QICOM: Open-Source Software for Reproducible Image Quality Evaluation, Blinded Assessment, and Radiomic Comparison

Fabio Mattiussi, Switinder Ghotra

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Abstract

QICOM (Quality Image COMparator) is open-source software for quantitative image comparison, paired radiomics, and blinded assessment. It saves processing settings, image identities, measurements, and response mappings so a comparison can be reopened and recalculated. We evaluate reference agreement, responses to controlled perturbations, and recovery of saved analyses. Synthetic experiments cover 37 quantitative descriptors using independent numerical references and separate upstream integration comparisons. The QICOM/MIRP backend meets the reference tolerances for all 482 numeric IBSI 1 digital-phantom entries and all 1,622 entries across configurations A–E of the public CT reference case; these comprise 1,804 feature/aggregation comparisons and 300 preprocessing diagnostics. Restricting masked SSIM and HFEN to complete kernel support removes the observed dependence on excluded pixels. In a synthetic filtering example, stronger smoothing increases SSIM while worsening MSE and HFEN relative to milder smoothing and changing GLCM contrast more strongly. Reopened radiomics projects reproduce all 43 feature pairs per comparison, and two blinded deliveries return four correctly mapped scripted responses. These experiments document the calculations and data transfers linking the three workflows.

1 Introduction

The evaluation of medical imaging technologies, image-processing methods, and acquisition protocols often combines quantitative analyses with qualitative assessments. Quantitative approaches characterize image properties such as intensity differences, noise, tissue contrast, and structural similarity. Region-of-interest (ROI) measurements include signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), whose interpretation requires appropriate noise estimation, particularly when reconstruction or post-processing changes noise characteristics. Reference-based metrics provide complementary measures of image fidelity. For example, the ErisNet CT-denoising study combined mean squared error (MSE), peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), visual information fidelity (VIF), and edge preservation measures with ROI measurements of attenuation and noise [1]. Azour et al. [2] similarly combined SSIM and Fréchet Inception Distance with radiologist assessments of image quality and lesion conspicuity.

Structured visual assessment examines anatomical visibility, perceived noise, artifacts, and confidence in image interpretation. Reader scores can be evaluated using visual grading analysis (VGA) and visual grading characteristics (VGC), alongside statistics that quantify agreement between observers [3]. These methods have been applied to compare optimized X-ray protocols [4] and AI-reconstructed knee MRI with standard imaging [5], and to evaluate MRI protocols for total hip arthroplasty planning [6]. When the objective concerns pathology detection, diagnostic accuracy studies compare reader decisions against a predefined reference standard. Sensitivity, specificity, and receiver operating characteristic (ROC) analysis characterize detection performance, while free-response ROC (FROC) methods additionally incorporate lesion localization. Analyses involving multiple readers and cases must account for both sources of variability and the dependence between observations [7, 8]. These task-specific evaluations address whether changes in image appearance affect diagnostic performance.

Radiomic comparison provides another perspective by quantifying changes in intensity distributions, texture, and other image-derived features. The Image Biomarker Standardization Initiative (IBSI) supplies standardized feature definitions and reference data to support consistent extraction [9]. Evaluating radiomic features alongside image-quality measurements can help characterize processing-related changes that may affect subsequent quantitative analyses.

Combining these evaluations requires numerical measurements, radiomic features, and reader responses to remain linked to the same images and processing conditions. However, the necessary methods are often distributed across separate applications and custom scripts. Researchers may need to learn different interfaces, adapt images and masks to tool-specific requirements, and reconcile exported results. A reproducible workflow must therefore preserve image identifiers, masks, analysis settings, and blinded response mappings. Relevant settings include slice or volume selection, intensity scaling, discretization, and the spatial support of ROI calculations. Recording these choices supports repeated calculation and examination of the relationship between numerical and observer-based outcomes.

Several open-source tools provide components of this workflow. scikit-image supplies general image-processing algorithms [10]; PyTorch Image Quality (PIQ) implements image-quality metrics [11]; PyRadiomics provides configurable radiomic feature extraction [12]; 3D Slicer supports extensible medical-image computing [13]; and ViewDEX supports blinded observer assessment of DICOM images [14]. Building on established computational libraries, QICOM (Quality Image COMparator) brings quantitative image evaluation and comparison, blinded assessment, and paired radiomics into a single application. Saved configurations, native projects, and response mappings preserve the connections between inputs, processing conditions, and results.

This study presents QICOM and evaluates three questions: whether its measurements agree with analytical and external references; how they respond to controlled perturbations and spatial restrictions; and whether saved inputs, settings, and outputs support re-execution of a complete comparison. Numerical phantom experiments assess reference agreement and metric behavior. IBSI digital-phantom and CT comparisons evaluate radiomic extraction, including five preprocessing configurations applied to the same acquired CT case. A synthetic filtering example follows the same images through all three workbenches, including export, reopening, and repeated calculation, while scripted responses test blinded delivery and recovery of source conditions. These experiments assess computational agreement and workflow reproducibility; human perceptual validity and diagnostic performance require subsequent evaluation.