Manuscript Offprint

arXiv:2610.41981

Design and Validation of a Robotic Knee Exoskeleton with Musculoskeletal Simulation-Derived Biological Torque Control

Rajiv V. Joshi, Jimin An, Ilseung Park, Dawit Lee, Inseung Kang

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Rajiv V. Joshi¹, Jimin An¹, Ilseung Park¹, Dawit Lee², and Inseung Kang¹

Abstract—Data-driven controllers have enabled lower-limb robotic exoskeletons to adapt assistance to the user across many tasks. A leading example is biological torque control, which estimates the user's own joint moment from onboard sensors and assists in proportion to that estimate. However, its estimators are trained on data recorded with the target device, so deploying the controller on new hardware requires collecting new data. Existing open-source efforts release hardware, software, or controllers separately, so this data collection must still be repeated on the assembled device. Here, we present an open-source knee exoskeleton, built from off-the-shelf components and widely available fabrication tools, with a biological torque controller derived entirely from simulation. A muscle-driven imitation policy reproduces squatting motions from an open biomechanics dataset, and the resulting knee moments are distilled into an estimator that uses only onboard sensors, so no training data are recorded on the device. In seven able-bodied participants, assistance reduced peak biological knee extension moment by 21.5% and peak vastus medialis activation by 26.7% during squatting, relative to squatting without the exoskeleton (p<0.05). Our results demonstrate that a biological torque controller derived from simulation can transfer to physical hardware without device-specific data collection. Because the hardware and training pipeline are released together, research groups can train and compare knee exoskeleton controllers on shared hardware without additional data collection overhead.

I. INTRODUCTION

Wearable robotic exoskeletons have emerged as a promising intervention for people with motor impairments [1]–[4]. The knee is a particularly important target joint, bearing the majority of the load during high-demand activities such as squatting and stair negotiation [5], [6]. Over the past decade, exoskeleton control has advanced rapidly, as data-driven methods now estimate gait phase and joint moments in real time from wearable sensors [7]–[10], allowing assistance to adapt to the user's movement across tasks. Among these data-driven controllers, biological torque control is a leading approach, in which a neural network estimates the user's joint moment from onboard sensors and the exoskeleton applies a scaled, delayed copy of this estimate [8]. Because assistance follows the user's own effort rather than a predefined profile, the approach has reduced metabolic cost during level and inclined walking [8] and has been extended to unstructured locomotor tasks with a single controller [9].

However, this generality depends on device-specific data. The estimators are trained on large datasets recorded with the target device [9] and tend to overfit to that device's sensing configuration [11]. Collecting and labeling such data is time consuming, requires expensive laboratory equipment, and must be repeated for each new device [12]. Because these controllers are typically developed and validated on custom hardware built by the same research group [13]–[17], each controller also remains coupled to the device on which it was developed, which limits its reproducibility across research groups.

To lower this barrier, open-source efforts have released individual components of the exoskeleton system, ranging from modified commercial orthoses and mechanical designs to embedded software and trained controllers [9], [18]–[20]. These components are released separately and were not designed to operate together, so new data must still be collected on the assembled device before a data-driven controller can be deployed. In parallel, simulation and data reuse offer a path to reduce this data burden, including reinforcement learning in musculoskeletal simulation, adaptation of existing biomechanics datasets, and device-independent kinematic representations [11], [12], [21], [22]. However, a key limitation remains. No current exoskeleton approach has paired open-source hardware on which other groups could reproduce the full pipeline, from hardware to a simulation deployed controller.

We address these limitations by deriving a knee biological torque controller entirely from simulation and deploying it on an exoskeleton designed for accessible fabrication. Reference squatting motions from an open biomechanics dataset are reproduced by a muscle-driven imitation policy, and the knee moments generated in simulation are distilled into an estimator that uses only the exoskeleton's onboard sensors. Because no training data are recorded on the device, the same pipeline can be applied to other builds of the hardware. The exoskeleton is built from off-the-shelf components and parts fabricated with widely available tools, which minimizes the manufacturing overhead for other groups. During development, this construction also allowed the physical interface to be refined alongside the controller during development.

In this work, we present the following: 1) a knee exoskeleton, including its mechanical design, electronics, embedded firmware, and a two-tier control architecture; 2) a biological torque controller trained on an open biomechanics dataset through muscle-driven motion imitation and distilled to onboard sensors, requiring no data recorded on the device; and 3) an experimental valida-

This work has been submitted to the IEEE for possible publication. Copyright may be transferred without notice, after which this version may no longer be accessible.

¹R. Joshi, J. An, I. Park, and I. Kang are with the Department of Mechanical Engineering, Carnegie Mellon University, Pittsburgh, PA, 15213 USA (e-mail: rvj@andrew.cmu.edu). ²D. Lee is with the Department of Bioengineering, Stanford University, Stanford, CA, 94305

Figure

Fig. 1. Overview of the robotic knee exoskeleton design. (a) Sagittal-plane view of the unilateral device, with straps, hooks, and cuffs at the thigh, shank, and ankle. (b) A participant wearing the bilateral device, shown left to right in sagittal standing, sagittal squatting, and frontal standing. (c) The high-level control board built around the embedded computer (top) and the low-level control board built around the microcontroller (bottom).

TABLE I DEVICE SPECIFICATIONS

Mechanical
Total mass, bilateral* (kg)3.5
Unilateral mass (kg)1.3
Actuator mass (kg)0.49
Rated / peak torque** (Nm)9 / 22
Range of motion (deg)0–120
Compute/Electrical
Inertial sensor communicationSPI
Motor communicationCAN
Low-to-High level communicationCANFD
Low-level loop execution (µs)~80
High/low-level round trip (ms)<0.5
Estimated run time (h)~2

*Includes the battery and embedded electronics.

**The stock motor driver limits output to 18 Nm per the manufacturer's safety constraint.

tion of this controller in seven participants during squatting. We hypothesized that the controller would reduce knee extensor effort, quantified by peak biological knee extension moment and vastus medialis activation, relative to squatting without the exoskeleton. All hardware designs, code, and training pipelines are publicly available at: https://github.com/MetaMobilityLabCMU/Knee-Exoskeleton-Project.

References (39)

  • The exoskeleton expansion: improving walking and running economy
  • State of the art and future directions for lower limb robotic exoskeletons
  • Lower extremity exoskeletons and active orthoses: Challenges and state-of-the-art
  • Opportunities and challenges in the development of exoskeletons for locomotor assistance
  • The biomechanical functional demand placed on knee and hip muscles of older adults during stair ascent and descent
  • Knee biomechanics of the dynamic squat exercise
  • Real-time gait phase estimation for robotic hip exoskeleton control during multimodal locomotion
  • Estimating human joint moments unifies exoskeleton control, reducing user effort
  • Task-agnostic exoskeleton control via biological joint moment estimation
  • Optimizing locomotor task sets for training a biological joint moment estimator
  • A Unified Kinematic Representation Enables Reusable Biological Joint Moment Estimation
  • Deep domain adaptation eliminates costly data required for task-agnostic wearable robotic control
  • Wearable knee assistive devices for kneeling tasks in construction
  • Controlling negative and positive power for efficiency enhancement and muscle strain mitigation during squatting with a portable knee exoskeleton
  • Modeling and stiffness-based continuous torque control of lightweight quasi-direct-drive knee exoskeletons for versatile walking assistance
  • The only way is up: Active knee exoskeleton reduces muscular effort in quadriceps during weighted stair ascent
  • A versatile knee exoskeleton mitigates quadriceps fatigue in lifting, lowering, and carrying tasks
  • Enhancing voluntary motion with modular, backdrivable, powered hip and knee orthoses
  • AI-driven universal lower-limb exoskeleton system for community ambulation
  • Epically powerful: An open-source software and mechatronics infrastructure for wearable robotic systems (opens arXiv in a new tab)
  • Learning hip exoskeleton control policy via predictive neuromusculoskeletal simulation (opens arXiv in a new tab)
  • Musculoskeletal motion imitation for learning personalized exoskeleton control policy in impaired gait (opens arXiv in a new tab)
  • 3D printing settings impacting part strength
  • Design and backdrivability modeling of a portable high torque robotic knee prosthesis with intrinsic compliance for agile activities
  • Design and validation of a modular, backdrivable ankle exoskeleton
  • A human lower-limb biomechanics and wearable sensors dataset during cyclic and non-cyclic activities
  • SCONE: Open source software for predictive simulation of biological motion
  • The Hyfydy simulation software
  • DeepMimic: Example-guided deep reinforcement learning of physics-based character skills
  • Soft actor-critic: Off-policy maximum entropy deep reinforcement learning with a stochastic actor
  • Temporal convolutional networks: A unified approach to action segmentation
  • OpenSim: Open-source software to create and analyze dynamic simulations of movement
  • Muscle coordination retraining inspired by musculoskeletal simulations reduces knee contact force
  • A single-source dual-drive flexible knee assistive exoskeleton for elderly daily locomotion
  • Knee exoskeleton reduces muscle effort and improves balance during sit-to-stand transitions after stroke: A case study
  • Physical interface dynamics alter how robotic exosuits augment human movement: implications for optimizing wearable assistive devices
  • Design and clinical implementation of an open-source bionic leg
  • OpenExo: An open-source modular exoskeleton to augment human function
  • A comprehensive, open-source dataset of lower limb biomechanics in multiple conditions of stairs, ramps, and level-ground ambulation and transitions