Physically Constrained Kinematic Calibration for High-Precision Robotic Needle Guidance

Prabhu Nithin Gollapudi, Andreas Maier, Franziska Mathis-Ullrich, Water Zhao, Jia Wang, Siming Bayer

Prabhu Nithin Gollapudi*, Andreas Maier*, Franziska Mathis-Ullrich*, Water Zhao†, Jia Wang†, and Siming Bayer*

  • Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany

Email: {prabhu.nithin.gollapudi, andreas.maier, franziska.mathis-ullrich, siming.bayer}@fau.de

†Precision Medtech Co., Ltd, Beijing, China

Email: {water.zhao, jia.wang}@precisionmedtech.com

Abstract—Percutaneous needle insertion is a common practice in minimally invasive procedures whose clinical success depends on accurate needle placement. However, achieving the necessary precision is a challenge for manual operators. Robotic assistance can improve targeting accuracy and reduce operator dependence, but residual millimeter-scale errors persist because nominal kinematic models do not fully represent the physical manipulator. This paper presents a robust bound-constrained kinematic calibration method designed to improve positional accuracy for a 6-Degree-of-Freedom (6-DoF) robotic needle-guidance manipulator. The approach uses physically plausible parameter ranges to regularize a compact 19-parameter geometric identification model. By treating these bounds as active optimization constraints rather than mere numerical safeguards, the framework effectively mitigates parameter coupling and overfitting, ensuring highly stable parameter identification. In simulation, the proposed method reduced end-effector Root Mean Square Error (RMSE) from 3.422 mm to 0.690 mm (a 79.8% improvement). In benchtop hardware experiments, the calibrated model achieved sub-millimeter validation accuracy, including 0.865 mm RMSE in experimental trajectory validation and 0.610 mm RMSE in a coordinate-lattice verification task. These results demonstrate that physically constrained kinematic calibration can enhance robotic needle-targeting accuracy, establishing a high-precision foundation for future clinical applications.

Index Terms—Kinematic calibration, Medical robotics, Needle guidance, Parameter identification, Percutaneous intervention, Positional accuracy.

I. INTRODUCTION

Accurate percutaneous needle insertion is a critical component of various minimally invasive interventions, including biopsies, localized drug delivery, and thermal ablation. Among these, percutaneous thermal ablation for liver tumors represents a particularly demanding application. While ablation serves as a primary therapeutic alternative for patients ineligible for surgical resection [1]–[3], its clinical efficacy relies heavily on precise targeting. Achieving complete tumor necrosis requires a strict 5–10 mm safety margin while carefully avoiding collateral damage to adjacent vasculature and bile ducts [4]–[7]. Even small targeting deviations can lead to incomplete ablation or severe complications. Consequently, robotic assistance has emerged as a promising solution for reducing operator variability and enabling consistent, high-precision workflows [3], [7], [8].

Despite these advantages, the absolute positioning accuracy of current robotic systems is still limited by discrepancies between the nominal kinematic model provided by the manufacturer and the actual physical manipulator. These discrepancies, caused by manufacturing tolerances, assembly errors, joint compliance, and encoder offsets, typically result in millimeter-scale Cartesian positioning errors at the end-effector [9], [10].

Kinematic calibration is the established method for compensating for such geometric inaccuracies in serial manipulators. However, existing approaches often face a fundamental trade-off between model fidelity and robustness. High-dimensional models suffer from parameter coupling and overfitting, while simplified models may miss critical error sources. Moreover, most studies emphasize in-sample fitting accuracy with limited attention to parameter stability, generalization performance, or the incorporation of physical constraints [9], [11], [12].

To address these limitations, this paper proposes a robust kinematic calibration framework for a 6-DoF serial manipulator tailored for percutaneous needle guidance. Rather than treating parameter bounds merely as numerical safeguards, our approach utilizes physically motivated bounds as an explicit regularization mechanism to ensure parameter stability and predictive accuracy.

The specific contributions of this work are:

  • A 19-parameter geometric identification model that captures the dominant error sources of the manipulator.

  • A calibration pipeline combining observability-optimized configuration sampling, Kabsch-based tracker-to-base initialization, and bound-constrained Trust-Region Reflective (TRF) optimization.

  • Comprehensive benchtop validation utilizing a 10-fold cross-validation methodology, simulation, and physical hardware experiments that demonstrates robust sub-millimeter positioning accuracy. This establishes the fundamental kinematic precision required prior to future phantom and ex vivo studies.

The remainder of this paper is organized as follows: Section II reviews related work. Section III details the kinematic modeling and the proposed calibration methodology. Section IV describes the experiments and results, and Section V concludes the paper.