arXiv:2610.37081cs.AIOct 2026Cited by 1 paper on PaperVantage

Manuscript Offprint

arXiv:2610.37081

OpenGray: a benchmark for evaluating the decisions of radiotherapy planning agents

Humza Nusrat, Kundan Thind

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Humza Nusrat 1,2,∗^{1,2,*} and Kundan Thind 1,2^{1,2}

1^{1}Department of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA

2^{2}Department of Radiology, College of Human Medicine,

Michigan State University, East Lansing, Michigan, USA

(Dated: October 26, 2026)

BACKGROUND AND PURPOSE: Radiotherapy planning agents based on large language models (LLMs) decide when to submit a plan or request review but are evaluated mainly by dose. This study developed OpenGray, a benchmark of planning tasks with known correct decisions.

MATERIALS AND METHODS: OpenGray adds tasks with known answers to the OpenKBP-Opt head-and-neck dataset, including two contradictions with dose goals proven incompatible and a feasible control with a stored plan meeting every hard goal. Four systems, from a rule-based planning script to a full LLM agent, were evaluated on 30 patients, the LLM-based ones with a commercial (Terra) and a local (Ministral 3 8B) LLM. Decisions, achieved dose, problem recognition, and technical failures were recorded and also read from logs of 39 stereotactic radiosurgery runs of SAGE, a clinical LLM agent.

RESULTS: With Terra, the script preceded by an LLM review and the full agent had decision accuracies of 95.4% and 92.1%, respectively, yet escalated only 50% (95% confidence interval, 25–75%) and 25% (10–40%) of overlap contradictions, which required arithmetic on the anatomy. All four systems submitted every feasible control, but 24% of these plans missed a hard goal that the stored plan had met. In four SAGE runs, one goal failed at every attempt; the agent identified it in every reply but had no action for stopping.

CONCLUSIONS: Favorable aggregate scores concealed specific decision failures. Planning agents should be evaluated on their decisions as well as their dose and given an explicit action for stopping and requesting review.

Keywords: Radiotherapy planning; Large language models; Benchmarking; Dose optimization; Quality assurance

1. INTRODUCTION

Automated radiotherapy planning methods have typically been evaluated by dosimetric quality, planning time, physician preference, and clinical acceptability [1–3]. More recently, planning agents based on large language models (LLMs) have been made responsible for interpreting the planning request, selecting optimization objectives, and deciding when a plan is complete [4–7], yet they have been evaluated in much the same way, by comparing the dose of their plans with that of clinical or reference plans. However, as these agents take on more of the planning process, they also make decisions that the final dose cannot reveal, such as whether a request can be met at all, whether to continue optimizing, and whether to submit a plan or request review by a physicist.

These decisions matter most when a request is flawed, for example when a required structure is missing, a physician note contradicts the formal planning goals, or two dose goals cannot be met together. A well-designed agent would stop and refer such a request for review, referred to here as escalation, whereas a poorly designed agent may continue to optimize a request that cannot be met or submit a plan that misses its goals. Given that such a plan cannot show whether the request was impossible or the optimization inadequate, judging the decision behind it requires knowing the correct answer in advance.

Several related safeguards have been described: automated plan-checking software identifies structural and dosimetric errors [8], some planning agents validate their own parameter updates or include supervisory review [6, 9, 10], and mathematical methods detect incompatible constraints in inverse planning [11]. Evaluating the decisions themselves, however, requires tasks for which the correct decision is known, which the public OpenKBP and OpenKBP-Opt datasets make possible [12, 13], given that their dose-influence matrices allow the dose of any plan to be recomputed exactly and some conflicting requests to be proven impossible from the anatomy alone. Interactive benchmarks such as AgentClinic already evaluate the sequential decisions of agents in other areas of medicine [14].

Agents that have already planned clinical cases are difficult to rerun on a benchmark, given that they usually work within a commercial treatment planning system; however, their planning logs record each request, each proposed change to the objectives, and the resulting dose, from which outcomes like those of a benchmark can be read. The purpose of this work was to develop OpenGray, a benchmark of planning tasks with known correct decisions, to determine with it whether favorable aggregate scores conceal specific decision failures, and to apply the same outcomes to the planning logs of SAGE, an LLM planning agent previously evaluated in prostate [6] and stereotactic radiosurgery (SRS) [15] planning.

References (16)