OpenULF: A Modular Toolkit Enables Data-Driven Ultra-Low-Field MRI with both Classical and Quantum Sensors

Hanhan Wu, Yuchen Yuan, Weiwen Jiang

Hanhan Wu, Yuchen Yuan, Weiwen Jiang

George Mason University, Fairfax, VA, USA

hwu28@gmu.edu, yyuan21@gmu.edu, wjiang8@gmu.edu

Abstract—Ultra-low-field (ULF) Magnetic Resonance Imaging (MRI) offers a portable, low-cost alternative to conventional scanners by operating at field strengths below 10 mT. At these fields, two sensor technologies compete: quantum SQUID (Superconducting Quantum Interference Device) sensors, which provide high sensitivity but require cryogenic cooling, and classical induction coils, which are inexpensive and room-temperature but produce weaker signals. With conventional physics-based imaging, both suffer from low signal-to-noise ratios (SNR). Data-driven reconstruction has overcome similar limits in related inverse problems, but its use for ULF-MRI has been hindered by a lack of open datasets, benchmarks, and flexible simulators. To close this gap, we present OpenULF, a modular, open-source simulation toolkit that generates paired training data and supports end-to-end machine-learning (ML) reconstruction. OpenULF integrates a Bloch-equation-validated physics simulator, calibrated SQUID and coil receiver models, reconstruction ML networks, and an automated design-space exploration engine. Architecturally, a receiver abstract base class, a typed HDF5 contract that decouples physics from learning, and a resumable sweep engine make sensors, models, and datasets independently substitutable. Using OpenULF, we train 774 models across four anatomical datasets and conduct three design-space explorations. The effectiveness of ML reconstruction depends strongly on the dataset and the sensor. We identify a SQUID-to-coil crossover near 450 µT and demonstrate that sensor hardware and ML model capacity can be traded off during system design. Sensor and reconstruction choices must therefore be co-designed: the best sensor is set by the dataset and field strength, not by SNR.

Index Terms—simulation toolchain, modular software architecture, design-space exploration, ultra-low-field MRI, co-design.

I. INTRODUCTION

Ultra-low-field (ULF) MRI operates at field strengths below 10 mT and is a portable, low-cost alternative to conventional 1.5–3 T scanners [1]–[3]. ULF systems can be built from lightweight permanent magnets rather than multi-ton superconducting coils, enabling deployment where conventional MRI is impractical [3], [4]. At these field strengths, two fundamentally different sensor technologies are available. Quantum sensors based on Superconducting Quantum Interference Devices (SQUIDs) provide high sensitivity independent of frequency but require cryogenic cooling with liquid helium. Classical induction coils are inexpensive and operate at room temperature, but their signal scales with the Larmor frequency ω0=γB0\omega_0 = \gamma B_0 and is therefore very weak at ULF. With conventional physics-based reconstruction, both sensing modalities suffer from low signal-to-noise ratios (SNR), resulting in limited image quality. In our simulation, even the SQUID achieves only about 11 dB in Peak Signal-to-Noise Ratio

TABLE I: Feature comparison of MRI simulation tools (✓ = supported, − = not). “Automated param. sweep” means a built-in resumable, deduplicated sweep engine, not the general scripting that MRILab and JEMRIS also provide; “Multi-model library” means bundled reconstruction models.

FeatureMRILabJEMRISfastMRIOpenULF
Bloch spin dynamics✓✓−✓
Pulse sequence sim.✓✓−✓
ULF sensor models−−−✓
SQUID + Coil in one−−−✓
ML reconstruction−−✓✓
Multi-model library−−✓✓
Automated param. sweep−−−✓
Modular composition−−−✓

(PSNR) and 0.16 in Structural Similarity Index (SSIM) on brain data, well below diagnostic quality.

Recent progress in data-driven reconstruction for inverse problems, including ultrasound computed tomography (USCT) [5], [6] and full-waveform inversion (FWI) [7], [8], suggests that machine learning (ML) can overcome the severe SNR limitations of ULF-MRI by learning to recover anatomical structure from noisy measurements. However, systematic exploration of data-driven ULF-MRI has been hindered by three gaps: the lack of open datasets, the absence of standardized benchmarks, and the shortage of flexible simulation frameworks. Existing MRI simulators such as MRILab [9] and JEMRIS [10] target clinical field strengths. Their receiver models do not support SQUID sensors or prepolarization, and their noise models assume sample-noise-dominated regimes that do not apply at ULF. ML reconstruction frameworks such as fastMRI [11] assume fixed, pre-acquired clinical-quality data. Table I summarizes the feature coverage. No existing tool integrates ULF-specific sensor physics, ML reconstruction, and automated parameter sweeps in a single pipeline.

To address these gaps, we present OpenULF, 1^{1} a modular, open-source simulation toolkit for data-driven ULF-MRI with both classical and quantum sensors (Fig. 1). OpenULF integrates a Bloch-equation-validated physics simulator ( ≤\leq0.01% error across six analytical tests), calibrated SQUID and coil receiver models with ULF-specific noise, five reconstruction networks spanning 0.2M to 10M parameters, and an automated design-space exploration engine. This work makes three contributions:

1^{1}Code available at https://github.com/Wu-Hanhan/OpenULF.

© 2026 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works. Accepted for publication in the 44th IEEE International Conference on Computer Design (ICCD), 2026.

  • A modular co-simulation architecture in which a receiver abstraction and a typed on-disk contract decouple calibrated ULF sensor physics (both SQUID and coil receivers, prepolarization, and sensor-noise-dominated regimes) from end-to-end ML reconstruction, so researchers can substitute their own architectures or anatomical datasets without modifying the physics simulator. To the best of our knowledge, no existing tool provides this combination.

  • An automated sweep engine that explores the B0×NEX×channels×model×sensor design space at scaleB_0 \times \text{NEX} \times \text{channels} \times \text{model} \times \text{sensor design space at scale}, training 774 models across four anatomical datasets (BrainWeb, OASIS, IXI, and a synthetic abdomen set).

  • Three quantitative design rules, derived from systematic experiments, that reveal interactions between sensor configuration and ML reconstruction not available from component-level metrics such as SNR.

Our experiments produce three main observations. First, the effect of ML reconstruction on the SQUID-coil quality gap depends on both the dataset and the sensor: ML compresses the gap on some anatomies and amplifies it on others. Second, a SQUID-to-coil crossover near 450 μ\muT separates a SQUID-favored low-field regime from a coil-favored high-field one. Third, sensor channel count and ML model capacity are interchangeable design variables. None of these follow from isolated SNR analysis; they require running sensor, acquisition, and ML choices together end-to-end.

The remainder of this paper provides background and motivation (Sections II and III), the OpenULF platform and its validation (Section IV), and the three design-space explorations with their resulting design rules (Section V).