arXiv:2611.01511OpenULF: A Modular Toolkit Enables Data-Driven Ultra-Low-Field MRI with both Classical and Quantum Sensors
Hanhan Wu, Yuchen Yuan, Weiwen Jiang
George Mason University, Fairfax, VA, USA
hwu28@gmu.edu, yyuan21@gmu.edu, wjiang8@gmu.edu
Abstract—Ultra-low-field (ULF) Magnetic Resonance Imaging (MRI) offers a portable, low-cost alternative to conventional scanners by operating at field strengths below 10 mT. At these fields, two sensor technologies compete: quantum SQUID (Superconducting Quantum Interference Device) sensors, which provide high sensitivity but require cryogenic cooling, and classical induction coils, which are inexpensive and room-temperature but produce weaker signals. With conventional physics-based imaging, both suffer from low signal-to-noise ratios (SNR). Data-driven reconstruction has overcome similar limits in related inverse problems, but its use for ULF-MRI has been hindered by a lack of open datasets, benchmarks, and flexible simulators. To close this gap, we present OpenULF, a modular, open-source simulation toolkit that generates paired training data and supports end-to-end machine-learning (ML) reconstruction. OpenULF integrates a Bloch-equation-validated physics simulator, calibrated SQUID and coil receiver models, reconstruction ML networks, and an automated design-space exploration engine. Architecturally, a receiver abstract base class, a typed HDF5 contract that decouples physics from learning, and a resumable sweep engine make sensors, models, and datasets independently substitutable. Using OpenULF, we train 774 models across four anatomical datasets and conduct three design-space explorations. The effectiveness of ML reconstruction depends strongly on the dataset and the sensor. We identify a SQUID-to-coil crossover near 450 µT and demonstrate that sensor hardware and ML model capacity can be traded off during system design. Sensor and reconstruction choices must therefore be co-designed: the best sensor is set by the dataset and field strength, not by SNR.
Index Terms—simulation toolchain, modular software architecture, design-space exploration, ultra-low-field MRI, co-design.
I. INTRODUCTION
Ultra-low-field (ULF) MRI operates at field strengths below 10 mT and is a portable, low-cost alternative to conventional 1.5–3 T scanners [1]–[3]. ULF systems can be built from lightweight permanent magnets rather than multi-ton superconducting coils, enabling deployment where conventional MRI is impractical [3], [4]. At these field strengths, two fundamentally different sensor technologies are available. Quantum sensors based on Superconducting Quantum Interference Devices (SQUIDs) provide high sensitivity independent of frequency but require cryogenic cooling with liquid helium. Classical induction coils are inexpensive and operate at room temperature, but their signal scales with the Larmor frequency and is therefore very weak at ULF. With conventional physics-based reconstruction, both sensing modalities suffer from low signal-to-noise ratios (SNR), resulting in limited image quality. In our simulation, even the SQUID achieves only about 11 dB in Peak Signal-to-Noise Ratio
TABLE I: Feature comparison of MRI simulation tools (✓ = supported, − = not). “Automated param. sweep” means a built-in resumable, deduplicated sweep engine, not the general scripting that MRILab and JEMRIS also provide; “Multi-model library” means bundled reconstruction models.
| Feature | MRILab | JEMRIS | fastMRI | OpenULF |
|---|---|---|---|---|
| Bloch spin dynamics | ✓ | ✓ | − | ✓ |
| Pulse sequence sim. | ✓ | ✓ | − | ✓ |
| ULF sensor models | − | − | − | ✓ |
| SQUID + Coil in one | − | − | − | ✓ |
| ML reconstruction | − | − | ✓ | ✓ |
| Multi-model library | − | − | ✓ | ✓ |
| Automated param. sweep | − | − | − | ✓ |
| Modular composition | − | − | − | ✓ |
(PSNR) and 0.16 in Structural Similarity Index (SSIM) on brain data, well below diagnostic quality.
Recent progress in data-driven reconstruction for inverse problems, including ultrasound computed tomography (USCT) [5], [6] and full-waveform inversion (FWI) [7], [8], suggests that machine learning (ML) can overcome the severe SNR limitations of ULF-MRI by learning to recover anatomical structure from noisy measurements. However, systematic exploration of data-driven ULF-MRI has been hindered by three gaps: the lack of open datasets, the absence of standardized benchmarks, and the shortage of flexible simulation frameworks. Existing MRI simulators such as MRILab [9] and JEMRIS [10] target clinical field strengths. Their receiver models do not support SQUID sensors or prepolarization, and their noise models assume sample-noise-dominated regimes that do not apply at ULF. ML reconstruction frameworks such as fastMRI [11] assume fixed, pre-acquired clinical-quality data. Table I summarizes the feature coverage. No existing tool integrates ULF-specific sensor physics, ML reconstruction, and automated parameter sweeps in a single pipeline.
To address these gaps, we present OpenULF, a modular, open-source simulation toolkit for data-driven ULF-MRI with both classical and quantum sensors (Fig. 1). OpenULF integrates a Bloch-equation-validated physics simulator ( 0.01% error across six analytical tests), calibrated SQUID and coil receiver models with ULF-specific noise, five reconstruction networks spanning 0.2M to 10M parameters, and an automated design-space exploration engine. This work makes three contributions:
Code available at https://github.com/Wu-Hanhan/OpenULF.
© 2026 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works. Accepted for publication in the 44th IEEE International Conference on Computer Design (ICCD), 2026.
-
A modular co-simulation architecture in which a receiver abstraction and a typed on-disk contract decouple calibrated ULF sensor physics (both SQUID and coil receivers, prepolarization, and sensor-noise-dominated regimes) from end-to-end ML reconstruction, so researchers can substitute their own architectures or anatomical datasets without modifying the physics simulator. To the best of our knowledge, no existing tool provides this combination.
-
An automated sweep engine that explores the , training 774 models across four anatomical datasets (BrainWeb, OASIS, IXI, and a synthetic abdomen set).
-
Three quantitative design rules, derived from systematic experiments, that reveal interactions between sensor configuration and ML reconstruction not available from component-level metrics such as SNR.
Our experiments produce three main observations. First, the effect of ML reconstruction on the SQUID-coil quality gap depends on both the dataset and the sensor: ML compresses the gap on some anatomies and amplifies it on others. Second, a SQUID-to-coil crossover near 450 T separates a SQUID-favored low-field regime from a coil-favored high-field one. Third, sensor channel count and ML model capacity are interchangeable design variables. None of these follow from isolated SNR analysis; they require running sensor, acquisition, and ML choices together end-to-end.
The remainder of this paper provides background and motivation (Sections II and III), the OpenULF platform and its validation (Section IV), and the three design-space explorations with their resulting design rules (Section V).
A. Ultra-Low-Field MRI
Clinical MRI operates at 1.5–3 T using superconducting magnets that weigh several tons and cost millions of dollars. Ultra-low-field MRI reduces to below 10 mT, typically 10–200 T, trading signal strength for portability and cost [1], [2]. ULF-MRI brain imaging has been demonstrated at national labs and in commercial scanners such as Hyperfine's Swoop [1], [3], [12]. Because the equilibrium magnetization is proportional to , a 50 T scanner produces a signal roughly 60,000 times weaker than a 3 T system. Prepolarization partially compensates for this deficit: the sample is magnetized at a stronger field before readout at [1].
Two fundamentally different sensor technologies compete at these field strengths. Both detect the precessing transverse magnetization through the reciprocity flux (Eq. 4), but they convert that flux into a measurable voltage differently.
SQUID sensors (Superconducting Quantum Interference Devices) sense magnetic flux via the Josephson effect. When operated in a flux-locked loop (FLL), the output voltage is
where [V/Wb] is the system flux-to-voltage coefficient. The key property is that does not depend on frequency: a SQUID responds equally well at 2 kHz ( ) and at 128 MHz ( ). This makes SQUIDs well suited to the very low Larmor frequencies of ULF-MRI, but the Josephson junction must be kept below its critical temperature, which requires a liquid-helium dewar.
Induction coils detect the time derivative of flux via Faraday's law:
where is the number of turns and is the Larmor frequency. The induced EMF grows linearly with , so at clinical fields coils are the standard sensor; but at the EMF falls by the same factor, placing the coil at a severe disadvantage. The benefit is that coils operate at room temperature and require no cryogenic infrastructure.
This split in the physics, frequency-independent gain for the SQUID versus -proportional gain for the coil, is the root of the sensor design problem that OpenULF addresses.
B. ML for MRI Reconstruction
Deep learning has become widely used in MRI image reconstruction, especially at low SNR where classical methods such as Wiener filtering, BM3D, and total variation underperform [13]. UNet [14] and DnCNN [15] are widely used; more recent architectures include NAFNet [16] and diffusion-based reconstructors [17]. Compressed sensing approaches [18] reduce acquisition time but still require adequate input SNR. At the noise levels typical of ULF-MRI (input SNR roughly 5–11 dB), ML reconstruction can recover several dB of PSNR depending on model capacity, making it increasingly important for ULF-MRI.
C. Related Simulation Tools
Several MRI simulation frameworks exist. MRiLab [9] and JEMRIS [10] simulate spin dynamics and k-space acquisition with configurable pulse sequences, but they do not include an ML reconstruction stage or sensor noise models calibrated for ULF-MRI. Their receiver models target clinical RF frequencies in sample-noise-dominated regimes and do not support SQUID sensors or prepolarization, both of which are specific to ULF-MRI. Virtual Scanner [19] is a browser-based educational simulator; BART [20] and SigPy [21] are reconstruction libraries that operate on pre-acquired k-space; and fastMRI [11] provides benchmark data and ML baselines but assumes fixed, clinical-quality input. As summarized in Table I, none integrates sensor-specific physics, ML reconstruction, and automated parameter sweeps in a single pipeline.
III. MOTIVATION
Designing a portable ULF-MRI system requires joint decisions across sensor hardware, acquisition protocol, and reconstruction software. Three observations motivate the need for an integrated simulation toolkit.
Physics simulators and ML frameworks are developed in isolation: physics tools such as MRiLab and JEMRIS stop at signal generation, while ML frameworks such as fastMRI assume pre-acquired data and never model the sensor physics.
insight-driven extension

(a) Component Library
(b) OpenULF Toolkit
(c) Design Insights
Fig. 1: Overview of OpenULF. (a) Component library with extensible physics, ML, data, and composition modules. (b) OpenULF toolkit: the co-simulation pipeline (left sub-box) transforms tissue maps through sensor-specific receivers and ML reconstruction; the design-space exploration engine (right sub-box) sweeps , training 774 models across four datasets. (c) Design insights: three quantitative design rules plus cross-dataset and compression-ratio validation. The top-level feedback arrow indicates that insights drive iterative extension of the component library.
that produced it. Combining them is not a one-time cost but a recurring one: every new sensor-or-model pairing means re-bridging data formats, coordinate systems, and noise models by hand, with glue code rewritten for every study and never reused. OpenULF instead fixes that boundary once, as a typed on-disk contract, so any sensor composes with any model.
Sensor SNR does not reliably predict post-ML reconstruction quality. The higher-SNR sensor is conventionally assumed to give the better image, but ML is a non-linear transform whose outcome depends on model capacity: as Section V shows, it can shrink or widen the sensor gap depending on the dataset, so SNR-based selection is unreliable once ML is in the loop.
The design space is also combinatorially large. A ULF-MRI system is parameterized by , NEX (number of signal averages), SQUID array channel count, and ML model choice. A modest grid of 13 values, 8 NEX settings, 7 channel counts, 5 models, and 3 sensor configurations already exceeds 10,000 design points. Automated exploration is needed to extract design rules from such a space.
IV. OPENULF PLATFORM
OpenULF is organized into three parts. The component library (Section IV-A) provides reusable physics, ML, and data-generation components that can be combined without writing custom integration code. The end-to-end co-simulation
pipeline (Section IV-B) runs these modules in sequence for any given sensor, model, and dataset configuration, producing reconstruction quality metrics rather than intermediate SNR values. The automated sweep engine (Section IV-C) runs the pipeline across large parameter grids, making it practical to explore thousands of design points. Architecturally, OpenULF rests on three reusable mechanisms, none of them MRI-specific: (i) a receiver abstract base class that reduces a new sensor to five enforced extension points over a shared acquisition core; (ii) a typed HDF5 contract that acts as a compiler-style intermediate representation between a physics front-end with no ML dependencies and a learned back-end, so the two halves are independently substitutable; and (iii) a make-like, idempotent sweep engine that turns design-space exploration into a resumable, deduplicated build. This work's two receivers, five reconstruction networks, and four dataset generators were each added through these interfaces alone, without modifying the shared acquisition core, the HDF5 contract, or the sweep engine.
A. Component Library
The component library provides physics, ML, and data modules, wired together by a thin composition layer; all four are shown on the left of Fig. 1.
- Physics Modules: The forward engine computes transverse magnetization [A/m] for each tissue class using a three-phase prepolarization model [1]. The sample is first polarized at for duration , then transferred to with a delay , and finally excited by a pulse for spin-echo readout. The magnetization available before excitation is
where . Prepolarization is essential for ULF-MRI: without it, the equilibrium magnetization at is too weak for imaging. With , the available magnetization increases by roughly three orders of magnitude, bringing the signal into the detectable range for both sensor types. Tissue-specific , , and proton density values are drawn from the BrainWeb atlas [22] with field-dependent relaxation following the Bottomley model [23]. The engine converts the image into k-space via 2-D FFT.
The receiver layer is the architectural core of OpenULF. Both sensors are concrete subclasses of a single abstract base class, ReceiverBase, which provides the shared acquisition path (reciprocity-based flux integration [24], NEX signal averaging, ADC quantization, GPU/CPU-portable colored-noise synthesis, and root-sum-of-squares multi-coil combination) in one core:
A concrete sensor supplies only five hooks: the per-unit-current sensitivity map and a per-readout voltage transform, plus three frequency responses (flux-to-voltage gain, transfer function , and noise power spectral density). The two shipped sensors differ only in these hooks: the room-temperature induction coil (Faraday EMF, Johnson-Nyquist noise) and the cryogenic SQUID gradiometer (frequency-independent gain, fT-scale noise floor [2], [25]). Both reuse the same acquisition base, so a new sensor touches only the five hooks, never the acquisition path. Adding a heterogeneous sensor thus means implementing one class against this interface, leaving the rest of the pipeline untouched; multi-channel arrays are produced by a factory that arranges hexagonal elements with decorrelated noise. Per-sensor geometry (the SQUID in a dewar, the coil on the scalp) is configuration, not a hard-wired assumption [2].
-
ML Modules: OpenULF includes five reconstruction architectures that span nearly two orders of magnitude in parameter count: SimpleCNN (0.22 M, 8-layer plain CNN), DnCNN (0.56 M, 17-layer denoiser [15]), UNet (7.76 M, encoder-decoder with skip connections [14]), ResUNet (10.3 M, UNet with residual blocks), and NAFNet (7.4 M, non-linear activation free network [16]). All expose the same single- or dual-channel input (SQUID, coil, or both) to single-channel magnitude output interface and share a unified training loop; a new architecture is added by dropping in a module that honors this interface, with no change to the data or physics layers. Spanning capacity this widely makes the model a tunable design resource in its own right, echoing model-compression work that trades network size against quantum and embedded hardware budgets [26].
-
Data Modules: Four dataset generators produce paired (noisy input, ideal target) HDF5 files from different anatomical sources: BrainWeb (20 fuzzy phantoms, 1,200 training slices), OASIS (414 subjects, 11,838 slices), IXI (583 subjects, 12,835 slices), and a synthetic abdomen set (500 phantoms). A new anatomy is added by writing a generator that emits the same (ideal, squid, coil) HDF5 contract; every generator is parameterized by , NEX, , and array channel count (1–7 hexagonal SQUID channels [27]), which enables the design-space sweeps in Section V.
-
Composition: A thin composition layer wires these modules into runs without custom integration code: a configuration interface (command-line parameters, captured per run as JSON manifests) and a typed HDF5 I/O contract: the (ideal, squid, coil) arrays plus the physics attributes that produced them. This layer is what makes the physics and learning halves composable and independently substitutable, as detailed next.
B. End-to-End Co-Simulation Pipeline
The co-simulation pipeline chains the full signal path (tissue atlas, forward engine, receiver, ML reconstruction, and evaluation via PSNR/SSIM) for any (sensor, model, dataset) configuration. The HDF5 dataset is the contract between the two halves of the toolkit: physics modules emit the arrays (ideal, squid, coil) with per-image normalization factors, and the ML modules import nothing from the physics layer, consuming only this file. Because the boundary is a stable on-disk schema rather than a shared in-memory API, the physics and learning stages evolve independently: a new architecture never touches the simulator, and a trained checkpoint can be re-evaluated against a regenerated dataset. A run is composed from a small set of command-line parameters (sensor, model, dataset, and physics settings); each dataset records its full physics configuration as HDF5 attributes and each training run emits a JSON manifest of hyperparameters, parameter count, and hardware, giving end-to-end provenance from physics parameter to trained checkpoint.
As its end-to-end metric the pipeline reports the compression ratio, which measures how much of the physical quality difference between sensors survives ML reconstruction. This metric cannot be recovered from SNR analysis and is defined in full in Section V.
C. Automated Parameter-Sweep Engine
The sweep engine turns the co-simulation pipeline into the inner loop of an automated design-space exploration over . It generates a parameterized HDF5 dataset per physics point and submits one cluster job per (model, sensor) training run. The sweep behaves like a build system: datasets and checkpoints are targets guarded by existence checks, so a preempted cluster job requeues and resumes where it stopped, and design points shared across studies are deduplicated rather than retrained. Per-run logs are then collected into a single index keyed by
TABLE II: Numerical-correctness check of the forward engine against closed-form Bloch solutions. Near-zero error is expected of a correct implementation; physical realism comes from measured parameters (Section IV-A). Here 0.00% denotes error below the two-decimal display threshold.
| # | Test description | Error |
|---|---|---|
| 1 | Single-voxel vs. spin-echo (6 tissues) | 0.00% |
| 2 | contrast: GM/WM ratio vs. TR (5 pts) | 0.00% |
| 3 | decay: signal vs. TE (6 points) | 0.00% |
| 4 | Prepolarization (12 config combinations) | <0.01% |
| 5 | Image: Parseval, k-space DC, tissue order | 0.00% |
| 6 | field dependence [23] | 0.00% |
all sweep dimensions. Rather than enumerate the full Cartesian product (over 10,000 points), each of the three studies fixes all axes but one and sweeps that axis at full resolution, so the 774 trained models are targeted, representative slices rather than exhaustive coverage: Study 1 varies sensor, model, and dataset at the default field; Study 2 sweeps (13 values, 30–10,000 T) and NEX (8 settings, 1–100); and Study 3 sweeps channel count (1–7). Design points shared across studies are trained once and reused.
D. Validation
We check the physics simulator in two complementary ways: numerical correctness and physical grounding. For correctness, Table II compares the forward engine against closed-form Bloch-equation solutions over six tests, all at or below 0.01% relative error, and an independent cross-check against the EPG simulator mri-sim-py, a different simulation method, gives consistent results. Near-zero error here is exactly what a correct implementation should produce: it confirms the equations are solved correctly. Physical realism is a separate question, answered by calibration that grounds every physical parameter (, receiver geometry, noise floors, standoff distances, and field dependence) in published measurements [1], [2], [23], [25].
V. EXPERIMENTS
A. Experimental Setup
We use four anatomical datasets generated by the sweep engine: BrainWeb (20 fuzzy phantoms, 1,200/171/336 train/val/test slices), OASIS (414 subjects, 11,838/1,684/3,458), IXI (583 subjects, 12,835/1,831/3,728), and a synthetic abdomen set (500 phantoms, 350/50/100). For OASIS and IXI the splits are subject-level; for BrainWeb and the synthetic abdomen set they are phantom-level. No subject or phantom contributes slices to both training and test, ruling out same-source leakage. All images are and normalized per-image to using the 99.5th percentile. The default physics configuration follows
TABLE III: Test-set PSNR (dB) and SSIM across 4 datasets, 5 models, and 3 sensor configurations. The Input row reports the raw sensor output (before ML) against the ideal image. Gap = SQUID – Coil in PSNR.
| Dataset | Model | SQUID | Coil | Both | Gap | |||
|---|---|---|---|---|---|---|---|---|
| dB | SSIM | dB | SSIM | dB | SSIM | |||
| BrainWeb | Input | 11.0 | .16 | 5.3 | .02 | – | – | +5.8 |
| SimpleCNN | 18.0 | .41 | 12.2 | .16 | 18.1 | .49 | +5.9 | |
| DnCNN | 18.9 | .66 | 13.8 | .37 | 19.0 | .68 | +5.2 | |
| UNet | 19.8 | .80 | 16.5 | .59 | 19.8 | .79 | +3.4 | |
| NAFNet | 19.8 | .78 | 16.5 | .61 | 19.8 | .78 | +3.3 | |
| ResUNet | 19.7 | .79 | 16.5 | .60 | 19.7 | .79 | +3.2 | |
| OASIS | Input | 10.6 | .13 | 4.5 | .01 | – | – | +6.1 |
| SimpleCNN | 31.9 | .98 | 12.9 | .38 | 31.8 | .98 | +19.0 | |
| DnCNN | 32.9 | .99 | 14.0 | .54 | 32.9 | .99 | +18.9 | |
| UNet | 33.3 | .99 | 17.6 | .80 | 33.4 | .99 | +15.7 | |
| NAFNet | 33.3 | .99 | 17.6 | .79 | 33.3 | .99 | +15.7 | |
| ResUNet | 33.4 | .99 | 17.7 | .81 | 33.4 | .99 | +15.7 | |
| IXI | Input | 9.4 | .14 | 5.7 | .02 | – | – | +3.7 |
| SimpleCNN | 12.9 | .28 | 9.3 | .10 | 13.3 | .35 | +3.6 | |
| DnCNN | 15.6 | .45 | 11.0 | .24 | 16.0 | .59 | +4.6 | |
| UNet | 18.2 | .70 | 13.3 | .43 | 18.3 | .76 | +4.9 | |
| NAFNet | 18.1 | .72 | 13.3 | .45 | 18.2 | .74 | +4.9 | |
| ResUNet | 18.2 | .74 | 13.2 | .43 | 18.2 | .76 | +4.9 | |
| Abdomen | Input | 4.6 | .04 | 4.7 | .02 | – | – | -0.1 |
| SimpleCNN | 16.8 | .49 | 12.7 | .16 | 17.2 | .54 | +4.1 | |
| DnCNN | 19.4 | .80 | 14.2 | .37 | 20.0 | .81 | +5.3 | |
| UNet | 22.4 | .91 | 16.7 | .77 | 22.7 | .91 | +5.6 | |
| NAFNet | 21.7 | .85 | 17.0 | .69 | 22.0 | .87 | +4.7 | |
| ResUNet | 22.3 | .89 | 16.9 | .75 | 22.6 | .88 | +5.5 | |
Vesanen et al. [1]: , , TE=50 ms, NEX=10, 7-channel hexagonal SQUID array whose elements are root-sum-of-squares combined into one input image. Five architectures are trained with MSE loss, Adam (, weight decay 0), batch size 16, early stopping (patience 20), and standard augmentation. Each model is trained separately for SQUID-only (1-ch), coil-only (1-ch), and both (2-ch) inputs. All runs use A100 GPUs. The sweep engine trained 774 models in total.
B. Sensor-ML Coupling
Can ML reconstruction close the quality gap between the expensive SQUID and the cheap coil, and is its effect on that gap consistent across anatomies? We answer this by running all five models on both sensors across all four datasets and reading reconstruction quality end-to-end rather than from input SNR.
The answer (Table III) is not the expected one. After reconstruction the SQUID leads on every dataset and every model, but the margin swings sharply with the anatomy: the same UNet that narrows the gap to +3.4 dB on BrainWeb blows it open to +15.7 dB on OASIS. The raw input gap does not predict this. SQUID and coil start within 3.7–6.1 dB on the brain datasets and roughly even on abdomen, yet ML can either compress that lead or amplify it several-fold. SSIM tells the same story, and the effect is a property of model capacity rather than of any one network: NAFNet stays within 0.2 dB of
Data were provided in part by OASIS-1 (Cross-Sectional; Principal Investigators D. Marcus, R. Buckner, J. Csernansky, J. Morris; grants P50 AG05681, P01 AG03991, P01 AG026276, R01 AG021910, P20 MH071616, U24 RR021382) [28].
The IXI data were provided by the Biomedical Image Analysis Group, Imperial College London, under CC BY-SA 3.0 [29].

Fig. 2: Reconstruction on a single representative BrainWeb test slice (per-slice PSNR shown; test-set means are in Table III). Top row: ideal target and noisy inputs from SQUID and coil. Bottom row: UNet outputs from each sensor configuration.

Fig. 3: Input vs. output SQUID-coil PSNR gap for four datasets (marker: 5-model mean, whisker: min-max across the five networks). The dashed CR=1 line is where ML leaves the gap unchanged; above it the gap is amplified, below compressed.
UNet and ResUNet on the brain datasets. Two-channel fusion pays off only where the sensors are genuinely complementary (a modest gain on BrainWeb and abdomen) and is redundant on OASIS, where the SQUID input is already near-perfect (SSIM 0.99).
Fig. 2 shows representative BrainWeb reconstructions. The SQUID input preserves gross brain anatomy while the coil input is dominated by noise. After UNet reconstruction, the SQUID-derived result retains finer detail.
To make this precise we define the compression ratio (CR) as the output gap divided by the input gap: CR=1 means ML preserves the physical difference, CR<1 means it compensates for the weaker sensor, and CR>1 means it magnifies the stronger sensor's lead. The ratio is meaningful only when the sensors differ at the input; where they start essentially tied (abdomen), we report the absolute dB change instead. Fig. 3 plots the output gap against the input gap for each dataset, with the five models shown as a min-max whisker.
The CR makes the dataset dependence concrete, spanning from compression on BrainWeb (UNet 0.59×, SimpleCNN

Fig. 4: PSNR gap (SQUID minus coil) vs. for three models at NEX=50. Positive values indicate SQUID is the superior sensor; negative values favor the coil. All three models cross zero near 450 T; the shaded band (418–473 T) spans the three models' individual zero-crossings, confirming that the crossover is driven by sensor physics rather than model choice.
1.02×) through near-preservation on IXI (0.99–1.35×) to strong amplification on OASIS (UNet 2.57×, SimpleCNN 3.11×). On abdomen the input gap is near zero, yet ML still creates a +4 to +6dB SQUID advantage.
These behaviors reflect how training-set size and anatomical regularity combine. When data is abundant and regular, the network recovers fine detail from the cleaner SQUID input (OASIS), so the gap widens; when data is limited (BrainWeb), neither sensor yields fine detail and the gap compresses. The abdomen case is the extreme one: the SQUID's distinct spatial sensitivity profile produces an advantage even though the two sensors start essentially tied.
Design Rule 1: The interaction between ML reconstruction and sensor quality is dataset-dependent. On data-rich problems ML can amplify the SQUID advantage; on data-limited problems it can compress the gap. System designers must evaluate sensor choices through end-to-end simulation rather than relying on SNR alone.
C. Field-Strength Sweep and Sensor Crossover
ULF spans two orders of magnitude in , and the coil's disadvantage shrinks as its EMF grows with frequency. At what field does the room-temperature coil become good enough to retire the cryogenics, and is that threshold a property of the sensors or of the reconstruction model? We sweep from 30 to 10,000 T at NEX=50 for three models (SimpleCNN, UNet, ResUNet).
The threshold is real, and it belongs to the sensors, not the network (Fig. 4). The SQUID-minus-coil gap starts positive and falls monotonically as the coil EMF grows with , crossing zero near 450 T; all three models cross within a narrow 418–473 T band, so the crossover is fixed by Faraday gain, not by ML capacity. Its location is robust to the number of averages: the NEX=50, 70, and 100 sweeps all cross within the same 418–473 T band, which reinforces that it is set by

Fig. 5: Design-space coverage: test PSNR (UNet) over the swept (, NEX) grid for the SQUID (left) and coil (right) sensors. Brighter is better; gray cells were not trained, exposing the sweep's goal-directed, non-exhaustive coverage. SQUID leads at low and the coil at high ; the crossover of Fig. 4 is where they switch.
sensor physics, not acquisition. Above it the coil, which needs no cryogenics and costs an order of magnitude less, is the sensor of choice. We run this sweep on BrainWeb, but unlike Rule 1's magnitude, which is data-bound, the crossover is set by the sensor transfer functions (Eqs. 1–2), which carry no anatomical assumptions, so we expect the field threshold to transfer across anatomies.
Fig. 5 places this crossover in the larger picture: it shows the reconstruction quality the sweep achieved across the full (, NEX) grid for each sensor. SQUID dominates the low-field corner and the coil the high-field corner, and the crossover is simply where the two panels exchange the lead; the blank cells expose the sweep's sparse, goal-directed coverage.
Design Rule 2: For the modeled sensor geometries and noise, SQUID is the superior sensor below roughly 450 T; above this field, the room-temperature induction coil suffices, eliminating the need for cryogenic infrastructure.
- NEX-model trade-off: Acquisition time is the other currency a designer spends on quality: NEX averaging buys input SNR as at a linear cost in scan time. Can model capacity buy the same quality for less time? Sweeping NEX from 1 to 100 at , it can: on SQUID, UNet at NEX = 4 (18.8 dB) matches SimpleCNN at NEX = 20 (19.0 dB): a 5 reduction in scan time bought with a larger network (Fig. 6, left).
On the coil the substitution is far less efficient: UNet at NEX=4 reaches 15.3 dB, while SimpleCNN requires NEX=50 to reach a comparable 14.9 dB, a ratio. The gap between SQUID and coil also widens with NEX: for UNet, from +3.6 dB at NEX = 1 to +4.7 dB at NEX = 100, because additional averaging benefits the SQUID signal (which contains recoverable anatomical detail) more than the coil signal (which remains noise-dominated).
These results add a temporal dimension to Rule 2: when scan time is limited, using a larger model is more effective

Fig. 6: Two faces of hardware–ML substitutability on SQUID (). Left: test PSNR vs. NEX, where UNet at NEX=4 matches SimpleCNN at NEX=20, trading model capacity for a 5 shorter scan. Right: test PSNR vs. SQUID channel count, where a 2-channel UNet matches SimpleCNN between its 3- and 4-channel points, trading model capacity for sensor channels.
than increasing NEX, particularly for SQUID-based systems.
D. Channel Count vs. Model Capacity
A SQUID array trades cryogenic channels for image quality; a larger network trades compute for the same. Are the two resources interchangeable, and at what exchange rate? We vary the SQUID channel count from 1 to 7 at and NEX=100 for three models (SimpleCNN, UNet, NAFNet).
They are, and the rate is concrete (Fig. 6, right). Quality rises monotonically with channel count but saturates above 4–5 channels, and model capacity buys the same quality as cryogenic channels: a 2-channel array with UNet (PSNR 20.5 dB) matches SimpleCNN only between its 3- and 4-channel points. A system designer can therefore trade sensor hardware for ML capacity in either direction, against constraints such as dewar size, cooling capacity, or available compute, though the exact exchange rate depends on the model pair and operating point.
Adding a coil channel provides a fusion gain at every channel count, confirming that the coil's complementary spatial profile adds value even alongside multiple SQUID channels. The fusion gain is largest at low channel counts (1.0 dB at 1 channel for UNet) and decreases as channels increase (0.4 dB at 6 channels), because additional SQUID channels progressively cover regions where the coil previously provided unique information.
Design Rule 3: In hardware/software co-design, channel count and model capacity are substitutable: for systems with strict hardware constraints, a more capable ML model can partially compensate for fewer SQUID channels. Returns diminish above 4–5 channels.
E. Discussion
Rule 1's per-dataset behavior (Section V-B) is itself the key finding: there is no universal rule for how ML interacts with sensor quality. The broader lesson is methodological: once a learned component sits in the signal path, component-level figures of merit such as SNR no longer predict system quality, and end-to-end co-simulation becomes the only reliable predictor. Nothing about that structure (a simulated acquisition front-end and a learned reconstruction back-end joined by a typed IR) is MRI-specific: the receiver base class, HDF5 contract, and idempotent sweep make no MRI assumptions, so any modality with a simulable forward model could reuse the tool chain by supplying a new front-end and dataset generator.
Our conclusions are simulation-based. The physics is validated for numerical correctness (Table II) and calibrated to published measurements [1], [2], [25], but hardware imperfections such as patient motion, environmental magnetic interference, and manufacturing tolerances are not yet modeled. PSNR and SSIM measure fidelity to the simulated ground truth rather than task-level quality such as lesion detectability; the four datasets are brain and abdomen simulants, leaving anatomies such as cardiac and spine untested; and each configuration is trained with a single seed. Validating the rules on scanner data (Rules 2 and 3 are swept on BrainWeb) is the natural next step, though paired ground-truth ULF-MRI data remains scarce.
VI. CONCLUSION
We presented OpenULF, a modular, open-source toolkit for data-driven ultra-low-field MRI with both classical and quantum sensors. Across four datasets and 774 trained models, three explorations yield findings that SNR analysis alone cannot provide: the ML-sensor interaction is dataset-dependent, compressing the SQUID-coil gap on some datasets and amplifying it on others; the SQUID-coil crossover near 450 µT is where the room-temperature coil takes over from the cryogenic SQUID; and channel count and model capacity are interchangeable design variables, with diminishing returns above four to five SQUID channels. The OpenULF code is released as open source, with the trained models to follow on publication; because the pipeline is configuration-driven, researchers can substitute their own architectures or datasets to explore questions beyond those reported here. Future work includes validation against real ULF-MRI hardware and support for sensors such as optically pumped magnetometers.
ACKNOWLEDGMENT
This work was supported in part by the U.S. National Science Foundation under Grants 2507948, 2513431, 2440637, and 2612445.
REFERENCES
[1] P. T. Vesanen et al., “Hybrid ultra-low-field mri and magnetoencephalography system based on a commercial whole-head neuromagnetometer,” Magnetic Resonance in Medicine, vol. 69, no. 6, pp. 1795–1804, 2013.
[2] V. S. Zotev et al., “Microtesla mri of the human brain combined with meg,” Journal of Magnetic Resonance, vol. 194, no. 1, pp. 115–120, 2008.
[3] Hyperfine, Inc., “Swoop portable MRI system,” 2020, https://www.hyperfine.io.
[4] K. N. Sheth et al., “Assessment of brain injury using portable, low-field magnetic resonance imaging at the bedside of critically ill patients,” JAMA neurology, vol. 78, no. 1, 2021.
[5] C. Kleman et al., “Full waveform inversion-based ultrasound computed tomography acceleration using two-dimensional convolutional neural networks,” Journal of Nondestructive Evaluation, Diagnostics and Prognostics of Engineering Systems, vol. 6, no. 4, 2023.
[6] Y. Yuan, H. Wu, J. Li, H. Wang, Y. Wu, Y. Lin, and L. Yang, “MUX-USCT: A noise-robust neural network for ultrasound computed tomography,” arXiv preprint arXiv:2607.10648, 2026.
[7] Y. Lin et al., “Physics-guided data-driven seismic inversion: recent progress and future opportunities in full waveform inversion,” IEEE Signal Processing Magazine, vol. 40, no. 1, pp. 115–133, 2023.
[8] Y. Yuan et al., “Two teachers better than one: Hardware-physics co-guided distributed scientific machine learning,” arXiv preprint arXiv:2603.09032, 2026.
[9] F. Liu et al., “MRILab: Fast realistic MRI simulations based on generalized multi-pool exchange tissue model,” IEEE Trans. Med. Imaging, vol. 36, no. 2, pp. 527–537, 2017.
[10] T. Stoecker et al., “High-performance computing MRI simulations,” Magnetic Resonance in Medicine, vol. 64, no. 1, pp. 186–193, 2010.
[11] J. Zbontar et al., “fastMRI: An open dataset and benchmarks for accelerated MRI,” arXiv preprint arXiv:1811.08839, 2018.
[12] V. S. Zotev et al., “Squid-based instrumentation for ultralow-field mri,” Superconductor Science and Technology, vol. 20, no. 11, pp. S367–S373, 2007.
[13] A. S. Lundervold and A. Lundervold, “An overview of deep learning in medical imaging focusing on mri,” Zeitschrift für medizinische Physik, vol. 29, no. 2, pp. 102–127, 2019.
[14] O. Ronneberger et al., “U-net: Convolutional networks for biomedical image segmentation,” in MICCAI. Springer, 2015, pp. 234–241.
[15] K. Zhang et al., “Beyond a gaussian denoiser: Residual learning of deep cnn for image denoising,” IEEE transactions on image processing, vol. 26, no. 7, pp. 3142–3155, 2017.
[16] L. Chen et al., “Simple baselines for image restoration,” in European conference on computer vision. Springer, 2022, pp. 17–33.
[17] Y. Song et al., “Solving inverse problems in medical imaging with score-based generative models,” in International Conference on Learning Representations (ICLR), 2022.
[18] M. Lustig et al., “Sparse mri: The application of compressed sensing for rapid mr imaging,” Magnetic Resonance in Medicine: An Official Journal of the International Society for Magnetic Resonance in Medicine, vol. 58, no. 6, pp. 1182–1195, 2007.
[19] G. Tong et al., “Virtual scanner: MRI on a browser,” Journal of Open Source Software, vol. 4, no. 43, p. 1637, 2019.
[20] M. Uecker et al., “Berkeley Advanced Reconstruction Toolbox,” in Proc. Intl. Soc. Mag. Reson. Med., vol. 23, 2015, p. 2486.
[21] F. Ong and M. Lustig, “SigPy: A Python package for high performance iterative reconstruction,” in Proc. Intl. Soc. Mag. Reson. Med., vol. 27, 2019, p. 4819.
[22] C. A. Cocosco et al., “BrainWeb: Online interface to a 3D MRI simulated brain database,” NeuroImage, vol. 5, no. 4, p. S425, 1997.
[23] P. A. Bottomley et al., “A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: Dependence on tissue type, NMR frequency, temperature, species, excision, and age,” Medical Physics, vol. 11, no. 4, pp. 425–448, 1984.
[24] D. I. Hoult and R. Richards, “The signal-to-noise ratio of the nuclear magnetic resonance experiment,” Journal of Magnetic Resonance (1969), vol. 24, no. 1, pp. 71–85, 1976.
[25] J. Clarke and A. I. Braginski, Eds., The SQUID Handbook: Fundamentals and Technology of SQUIDs and SQUID Systems. Wiley-VCH, 2004.
[26] Z. Hu et al., “Quantum neural network compression,” in Proc. 41st IEEE/ACM International Conference on Computer-Aided Design (IC-CAD), 2022.
[27] V. S. Zotev et al., “Multi-channel squid system for meg and ultra-low-field mri,” IEEE transactions on applied superconductivity, vol. 17, no. 2, pp. 839–842, 2007.
[28] D. S. Marcus, T. H. Wang, J. Parker, J. G. Csernansky, J. C. Morris, and R. L. Buckner, “Open access series of imaging studies (OASIS): Cross-sectional MRI data in young, middle aged, nondemented, and demented older adults,” Journal of Cognitive Neuroscience, vol. 19, no. 9, pp. 1498–1507, 2007.
[29] Biomedical Image Analysis Group, Imperial College London, “IXI dataset,” https://brain-development.org/fixi-dataset/, cC BY-SA 3.0.