arXiv:2610.39879cs.ROOct 2026

Manuscript Offprint

arXiv:2610.39879

Compressed Orientation Reachability Maps for Sub-Second Robotic Needle Entry Point Planning

Prabhu Nithin Gollapudi, Andreas Maier, Franziska Mathis-Ullrich, Water Zhao, Jia Wang, Siming Bayer

Prabhu Nithin Gollapudi*, Andreas Maier*, Franziska Mathis-Ullrich*, Water Zhao†, Jia Wang†, and Siming Bayer*

  • Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany

Email: {prabhu.nithin.gollapudi, andreas.maier, franziska.mathis-ullrich, siming.bayer}@fau.de

†Precision Medtech Co., Ltd, Beijing, China

Email: {water.zhao, jia.wang}@precisionmedtech.com

Abstract—In the context of multi-stage trajectory planning for robotic percutaneous liver interventions, identifying a kinematically feasible entry point remains a critical computational bottleneck. Existing methods rely on exhaustive inverse kinematics or randomized searches, incurring delays of 13 to 27 seconds per target and limiting real-time intraoperative adaptation. This paper introduces an Orientation-Based Reachability Map (ORM) that precomputes the kinematic feasibility of a 6 Degrees of Freedom (DoF) serial manipulator across 188,873 workspace positions and encodes orientation feasibility into a compact, indexed 83MB map. A Just-In-Time (JIT) compiled runtime pipeline replaces live optimization with near-instantaneous array lookups. Experimental evaluation demonstrates a reduction in planning latency from 13.5–26.9 s to 0.289–0.349 s per target while maintaining high geometric fidelity, achieving a mean convex hull volume overlap ratio of 0.933 and a mean alpha-shape volume ratio of 0.976 relative to the baseline reachable point cloud. The proposed ORM framework enables sub-second entry point planning and provides an efficient computational framework for robotic needle insertion in percutaneous liver ablation.

Index Terms—Medical Robotics, Needle Placement, Reachability Maps, Robotic Surgery.

I. INTRODUCTION

Liver cancer remains one of the leading causes of cancer-related mortality worldwide, with approximately 865,000 new cases and 757,948 deaths reported in 2022 [1], and a 5-year survival rate of only 22% [2]. For patients with underlying liver disease who are not eligible for surgical resection, percutaneous thermal ablation techniques such as radiofrequency ablation (RFA) and microwave ablation (MWA) are established first-line therapies for small hepatocellular carcinoma (HCC) and selected liver metastases [3]–[5]. These minimally invasive procedures offer reduced morbidity, shorter hospital stays, and tissue preservation, but their effectiveness depends critically on accurate needle placement to ensure complete tumor coverage with an adequate safety margin (typically 5–10 mm) while avoiding major vessels and bile ducts [5]–[8].

Robotic assistance offers clear advantages over freehand techniques by reducing tremor, operator variability, and radiation exposure [4], [6]–[8]. In image-guided ablation, treatment success depends on a multi-stage planning pipeline that includes tumor targeting, ablation volume simulation, and needle trajectory selection [9], [10]. Within this pipeline, identifying a kinematically feasible skin entry point is a necessary prerequisite, since even a clinically optimal anatomical trajectory cannot be executed if the manipulator cannot reach it [9], [11].

However, existing intraoperative entry point planning methods often rely on exhaustive inverse kinematics (IK) verification or randomized search to evaluate manipulator reachability over large candidate sets. This leads to latencies of 13–27 s per target [4], [10], which is slow for responsive intraoperative re-planning under respiratory motion and soft-tissue deformation.

To address this limitation, we propose an Orientation-Based Reachability Map (ORM) framework for fast kinematic feasibility evaluation in robotic needle guidance. The specific contributions of this paper are:

• An ORM representation that precomputes reachability for a 6 Degrees of Freedom (DoF) serial manipulator over 188,873 workspace positions, storing orientation feasibility in a compact 83 MB indexed structure.

• A Just-In-Time (JIT) compiled query pipeline that replaces repeated online inverse kinematics (IK) evaluations with indexed lookups.

• An experimental evaluation demonstrating a reduction in entry point planning latency from 13.5–26.9 s to 0.289–0.349 s per target (a 60–90× speedup), while preserving high geometric fidelity with a mean alpha-shape volume ratio of 0.976 and a mean convex hull volume overlap ratio of 0.933.

These results support the use of the proposed ORM as a practical foundation for sub-second, workspace-aware entry point planning in robotic percutaneous liver ablation.

The remainder of this paper is organized as follows. Section II reviews related work in reachability analysis and robotic surgery. Section III details the proposed ORM generation and runtime pipeline. Section IV presents the experimental setup and results. Finally, Section V concludes the paper and discusses future work.

References (22)