arXiv:2611.03446cs.RONov 2026

Manuscript Offprint

arXiv:2611.03446

Clinician-AI Collaboration in Endovascular Neurointerventions: In-Vitro User Study of Copilot Assistance Modes for Robotic Navigation Across the Autonomy Spectrum

Nikola Fischer, Harry Robertshaw, S. M. Hadi Sadati, Xingyu Chen, Weiyuan Deng, Benjamin Jackson, Michael Adofo Kwakye, Sara Sciacca, Naga Kandasamy, Parthiban Balasundaram, Sahan Guruge, Charles Cummins, Joe Leyon, Nasr Abdelsalam, Matthew Benger, Andrew Falzon, Alejandro Granados, Christos Bergeles, Thomas Booth

Nikola Fischer¹, Harry Robertshaw¹, S.M.Hadi Sadati¹,², Xingyu Chen¹, Weiyuan Deng¹, Benjamin Jackson¹, Michael Adofo Kwakye¹, Sara Sciacca³, Naga Kandasamy³, Parthiban Balasundaram³, Sahan Guruge³, Charles Cummins³, Joe Leyon³, Nasr Abdelsalam⁴, Matthew Benger⁵, Andrew Falzon⁵, Alejandro Granados¹, Christos Bergeles¹†, and Thomas Booth¹,³*†

*Abstract—Mechanical thrombectomy (MT) is an effective treatment for acute ischemic stroke, yet access remains limited by the availability of specialist neurointerventional expertise. Telerobotic and AI-assisted systems may help address this challenge, but it remains unclear how clinicians interact with different levels of autonomy and how autonomous navigation systems perform in clinically realistic environments. This work introduces Copilot Assistance Modes (CAMs), a framework for clinician-AI collaboration in robotic endovascular navigation spanning the autonomy spectrum from manual operation to full autonomy. In the first in-vitro user study systematically comparing clinician-AI collaboration strategies, 11 expert and two student participants evaluated six navigation modes in a patient-specific vascular phantom under fluoroscopic guidance. Objective performance metrics were combined with subjective assessments, including workload, usability, and operator preference. Autonomous performance was additionally evaluated both in silico and in vitro to investigate long-horizon sim-to-real transfer. Hands-on manual navigation achieved the strongest overall performance (100% success). While fully autonomous navigation achieved 81% success in simulation, performance dropped to 23% in vitro. In contrast, several clinician-AI shared-autonomy modes achieved success rates of 92–97%, approximately four times higher than full autonomy and comparable to direct robotic control (95%), potentially demonstrating mitigation to sim-to-real domain shift when using imperfect AI through clinician-AI collaboration. Among the evaluated AI-driven approaches, user-directed task delegation and rotational shared control demonstrated the most favorable balance between performance and clinician acceptance. These findings plausibly suggest that, for robotic neurointerventional navigation, effective clinician-AI collaboration and allocation of control authority may currently be more important for reliable performance than increasing autonomy alone.

Figure

Figure 1. Clinician-AI Collaboration for Neurointerventions: Introducing the Copilot Assistance Modes (CAMs) including the conventional Manual (0) and Direct Robot Control (1) as well as the AI-driven modes (blue) AI Advice (2), AI Assistance (3), Full Copilot Assistance (4), and Fully Autonomous Navigation (5). The blue color gradient indicates increasing levels of autonomy, ranging from white (manual) over grey (teleoperation) to dark blue (highest autonomy).

Index Terms—human-AI, human-machine, shared autonomy, autonomous navigation, zero-shot sim-to-real transfer

I. INTRODUCTION

STROKE is one of the leading causes of death and long-term disability worldwide, with a substantial socioeconomic burden on healthcare systems [1]. Acute ischemic stroke (AIS) remains a major global health burden, with approximately 12 million new cases and 7 million deaths annually [2]. Large vessel occlusion (LVO) strokes are particularly time critical, as reduced cerebral perfusion in a large territory accelerates considerable neuronal loss, making rapid recanalization essential for preserving neurological function and improving clinical outcomes. Mechanical thrombectomy (MT) has emerged as the reference standard treatment for eligible patients, significantly improving functional outcomes and reducing mortality compared to medical therapy alone [3]–[5].

¹School of Biomedical Engineering & Imaging Sciences, King's College London, UK; ²School of Engineering and Materials Science, Queen Mary University of London, UK; ³Department of Neuroradiology, King's College Hospital NHS Foundation Trust, London, UK; ⁴The Walton Centre NHS Foundation Trust, Liverpool, UK; ⁵St George's University Hospitals NHS Foundation Trust, London, UK *Corresponding author: thomas.booth@kcl.ac.uk †These authors contributed equally as senior authors. This work was supported by the Medical Research Council (MRC) Impact Acceleration Account (IAA) 2021 at King's College (Grant No. MR/X502923/1), by EPSRC [VITRRO / EP/Y024281/1 - ERC Proof of Concept] under the Horizon Europe Guarantee Extension, the King's-funded Centre for Doctoral Training on Surgical & Interventional Engineering, and the King's-China Scholarship Council [CSC No.: 202408330072]. Work in the Surgical & Interventional Engineering Validation Suite (mock OR) was supported by core funding from the Wellcome/EPSRC Centre for Medical Engineering [WT203148/Z/16/Z], a multi-user equipment grant for post-mortem evaluation of medical devices from Wellcome [218286/Z/19/Z], and the Wolfson Foundation [PR/ylr/md/21896].

Despite its clinical effectiveness, access to MT remains highly unequal. Because treatment success is critically dependent on minimizing the time from symptom onset to reperfusion, the limited availability of specialized neurointerventional centers and trained operators prevents many eligible patients from receiving timely treatment [6], thereby reducing the likelihood of favorable clinical outcomes [7]. The workforce shortage [8] is further exacerbated by the technical complexity of endovascular interventions, in which operators must navigate instruments (i.e., guidewires and catheters) through highly tortuous vasculature while achieving predictable control of the distal instrument tip through indirect manipulation of the two degrees of freedom (DoF) rotation φ\varphi and translation dd at the proximal end. As a result, accurate navigation places considerable demands on clinician dexterity, hand-eye coordination, and procedural experience. To facilitate instrument handling, robotic (neuro)endovascular systems are emerging [9]. These systems enable precise teleoperated manipulation of endovascular instruments, potentially also allowing expert clinicians to perform procedures remotely from centralized stroke centers [10]. This paradigm may reduce geographic disparities by concentrating expertise while extending procedural reach to peripheral medical centers. In addition, robotic actuation may reduce radiation exposure and improve operator ergonomics [11], [12], even for clinicians performing procedures locally.

Beyond pure teleoperation, robotic systems also create opportunities for enhancement with artificial intelligence (AI) of the challenging navigation from the vascular access site to the intracranial occlusion. As this navigation requires operators to achieve predictable instrument tip behavior through indirect control and continuous adaptation, it remains a major procedural bottleneck and a potential source of serious complications (risks), including procedure-related vessel perforation (1%), dissection (2%), and distal embolization of the thrombus (9%) [13].

Outside of neurointervention, early efforts focused on automating constrained procedural subtasks [14]. Chen et al. presented an autonomous vascular access system using deep-learning-based robotic guidance, demonstrating the feasibility of AI-assisted vascular procedures [15]. Building on this for neurointervention, Abah et al. proposed a semi-autonomous catheter steering framework based on path-planar algorithms and preplanned optimal catheter configurations [16]. Yet, maintaining robust performance over long-horizon navigation tasks remains a key challenge in autonomous endovascular navigation, with recent work highlighting limitations in the generalization of reinforcement learning (RL) policies as task complexity and navigation duration increase [17].

Recent developments in AI-driven robotic control systems, also described as agentic control systems or AI copilots [18], have increasingly shifted attention toward collaborative paradigms that combine AI capabilities with clinician expertise, reflecting the emerging vision of surgical teams in which clinicians, AI systems, and robotic platforms operate as collaborative partners [19]. However, effective integration of human-AI collaboration remains an open challenge. In the endovascular domain, Yan et al. introduced a machine-

learning-based human-machine collaborative control strategy for guidewire insertion that dynamically adapts to varying surgical conditions and incorporates a trust evaluation model to improve procedural safety and efficiency [20]. Such systems are increasingly viewed as shared autonomy agents designed to enhance procedural precision, reduce cognitive workload, and improve navigation efficiency. This way, AI-assisted robotic systems can help broaden access to care by supporting less experienced operators (i.e. novices) or even clinicians without specialized training during demanding procedural steps.

Despite these advances shown in [14]–[17], [19]–[22], existing AI research for robotic endovascular use cases such as MT, remains primarily focused on the development of increasingly autonomous navigation systems and improvements in navigation performance. In contrast, considerably less attention has been paid to understanding how clinicians and AI systems should collaborate during robotic endovascular procedures. As a result, it remains largely unclear how clinician-AI teams perform in robotic endovascular navigation settings, how operators prefer to interact with AI copilots in shared-control scenarios, and how different distributions of control authority influence navigation performance, workload, usability, and clinician perception.

Addressing these questions is critical for the clinical translation of AI-assisted robotic intervention, where successful adoption will likely depend not only on autonomous capabilities, but also on effective mechanisms for clinician-AI collaboration. Accordingly, this study investigates how different levels of AI-driven autonomy [23] influence objective navigation performance, operator workload, usability, and clinician perception during robotic endovascular navigation for MT (Figure 1).

Specifically, this work makes the following contributions:

  • We introduce the Copilot Assistance Modes (CAMs), a framework for systematically investigating clinician-AI collaboration in robotic neurointerventional navigation. We instantiate this framework through multiple assistance paradigms representing guidance, shared control, delegated control, and fully autonomous navigation.

  • We developed a comprehensive in-vitro evaluation framework that combines objective navigation outcomes with subjective assessments of workload, usability, and clinician perception for the evaluation of AI-assisted robotic interventions.

  • We present, to our knowledge, the first in-vitro user study with clinical experts systematically comparing clinician-AI collaboration modes for robotic neurointerventional navigation under fluoroscopic guidance.

  • We provide one of the first empirical investigations into how different allocations of control authority between clinicians and AI systems influence subjective and objective metrics, while also highlighting challenges associated with long-horizon autonomous navigation and zero-shot sim-to-real transfer.

References (35)