as CAT – traditional. Below: Hodges–Lehmann (HL) estimate with exact 96.9% CI. Open rings mark outlying view accuracy values (robust |z| above 3.5). (b) Recall of reference frames by view; bars, medians. (c) Clip S: the expert reference and each student's labels, with per-student view accuracy on the clip (as in a). Shade gives image quality; NT has none. View and quality abbreviations as in Figure 2.
available only for CAT. The expert reference was prepared using CAT, sharing its annotation granularity. Finally, CAT was developed and evaluated by the same research group.
6 | Conclusion
CAT streamlines frame-level annotation of fetal echocardiography, allowing fetal cardiologists to focus on image interpretation rather than manual spreadsheet entry. Our evaluations showed that CAT enabled finer-grained view annotations, avoided structural data-entry errors, and improved annotation efficiency among medical students without a detectable difference in view-label accuracy. These findings support CAT as a practical tool for preparing fetal echocardiography datasets for AI research. Future work will focus on two extensions: integrating a machine-learning module that predicts standard cardiac views and adapts
to annotator corrections, and developing a study-level dashboard to track annotated views and identify remaining annotation gaps.
Funding
Research reported in this publication was supported by the National Institute of General Medical Sciences of the National Institutes of Health under Award Number P20GM152326, the Child Health Research Institute AI pilot, and the University of Nebraska Collaboration Initiative Grant from the Nebraska Research Initiative (NRI). This project is also supported by the U.S. National Science Foundation under Award Numbers 2500836 and 2614824, the National Cancer Institute of the National Institutes of Health under Award Number R03CA317707, and the National Institute on Alcohol Abuse and Alcoholism of the National Institutes of Health under Award Number R21AA032098. This research was supported by the State of Nebraska through
9 of 12 the Pediatric Cancer Research Group, part of the Child Health Research Institute. The content is solely the responsibility of the authors and does not necessarily represent the official views of the funding organizations.
Ethics Statement
This study was conducted under an Institutional Review Board (IRB)-approved protocol (#0342-24-EP) and in accordance with the principles of the Declaration of Helsinki. The approved protocol encompassed both the retrospective use of fetal ultrasound images and image annotation by clinical experts and medical students as research personnel. Retrospective fetal ultrasound images were de-identified and assigned unique study identification numbers. To ensure confidentiality, annotators were identified using pseudonyms (Expert 1–4 and Student 1–6), and patient identifiers were excluded from all annotation datasets and analytical outputs.
Consent
The Institutional Review Board waived the requirement for informed consent for the analysis of de-identified imaging data.
Conflicts of Interest
The authors declare no conflicts of interest.
Data Availability Statement
The fetal echocardiography studies contain protected health information and cannot be shared. CAT is available from the corresponding author for non-commercial research use on reasonable request.
References
-
Denise van der Linde, Elisabeth E. M. Konings, Maarten A. Slager, Maarten Witsenburg, Willem A. Helbing, Johanna J. M. Takkenberg, and Jolien W. Roos-Hesselink. Birth prevalence of congenital heart disease worldwide: A systematic review and meta-analysis. Journal of the American College of Cardiology, 58 (21):2241–2247, 2011. doi: 10.1016/j.jacc.2011.08.025.
-
Yingjuan Liu, Sen Chen, Liesl Zühlke, Graeme C. Black, Mun-kit Choy, Ningxiu Li, and Bernard D. Keavney. Global birth prevalence of congenital heart defects 1970–2017: updated systematic review and meta-analysis of 260 studies. International Journal of Epidemiology, 48(2):455–463, 2019. doi: 10.1093/ije/dzy009.
-
J. S. Carvalho, R. Axt-Fliedner, R. Chaoui, J. A. Copel, B. F. Cuneo, D. Goff, L. Gordin Kopylov, K. Hecher, W. Lee, A. J. Moon-Grady, H. A. Mousa, H. Munoz, D. Paladini, F. Prefumo, E. Quarello, J. Rychik, B. Tutschek, M. Wiechec, and S. Yagel. ISUOG practice guidelines (updated): fetal cardiac screening. Ultrasound in Obstetrics & Gynecology, 61(6):788–803, 2023. doi: 10.1002/uog.26224.
-
GBD 2017 Congenital Heart Disease Collaborators. Global, regional, and national burden of congenital heart disease, 1990–2017: a systematic analysis for the global burden of disease study 2017. The Lancet Child & Adolescent Health, 4(3):185–200, 2020. doi: 10.1016/S2352-4642(19)30402-X.
-
B. J. Holland, J. A. Myers, and C. R. Woods, Jr. Prenatal diagnosis of critical congenital heart disease reduces risk of death from cardiovascular compromise prior to planned neonatal cardiac surgery: a meta-analysis. Ultrasound in Obstetrics & Gynecology, 45(6):631–638, 2015. doi: 10.1002/uog.14882.
-
Marian K. Bakker, Jorieke E. H. Bergman, Sergey Krikov, Emmanuelle Amar, Guido Cocchi, Janet Cragan, Hermien
E. K. de Walle, Miriam Gatt, Boris Groisman, Shiliang Liu, Wendy N. Nembhard, Anna Pierini, Anke Rissmann, Shanthi Chidambarathanu, Jr Sipek, Antonin, Elena Szabova, Giovanna Tagliabue, David Tucker, Pierpaolo Mastroiacovo, and Lorenzo D. Botto. Prenatal diagnosis and prevalence of critical congenital heart defects: an international retrospective cohort study. BMJ Open, 9(7):e028139, 2019. doi: 10.1136/bmjopen-2018-028139.
-
Michael D. Quartermain, Sara K. Pasquali, Kevin D. Hill, David J. Goldberg, James C. Huhta, Jeffrey P. Jacobs, Marshall L. Jacobs, Sunghee Kim, and Ross M. Ungerleider. Variation in prenatal diagnosis of congenital heart disease in infants. Pediatrics, 136(2): e378–e385, 2015. doi: 10.1542/peds.2014-3783.
-
Mary T. Donofrio, Anita J. Moon-Grady, Lisa K. Hornberger, Joshua A. Copel, Mark S. Sklansky, Alfred Abuhamad, Bettina F. Cuneo, James C. Huhta, Richard A. Jonas, Anita Krishnan, Stephanie Lacey, Wesley Lee, Erik C. Michelfelder, Gwen R. Rempel, Norman H. Silverman, Thomas L. Spray, Janette F. Strasburger, Wayne Tworetzky, and Jack Rychik. Diagnosis and treatment of fetal cardiac disease: A scientific statement from the american heart association. Circulation, 129(21):2183–2242, 2014. doi: 10.1161/01.cir.0000437597.44550.5d.
-
Mahmoud Abdelnaby, Abdulmeneam Fawzi, Manal Swelem, Hisham Elgammal, and Hamdy Elgammal. SonoCAD: a software to solve the dilemma of the prenatal diagnosis of fetal syndromes. Prenatal Diagnosis, 37(8):793–801, 2017. doi: 10.1002/pd.5088.
-
Christian F. Baumgartner, Konstantinos Kamnitsas, Jacqueline Matthew, Tara P. Fletcher, Sandra Smith, Lisa M. Koch, Bernhard Kainz, and Daniel Rueckert. SonoNet: Real-time detection and localisation of fetal standard scan planes in freehand ultrasound. IEEE Transactions on Medical Imaging, 36(11):2204–2215, 2017. doi: 10.1109/TMI.2017.2712367.
-
Rima Arnaout, Lara Curran, Yili Zhao, Jami C. Levine, Erin Chinn, and Anita J. Moon-Grady. An ensemble of neural networks provides expert-level prenatal detection of complex congenital heart disease. Nature Medicine, 27(5):882–891, 2021. doi: 10.1038/s41591-021-01342-5.
-
Maria Chiara Fiorentino, Francesca Pia Villani, Mariachiara Di Cosmo, Emanuele Frontoni, and Sara Moccia. A review on deep-learning algorithms for fetal ultrasound-image analysis. Medical Image Analysis, 83:102629, 2023. doi: 10.1016/j.media.2022.102629.
-
Martin J. Willemink, Wojciech A. Koszek, Cailin Hardell, Jie Wu, Dominik Fleischmann, Hugh Harvey, Les R. Folio, Ronald M. Summers, Daniel L. Rubin, and Matthew P. Lungren. Preparing medical imaging data for machine learning. Radiology, 295(1): 4–15, 2020. doi: 10.1148/radiol.2020192224.
-
Junmin Zhang, Sushan Xiao, Ye Zhu, Zisang Zhang, Haiyan Cao, Mingxing Xie, and Li Zhang. Advances in the application of artificial intelligence in fetal echocardiography. Journal of the American Society of Echocardiography, 37(5):550–561, 2024. doi: 10.1016/j.echo.2023.12.013.
-
Rina Kim, Mi-Young Lee, Yoo Jin Lee, Hye-Sung Won, Jinki Park, Jihoon Lee, and Kwangyeon Choi. Artificial intelligence based automatic classification, annotation, and measurement of the fetal heart using HeartAssist. Scientific Reports, 15(1):13055, 2025. doi: 10.1038/s41598-025-97934-z.
-
Caroline A. Taksoee-Vester, Kamil Mikolaj, Zahra Bashir, Anders N. Christensen, Olav B. Petersen, Karin Sundberg, Aasa Feragen, Morten B. S. Svendsen, Mads Nielsen, and Martin G. Tolsgaard. AI supported fetal echocardiography with quality assessment. Scientific Reports, 14(1):5809, 2024. doi: 10.1038/s41598-024-56476-6.
-
Siti Nurmaini, Muhammad Naufal Rachmatullah, Ade Iriani Sapitri, Annisa Darmawahyuni, Bambang Tutuko, Firdaus Firdaus, Radiyati Umi Partan, and Nuswil Bernolian. Deep learning-based computer-aided fetal echocardiography: Application to heart standard view segmentation for congenital heart defects detection. Sensors, 21(23):8007, 2021. doi: 10.3390/s21238007.
10 of 12
-
Bharath Srinivas Prabakaran, Paul Hamelmann, Erik Ostrowski, and Muhammad Shafigue. FPUS23: An ultrasound fetus phantom dataset with deep neural network evaluations for fetus orientations, fetal planes, and anatomical features. IEEE Access, 11: 58308–58317, 2023. doi: 10.1109/ACCESS.2023.3284315.
-
Bin Pu, Jiewen Yang, Liwen Wang, Ying Tan, Guannan He, Xingbo Dong, Qika Lin, Jiaring Guo, Lixian Yang, Zuozhu Liu, Shengli Li, and Kenli Li. FUSEP: A multi-center benchmark for diverse tasks in fetal ultrasound screening early pregnancy. In Proceedings of the 32nd ACM SIGKDD Conference on Knowledge Discovery and Data Mining V.2, pages 9695–9706. ACM, 2026. doi: 10.1145/3770855.3817453.
-
Xavier P. Burgos-Artizzu, David Coronado-Gutiérrez, Brenda Valenzuela-Alcaraz, Elisenda Bonet-Carne, Elisenda Eixarch, Fatima Crispi, and Eduard Gratacos. Evaluation of deep convolutional neural networks for automatic classification of common maternal fetal ultrasound planes. Scientific Reports, 10(1):10200, 2020. doi: 10.1038/s41598-020-67076-5.
-
Bryan C. Russell, Antonio Torralba, Kevin P. Murphy, and William T. Freeman. LabelMe: A database and web-based tool for image annotation. International Journal of Computer Vision, 77 (1-3):157–173, 2008. doi: 10.1007/s11263-007-0090-8.
-
Kentaro Wada. Labelme: Image polygonal annotation with Python. GitHub repository, 2026. URL https://github.com/wkentaro/labelme. Version 7.9.0. Accessed October 24, 2026.
-
Abhishek Dutta and Andrew Zisserman. The VIA annotation software for images, audio and video. In Proceedings of the 27th ACM International Conference on Multimedia, MM '19, pages 2276–2279, Nice, France, 2019. ACM. doi: 10.1145/3343031.3350535.
-
CVAT.ai Corporation. CVAT documentation. Official documentation, 2026. URL https://docs.cvat.ai/docs/. Release v2.80.0; source code at https://github.com/cvat-ai/cvat. Accessed October 24, 2026.
-
HumanSignal, Inc. Label Studio documentation: Tags. Official documentation, 2026. URL https://labelstud.io/tags/. Accessed October 24, 2026.
-
Maxim Tkachenko, Mikhail Malyuk, Andrey Holmanyuk, and Nikolai Liubimov. Label Studio: Data labeling software. Open source software, 2020. URL https://github.com/HumanSignal/label-studio. Version 1.23.2. Accessed October 24, 2026.
-
Erik Smistad, Andreas Østvik, and Lasse Lovstakken. Annotation web – an open-source web-based annotation tool for ultrasound images. In 2021 IEEE International Ultrasound Symposium (IUS), pages 1–4, Xi'an, China, 2021. IEEE. doi: 10.1109/IUS52206.2021.9593336.
-
Gilles Van De Vyver, Svein-Erik Måsoy, Hävard Dalen, Bjørnar Leangen Grenne, Espen Holte, Sindre Hellum Olaisen, John Nyberg, Andreas Østvik, Lasse Lovstakken, and Erik Smistad. Regional image quality scoring for 2-D echocardiography using deep learning. Ultrasound in Medicine & Biology, 51(4): 638–649, 2025. doi: 10.1016/j.ultrasmedbio.2024.12.008.
-
SlicerUltrasound contributors. SlicerUltrasound: 3D Slicer extension for ultrasound data processing. GitHub repository, 2026. URL https://github.com/SlicerUltrasound/SlicerUltrasound. Commit e1eb38d. Accessed October 24, 2026.
-
Christopher P. Bridge. heart_annotation_tool: Two small GUIs for annotating fetal heart ultrasound videos. GitHub repository, https://github.com/CPBridge/heart_annotation_tool, 2021. GPL-3.0. Accessed October 24, 2026.
-
Christopher Bridge. Computer-aided analysis of fetal cardiac ultrasound videos. DPhil thesis, University of Oxford, 2017.
-
Lorenzo Venturini, Samuel Budd, Alfonso Farruggia, Robert Wright, Jacqueline Matthew, Thomas G. Day, Bernhard Kainz, Reza Razavi, and Jo V. Hajnal. Whole examination AI estimation of fetal biometrics from 20-week ultrasound scans. npj Digital Medicine, 8(1):22, 2025. doi: 10.1038/s41746-024-01406-z.
-
Harshita Sharma, Lior Drukker, Pierre Chatelain, Richard Droste, Aris T. Papageorghiou, and J. Alison Noble. Knowledge representation and learning of operator clinical workflow from full-length routine fetal ultrasound scan videos. Medical Image Analysis, 69: 101973, 2021. doi: 10.1016/j.media.2021.101973.
-
Lior Drukker, Harshita Sharma, Richard Droste, Mohammad Alsharif, Pierre Chatelain, J. Alison Noble, and Aris T. Papageorghiou. Transforming obstetric ultrasound into data science using eye tracking, voice recording, transducer motion and ultrasound video. Scientific Reports, 11(1):14109, 2021. doi: 10.1038/s41598-021-92829-1.
-
Alvaro E. Ulloa Cerna, Linyuan Jing, Christopher W. Good, David P. vanMaanen, Sushravya Raghunath, Jonathan D. Suever, Christopher D. Nevius, Gregory J. Wehner, Dustin N. Hartzel, Joseph B. Leader, Amro Alsaid, Alpen A. Patel, H. Lester Kirchner, John M. Pfeifer, Brendan J. Carry, Marios S. Pattichis, Christopher M. Haggerty, and Brandon K. Fornwalt. Deep-learning-assisted analysis of echocardiographic videos improves predictions of all-cause mortality. Nature Biomedical Engineering, 5(6):546–554, 2021. doi: 10.1038/s41551-020-00667-9.
-
Thomas G. Day, Lorenzo Venturini, Samuel F. Budd, Alfonso Farruggia, Robert Wright, Jackie Matthew, Vita Zidere, Trisha Vigneswaran, Ilaria Bo, Alex Savis, Jo Wolfenden, John Simpson, Jo Hajnal, Bernhard Kainz, and Reza Razavi. Video clip extraction from fetal ultrasound scans using artificial intelligence to allow remote second expert review for congenital heart disease. Prenatal Diagnosis, 45(4):531–538, 2025. doi: 10.1002/pd.6757.
-
Jeremy Tan, Anselm Au, Qingjie Meng, Sandy FinesilverSmith, John Simpson, Daniel Rueckert, Reza Razavi, Thomas Day, David Lloyd, and Bernhard Kainz. Automated detection of congenital heart disease in fetal ultrasound screening. In Medical Ultrasound, and Preterm, Prenatal and Paediatric Image Analysis (ASMUS 2020, PIPPI 2020), volume 12437 of Lecture Notes in Computer Science, pages 243–252, Cham, 2020. Springer. doi: 10.1007/978-3-030-60334-2_24.
-
Jacob Cohen. A coefficient of agreement for nominal scales. Educational and Psychological Measurement, 20(1):37–46, 1960. doi: 10.1177/001316446002000104.
-
Kilem Li Gwet. Computing inter-rater reliability and its variance in the presence of high agreement. British Journal of Mathematical and Statistical Psychology, 61(1):29–48, 2008. doi: 10.1348/000711006X126600.
-
Jacob Cohen. Weighted kappa: Nominal scale agreement with provision for scaled disagreement or partial credit. Psychological Bulletin, 70(4):213–220, 1968. doi: 10.1037/h0026256.
-
Edwin B. Wilson. Probable inference, the law of succession, and statistical inference. Journal of the American Statistical Association, 22(158):209–212, 1927. doi: 10.1080/01621459.1927.10502953.
-
Bradley Efron and Robert J. Tibshirani. An Introduction to the Bootstrap. Number 57 in Monographs on Statistics and Applied Probability. Chapman & Hall, New York, 1993. ISBN 9780412042317. doi: 10.1007/978-1-4899-4541-9.
-
Kung-Yee Liang and Scott L. Zeger. Longitudinal data analysis using generalized linear models. Biometrika, 73(1):13–22, 1986. doi: 10.1093/biomet/73.1.13.
-
Michael D. Ernst. Permutation methods: A basis for exact inference. Statistical Science, 19(4):676–685, 2004. doi: 10.1214/088342304000000396.
-
J. L. Hodges and E. L. Lehmann. Estimates of location based on rank tests. The Annals of Mathematical Statistics, 34(2):598–611, 1963. doi: 10.1214/aoms/1177704172.
-
Myles Hollander, Douglas A. Wolfe, and Eric Chicken. Non-parametric Statistical Methods. Wiley Series in Probability and Statistics. John Wiley & Sons, Hoboken, NJ, 3rd edition, 2013. ISBN 9780470387375. doi: 10.1002/9781119196037.
-
Daniela Vega, Hannah V. Ceballos, Javier S. Vera, Santiago Rodríguez, Alejandra Perez, Angela Castillo, Maria Escobar, Dario
11 of 12
Londoño, Luis A. Sarmiento, Camila I. Castro, Nadiezhda Rodriguez, Juan C. Briceño, and Pablo Arbeláez. CARDIUM: Congenital anomaly recognition with diagnostic images and unified medical records. In Proceedings of the IEEE/CVF International Conference on Computer Vision (ICCV) Workshops, pages 1204–1213, 2025.