arXiv:2611.01536cs.CVNov 2026

Manuscript Offprint

arXiv:2611.01536

CUFormer: Modality-Aware CT-Ultrasound Point Cloud Registration for Orthopedic Navigation

Lixing Tan, Qing Xia, Yuting Guo, Shuai Li, Aimin Hao

Lixing Tan¹ Qing Xia¹,* Yuting Guo² Shuai Li¹ Aimin Hao¹

¹Beihang University ²Beijing Information Science and Technology University

Project page: https://lextran.github.io/CUFormer/

*Corresponding author: xiaqing@buaa.edu.cn

Figure

Figure 1. CUFormer registers intraoperative US bone surfaces to preoperative CT for radiation-free orthopedic navigation. Tracked B-mode images are reconstructed into a partial point cloud and aligned with the CT model. CT-US results span the tibia, fibula, foot, and femur. The spine example adds CT-RGB-D registration. Both US and RGB-D point clouds provide partial bone-surface observations.

Abstract

Registering intraoperative Ultrasound (US) bone-surface point clouds to preoperative Computed Tomography (CT) models can support radiation-free orthopedic navigation. Differences in imaging physics produce distinct surface sampling patterns, complicating cross-modal feature matching. Bone strongly reflects ultrasound and casts an acoustic shadow, so US captures only the probe-facing bone surface and leaves most CT points unmatched. This partial coverage makes correspondences ambiguous on near-symmetric bones. We propose CUFormer, a modality-aware registration framework. We introduce a Modality-Aware Transformer (MAT) that encodes input modality through identity embeddings and updates features through modality-specific adapters within a shared geometric matcher. We model imaging visibility and use predicted reference-side visibil-

ity to weight coarse and fine correspondences. CUFormer achieves the lowest mean registration errors among evaluated methods on two CT-US datasets, UltraBones100k and UltraBonesHip, and the CT-RGB-D dataset SpineDepth. On UltraBones100k, it reaches mean rotation error 4.53°, translation error 3.19 mm, and Root Mean Squared Error (RMSE) 2.54 mm, reducing RMSE by 65.8% relative to the best-performing baseline by this metric. These results demonstrate CUFormer's effectiveness in aligning partial bone surfaces with preoperative CT across anatomical sites.

1. Introduction

Image-guided orthopedic surgery requires millimeter-level localization of bony anatomy. Preoperative Computed Tomography (CT) provides a complete, high-resolution bone model. Using this model for intraoperative guidance re-

Figure

Figure 2. Comparison of a previous method and CUFormer. CT-US correspondence ambiguity on a fibula case. US is orange. Lines connect sampled US points to estimated CT positions: green indicates errors within 5 mm of the Ground-Truth (GT) counterpart, and red indicates larger errors. (a) CoFiNet produces an end-for-end flip. (b) CUFormer uses MAT and visibility-guided matching, with CT colored by predicted visibility. Nearest-CT assignment places 34.4% of all US points in negative regions for CoFiNet, 11.8% for CUFormer, and 11.7% for GT.

quires aligning it with the patient's current anatomy. Repeated intraoperative CT or fluoroscopy exposes patients and clinicians to ionizing radiation, motivating radiation-free intraoperative imaging. Ultrasound (US) provides real-time, non-invasive observations of local bone surfaces, which can be reconstructed as point clouds from tracked acquisitions. Registering these surfaces to the preoperative CT model establishes the spatial correspondence between the intraoperative anatomy and the preoperative reference for orthopedic navigation (Figure 1).

Representing both modalities as bone-surface point clouds removes image intensity differences [26], but imaging physics still shapes the observed geometry. CT-US registration therefore faces two major challenges: modality-dependent surface sampling and highly asymmetric coverage. Modality-dependent sampling changes the local neighborhoods used to construct descriptors, making corresponding CT and US surface points harder to match. Coverage asymmetry also originates in imaging physics. Bone strongly reflects ultrasound waves and casts an acoustic shadow, so a typical tibial sweep captures only a thin, probe-facing surface strip and leaves most of the CT reference unobserved. This partial coverage makes correspondence estimation ambiguous along near-symmetric bone structures. Figure 2(a) shows a matcher placing the US source at the wrong end of the CT reference, producing an end-for-end flip.

Existing strategies can hardly address these challenges. Local refinement such as ICP [5] requires a close initializa-

tion, and RANSAC-based estimation [10] relies on putative correspondences, whose inlier ratio drops under partial observations. NeuralBoneReg (NBReg) [26] optimizes the transformation per case at test time, which limits real-time use, and sparse or limited surface observations may reduce its performance.

By contrast, learning-based matchers learn features across cases and register a new pair without per-case optimization, performing well on same-modality RGB-D benchmarks [19, 34]. These methods treat the two inputs as scans from the same sensor. GeoTransformer [19] encodes both clouds with shared weights, leaving no mechanism to account for modality-dependent sampling. Low-overlap methods [13, 33] model overlap as a symmetric relation between two partial scans, where non-overlap stems from viewpoint and scan range. In CT-US registration, the CT reference is complete, whereas US observes only part of the reference surface owing to the acoustic properties of bone. A CT-US matcher therefore needs features that distinguish the two modalities and correspondence selection that reflects reference-side coverage.

To address these challenges, we propose CUFormer. Its Modality-Aware Transformer (MAT) preserves a shared geometric matcher while providing explicit input identities and modality-specific feature updates. Learnable embeddings encode CT and US stream identities at coarse and fine resolutions. Gated bottleneck adapters provide modality-specific residual updates after shared attention blocks. A visibility module models imaging visibility through geometric overlap, predicting which CT regions are observed by US under overlap-derived supervision. The predicted visibility scores weight coarse and fine correspondence scores, guiding correspondence selection toward CT regions supported by the partial US observation. Figure 2(b) shows the result.

Our contributions are summarized as follows:

  • We propose CUFormer, a modality-aware framework for CT-US bone-surface registration in orthopedic navigation. It addresses modality-dependent surface sampling and asymmetric coverage within a single learned matcher that requires no per-case optimization.
  • We introduce MAT, which integrates modality identity encoding and modality-specific residual adaptation into a shared geometric matcher.
  • We develop reference-side visibility-guided matching, which complements MAT by modeling US imaging visibility on CT through geometric overlap and using the predicted visibility to weight coarse and fine correspondence scores.
  • We evaluate CUFormer against state-of-the-art methods on (i) the public CT-US datasets UltraBones100k and UltraBonesHip, covering five structures: fibula, tibia, foot, femur, and pelvis; and (ii) the public CT-RGB-D dataset SpineDepth, covering multiple vertebrae. Ablations fur-

ther assess the contribution of each component.