iMED: A Multi-Endoscope Dataset for Surgical 3D Perception

Sierra Bonilla, Fengyi Jiang, Chinedu Nwoye, Jingpei Lu, Kailey Reardon, Humphrey W. Chow, Francisco Vasconcelos, Sophia Bano, Adam Schmidt, Omid Mohareri

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Sierra Bonilla¹, Fengyi Jiang², Chinedu Nwoye², Jingpei Lu², Kailey Reardon², Humphrey W. Chow², Francisco Vasconcelos¹, Sophia Bano¹, Adam Schmidt², and Omid Mohareri²

¹ University College London, London WC1E 6BT, United Kingdom ² Advanced Research, Intuitive, Sunnyvale, CA 94086, USA sierra.bonilla.21@ucl.ac.uk

Abstract. We introduce iMED, the first synchronized multi-endoscope dataset for robot-assisted minimally invasive surgery (RAMIS), designed to address a fundamental limitation in surgical vision benchmarking: the absence of independent held-out viewpoints. Existing surgical datasets capture single-trajectory sequences from a single endoscope, making it impossible to distinguish geometric generalization from photometric interpolation along a narrow forward-facing path. iMED provides 340 sequences (≈170K synchronized timepoints with 4 views per timepoint) recorded simultaneously from two independent stereo endoscopes across ex vivo, postmortem, and live surgical settings, spanning 14 specimens with diverse anatomical regions and motion regimes. The dataset includes calibrated camera intrinsics, ArUco-based frame-wise pose estimates with uncertainty quantification, instrument segmentation masks, and rich clinical metadata. Using a train-on-one-endoscope, test-on-another evaluation protocol, we benchmark 23 state-of-the-art methods across rigid and deformable novel view synthesis, pose estimation, feature matching, and monocular depth estimation. Our experiments show that methods relying primarily on photometric supervision degrade substantially under this held-out-endoscope setting, while methods with explicit geometric regularization are more robust across views in our benchmark. Notably, large-scale foundation models transfer surprisingly well to surgical imagery, often outperforming domain-specific models, suggesting pitfalls in current fine-tuning protocols. iMED establishes a new evaluation protocol for geometric generalization in surgical vision. Dataloader and dataset links can be found at github.com/surgical-vision/imed.

Keywords: Surgical vision · Robot-assisted minimally invasive surgery · 3D from multi-view

1 Introduction

Non-rigid scenes with constrained camera motion are among the most challenging regimes for image-based 3D perception tasks (i.e. pose estimation, feature matching, optical flow, depth estimation, and novel view synthesis (NVS)).

S. Bonilla et al.

Figure

Fig. 1: Data platform with two stereo endoscopes (top-left), common camera trajectories (top-right), and pipeline outputs including session/anatomy/instrument/motion labels, camera intrinsics, ArUco pose estimates, inferred tool masks, stereo depth (code provided), and derived 3D geometry. See Tab. 2 for provenance.

Although challenging, such conditions often arise in real-world robotic applications like digital human modeling of clothed bodies [7], submarine exploration [10], and robot-assisted minimally invasive surgery (RAMIS) [54]. Among these, RAMIS represents an extreme regime. Cameras operate inside the body under severe viewpoint constraints, typically following narrow, forward-facing trajectories through deformable tissue with strong specularities and transient phenomena (smoke, bleeding, topology changes). These conditions compress multiple failure modes for 3D perception methods, making RAMIS an informative stress test for visual geometry in real-world robotic systems.

Computer vision benchmarks have historically advanced 3D perception through standardized evaluation protocols [20]. In rigid environments, datasets capturing scenes from many surrounding viewpoints with known camera poses enabled progress in multi-view reconstruction and mapping [1,12,31,75]. Dynamic-scene benchmarks later extended these ideas to time-varying NVS and reconstruction [28,45,46,51], typically relying on wide-baseline or 360°/180° capture where cameras observe scenes from diverse directions. Surgical vision datasets have enabled progress in reconstruction, camera tracking, and tool analysis [3,22,23,54]. Unlike the surrounding-view capture in general benchmarks, endoscopic cameras follow narrow forward-facing paths through tissue, producing sequences with significant image overlap and limited viewpoint diversity.

In NVS and 3D scene representation, models are typically evaluated using images at held-out camera poses and reporting metrics such as PSNR and SSIM against held-out views [5,24,30,56,73]. These metrics quantify photometric fidelity and have become standard quantitative benchmarks. However, they are proxies for reconstruction quality rather than direct measures of geometric consistency: high PSNR/SSIM can reflect local interpolation between nearby view-

Table 1: Comparison of stereo-endoscopic datasets for 3D perception in minimally invasive surgery. Multi-Camera* defined as more than one stereo camera pair. **only for 2 phantom validation sequences with CT scans.

DatasetDataset AttributesScene TypesInferred LabelsProvided Reference Signals
Seq. SpecimensModalitySettingLiveStatic Dyn.Rigid Deform. Obj.3D Model Tool MasksCamera Calib.Multi-Camera*
EndoVis 17/18[2,4]2929Da Vinci XiDa Vinci XIn vivo✓✓✓
MITI [22]-1Karl Storz3D TipcamIn vivo✓✓✓
STIR [54]5724Da Vinci XiEx vivoIn vivo✓✓✓
Hamlyn Centre[42,50,59]--Da Vinci SiPhantomEx vivoIn vivo✓✓✓**
SERV-CT [16]162Da VinciPostmortem✓✓✓
SCARED [3]99Da Vinci XiPostmortem✓✓✓
StereoMIS [23]166Da Vinci XiIn vivo✓✓✓✓
iMED (Ours)34014Da Vinci 5Ex vivoIn vivoPostmortem✓✓✓✓✓✓

points, without uniquely determining whether the model has learned the underlying 3D structure [68]. This limitation is amplified when train and test poses lie along the same forward-facing trajectory, where test views are typically sampled from nearby frames in time so differ only slightly in view. A complementary evaluation is to test generalization under larger viewpoint shifts. In classic multi-camera dynamic-scene benchmarks, this protocol is well established: models are trained on a subset of cameras and evaluated on held-out cameras [8,28,34].

To address this limitation in surgical vision, we introduce iMED, the first multi-endoscope dataset for RAMIS, providing synchronized recordings from two physically independent stereo endoscopes, four camera streams with overlapping fields of view, enabling cross-trajectory geometric validation (see Fig. 1). The dataset comprises 340 sequences totaling ~170K synchronized timepoints (4 frames×60 FPS) across diverse surgical scenarios: 10 ex vivo organ specimens (chicken, porcine, and bovine; ≈7 sequences each), 3 human cadaver subjects (≈6 anatomical sites per subject, ≈7 sequences per site), and 1 live porcine subject (10 anatomical sites, ≈7 sequences per site). Data collection was performed in a USDA-licensed clinical lab using one porcine model and three human cadaver models sourced from a willed body program. Each 10-15 second sequence captures varied conditions including static scenes, respiratory motion, tool-tissue interactions with 20 common robotic instruments (see both Tab. 2 and Suppl.Tab.7, contact-based tissue deformations, lighting variations, and specularities. All sequences include comprehensive metadata: anatomical regions labeled by clinical specialists with surgical robotics expertise, robot tool types, powered instrument usage (e.g., electrocautery), and motion categories (e.g., camera sweep, robot tool-tissue interaction, calibration). All frames include inferred filtered instrument segmentation masks generated by [9]. Refer to Tab. 2 for label provenance. To filter sequences towards specific applications (e.g. deformable vs. rigid), refer to Suppl.Tab.10. Recordings include intrinsic calibrations and frame-wise ArUco-based [18] camera poses refined through reprojection-error optimization with uncertainty estimates.

S. Bonilla et al.

Figure

Fig. 2: Example geometric reconstructions and label projection enabled by iMED. Left: synchronized stereo views from both endoscopes, stereo depth estimationo [67], aligned point clouds from provided camera calibrations, Poisson-reconstruction, and reprojected tool masks. Right: Representative anatomical regions with surface reconstructions from ex vivo specimens (chicken, porcine, bovine), human cadaver sites, and live porcine tissue. Inference code to recreate geometry and 3D tool labels is provided.

Using iMED, we evaluate 23 SOTA algorithms across tasks central to 3D perception: rigid NVS, deformable (online & not) NVS, camera pose estimation, feature matching, and depth estimation. Our analysis suggests that, under this held-out-endoscope protocol, methods relying primarily on photometric supervision degrade under physically independent viewpoints, while methods with explicit geometric regularization are more robust across views. Depth supervision has representation-dependent effects, and large multi-domain foundation models often transfer well to surgical imagery compared with surgical-only training or aggressive fine-tuning. We define train-on-one-endoscope, test-on-another evaluation protocols that eliminate trajectory overlap and measure geometric generalization across physically independent cameras, a standard in multi-view human capture datasets that is absent from surgical benchmarks. Motivated by the lack of independent test views in surgical datasets, we introduce iMED and make the following contributions:

  1. A train-on-one-endoscope, test-on-another evaluation protocol providing held-out viewpoints with large baseline and viewing-direction differences, as a better proxy for geometric generalization beyond photometric interpolation
  2. A synchronized multi-endoscope dataset for RAMIS with 340 sequences (≈170K timepoints with 4 images per timepoint) across ex vivo, in vivo postmortem, and in vivo settings (See Fig. 3)
  3. Calibrated multi-view images and structured annotations (cam intrinsics, reference trajectories, anatomical labels, motion categories, instrument masks)
  4. A benchmark of 23 methods across NVS, pose estimation, feature matching, and monocular depth estimation that exposes previously unobserved behaviors: models trained using only image supervision perform poorly on new viewpoints, methods with explicit 3D regularization generalize more reliably, and foundation models often transfer better than surgical-only training

Figure

Fig. 3: A: Distribution of seqs. by anatomical region and organ type including number of timepoints per category. B: Distribution of seqs. by motion regime (camera motion vs. scene motion). Calibration-only sequences are excluded. C: Distribution of seqs. by session type. D: Distribution of seqs. by cam spatial relationships, showing inter-cam separation distance and orientation difference.