Manuscript Offprint
arXiv:2610.32468OPTARI: An open-source software framework for clinical translation of optoacoustic imaging
Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.
Abstract
Significance: Combined optoacoustic (OptA) and ultrasound imaging is a promising modality across a wide range of pathologies. Its use in clinical research, however, is limited by a lack of analysis software accessible to users without programming expertise. Existing tools are either script-based engineering frameworks or proprietary vendor applications.
Aim: We present OPTARI, an open-source graphical software application to facilitate large-scale clinical optoacoustic-ultrasound (OPUS) studies.
Approach: OPTARI builds on the napari image viewer and the PATATO toolkit for OptA data processing. It supports native vendor and open HDF5 data formats, aligns with the IPASC standard, and is distributed as a standalone executable. Features include region-of-interest annotation, reconstruction, chromophore unmixing, deep-learning-based tissue segmentation, and temporal and spectral analysis modules.
Results: OPTARI offers a combination of features for clinical study analysis that none of the compared tools provide. The automated tissue segmentation produced results comparable to expert labeling for muscle tissue, and batch processing reduced analysis time from minutes to under one second per scan. The tool is used in ongoing clinical OPUS studies.
Conclusions: OPTARI connects engineering frameworks with medical research and provides an open-source basis for reproducible and standardized analysis in clinical OptA imaging studies.
Keywords: optoacoustic imaging, photoacoustic imaging, clinical translation, standardization, open-source, image analysis.
*Corresponding author: Moritz Schillinger, moritz.schillinger@fau.de
1 Introduction
By combining the strengths of both optoacoustic (OptA) and ultrasound (US) imaging, dual-modal optoacoustic-ultrasound (OPUS) imaging has shown great potential across a wide range of clinical applications. While conventional US visualizes the mechanical and structural properties of tissue, OptA imaging provides functional and molecular information, such as blood oxygenation and chromophore concentrations. Since OptA and US share the same sensor geometry and acoustic detection hardware, dual-modal scanners can acquire both images simultaneously, generating inherently co-registered datasets with minimal acquisition overhead.
Extracting quantitative information from these datasets requires dedicated software for reconstruction, spectral unmixing, and analysis. OptA signal processing has advanced substantially in recent years. Yet, despite this engineering progress, clinical research studies rarely make use of the latest algorithms, largely because researchers outside of engineering lack a straightforward way to access them. Based on recent clinical OptA reviews and feedback from our clinical collaborators, we identified three challenges currently limiting clinical research in OptA on the software and data-processing side:
-
Data Standardization and Interoperability: Clinical OptA scans are typically saved in proprietary, manufacturer-specific formats. This fragmentation makes it difficult to migrate older datasets, share data between research centers, or perform custom downstream analysis. The International Photoacoustic Standardisation Consortium (IPASC) has proposed unified, open formats to store and share OptA imaging data. However, this has not yet seen widespread adoption in clinical studies
-
Workflow Subjectivity and Operator Variability: The quantitative analysis of OptA data involves tuning a long list of parameters. Even a standard clinical analysis requires choosing a reconstruction algorithm, defining unmixing spectra, choosing the right frame, and selecting regions of interest (ROIs) to extract chromophore concentrations. Many of these steps depend on subjective expert judgment. This manual approach introduces intra- and inter-operator variability. To make OPUS studies reproducible and comparable, it is important to move from user-dependent steps to objective workflows.
-
Technical Barriers and Parameter Abstraction: Large clinical studies need fast, intuitive analysis, but manually configuring complex reconstruction and unmixing parameters is impractical in this setting, so clinical users benefit from well-documented default settings that can be saved and shared easily. Many studies also do not need the full content of research scans, such as raw time series or long frame sequences, so analysis tools should abstract this complexity and focus on the relevant data and metrics.
Existing software falls into two groups, each with its own limitations. Proprietary vendor software, like iLabs (iThera Medical, Munich, Germany) or Imagio/SenoGram (Seno Medical Instruments, San Antonio, TX, USA), is typically locked to specific hardware or data formats, lacks capacity for custom extensions, and depends on the manufacturer for maintenance. In contrast, a small number of open-source frameworks exist for clinical OptA data, ranging from simulation-focused toolkits to script-based reconstruction and unmixing libraries with limited annotation support. However, none of these open-source tools combine a graphical interface with the structured, multi-patient workflow required for large-scale clinical studies.
To bridge this gap between engineering frameworks and clinical OPUS studies, we introduce OPTARI (OPToacoustic imaging toolkit based on napARI), an open-source and platform-agnostic software application . Rather than reimplementing existing functionality, OPTARI extends established open-source tools, like the Photoacoustic Tomography Analysis Toolkit (PATATO) for core
https://mo-sc.github.io/OPTARI data management and napari for the graphical interface, and combines them with an analysis framework optimized for a standard clinical OPUS study workflow. Deterministic, shareable configurations are included to reduce operator variability, while automated ROI placement and batch processing reduce the time required to process large datasets. OPTARI does not define standards or specific parameters itself. Instead, it lets clinical researchers adjust processing configurations, test them, and share them with collaborators, which may help the field to establish common defaults.
The next section defines the scope of OPTARI and provides an overview of its core features. Section 3 outlines the underlying software architecture and implementation details. Section 4 compares OPTARI to related tools, describes its use in clinical studies, and evaluates the automated analysis in terms of segmentation performance and processing time, followed by discussion and limitations.
2.1 System Requirements
We formulated six requirements that follow directly from the challenges identified in Section 1, summarized in Table 1: data standardization (R6), operator variability (R4), and technical barriers (R1, R2, R3, R5).
2.2 Feature Overview
The following sections break down the main features of OPTARI. Figures 1 and 2 show screenshots of the user interface (OPTARI v0.8.0), with numbered labels matching the specific features described in the text.
2.2.1 Viewing and Navigation
OPTARI’s interface uses a Scan Browser that makes navigating multi-scan studies fast and straightforward (1, Fig. 1). Sliders give users direct control over multi-dimensional image stacks, making it easy to scroll through wavelengths, frames, and unmixed tissue components (2, Fig. 1). A motion-scoring algorithm automatically selects the frame with the least motion in the sequence. The main viewer displays US, OptA, and annotation layers in a flexible Layer List (3, Fig. 1). Users can adjust opacity, contrast limits, and colormaps to overlay OptA data onto US images. Available metadata, such as timestamps and acquisition settings, are displayed alongside the image using IPASC-compliant tag names and units (4, Fig. 1).
Table 1 System requirements for OPTARI.
| Requirement | Description |
|---|---|
| R1 Easy Deployment | Installation and execution require no command-line interaction or script execution. Distribution as a standalone executable that runs without administrative privileges, so it can be deployed in restricted hospital IT environments. |
| R2 Graphical Interface | Provide a click-based graphical user interface (GUI) for intuitive navigation, paired with keyboard shortcuts to speed up repetitive workflows. |
| R3 Study Navigation | Support analysis across multiple subjects and scans, easy switching between scans, and fast scrolling through temporal and spectral layers. |
| R4 Objective Processing | Replace subjective parameter selection with deterministic procedures as much as reasonably possible, and allow easy sharing of parameters and configurations. |
| R5 Documentation | Provide feature documentation and step-by-step guides for common clinical study workflows. |
| R6 Interoperability | Read and write both vendor-native and open data formats, and keep the underlying code accessible and documented so that custom modules or data formats can be integrated. |
2.2.2 Annotation, ROI Management, and Automated Segmentation
For quantitative evaluation, users can draw or edit custom and predefined ROIs, like polygons and ellipses, directly on the image. OPTARI automatically extracts customizable statistical features from these regions, which can correspond to chromophore concentrations or derived parameters like tissue oxygenation. Data collection relies on a dual-table system: the Live Analysis table dynamically updates measurements whenever an ROI is moved, resized, or switched to another frame (5, Fig. 1), while the Saved Analysis table freezes selected data points into a permanent log (6, Fig. 1).
A deep-learning-based tissue Segmentation module can segment key tissues for fully automatic ROI placement (7, Fig. 2), with pretrained models provided for common examination sites in clinical OPUS studies. Users can also save custom ROI presets to an ROI Library, backed by a JSON file, making it easy to share ROI configurations across team members or workstations (8, Fig. 2). These presets scale to each scan's field of view and can be placed at their saved coordinates or anchored automatically onto a matching tissue segmentation class. Applying a preset across multiple frames groups the ROIs into a single track (a set of ROIs linked as one region across frames), where each frame is anchored independently on its own segmentation mask, so the track follows the target tissue.

Fig 1 Main viewer window of OPTARI. (1) Scan Browser for study navigation. (2) Sliders for temporal and spectral scrolling. (3) Layer List, showing image data and annotations. (4) Metadata viewer. (5) Live Analysis table, showing dynamic updates of ROI data. (6) Saved Analysis table to collect data across scans.
2.2.3 Temporal and Spectral Analysis
OPTARI provides interactive plotting tools to help inspect temporal dynamics and spectral properties directly within the GUI for exploratory data analysis. The Temporal Analysis module uses timestamps from the scan metadata to plot intensities across groups of ROIs, enabling tracking of chromophore concentrations or tissue oxygenation over time, even in scans with motion artifacts (9, Fig. 2). The Spectral Analysis module displays signal intensities as a function of wavelength to show the spectral distribution (10, Fig. 2).
2.2.4 Data Processing and Conversion
In addition to the PATATO data loaders for HDF5 files and MSOT Acuity scans (iThera Medical GmbH, Munich, Germany), we implemented support for the newer MSOT Frontier scanner (iTheraNEXT Imaging GmbH, Munich, Germany).
OPTARI wraps PATATO’s processing features into GUI widgets so users can run Reconstruction (11, Fig. 2) and Unmixing (12, Fig. 2). Reconstruction settings include signal preprocessing steps such as band-pass filtering and impulse response correction. Results are automatically added to the viewer as interactive layers. In addition to the included backprojection (BP)-based reconstruction algorithms, we added an adapter for DeepMB, a state-of-the-art deep-learning model trained on data from MSOT Acuity scanners. Chromophore concentrations are calculated using linear spectral unmixing (LU), with reference spectra available for water ( ), oxy-genated hemoglobin (HbO₂), deoxygenated hemoglobin (Hb), indocyanine green (ICG), lipid, and melanin. These can also be used to calculate derived features like oxygen saturation (SO₂) or total hemoglobin (THb). Processing parameters can be loaded from and stored in JSON files to re-use or share them with collaborators.
2.2.5 Data Export
Exporting data in OPTARI automatically wraps processed scans, unmixed layers, drawn ROIs, and segmentation masks into HDF5 files. This provides an easy way to convert proprietary vendor data into open HDF5 files that include IPASC-compliant metadata. All extracted metrics and metadata from the Saved Analysis table can be exported with a single click into standard spreadsheet formats, which can also be re-imported to resume an analysis or combine results from several sessions or machines. Analytical plots like histograms or spectral curves can be exported directly as image files. Additionally, the current viewer contents can be saved as figures or compiled across frame sequences into video exports.
2.2.6 Batch Mode
Batch Mode runs a previously defined pipeline of reconstruction, unmixing, segmentation, and ROI placement across an entire dataset without manual interaction. Each step is tuned once on a representative scan and saved as a preset. A batch preset then references these presets and the full analysis plan is validated so that incompatibilities or missing requirements are flagged before the run starts. Individual scan failures are logged and skipped. The output is a measurement table with the same structure as the Saved Analysis table in the GUI, along with optional overlay images and HDF5 conversion.
3 Implementation and Architecture
This section describes how OPTARI is built, some of the external tools we integrated, and how it is deployed.
3.1 Software Architecture
OPTARI uses a Model-View-Controller (MVC)-inspired architecture to separate the graphical interface from the data processing logic. As shown in Fig. 3, the application is divided into four main layers: a UI Layer, a Control Layer, a Utilities Layer, and an External Layer. Rather than running as a standard napari plugin, OPTARI embeds the napari viewer directly into a custom Qt application. The UI Layer contains only graphical widgets that hold no internal logic. Instead, a stateless UIManager builds these widgets and wires UI signals to backend routines, while a separate ShortcutManager handles custom keyboard commands.
The core of the application is the OptariController. It manages global application state, like tracking open widgets, the active study, and the loaded scan, while delegating specific tasks to modular task controllers. For example, the ROI annotation subsystem uses a dedicated controller to monitor napari's shape layers. Whenever a user draws or resizes an ROI, the controller automatically recalculates tissue metrics and refreshes the live data tables. The SegController manages
https://doc.qt.io/qtforpython-6; v6.11.2

Fig 2 Selection of processing and analysis widgets in OPTARI. (7) Segmentation module to automatically place an ROI in the target tissue. (8) ROI Library for loading and saving ROI presets. (9) Temporal Analysis module to plot signal intensity changes over time. (10) Spectral Analysis module to visualize signal magnitude across wavelengths. (11) Reconstruction module for BP- and deep-learning-based image reconstruction. (12) Unmixing module for spectral unmixing of chromophore concentrations.
tissue segmentation by running inference via pre-trained deep learning models. Based on the generated masks, ROIs can be placed automatically in the target tissue and added to the analysis tables. We include model weights pre-trained or fine-tuned on US images from three examination sites common in clinical OPUS studies: the gastrocnemius muscle, the tibialis anterior muscle, and the vastus lateralis muscle. Model weights are automatically downloaded during their first execution to keep the base installation size small. New models can easily be plugged in via the built-in ONNX adapter. Other task controllers wrap or extend processing algorithms provided by external frameworks and automatically convert the external outputs into interactive napari layers. Long-running operations such as reconstruction, segmentation, and unmixing run in a background task slot so the event loop stays responsive while a scan processes. The same slot is also used by Batch Mode (Section 2.2.6).
For real-time data inspection, we used pyqtgraph to build analysis widgets docked below the
https://github.com/pyqtgraph/pyqtgraph; v0.14.0

Fig 3 Overview of the modular software architecture and selected components of OPTARI: UI Layer (dock widgets), Control Layer (central OptariController and specialized task modules), Utilities Layer, and External Layer.
viewer that can dynamically plot intensity distributions, wavelength spectra, or temporal signal changes in the ROIs.
3.2 External Integrations
OPTARI integrates several external tools to handle data I/O, image reconstruction, and quality control. For data management, a custom patato-bridge utility connects the software to an extended version of the PATATO framework (based on v0.7.0), which natively handles both proprietary vendor files (e.g., iThera .msot datasets) and standardized HDF5 datasets. Through this integration, users can also run PATATO's JAX-accelerated BP algorithm. We choose a PATATO-inspired HDF5 schema over the format provided directly by the IPASC, because the native IPASC format is scoped to raw time-series data and does not accommodate dual-modality ultrasound or processed/reconstructed layers. Nevertheless, OPTARI stores IPASC's consensus metadata tags alongside its own HDF5 schema. Native IPASC files can be imported and exported directly from OPTARI using PACFISH , which also performs metadata consistency checks.
Out of the increasing number of deep-learning-based OptA image reconstruction algorithms, none has been widely adopted in published clinical OPUS studies. Apart from general concerns over black-box predictions in OptA imaging, this is largely a problem of training on simulated data or non-standard scanner geometries. To our knowledge, DeepMB is currently the only deep-learning reconstruction framework trained specifically on clinical OptA data recorded with the widely deployed MSOT Acuity scanner and has been integrated as the standard reconstruction
https://github.com/IPASC/PACFISH; v0.4.4 software in the latest MSOT Frontier scanner. It achieved near-identical image quality to iterative model-based algorithms at an approximately 1,000-fold acceleration. Like the segmentation models, DeepMB runs on the ONNX Runtime within OPTARI. This balances performance with hardware compatibility, since it uses GPU acceleration where available and falls back to CPU execution if the required hardware or drivers are not present.
To reduce operator bias during frame selection, OPTARI incorporates a motion scoring algorithm that quantifies movement in sequences of US frames based on their structural similarity index and zero-mean normalized cross-correlation, adapted from the frame-selection approach of Kukačka et al. The software then automatically suggests the frame with the lowest motion score for subsequent analysis.
3.3 Deployment and Installation
To simplify deployment in hospital environments, OPTARI is packaged as a standalone desktop application using PyApp . PyApp wraps the project into a single launcher executable that sets up an isolated runtime on its first launch and requires no administrative rights or terminal commands. Because certain PATATO modules depend on native C/C++ code, we host and distribute precompiled PATATO binaries for Windows, macOS, and Linux (Ubuntu), removing any requirement for local compilers. We use an automated GitHub Actions CI/CD pipeline to handle unit testing, tag-based versioning, cross-platform release builds and documentation builds. OPTARI can be customized through a GUI dialog that covers viewer defaults, dock visibility, and ROI table display. The presets for reconstruction, segmentation, unmixing, and ROI placement are managed separately from their own dock. All settings and presets are stored as human-readable JSON files in the user's home directory.
4 Evaluation
We compared OPTARI to existing software, described how it was integrated into clinical study workflows, and evaluated the performance of its automated segmentation and ROI placement.
4.1 Feature Comparison
We included only tools that offer some form of graphical access to OptA data processing (Table 2). iLabs, iThera Medical’s proprietary clinical analysis software for OptA data, offers a mature interface with clinical documentation, but remains closed-source and locked to iThera’s own hardware and data formats. PATATO is currently the closest open-source tool to clinical utility, offering reconstruction and unmixing capabilities alongside minimal GUI and annotation features, but remains largely script-based and lacks a structured, multi-patient study workflow. The MITK platform offers an OptA plugin that integrates OptA data visualization but it misses important features like spectral unmixing or advanced clinical annotation tools, and only provides limited data format support. PATLAB offers a GUI-based reconstruction and simulation environment aimed at system and algorithm development, but lacks spectral unmixing, clinical annotation tools, a multi-patient workflow, and requires a MATLAB license. ImageJ/Fiji is a general-purpose, extensible image analysis platform and is included as a common fallback tool clinical researchers use in the absence of dedicated software. However, it does not offer any OptA-specific reconstruction, unmixing, or annotation features.
https://github.com/ofek/pyapp; v0.29.0
Table 2 Feature comparison of OPTARI against existing software tools used in clinical OptA imaging. Evaluation criteria correspond to the requirements (R1–R6) defined in Section 2. Assessment based on the software versions and documentation available in September 2026.
| Requirement / Feature | iLabs | PATATO | MITK-PAI | PATLAB | ImageJ/Fiji | OPTARI |
|---|---|---|---|---|---|---|
| R1: Easy Deployment | ||||||
| – Standalone executable | + | - | + | - | + | + |
| – No admin rights required | + | ○ | ○ | - | + | + |
| R2: Graphical Interface | ||||||
| – GUI-driven analysis | + | ○ | + | ○ | + | + |
| – Keyboard shortcuts | - | - | ○ | - | + | + |
| R3: Study Navigation | ||||||
| – Multi-subject workflow | + | - | - | - | - | + |
| – Temporal/spectral scrolling | + | ○ | ○ | ○ | + | + |
| R4: Objective Processing | ||||||
| – Shareable parameters | ○ | + | - | ○ | - | + |
| – Automated ROI placement | - | - | - | - | - | + |
| R5: Documentation | ||||||
| – Clinical OptA documentation | + | ○ | - | - | - | + |
| – Code documentation | - | + | + | + | + | + |
| R6: Interoperability | ||||||
| – Vendor & open data support | ○ | + | ○ | ○ | ○ | + |
| – Open-source codebase | - | + | + | + | + | + |
| – fully supported | ○ partially supported | – not supported | ||||
Beyond these, several other tools exist outside the scope of this comparison. SIMPA (Simulation and Image Processing for Photonics and Acoustics) is one of the most widely used frameworks for OptA simulation and covers the full pipeline from synthetic data generation to reconstruction. However, it does not offer a GUI, annotation, or clinical workflow features. PHANTOM is a MATLAB tool for ultrasound-guided fluence compensation in preclinical OptA imaging, not a general clinical analysis platform. MSOT-RAFT is a script-based reconstruction and filtering toolbox for preclinical OptA data with no graphical interface.
4.2 Integration into Clinical Studies
OPTARI was developed in close collaboration with clinical professionals experienced in OPUS imaging studies, including researchers from the Departments of Vascular Surgery and Pediatrics and Adolescent Medicine at University Hospital Erlangen, one of the leading centers for clinical OptA research. Requirements R1–R6 were derived from this collaboration, and the interface was revised iteratively in response to their feedback. This resulted in default workflows that closely align with clinical OPUS study procedures. Given that structured training has been shown to improve MS-OPUS measurement reproducibility, OPTARI provides user-friendly documentation that also includes step-by-step guides and training videos. The tool is actively being used in ongoing clinical studies, with corresponding manuscripts currently in preparation.
4.3 Performance of the Automated Analysis
The automated ROI placement implemented in OPTARI's segmentation module is based on a muscle segmentation approach that was previously developed and validated. In that work, a UNet-based convolutional neural network trained to delineate the target muscle on co-registered ultrasound achieved a Dice score of 0.86 and a 95th-percentile Hausdorff distance of 1.92 mm across 5-fold cross-validation, outperforming Segment Anything Model (SAM)- and MedSAM-based alternatives. Classification of peripheral artery disease (PAD) patients versus healthy volunteers based on the automatically generated ROIs was comparable to manual expert annotation, with an area under the receiver operating characteristic curve (AUROC) of 0.76 (automatic) vs. 0.77 (manual).
Subsequent work further validated this approach on a post-exercise oxygenation recovery dataset, comparing automatically and physician-annotated MS-OPUS-derived oxygen saturation parameters. Agreement between the two annotation methods was good to excellent across most summary parameters (e.g. mean recovery: intraclass correlation coefficient (ICC) 0.934 (95% CI [0.834, 0.967]), concordance correlation coefficient (CCC) 0.933 (95% CI [0.906, 0.953]), with near-zero bias. Classification performance for separating PAD patients from healthy volunteers was likewise comparable between methods (e.g. peak AUROC 0.966 (95% CI [0.896, 0.997]) automatic vs. 0.981 (95% CI [0.958, 1.000]) manual).
Figure 4 benchmarks the batch-mode pipeline on one representative scan (13 acquired wavelengths, each reconstructed into a px image), comparing two configurations at each step while holding the rest of the pipeline fixed. Timings were measured on a standard machine (Intel Core i7-1255U, 1.7 GHz, 16 GB RAM), rather than high-performance hardware. With the default configuration (HDF5-loaded data, BP reconstruction, two-chromophore LU, single-class segmentation, single-frame ROI extraction), the complete sequence took s per scan, of which reconstruction accounts for approximately 67%, whereas segmentation and automatic ROI placement together contributed only s. For comparison, manual delineation of a single complex ROI can take up to one minute, and a complete multi-class segmentation mask took min (10 timed annotations by an experienced operator). At study scale, the default configuration processes the dataset from (1,977 scans from 102 subjects) in roughly 31 min of unattended computation, excluding manual quality checks.
Together, these results suggest that automated segmentation is a viable alternative to manual ROI annotation in MS-OPUS-based PAD studies without a clear loss in group discrimination, while removing the operator-dependence of the annotation step and substantially reducing the required analysis time.
5 Discussion
To our knowledge, OPTARI is the only open-source tool that combines a graphical interface with the reconstruction, unmixing, and multi-patient workflow capabilities needed for large-scale clinical OPUS studies. Rather than being developed from scratch, OPTARI builds on existing open-source infrastructure such as napari and PATATO, adding a clinically oriented workflow and automated parameter selection.
OPTARI also aligns closely with the priorities identified in the IPASC community’s standardization roadmap. The consortium named automated ROI delineation as a priority for overcoming segmentation and interpretation challenges, which is directly addressed by OPTARI’s automated
(a1) Data Loading

HDF5 (13λ): ms
iThera (13λ): ms
(a2) Reconstruction

BP (13λ): ms
DeepMB (13λ): s
(a3) Unmixing

LU (2λ/2chro): ms
LU (9λ/4chro): ms
(b1) Ultrasound
(b2) Segmentation

(c) ROI Data Extraction

1 class: ms
9 classes: ms

1 frame: ms
26 frames: ms
Fig 4 Automated batch-mode analysis applied to one representative scan (13 wavelengths, ). Each panel reports two configuration choices for its step. Times are reported as mean standard deviation over 100 repetitions, measured on a standard laptop (Intel Core i7-1255U, 1.7 GHz, 16 GB RAM). HDF5 data are loaded lazily, so the read cost may fall into the following steps. The default configuration (HDF5 loading, BP, two-chromophore LU, single-class segmentation, single-frame extraction) took s in total.
segmentation module (R4). Additionally, the roadmap for clinical adoption called for easy-to-use and openly available software for reconstruction and analysis, as well as for educational resources, both reflected by OPTARI's GUI (R1, R2) and documentation (R5). Their roadmap for data management called for consolidating the IPASC data format across research groups, which OPTARI supports by reading and writing IPASC files and by including its metadata schema (R6).
Several limitations remain. OPTARI specifically targets clinical two-dimensional handheld OPUS data and it does not support three-dimensional/volumetric scans or data from preclinical systems. The GUI also lacks a true multi-viewer layout with synchronized annotations across layers, a limitation of napari's grid-view mode. Automated ROI placement performance is further limited by the similarity between target data and the training distribution, and may require fine-tuning for new anatomical sites or scanner types. In practice, automatically placed ROIs should be reviewed rather than accepted unchecked, which reduces but does not eliminate the time saved.
OPTARI is research software intended for the analysis of data from clinical studies and has not undergone conformity assessment as a medical device. It is not intended for diagnosis or treatment decisions. The tool remains under active development, and we welcome further community contributions.
Disclosures
V.N. is a founder and equity owner of iTheraNEXT Imaging GmbH, Maurus OY, sThesis GmbH, Spear UG, Biosense Innovations P.C. and I3 Inc. F.K. holds a patent on an OptA imaging system together with iThera Medical GmbH. All other authors declare that they have no financial interests, commercial affiliations, or other potential conflicts of interest that could have influenced the objectivity of this research or the writing of this paper.
Code, Data, and Materials Availability
The OPTARI source code and executables are available at https://github.com/Mo-Sc/OPTARI under the BSD-3-Clause license, with documentation at https://mo-sc.github.io/OPTARI. The analyses in this paper used version 0.8.0. The extended PATATO version used by OPTARI is available at https://github.com/Mo-Sc/patato. Pretrained segmentation models are downloaded automatically on first use. The DeepMB adapter requires model weights, which are available from the original authors upon reasonable request. The clinical PAD dataset used in Section 4.3 is available on Zenodo.
Acknowledgments
This research received funding from the Bavarian Ministry of Economic Affairs, Regional Development and Energy (StMWi) (DIK-2208-0014/ DIK0464/02, SENDERO, granted to K.B. and iThera Medical GmbH), from the Munich Institute of Robotics and Machine Intelligence (MIRMI) of the Technical University of Munich (TUM) (RoMSOT, granted to V.N.); from the Deutsche Forschungsgemeinschaft (DFG) - 455422993 as part of the Research Unit FOR 5298 (iMAGO, subproject TP2, GZ: NT 3/32-1, granted to V.N.); from the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation programme under grant agreement No 694968 (PREMSOT, granted to V.N.); and from the European Union's Horizon Europe research and innovation programme under Grant Agreement No. 101291160 (QMSOT, granted to V.N.).
LLM tools were used to assist with language refinement during manuscript preparation and to refactor portions of the OPTARI codebase. All content was reviewed and verified by the authors.
References
1 Jeongwoo Park, Seongwook Choi, Ferdinand Knieling ... Chulhong Kim, “Clinical translation of photoacoustic imaging,” Nature Reviews Bioengineering 3(3), 193–212 (2024).
2 Vasilis Ntziachristos, “Addressing unmet clinical need with optoacoustic imaging,” Nature Reviews Bioengineering 3(3), 182–184 (2024).
3 Paul Beard, “Biomedical photoacoustic imaging,” Interface Focus 1(4), 602–631 (2011).
4 Kaushik Basak Chowdhury, Maximilian Bader, Christoph Dehner ... Vasilis Ntziachristos, “Individual transducer impulse response characterization method to improve image quality of array-based handheld optoacoustic tomography,” Optics Letters 46(1), 1 (2021).
5 Jan-Hinrich Nölke, Tim J. Adler, Melanie Schellenberg ... Lena Maier-Hein, “Photoacoustic quantification of tissue oxygenation using conditional invertible neural networks,” IEEE Transactions on Medical Imaging 43(9), 3366–3376 (2024).
6 Maximilian Bader, Philipp Haim, Lukas Imanuel Scheel-Platz ... Dominik Jüstel, “A method for accurate spatial focusing simulation via numerical integration and its application in optoacoustic tomography.” arXiv:2409.10138 (2024).
7 Christoph Dehner, Ledia Lilaj, Vasilis Ntziachristos ...Dominik Jüstel, “Scale-equivariant deep model-based optoacoustic image reconstruction,” Photoacoustics 44, 100727 (2025).
8 Ferdinand Knieling, Serene Lee, and Vasilis Ntziachristos, “A primer on current status and future opportunities of clinical optoacoustic imaging,” npj Imaging 3(1), 4 (2025).
9 Janek Gröhl, Lina Hacker, Ben T. Cox ...Sarah E. Bohndiek, “The IPASC data format: A consensus data format for photoacoustic imaging,” Photoacoustics 26, 100339 (2022).
10 Hisham Assi, Rui Cao, Madhura Castelino ...Sarah E. Bohndiek, “A review of a strategic roadmapping exercise to advance clinical translation of photoacoustic imaging: From current barriers to future adoption,” Photoacoustics 32, 100539 (2023).
11 Janek Gröhl, Kris K. Dreher, Melanie Schellenberg ...Lena Maier-Hein, “SIMPA: an open-source toolkit for simulation and image processing for photonics and acoustics,” Journal of Biomedical Optics 27(08), 083010 (2022).
12 Thomas R. Else, Janek Gröhl, Lina Hacker ...Sarah E. Bohndiek, “PATATO: a Python photoacoustic tomography analysis toolkit,” Journal of Open Source Software 9(93), 5686 (2024).
13 Chi-Li Chiu, Nathan Clack, and the napari community, “napari: a Python multi-dimensional image viewer platform for the research community,” Microscopy and Microanalysis 28(S1), 1576–1577 (2022).
14 Ferdinand Knieling, Clemens Neufert, Arndt Hartmann ... Maximilian J. Waldner, “Multispectral optoacoustic tomography for assessment of Crohn’s disease activity,” New England Journal of Medicine 376(13), 1292–1294 (2017).
15 Adrian P. Regensburger, Markus Eckstein, Matthias Wetzl ...Ferdinand Knieling, “Multispectral optoacoustic tomography enables assessment of disease activity in paediatric inflammatory bowel disease,” Photoacoustics 35, 100578 (2024).
16 Erin I. Neuschler, Reni Butler, Catherine A. Young ...Basak E. Dogan, “A pivotal study of optoacoustic imaging to diagnose benign and malignant breast masses: A new evaluation tool for radiologists,” Radiology 287(2), 398–412 (2018).
17 Gisela L. G. Menezes, Ruud M. Pijnappel, Carla Meeuwis ...Ritse M. Mann, “Downgrading of breast masses suspicious for cancer by using optoacoustic breast imaging,” Radiology 288(2), 355–365 (2018).
18 Adrian P. Regensburger, Lina M. Fonteyne, Jörg Jüngert ...Ferdinand Knieling, “Detection of collagens by multispectral optoacoustic tomography as an imaging biomarker for Duchenne muscular dystrophy,” Nature Medicine 25(12), 1905–1915 (2019).
19 Adrian P. Regensburger, Alexandra L. Wagner, Vera Danko ...Ferdinand Knieling, “Multispectral optoacoustic tomography for non-invasive disease phenotyping in pediatric spinal muscular atrophy patients,” Photoacoustics 25, 100315 (2022).
20 Anna P. Träger, Josefine S. Günther, Roman Raming ...Ulrich Rother, “Hybrid ultrasound and single wavelength optoacoustic imaging reveals muscle degeneration in peripheral artery disease,” Photoacoustics 35, 100579 (2024).
21 Milenko Caranovic, Julius Kempf, Yi Li ...Ulrich Rother, “Derivation and validation of a non-invasive optoacoustic imaging biomarker for detection of patients with intermittent claudication,” Communications Medicine 5(1), 88 (2025).
22 Jan Kukačka, Maximilian Bader, Dominik Jüstel ...Vasilis Ntziachristos, “Motion score for spectral quality control of optoacoustic-ultrasound data,” Scientific Reports 16(1), 1325 (2026).
23 Christoph Dehner, Guillaume Zahnd, Vasilis Ntziachristos ... Dominik Jüstel, “A deep neural network for real-time optoacoustic image reconstruction with adjustable speed of sound,” Nature Machine Intelligence 5(10), 1130–1141 (2023).
24 Scott Prahl, "Assorted spectra." https://omlc.org/spectra/index.html (2018).
25 Marcel Knopp, Christoph J. Bender, Niklas Holzwarth ...Kris K. Dreher, “Shortcut learning leads to sex bias in deep learning models for photoacoustic tomography,” International Journal of Computer Assisted Radiology and Surgery 20(7), 1325–1333 (2025).
26 Thomas Kirchner, Janek Gröhl, Franz Sattler ...Lena Maier-Hein, “An open-source software platform for translational photoacoustic research and its application to motion-corrected blood oxygenation estimation.” arXiv:1901.09781 (2019).
27 P. Omidi, L.C.M. Yip, E. Rascevska ...J.J.L. Carson, “PATLAB: A graphical computational software package for photoacoustic computed tomography research,” Photoacoustics 28, 100404 (2022).
28 Johannes Schindelin, Ignacio Arganda-Carreras, Erwin Frise ...Albert Cardona, “Fiji: an open-source platform for biological-image analysis,” Nature Methods 9(7), 676–682 (2012).
29 Allison Sweeney, Aayush Arora, Skye A. Edwards ...Srivalleesha Mallidi, “Ultrasound-guided photoacoustic image annotation toolkit in MATLAB (PHANTOM) for preclinical applications,” Photoacoustics 41, 100662 (2025).
30 Devin O'Kelly, James Campbell, Jeni L. Gerberich ...Ralph P. Mason, "A scalable open-source MATLAB toolbox for reconstruction and analysis of multispectral optoacoustic tomography data," Scientific Reports 11(1), 19872 (2021).
31 Yi Li, Janek Gröhl, Briain Haney ... Ulrich Rother, “Teachability of multispectral optoacoustic tomography,” Journal of Biophotonics 17(7), e202400106 (2024).
32 Moritz Schillinger, Maja Schlereth, Gaspard Boraud ...Katharina Breininger, “Automatic muscle segmentation for the diagnosis of peripheral artery disease using multispectral optoacoustic tomography,” in Photons Plus Ultrasound: Imaging and Sensing 2025, Alexander A. Oraevsky and Lihong V. Wang, Eds., 13319, 133191K, International Society for Optics and Photonics, SPIE (2025).
33 Adrian Buehler, Moritz Schillinger, Angelos Karlas ... Briain Haney, “Beyond the vessel: Muscle tissue oxygenation kinetics as a biomarker in peripheral arterial disease.” manuscript under review (2026).
34 Adrian Buehler, Moritz Schillinger, Angelos Karlas ... Briain Haney, “Imaging dataset for: Beyond the vessel: Muscle tissue oxygenation kinetics as a biomarker in peripheral arterial disease.” Zenodo, doi:10.5281/zenodo.22044239 (2026).