Task-driven sparse-view planar X-ray computed tomography for axial metrology

Yibin Tian, Shuming Yang, Wolfgang Heidrich

A PREPRINT

Yibin Tian *

SKL-Radio Frequency Heterogeneous Integration

Shenzhen University, Shenzhen 518060, China

ybtian@szu.edu.cn

Shuming Yang †

SKL-Precision Micro-Nano Manufacturing Technology

Xi'an Jiaotong University, Xi'an 710000, China

shuming.yang@mail.xjtu.edu.cn

Wolfgang Heidrich

Visual Computing Center

King Abdullah University of Science and Technology, Thuwal 23955-6900, Saudi Arabia

wolfgang.heidrich@kaust.edu.sa

November 3, 2026

ABSTRACT

Computational imaging can reduce the acquisition burden when the system is designed around task-relevant information rather than complete image recovery. We investigate this framework for sparse-view planar X-ray tomography (XCT) of layered objects, using measurement of the axial termination gap (ATG) of back-drilled printed-circuit-board (PCB) vias as a representative task. Instead of using 3D reconstruction, the proposed method estimates ATG directly from polychromatic projection data through a physics-based sensitivity model. A gated hierarchical estimator first fits ATG together with radiometric perturbations and selectively introduces global geometry and specimen-warp models only when a reduced-χ2\chi^2 adequacy test and Bayesian information criterion justify the added degrees of freedom. The same low-dimensional perturbation model is incorporated into marginalized Fisher-information analyses, yielding task-specific conditional and calibration-aware Cramér–Rao bounds. Numerical experiments show that, under a fixed total photon budget, task-driven metrology changes little over 12–64 views, while reconstruction-driven metrology is strongly view-dependent with much larger errors. The direct estimator also follows the expected photon-limited N0−1/2N_0^{-1/2} scaling. These results demonstrate a computational sensing strategy in which XCT acquisition, physical modeling, and inference are evaluated according to the quantitative information required by the task rather than for general-purpose volumetric image fidelity.

Keywords Computed tomography · 3D reconstruction · Axial metrology · Fisher information · Back-drilled via

1 Introduction

Computational sensing increasingly treats hardware, measurement encoding, physical forward models, and inference as components of a jointly designed system Mait et al. [2018], Sitzmann et al. [2018], Wang et al. [2022]. This co-design perspective allows measurements to be tailored to the information required by a reconstruction or inference task, trading optical or sampling complexity for computation Arguello et al. [2023]. Recent task-oriented approaches extend this idea further: task-driven lens design optimizes optical parameters for downstream performance Yang et al. [2026], while latent-space imaging encodes measurements into compact task-relevant representations Souza et al. [2025]. These studies motivate imaging systems designed for the final task rather than for faithful intermediate image formation.

*Corresponding author.

†Corresponding author.

A PREPRINT

Conventional X-ray computed tomography (XCT) is typically designed for volumetric reconstruction, with geometry, angular sampling, exposure, and reconstruction selected to achieve acceptable spatial resolution and artifact levels Kak and Slaney [1988], Feldkamp et al. [1984], Tuy [1983], Okawa and Furukawa [2026]. Sparse sampling reduces acquisition time but makes the full-volume inverse problem increasingly ill posed. Algebraic and regularized methods, such as SART and reconstruction constrained by total variations, mitigate sparse-view artifacts Andersen and Kak [1984], Sidky and Pan [2008], while learned methods seek visually faithful recovery from severe undersampling Lahiri et al. [2023], Lin et al. [2023], Chai et al. [2026]. These approaches remain appropriate when the reconstructed image itself is the desired output, but they do not directly address how measurement number and allocation determine the uncertainty of a specified physical parameter.

This distinction is important in industrial metrology, where a 3D volume may contain millions of unknown voxels while the desired output is only one or a few quantities, such as depth, spacing, interface position, wall thickness, or defect size. Task-based imaging and statistical estimation theory indicate that the relevant quantity is the sensitivity of measurements to the parameter of interest after accounting for noise and nuisance perturbations Barrett and Myers [2004], Kay [1993]. Consequently, a measurement set may be insufficient for a stable recovery of the entire object while retaining substantial Fisher information along a particular parameter direction Ly et al. [2017]. Related work in optical parameter estimation similarly emphasizes task sensitivity over global pixel fidelity Barbastathis [2024].

We investigate this framework using sparse-view planar XCT of layered printed circuit boards (PCBs) (Fig. 1), with the axial termination gap (ATG) of a back-drilled via as representative task. The Back-drilling removes the unused conductive portion of a via, which makes the remaining axial separation important for high-speed interconnect performance. The plate-like PCB geometry naturally motivates an oblique planar XCT trajectory similar to that of computed laminography Helfen et al. [2005], Marcus [2022]. This application provides a controlled test bed in which the desired output is low-dimensional, while calibration errors and board deformation can still bias axial metrology.

Figure

Figure

Figure 1: Illustration of (a) planar XCT (SOD: source-object distance; SDD: source-detector distance; FPD: flat panel detector) and (b) a layered PCB with back-drilled vias. ZrefZ_{\mathrm{ref}} and ZtermZ_{\mathrm{term}} denote the axial coordinates of the reference plane and hole termination, and LL is the axial termination gap (ATG).

The proposed framework combines sparse oblique XCT acquisition, a differentiable physical projection model for direct axial metrology, and a gated estimator that expands the perturbation model only when supported by the data. ATG is estimated directly from projection measurements rather than through an intermediate 3D reconstruction, while perturbation-marginalized Fisher analyses quantify the task information preserved by a given view and photon allocation. The main contributions of this study are as follows.

(1) a task-driven formulation of sparse-view planar XCT with projection-domain sensitivity and Fisher information for quantitative axial metrology;

(2) a low-dimensional hierarchical perturbation model in which radiometric effect is fitted first, followed only when necessary by global geometry and specimen warp, with Bayesian information criterion used to limit unnecessary model expansion Schwarz [1978];

A PREPRINT

(3) perturbation-marginalized Cramér–Rao bounds that incorporate both task sensitivity and calibration uncertainty; (4) numerical comparisons of task-driven and reconstruction-driven metrology under fixed-photon-budget view sweeps, photon-budget variation, and controlled model perturbations. The central result is that task-driven and reconstruction-driven metrology can have substantially different angular-sampling requirements.