EUReCA: End-to-End Unified CBCT Reconstruction Across Clinical Acquisitions

Jiening Zhu, Chengzhu Zhang, Jason Fan, LiCheng Kuo, Weixing Cai, Xiuxiu He, Laura Cervino, Jean Moran, Xiang Li, Tianfang Li, Yabo Fu

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Abstract

Cone-beam computed tomography (CBCT) is widely used for image-guided radiotherapy, but reducing the number of projections to lower dose and scan time makes reconstruction increasingly ill-posed. We present EUReCA (End-to-end Unified REconstruction across Clinical Acquisitions), a geometry-aware feed-forward model for variable-view clinical linac CBCT. For a given acquisition trajectory, EUReCA accepts 8–100 projections with a single set of weights, without view-count-specific retraining; the same architecture and training recipe extend unchanged to other clinical trajectories. It combines learned projection features with an in-frame Feldkamp–Davis–Kress (FDK) reconstruction. The model encodes each projection with a shared multiscale backbone, back-projects features into multiresolution 3D grids using the acquisition geometry, aggregates variable numbers of views at fixed computational cost, and refines the volume with a global 3D transformer and multilevel decoder. On simulated data, EUReCA achieved 29.9–33.2 dB PSNR with 10–100 views and outperformed all applicable feed-forward baselines; it also exceeded iterative optimization-based methods while running orders of magnitude faster. On 15 measured patient CBCT scans evaluated against vendor reconstructions, EUReCA achieved the best performance at every tested view count. These results suggest that geometry-aware, measurement-anchored feed-forward reconstruction can provide a practical and flexible foundation for variable-view linac CBCT.

Keywords: cone-beam computed tomography, sparse-view reconstruction, deep learning, image-guided radiotherapy.

1 Introduction

Cone-beam computed tomography (CBCT) provides three-dimensional patient anatomy for image-guided radiotherapy. Conventional linac-based CBCT acquires hundreds of projections over a full or partial gantry rotation. Reducing the number of projections can lower imaging dose and shorten acquisition time, but it also makes reconstruction increasingly ill-posed: the problem shifts from well constrained under dense angular sampling to severely underdetermined under sparse or incomplete sampling, producing structured streaks and missing-wedge artifacts in conventional reconstructions. A clinically useful reconstruction method must therefore remain reliable across a wide range of sampling conditions rather than at a single fixed operating point.

Measurement-driven methods reconstruct each volume primarily from its acquired projections. Feldkamp–Davis–Kress (FDK) reconstruction [1] is fast and deterministic but degrades substantially under angular undersampling. Algebraic and regularized iterative methods [2, 3] combine

*This work was supported in part by the National Cancer Institute of the National Institutes of Health under Award P30 CA008748. (Corresponding author: Yabo Fu.)

†The authors are with the Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY 10065 USA (e-mail: fuy3@mskcc.org).

projection fidelity with hand-crafted priors, but require per-scan optimization and careful parameter tuning. Neural fields and Gaussian representations, including NAF [4], IntraTomo [5], SAX-NeRF [6], and R2-Gaussian [7], instead fit a parameterized volume to each acquisition, with the representation itself acting as an implicit regularizer. These approaches remain closely anchored to the measurements, but they typically require minutes to hours per volume, and their constraints weaken as the number of views decreases.

Population-trained methods learn anatomical priors across datasets and apply them through either iterative inference or direct feed-forward reconstruction. Diffusion methods [8, 9] follow the first strategy, combining a learned prior with the acquired projections through iterative posterior sampling. In principle, they can represent multiple plausible reconstructions, although whether this variability corresponds to clinically meaningful uncertainty remains insufficiently validated. Their sampling procedures also require many network evaluations, and memory constraints led early CBCT implementations to slice-wise priors [10]; recent samplers enforce consistency across slices and reconstruct full volumes [11, 12], but per-volume reconstruction still takes far longer than a single feed-forward pass.

The second strategy amortizes reconstruction in a population-trained network, typically producing a volume without per-scan optimization. DIF-Net [13] estimates attenuation at arbitrary 3D points from pixel-aligned multiview features, while geometry-aware attenuation learning [14] explicitly back-projects 2D features into a voxel grid. C2RV [15] extends this idea with multi-scale volumetric features and cross-view attention. Later methods strengthen global 3D reasoning through more structured representations: DIF-Gaussian [16] models spatial feature distributions with 3D Gaussians and adds test-time refinement; X-LRM [17] uses a multiview transformer to construct an implicit X-triplane radiodensity field; X-GRM [18] predicts voxel-anchored Gaussians that support both CT extraction and differentiable projection rendering; and ILV [19] repeatedly updates an explicit latent volume with multiview evidence before Gaussian decoding and volumetric refinement.

Despite this progress, existing feed-forward methods have not demonstrated a single framework that spans the broad range of sampling conditions encountered in linac CBCT. Many architectures are tied to a fixed number of views or rely on attention mechanisms whose cost grows rapidly with projection count and resolution. Under sparse sampling, learned anatomical priors may also produce plausible but measurement unsupported structures. In addition, most methods are developed and evaluated using projections simulated under simplified acquisition trajectories rather than clinically realistic linac CBCT geometries. Beyond reducing the number of views in a routine half-scan, clinically constrained imaging may require qualitatively different sparse acquisitions, such as two orthogonal views, four fixed gantry views, or short arcs when scan time, dose, or clearance constraints limit angular coverage. Comparisons across mismatched geometries, view counts, datasets, and metric implementations provide limited evidence of relative performance [20], and transfer to measured clinical projections remains insufficiently studied.

We address these limitations with EUReCA (End-to-end Unified REconstruction across Clinical Acquisitions), a geometry-conditioned feed-forward model for variable-view linac CBCT. EU-ReCA encodes individual projections, back-projects multiscale features into volumetric grids, and aggregates them into a fixed-size representation for global 3D refinement. A multiscale decoder then combines learned projection features with an FDK reconstruction derived from the same measurements. Our contributions are:

  1. Scalable geometry-aware reconstruction. We introduce an end-to-end model that explicitly incorporates acquisition geometry while efficiently supporting variable view counts and trajectories.

  2. Adaptive FDK-reference fusion. We incorporate a full 3D FDK reconstruction from the same projections as an analytic reference, using voxel-wise gates to adapt its contribution at each decoder scale.

  3. Expanded range and acquisition flexibility. For a given trajectory, one set of weights spans ultra-sparse to well-sampled regimes (8–100 views). The same architecture and training recipe, trained once per trajectory, also cover two-view, four-view, and short-arc acquisitions without architectural changes.

  4. Controlled simulated and clinical evaluation. We compare feasible methods under matched data, geometries, splits, and metrics, and evaluate them on measured patient projections against corresponding clinical reference reconstructions.

Code, including the DRR projector (Apache-2.0), is available at https://github.com/Jiening666/EUReCA, and the trained weights of all reported models (CC BY-NC 4.0) at https://huggingface.co/jzhu35/EUReCA. An interactive visualization of the reconstructions is available at https://jiening666.github.io/cbct_recon/. The internal prostate cohort can be shared on request.