A Transparent 3-DoF Assistive Device for Upper-Limb Tremor and Ataxia Compensation: Design and Experimental Validation

Thomas Dobosz, Veronika Hofmann, Ahmed Barhoma, Ann-Kathrin Henke, Johannes A. Gaus, Hans Mees, Winfried Ilg, Isabell Wochner, Martin Giese, Andreas Bulling, Syn Schmitt, Christophe Maufroy, Urs Schneider, Thomas Bauernhansl, Daniel Haeufle

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Thomas Dobosz ∗1,2^{*1,2} and Veronika Hofmann ∗1,2^{*1,2}, Ahmed Barhoma 1^{1}, Ann-Kathrin Henke 1^{1}, Johannes A. Gaus 3,6^{3,6}, Hans Mees 3^{3}, Winfried Ilg 3,6^{3,6}, Isabell Wochner 3,6^{3,6}, Martin Giese 3,6^{3,6}, Andreas Bulling 4,6^{4,6}, Syn Schmitt 5,6^{5,6}, Christophe Maufroy 1,2^{1,2}, Urs Schneider 1,2,6^{1,2,6}, Thomas Bauernhansl 1,6^{1,6}, and Daniel Haeufle 3,6^{3,6}

Abstract—Assistive devices for patients with neurodegenerative movement disorders must combine high mechanical transparency with the ability to generate corrective interaction forces. Achieving both properties simultaneously remains challenging. This paper presents the completed iAssistADL platform, a 3-DoF assistive device for future tremor and ataxia compensation during activities of daily living. The system employs a transparency-oriented mechanical architecture with all actuators located at the base, resulting in low moving mass, decoupled kinematics, and high mechanical transparency. To compensate for friction-induced transparency limitations along the vertical axis, a hybrid control strategy is proposed that combines open-loop force rendering in the horizontal plane with closed-loop force control in the vertical direction. The platform is experimentally evaluated through transparency measurements and static and dynamic force-rendering experiments at representative configurations. Results demonstrate resistive torques below 0.1Nm in the horizontal plane, end-effector forces of up to approximately 10N throughout the workspace, and accurate horizontal force rendering up to 4 Hz. While the vertical axis remains limited by friction-induced dynamics at higher frequencies, the proposed system successfully combines high transparency with controlled force generation, establishing a promising hardware platform for future tremor and ataxia compensation.

I. INTRODUCTION

Activities of daily living (ADLs) such as drinking or eating are essential for independent living and quality of life. In patients affected by neurodegenerative movement disorders, including Parkinson's disease or cerebellar ataxia, these activities are often impaired by pathological movement components such as tremor or dysmetric over- or undershooting. While voluntary movement generation is largely preserved, involuntary motions interfere with task execution [1], [2], [3], [4], [5].

In contrast to stroke rehabilitation devices, assistive technologies for this patient group should not replace movement

Figure

Fig. 1: User interacting with the iAssistADL platform. The device applies assistive forces to the forearm while preserving natural upper-limb motion.

generation but rather support the user by selectively compensating for pathological motion while preserving voluntary motor control [6]. Existing assistive devices often face a trade-off between transparency and force capability, as force-capable systems often introduce friction, inertia, and resistance that interfere with voluntary user motion. Achieving high transparency while maintaining sufficient force rendering capability remains a key challenge. Related end-effector robots such as MIT-MANUS and H-Man similarly address the trade-off between low mechanical impedance and active force rendering, but target rehabilitation and motor-control applications rather than compensation of pathological movement components [7], [8].

Two families of devices exist for suppressing pathological upper-limb motion. Grounded systems are mounted to a table or wheelchair and slow the limb with velocity-dependent resistance [9], [10]. Wearable orthoses act at single joints instead, using dampers [11], driven transmission [12], [13], or electrical stimulation [14]. In either case, whatever resists the tremor also resists the user. A systematic review of wearable orthoses found that some residual resistance always remains, whether the mechanism is passive, active, or semi-active. The best-performing device in the review also removes 17% of the voluntary movement along with the tremor [15]. Users perceive this resistance and expend additional muscle effort to maintain the intended movement [16].

∗^{*}These authors contributed equally to this work.

This work was financed by the Baden-Wurttemberg Stiftung in the scope of the AUTONOMOUS ROBOTICS project iAssistADL granted to WI, MG, SS, and DH. 1^{1} Institute of Industrial Manufacturing and Management, University of Stuttgart, Germany; 2^{2} Biomechatronic Systems, Fraunhofer IPA, Stuttgart, Germany; 3^{3} Hertie Institute for Clinical Brain Research and Centre for Integrative Neuroscience, University of Tubingen, Germany; 4^{4} Institute for Visualization and Interactive Systems, University of Stuttgart, Germany; 5^{5} Institute for Modelling and Simulation of Biomechanical Systems, University of Stuttgart, Germany; 6^{6} Center for Bionic Intelligence Tubingen Stuttgart (BITS), Germany. The authors thank the International Max Planck Research School for Intelligent Systems (IMPRS-IS) for supporting Johannes A. Gaus.

Unlike wearable tremor-suppression devices, iAssistADL is intended as a stationary assistive platform mounted to a wheelchair or table environment, targeting upper-limb activities of daily living such as eating and drinking (see Fig. 1). This platform is grounded like CEDO [9], but it generates corrective forces instead of only removing energy. Damping can slow oscillation, but it has no preferred direction and therefore cannot correct a reach that overshoots its target. Dysmetric overshoot in cerebellar ataxia requires a force directed towards the intended endpoint, which only active force rendering can provide. The ataxia literature supports this: added limb mass improves some single-joint reaches but does not carry over to multi-joint movement [17], whereas forces applied through a laboratory exoskeleton and tailored to the individual patient did reduce reaching error [1].

This paper presents and experimentally validates a method for combining high mechanical transparency with effective force rendering in an assistive ADL platform. Building upon our previous work [18], which introduced the system concept and passive transparency characterization, the present work adds the completed 3-DoF platform, the hybrid force-control strategy, and its experimental force-rendering validation. The resulting system is experimentally evaluated through transparency measurements and force-rendering experiments at representative ADL configurations, assessing its suitability as a hardware platform for future compensation of pathological movement components, such as dysmetric overshoot or tremor.