IFC-SSL-Bench: Benchmarking Self-Supervised Learning for Annotation-Efficient Multispectral Imaging Flow Cytometry

Angelo Jovin Yamachui Sitcheu, Roxana-Teodora Fetescu, Thomas Lautenschlager, Nils Friederich, Linus Schilpp, Stephan Allgeier, Luca Deininger, Daniel Hübschmann, Ralf Mikut

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Angelo Jovin Yamachui Sitcheu 1,⊗^{1,\otimes}, Roxana-Teodora Fetescu 1^{1}, Thomas Lautenschlager 1^{1}, Nils Friederich 1,2^{1,2}, Linus Schilpp 1^{1}, Stephan Allgeier 1^{1}, Luca Deininger 1,3^{1,3}, Daniel Hübschmann 4,5,6,7,8^{4,5,6,7,8}, and Ralf Mikut 1^{1}

1^{1}Institute for Automation and Applied Informatics, Karlsruhe Institute of Technology, Eggenstein-Leopoldshafen, Germany

2^{2}Institute of Biological and Chemical Systems, Karlsruhe Institute of Technology, Eggenstein-Leopoldshafen, Germany

3^{3}Department of Biosystems Science and Engineering, ETH Zürich, Basel, Switzerland

4^{4}Heidelberg Institute for Stem Cell Technology and Experimental Medicine, HI-STEM gGmbH, Heidelberg, Germany

5^{5}Innovation and Service Unit for Bioinformatics and Precision Medicine, German Cancer Research Center, Heidelberg, Germany

6^{6}German Cancer Consortium, DKTK, Heidelberg, Germany

7^{7}Division of Translational Precision Medicine, Institute of Human Genetics, Heidelberg University, Heidelberg, Germany

8^{8}Computational Oncology Group, Molecular Precision Oncology Program, National Center for Tumor Diseases Heidelberg and German Cancer Research Center, Heidelberg, Germany

Imaging Flow Cytometry (IFC) generates large multispectral datasets, yet the need for expert annotations bottlenecks the development of supervised deep learning models for cellular image analysis. While Self-Supervised Learning (SSL) offers a solution, its application to IFC remains fragmented and lacks systematic benchmarking. To address this, we present IFC-SSL-Bench, a modular framework for evaluating SSL in IFC. We benchmark six SSL approaches across four backbones on two heterogeneous datasets. By assessing downstream classification performance across various annotation regimes of the training sets (1%–100%), we quantify annotation efficiency and identify optimal configurations for high-dimensional biological imaging. We demonstrate that SSL pretraining reduces annotation dependency. SSL matches the performance of fully supervised baselines trained on 100% of annotated data while using as little as 5% of annotations for morphological classification, with performance advantages in low data regimes under 5% of annotated data. Across both datasets, the contrastive approach MoCo-v2 paired with ConvNeXt backbones emerged as the top-performing configuration, with ConvNeXtV1–T performing best for immunological synapse classification and ConvNeXtV2–T for morphological classification. We identify a clear pattern in transfer dynamics: ImageNet initialization benefits morphological classification, consistent with its visual similarity to natural images, whereas pretraining from random initialization better captures the more domain-specific immunological synapse classification. SSL elevates class-specific F1-scores by over 35 percentage points for structurally complex classes at 1% of annotated data. We synthesize these empirical insights into the first unified set of guidelines for deploying self-supervised workflows in high-throughput IFC, establishing a benchmark and resource suite that lowers the annotation barrier for future work in IFC.

Imaging Flow Cytometry | Self-Supervised Learning | Annotation Efficiency | Cell Analysis | Deep Learning

Correspondence: angelo.sitcheu@kit.edu

Yamachui Sitcheu et al. | October 26, 2026

1-35

Introduction

Imaging Flow Cytometry (IFC) is a modern technology that combines the quantitative, high-throughput capabilities of Flow Cytometry (FC) with the spatial and morphological detail provided by microscopy [1]. In practice, this is achieved by hydrodynamically focusing cells into a single-file stream that passes through a laser-illuminated imaging zone. High-speed detectors, synchronized with the cell flow, capture multispectral images using combinations of brightfield, darkfield, phase-contrast, and fluorescence channels, to achieve sharp subcellular resolution at high velocities [2, 3]. By capturing multiple images of each cell as it passes through the imaging region, IFC produces high-dimensional image datasets that provide unprecedented single-cell depth [3]. This hybrid technique has advanced research in immunology, oncology, cell biology, and environmental science by allowing detailed examination of heterogeneous cell populations and subtle phenotypic variations with robust statistical confidence [2, 4–6].

Despite its powerful capabilities, IFC faces significant challenges that stem not only from the inherent biological heterogeneity of cell populations [7], but also from the technical constraints, and therefore hinders effective data analysis. The primary technical hurdle in IFC is the fundamental trade-off to set between throughput and image resolution. While increasing magnification or exposure time gradually improves spatial resolution and fine morphological detail, it inherently reduces acquisition speed. Conversely, high-throughput demands shorter exposure times, which results in reduced image quality with lower magnification and/or low Signal-to-noise ratio (SNR) [8]. In multichannel acquisitions, fluorescence signals are additionally subject to background and to spectral crosstalk between detectors, which can obscure weak markers and distort the measured intensity distributions [9]. The appearance of a given cell type therefore depends not only on its biology but also on how it was acquired, so that the morphological differences of interest are superimposed on variation of purely technical origin.

Beyond these acquisition challenges, the analysis of IFC data also faces various bottlenecks. Traditionally, IFC data analysis has relied on manual gating or software-assisted pipelines using handcrafted feature extraction. Tools such as CellProfiler [10] improved throughput by automating cell feature extraction of hundreds of morphological and intensity-based descriptors. However, these classical approaches still require extensive domain expertise to select and tune relevant features for specific biological assays, making the analysis labor-intensive and susceptible to operator subjectivity [11, 12]. This reliance on predefined features introduces variability and limits scalability, as manual or semi-automated gating is not only time-consuming but also prone to inconsistencies across different operators [2, 13].

With the advent of Deep Learning (DL), many of these challenges have been partially mitigated. DL models can automatically learn complex features from raw data, reducing the dependence on manual feature engineering. This transition from manual feature engineering to DL-based image analysis has also occurred in the field of IFC [14–16]. Several specialized pipelines have emerged to facilitate this transition, including PXPermute [17] for feature importance analysis, Deepometry [18] for multi-channel phenotyping, Deep Cytometry [19] for real-time sorting applications, and Small U-Net [20] for high-throughput segmentation.

Despite these successes, the models still require large amounts of accurately annotated data for supervised training [21, 22]. Given the high cost and expertise required to annotate images, obtaining sufficiently large and diverse annotated datasets remains a significant bottleneck, leading to scarce datasets, limiting generalization performance, and creating a need for annotation-efficient strategies [21]. To address this, Self-Supervised Learning (SSL) methods have emerged as a promising alternative [23, 24]. SSL leverages unannotated data to learn useful representations, reducing the need for costly manual annotations. SSL is a paradigm well suited to IFC, which naturally generates vast amounts of unannotated cellular images at high-throughput scales. The robust performance of SSL in processing high-dimensional data, most notably in natural language [25] and video analysis [26], has provided a strong impetus for its application to other fields such as transcriptomics [27, 28] and visual geo-localization [29]. In medical imaging, SSL has reached a high level of maturity: in digital histopathology, for example, the SSL method Barlow Twins (BT) was used to pretrain a model on whole slide images and demonstrated remarkable robustness to cohorts variations and tissue types [30]. Similarly, on widely used benchmarks like MedMNIST [31], SSL has been investigated to provide clear domain guidelines. Bundele et al. [32] benchmark eight SSL methods across eleven MedMNIST datasets, varying annotation fraction and initialization, and find that ImageNet initialization consistently benefits performance, especially at low annotation fractions. Boukdir similarly reports transferable SSL representations across diagnostic domains [33]. More recently in radiology with nnFoundation, self-supervised pretraining at scale has produced general purpose encoders that transfer across more than a hundred downstream tasks [34].

Cellular imaging has followed the same trajectory. On Cell Painting, Kraus et al. trained Masked Autoencoders (MAEs) on millions of images and recovered known relationships between genetic and chemical perturbations [35], while Kim et al. compared self-supervised objectives directly on more than 10,000 compounds and found that DINO outperformed MAE, contrastive learning and the engineered CellProfiler pipeline [36]. Self-distillation has since been scaled across assays and cell types: Moutakanni et al. trained DINOv2 on several million multichannel images from the Human Protein Atlas and Cell Painting, matching supervised performance on protein localization tasks with 1%

Yamachui Sitcheu et al. | IFC-SSL-Bench of the annotations [37], and Zedda et al. assembled 1.25 million single-cell images from 18 datasets to build a dedicated foundation model for red blood cell morphology [38].

However, these successes do not seamlessly translate to IFC, which presents a unique set of challenges. Current literature focuses on isolated methods, demonstrating, for example, that a specific method like MoCo-v2 [39] can be used for leukocyte differential analyses [40], or BT [41] for investigating immunological synapse formation and characterizing antibodies [42]. These studies fail to address whether these choices are optimal, leaving the comparative potential of other paradigms unexplored.

Collectively, these limitations underscore a fundamental gap in the principled evaluation of SSL approaches for IFC, raising the following key research question: How can SSL be effectively applied and optimized for multispectral IFC data?

This central inquiry is decomposed into two specific research questions (RQs):

• RQ1: Methodological and architectural effectiveness: Which state-of-the-art SSL methods and backbone architectures are most effective for learning reliable and discriminative embeddings from multispectral IFC data?

• RQ2: Annotation efficiency: How does SSL reduce annotation needs on downstream classification?

By addressing these questions, this work aims to provide a comprehensive benchmark for SSL in the IFC domain, offering insights into the optimal combination of algorithms, architectures, and preprocessing strategies for high-dimensional biological imaging. Our primary contributions are:

  1. A unified benchmarking pipeline for multispectral IFC: We introduce IFC-SSL-Bench, a standardized, modular, reproducible and extensible framework that unifies data processing, pretraining, and downstream evaluation to ensure fair comparisons across diverse SSL methods for IFC.

  2. Comprehensive empirical analysis: We systematically evaluate six state-of-the-art invariance-based SSL methods across four backbone architectures on two distinct SSL datasets, providing the first structured comparison of self-supervised representation learning efficacy in this domain, identifying superior performance regimes.

  3. Annotation-efficient learning: We deliver empirical evidence on the utility of SSL under annotation scarcity. By quantifying performance gains across varying fractions of annotated data, we provide insights in optimizing SSL in annotation-constrained scenarios.

  4. Actionable guidelines & open resource: We derive best-practice recommendations for models, SSL methods, and initialization strategies tailored to IFC data and release our full codebase under an open license at https://github.com/hciays/IFC-SSL-Bench to foster reproducibility and advance future research in SSL applications for IFC.

Yamachui Sitcheu et al. | IFC-SSL-Bench

References (51)