arXiv:2610.31302Whole-Body Motion Guided Reference Generation for Uncertainty-Aware Compliant Assistance
Muyuan Ma, Long Cheng, Lijun Han, Haotian Zhai, Xinpan Meng, Xiuze Xia, Houcheng Li and Zeyu Liu
Abstract—Exoskeletons have shown great potential for augmenting human physical capabilities, yet generating reliable assistance remains challenging because human motion patterns and interaction requirements vary substantially in practice. To address this problem, a dual-loop assistance strategy is proposed to couple multimodal exoskeleton trajectory generation with variable admittance control, where trajectory variance links the outer-loop generator and the inner-loop controller. The generator synthesizes future exoskeleton trajectories through denoising, primarily using whole-body human motion and exoskeleton proprioception, with task descriptions as auxiliary semantic cues. To reduce cross-modal representation mismatch, a three-stage training pipeline is introduced to progressively optimize the human motion encoder, the exoskeleton proprioception encoder, and the trajectory generator. In the inner loop, multiple future trajectories are sampled from the generator to estimate the trajectory mean and variance. The mean is used as the control reference, while the variance provides an uncertainty measure for prediction reliability. This measure is further incorporated into a stiffness regulation law, enabling compliant assistance with input-to-state stability under bounded lumped interaction input. Experimental results demonstrate that the proposed method significantly reduces muscle activation levels compared with standard baselines, while also decreasing users' metabolic energy expenditure relative to the unassisted condition.
Index Terms—Human motion guidance, multimodal trajectory generation, stiffness modulation, compliant assistance, wearable exoskeleton
I. INTRODUCTION
EXOSKELETONS can augment human strength, reduce physical effort, and assist individuals with motor impairments [1]. These capabilities make such systems broadly applicable to rehabilitation [2], military operations [3], and physically demanding industrial tasks [4]. However, designing exoskeleton assistance strategies that provide effective support while maintaining safe and comfortable physical interaction remains challenging, because the human and the exoskeleton form a tightly coupled dynamical system whose behavior varies across users and tasks [5]. Therefore, effective exoskeleton assistance requires reliable assistive input generation and compliant physical interaction under varying motion and interaction conditions [6].
Existing studies on exoskeleton assistance commonly generate assistive inputs in the form of either joint torque trajectories [7], [8] or reference motion trajectories [9], [10]. Torque trajectories directly prescribe joint-level assistance, whereas motion trajectories provide desired joint references that are converted into assistive torques through low-level control. For torque-based assistance, [11] estimated human joint torque from electromyographic (EMG) signals and generated exoskeleton

Fig. 1. Overview of the proposed dual-loop framework with uncertainty coupling for multimodal exoskeleton assistance.
assistive torque, validating the effectiveness of assistance in flexion and extension tasks. Moreover, [12] constructed an EMG-torque estimator based on radial basis function neural networks and integrated it with a Slotine–Li controller for exoskeleton assistance. To improve the efficiency of torque estimation, [13] introduced a physics-informed neural network for multiple joint angle and torque estimation. Furthermore, [14] designed a task-agnostic controller that instantaneously estimates lower-limb biological joint moments using a temporal convolutional network, assisting users across 28 lower-limb tasks. Despite these advances, torque-based methods often rely on human joint torque data collected under controlled laboratory conditions. Such data are difficult to obtain in practical settings, which limits the scalability and real-world applicability of these methods. For reference-based assistance, [15] generated ankle joint reference trajectories through cubic spline interpolation to assist human locomotion. In addition, [16] leveraged dynamic movement primitives (DMPs) to generate reference trajectories online. Furthermore, [17] employed a graph spiking neural network to generate globally optimal desired trajectories for lower limb exoskeletons, enabling safe and stable walking in complex indoor environments. For upper limb exoskeletons, [18] used a diffusion model to generate future motion trajectories and validated its effectiveness on four motion tasks. Although reference-based methods can provide effective assistance within specific task categories [19], [20], most of them generate trajectories mainly from limited inertial measurement unit (IMU) or EMG signals. The absence of richer multimodal information, such as visual cues, human motion states, and task-related context, limits reliable reference generation under varying assistance conditions [21].
The generated reference trajectories may not always match individual motion habits and preferred movement patterns across users [22]. To mitigate the resulting interaction issues, compliant control has become a commonly used low-level control strategy for exoskeletons [23], enabling dynamic regulation of the relationship between human-exoskeleton interaction forces and reference trajectories [24], [25]. This mechanism allows the exoskeleton to balance accurate reference tracking with compliance to the user's voluntary motion [26]. Based on this principle, admittance control has been employed to reshape reference trajectories online according to interaction forces, showing favorable performance in level-ground walking experiments with two subjects [27]. An event-triggered hybrid position-force control method was further introduced to adapt the control strategy to varying interaction forces [28]. To improve tracking safety, a tangent-type barrier Lyapunov function (BLF) controller was designed to constrain position tracking errors within predefined closed-loop boundaries, thereby reducing human muscular effort in experiments with eight subjects [29]. In addition, an adaptive modular neural control framework improved position tracking performance using only the robot's embedded position encoders, and its effectiveness was validated on five subjects [30]. Beyond these compliant control schemes, interaction forces have also been incorporated into the weighting matrix of the cost function in model predictive control, which further reduces human-exoskeleton interaction forces during assistance [31]. However, most existing control strategies regulate compliance mainly in response to interaction forces or tracking errors [32]–[34]. These feedback signals mainly indicate human-exoskeleton disagreement after it occurs, but provide limited information about whether the reference trajectory is reliable. As a result, the controller may still enforce tracking of an unsuitable reference, leading to unnecessary interaction forces and reduced comfort.
In this article, a dual-loop exoskeleton assistance framework is proposed that couples multimodal trajectory generation with compliant interaction control, using trajectory uncertainty as the coupling variable between the outer-loop generator and the inner-loop variable admittance controller. To achieve reliable trajectory generation, the generator produces future exoskeleton trajectories primarily conditioned on whole-body human motion and exoskeleton proprioceptive states, with task descriptions providing auxiliary semantic cues. A flow-matching objective is used to learn a conditional motion field that progressively maps Gaussian noise to future trajectory samples [35]. In addition, to alleviate representation misalignment among different modalities, a three-stage training pipeline is applied to sequentially learn human motion representations, align exoskeleton proprioceptive representations, and optimize the trajectory generator. To the best of our knowledge, this is the first work to explicitly leverage whole-body human motion as guidance for exoskeleton trajectory generation, enabling the generated reference trajectories to incorporate global motion context beyond local exoskeleton proprioception.
For the inner-loop admittance controller, the reference trajectory of the exoskeleton is derived from the mean of multiple sampled future trajectories. The corresponding trajectory variance is regarded as a measure of trajectory uncertainty and serves as an input to the stiffness regulation law subject to stability constraints. When the estimated trajectory uncertainty increases, the regulation law decreases the interaction stiffness to improve compliance with the user's voluntary motion. When the estimated trajectory uncertainty decreases, the stiffness is increased to enhance tracking of the generated reference trajectory. The proposed framework is illustrated in Fig. 1, including the dual-loop assistance structure composed of the trajectory generator and the variable admittance controller. The contributions of this article are as follows.
-
A dual-loop exoskeleton assistance framework is proposed to bridge generative motion assistance and compliant physical interaction control through trajectory uncertainty. By using uncertainty as an explicit coupling variable, the framework coordinates trajectory generation with compliant admittance regulation within a unified assistance architecture.
-
A multimodal exoskeleton trajectory generation method is developed to synthesize reliable reference trajectories under multimodal conditions. The method primarily leverages whole-body human motion and exoskeleton proprioceptive states as dynamic physical conditions, while incorporating task descriptions as auxiliary semantic cues. A three-stage training pipeline is further introduced to reduce cross-modal representation mismatch.
-
An uncertainty-aware variable admittance control strategy is designed for compliant physical interaction. The mean of sampled future trajectories defines the controller reference, and the temporal variation of the filtered trajectory uncertainty guides stiffness regulation. This strategy adjusts interaction compliance under stability constraints to balance active assistance and the user's voluntary motion.
The rest of this article is organized as follows. Section II presents the multimodal exoskeleton trajectory generator, including its network architecture, three-stage training pipeline, and optimization objectives. Section III introduces the uncertainty-aware variable admittance controller and the associated stiffness regulation law. Section IV reports the experimental results and comparative analyses. Section V concludes this article.
Future Exoskeleton Motion Trajectory
Gaussian Noise

Human–Exoskeleton Cooperative Motion Transformer Decoder
Human Motion Encoder
Language Encoder
3D Human Motion
Arm flexion while
wearing an exoskeleton
and holding a dumbbell.
Task Description
Proprioception Encoder
Exoskeleton Proprioception
Fig. 2. Architecture of the proposed multimodal exoskeleton trajectory generator.
A. Generator Architecture
The generator is designed to model future exoskeleton trajectory distributions under heterogeneous multimodal conditions. As shown in Fig. 2, the generator consists of a human motion encoder, a language encoder, a proprioception encoder, and a human–exoskeleton cooperative motion Transformer decoder. The three encoders transform heterogeneous inputs into token sequences with a unified latent dimension. These multimodal tokens are then used as conditional memory, while the decoder takes the encoded noisy trajectory as queries. This design separates condition encoding from trajectory denoising, allowing the generator to learn a conditional velocity field under the guidance of whole-body human motion, task descriptions, and exoskeleton proprioception.
The historical human motion is estimated from binocular RGB-D image sequences through a human pose estimation procedure [36]. Specifically, 2D skeletal keypoints are first detected from RGB images, lifted to 3D using depth information, and then temporally tracked and filtered to obtain a joint coordinate representation following the Skinned Multi-Person Linear (SMPL) kinematic structure [37]. The resulting historical human motion is denoted as , where is the number of historical frames and denotes the number of human joints. is then encoded into latent motion tokens as
where is the human motion encoder and specifies the unified latent dimension.
The task description provides semantic information about the intended assistance task. Given a task description sentence s with length , DistilBERT [38] is first used to extract token-level semantic features, which are then projected into latent task tokens as
where is the DistilBERT encoder and is the task projection module.
The historical exoskeleton proprioception is defined as a sequence of joint positions, joint velocities, and human–exoskeleton interaction forces, where is the number of historical proprioceptive frames and is the number of exoskeleton joints. is then encoded into latent proprioceptive tokens as
where is the proprioception encoder. The sampled timestep is also embedded into a timestep token . The encoded tokens from the three modalities and are concatenated along the token dimension to form the multimodal conditional memory :
The noisy future trajectory is mapped into latent trajectory tokens , where is the length of the future trajectory to be generated. The human–exoskeleton cooperative motion Transformer decoder consists of four stacked layers. The first layer takes as input, while each subsequent layer takes the output of its preceding layer. In each layer, self-attention is applied to the input trajectory tokens to model temporal dependencies, followed by cross-attention with the multimodal conditional memory . After four layers, the final output tokens are projected back to the exoskeleton trajectory space and decoded into the estimated velocity field .
During inference, trajectory generation follows the learned flow-matching dynamics over a decreasing timestep sequence , where and . The process starts from an initial Gaussian noise sample . At the -th generation step, the current noisy trajectory is encoded and fed into the decoder together with the multimodal conditional memory constructed using the timestep embedding . The decoder estimates the velocity field , and the noisy future trajectory is updated by numerical integration:
Dataset Scale Gradually Decreases

Task Relevance Gradually Decreases
Fig. 3. Three-stage training pipeline for multimodal exoskeleton trajectory generation.
where . After integration steps, the generated future exoskeleton trajectory is obtained as .
B. Three-Stage Training Pipeline
To train the generator under limited real exoskeleton data, a three-stage training pipeline is designed, as shown in Fig. 3. The training process follows a progressive data adaptation strategy, where the dataset scale gradually decreases while the task relevance gradually increases. Specifically, the first stage pretrains the human motion representation using large-scale text-motion data, the second stage introduces simulated human–exoskeleton interaction data to align multimodal conditions, and the third stage adapts the human–exoskeleton cooperative motion Transformer decoder to real exoskeleton trajectory generation.
- Stage 1: Human Motion Pretraining: In the first stage, the human motion encoder is pretrained using HumanML3D, a publicly available large-scale dataset of paired task descriptions and human motions [39]. The dataset contains 14,616 motion sequences and 44,970 natural language descriptions, covering a wide range of daily human activities.
The pretraining procedure follows the architecture of the Diffusion Planner (DiP) in [40]. As shown in Stage 1 of Fig. 3, the task description is encoded by a frozen DistilBERT to provide language guidance. The historical human motion and a noisy future human motion segment are concatenated and then processed by a trainable human motion encoder and an auxiliary Transformer decoder to estimate the velocity field for future human motion generation. Following the flow-matching formulation [35], given a future human motion segment and an initial Gaussian noise sample , the noisy future human motion at timestep is constructed as
where corresponds to the human motion data distribution and corresponds to the Gaussian noise distribution. The target velocity field for human motion generation is defined as
The human motion pretraining loss is then formulated as
where denotes the velocity field estimated by the auxiliary Transformer decoder. Through this pretraining stage, the human motion encoder learns temporal motion representations from large-scale text-motion data. The pretrained encoder is then transferred to the following stages as a human motion prior, while the auxiliary Transformer decoder is discarded after pretraining.
- Stage 2: Simulated Proprioception Alignment: In the second stage, simulated human–exoskeleton interaction data are introduced to align human motion, task descriptions, and exoskeleton proprioception. This stage bridges the gap between large-scale human motion pretraining and exoskeleton trajectory generation, since the text-motion data used in Stage 1 do not contain exoskeleton proprioception or trajectories.

Fig. 4. Simulation framework for human–exoskeleton interaction data generation.
As illustrated in Fig. 4, the simulated human–exoskeleton interaction data are generated in MuJoCo [42] using a coupled simulation system composed of a musculoskeletal human model and an elbow exoskeleton model. The musculoskeletal model is actuated by muscle activations and trained with Proximal Policy Optimization (PPO) to track reference human

Fig. 5. Motion tasks used for simulated human–exoskeleton interaction data generation.
motions. At each simulation step, the observation provided to the PPO policy is constructed as
where is the reference joint configuration, and are the joint position and velocity of the musculoskeletal human model, respectively, denotes the tracking error, and is the external joint torque induced by exoskeleton assistance. The PPO reward is defined as
where denotes the muscle activation and is the weighting coefficient for the activation penalty. This reward encourages the musculoskeletal human model to track the reference motion while avoiding excessive muscle activation. The exoskeleton model adopts a theoretically optimal assistance strategy, in which the assistance is generated by directly compensating the biological elbow joint torque of the musculoskeletal human model. The assistive torque is computed as
where is the biological elbow joint torque and denotes the assistance ratio. The compensated torque is applied to the exoskeleton joint and transmitted to the human elbow joint through the contact dynamics between the human and the exoskeleton, thereby forming a closed-loop human–exoskeleton interaction process.
To improve the diversity of the simulated human–exoskeleton interaction data, domain randomization is applied to both the musculoskeletal model and the assistance condition. Specifically, the body mass of the musculoskeletal human model is selected from 60 kg, 70 kg, and 80 kg, and the assistance ratio is selected from 0, 0.5, and 1. In addition, both left-side and right-side assistance conditions are considered. Finally, the simulation procedure yields a dataset covering the 10 motion tasks shown in Fig. 5. Each sample consists of a task description, a human motion sequence, exoskeleton proprioceptive states, and the corresponding exoskeleton trajectory.
Based on the simulated dataset, the proprioception encoder and the human–exoskeleton cooperative motion Transformer decoder are trained while keeping the human motion encoder and the DistilBERT text encoder fixed. This freezing strategy preserves the general human motion and language representations learned from large-scale text-motion data, allowing the trainable modules to focus on exoskeleton-specific proprioceptive alignment and trajectory generation under simulated multimodal conditions. For each future simulated exoskeleton trajectory , the noisy input , the target velocity field , and the training objective are constructed following the flow-matching formulation in (6), (7), and (8), respectively, with the human motion trajectory replaced by the simulated exoskeleton trajectory .
After this stage, the proprioception encoder provides exoskeleton-state tokens aligned with human motion and task semantics, and the decoder obtains a simulation-trained initialization for real trajectory adaptation.
- Stage 3: Real Trajectory Adaptation: In the third stage, real human–exoskeleton interaction data are used to adapt the simulation-trained generator to real-world sensing and physical coupling conditions. The real dataset is collected from three users wearing the exoskeleton in unassisted mode while performing 13 tasks, with each task repeated five times. The first four repetitions are used for training, resulting in 156 trajectory samples, while the last repetition is reserved for evaluation. Each sample contains the task description, human motion sequence, exoskeleton proprioception, and the corresponding exoskeleton trajectory.
Since the amount of real exoskeleton data is limited, low-rank adaptation (LoRA) is applied to fine-tune the human–exoskeleton cooperative motion Transformer decoder in a parameter-efficient manner [43]. The decoder weights learned from simulated interaction data are kept frozen, and trainable low-rank update matrices are inserted into the linear projections of the Transformer decoder. The effective weight update is expressed as
where W denotes the frozen weight, and A and B are trainable low-rank matrices. During real-data adaptation, the human motion encoder, DistilBERT text encoder, and proprioception encoder are kept fixed, while only the LoRA parameters in the decoder are optimized. For each future real exoskeleton trajectory , the noisy input and the target velocity field are constructed following the flow-matching formulation in (6) and (7), with the human motion trajectory replaced by the real exoskeleton trajectory . The real-data adaptation objective follows (8) accordingly.
This strategy preserves the coordination knowledge learned from simulation and adapts the generator to real-world sensing and interaction conditions. After the third stage, the adapted generator produces future exoskeleton trajectories from historical human motion, task descriptions, and proprioceptive feedback under real-world assistance conditions.
III. UNCERTAINTY-AWARE VARIABLE ADMITTANCE CONTROL
In this section, an uncertainty-aware variable admittance controller is designed to translate multimodal generated trajectories into compliant interaction behavior. The sampled trajectory mean defines the exoskeleton reference, whereas the averaged variance regulates the uncertainty-dependent interaction stiffness. In this manner, trajectory uncertainty explicitly couples generative motion assistance with physical interaction control.
At each trajectory update instant of the outer loop, the trajectory generator is sampled N times using the same historical human motion, task description, and exoskeleton proprioceptive states. The set of generated future exoskeleton trajectories is represented as
The reference trajectory is defined by the sample mean:
The trajectory uncertainty is characterized by the averaged sample variance of the generated trajectories:
where and are the j-th joint positions at the r-th future step in the i-th sampled trajectory and the mean trajectory, respectively.

(a)

(b)
Fig. 6. Elbow exoskeleton platform and geometric illustration for gravity compensation. (a) Elbow exoskeleton with the interaction force interface and AK60-6 actuator. (b) Geometric illustration for the gravity compensation algorithm.
The proposed controller is implemented on the elbow exoskeleton proposed in [44], as shown in Fig. 6a. The exoskeleton consists of a human–robot interaction interface for measuring the interaction force and an AK60-6 actuator (T-MOTOR, Nanchang, China) for active assistance. Each side is modeled as a single-degree-of-freedom rotational joint about the elbow flexion-extension axis. The dynamics of the elbow exoskeleton are formulated in joint space as
where , , and are the elbow joint position, velocity, and acceleration, respectively. and denote the inertia and Coriolis terms, respectively. is the overall gravity term of the human–exoskeleton system with the external load, given by
where , , and represent the gravity terms of the exoskeleton, the human forearm, and the external load, respectively. is the actuator torque and is the perpendicular moment arm from the elbow joint axis to the line of action of .
Admittance control aims to regulate the dynamic relationship between the human–robot interaction torque and the motion deviation from the reference trajectory. Instead of enforcing rigid position tracking, the interaction torque is used to reshape the exoskeleton motion, thereby allowing compliant behavior around the generated reference. For the elbow exoskeleton, the desired admittance behavior is specified as
where , , and denote the position, velocity, and acceleration tracking deviations, respectively. is the desired joint position obtained from the generated reference trajectory at the current inner-loop control instant. and are the virtual inertia and damping of the admittance model, respectively. is the time-varying interaction stiffness adjusted according to . The interaction stiffness regulation law is given by
where and denote the lower and upper bounds of the admissible interaction stiffness, respectively, satisfying
The term denotes the unconstrained stiffness update input before projection and is defined as
where is the stiffness adaptation coefficient satisfying
and is the bounded uncertainty-variation signal. For , it is defined as
where denotes the -th trajectory update instant, and is the saturation bound. It follows directly from (23) that . In (23), denotes the filtered normalized trajectory standard deviation, which is updated as
(24) where is the filtering coefficient. The normalized trajectory standard deviation is computed as
where is the normalization constant, and .
According to (21), when the generated trajectory uncertainty increases, one has , which yields and . Therefore, decreases unless it reaches , improving compliance with the user's voluntary motion. Conversely, when the generated trajectory uncertainty decreases, one has , which yields and . Therefore, increases unless it reaches , enhancing tracking of the generated reference trajectory.
To transform the exoskeleton dynamics in (16) into the desired admittance behavior in (18), the actuator torque is specified by
where is the estimated gravity compensation term, which is decomposed as
where , , and are the masses of the exoskeleton, human forearm, and external load, respectively. , , and are the distances from the elbow joint axis to the centers of mass of the exoskeleton, human forearm, and external load, respectively. g is the gravitational acceleration, and is the angle between the upper-limb motion plane and the sagittal plane illustrated in Fig. 6b. The angle is obtained from the normal directions of the two planes as
where and are the unit normal directions of the upper-limb motion plane and the sagittal plane, respectively. is constructed as
where , , and are the shoulder, elbow, and wrist keypoints extracted from the SMPL-format joint coordinates described in Section II, respectively. The sagittal plane is spanned by the body height direction and the body forward direction. In the adopted SMPL coordinate system, its unit normal direction is
Substituting (26) into (16), the closed-loop dynamics are derived as
where
is the lumped interaction input, which consists of the human–exoskeleton interaction torque and the gravity compensation error. Different from a purely exogenous disturbance to be rejected, is the physical human–exoskeleton interaction input involved in the desired admittance behavior. In practice, is bounded because the interaction force is physically limited by the human–exoskeleton contact and safety constraints, and the gravity compensation error remains bounded for bounded joint motion and finite parameter errors. Therefore, the following result establishes input-to-state stability of the tracking error with respect to .
Theorem 1: Given the single-degree-of-freedom elbow exoskeleton dynamics in (16), the desired admittance behavior in (18), the stiffness regulation law in (19), and the actuator torque in (26), the closed-loop error system in (31) is input-to-state stable (ISS) with respect to . Consequently, bounded human–exoskeleton interaction force and bounded gravity compensation error imply bounded tracking errors. Moreover, when , the tracking error exponentially converges to the equilibrium point.
Proof: According to (19) and , one has
Due to the projection operator in (19), is piecewise absolutely continuous. Hence, the following Lyapunov analysis is conducted in the almost-everywhere sense. Consider the Lyapunov function proposed in [45]:
(34)
Taking the time derivative of (34) and substituting (31) yield
(35)
According to (21), when the projection is inactive,
Thus,
(37)
When the projection is active, , and hence
(38)
Therefore,
(39)
TABLE I TASK LIST OF THE REAL EXOSKELETON DATASET.
| ID | Task | ID | Task |
|---|---|---|---|
| Task 1 | Lift and carry a box from the right-side floor to the table in front | Task 8 | OK gesture |
| Task 2 | Move a box horizontally between floor positions | Task 9 | Stop gesture |
| Task 3 | Carry a box while walking forward | Task 10 | Quiet gesture |
| Task 4 | Lift a box from the table to the floor | Task 11 | Wave gesture |
| Task 5 | Lift a box from the floor to the table | Task 12 | Salute gesture |
| Task 6 | Repetitive box lifting and placing | Task 13 | Six-box shelf stacking |
| Task 7 | Dumbbell weightlifting |
From (20) and (22), there exists a positive constant such that
where
Let
By the triangle inequality and , it follows that
where . Then, according to Young's inequality, for any ,
By choosing , substitution of (43) into (40) gives
where
Next, define
Since is nonsingular, the norms of and are equivalent. In addition, (34) can be written as
where
is positive definite and its eigenvalues are uniformly bounded. Hence, there exist positive constants and such that
The upper bound in (46) gives
Substituting (47) into (44) yields
where
Therefore, the closed-loop error system is input-to-state stable with respect to . If is bounded, the tracking error remains bounded. Moreover, when , (48) reduces to , which implies that the tracking error exponentially converges to the equilibrium point. This completes the proof.
IV. EXPERIMENTAL VALIDATION
This section evaluates the effectiveness of the proposed uncertainty-aware multimodal exoskeleton assistance framework. The experiments are organized into two groups. The first group consists of ablation studies, which analyze the contributions of the three-stage training pipeline and the multimodal conditional inputs. The second group consists of human-subject comparative experiments, in which the proposed generator and controller are compared with representative baselines to evaluate trajectory generation, interaction control, and overall assistance performance.
A. Ablation Studies
The ablation studies are conducted on the evaluation set of the real exoskeleton dataset to assess the exoskeleton trajectory generation capability of the proposed generator. The evaluation set covers 13 real exoskeleton tasks, including object manipulation, gesture, and weightlifting tasks, as summarized in Table I. The generated future exoskeleton trajectory is compared with the true trajectory . The evaluation metrics
TABLE II ABLATION RESULTS OF THE THREE-STAGE TRAINING PIPELINE. THE RESULTS ARE REPORTED AS MEAN STANDARD DEVIATION.
| Training Scheme | Exoskeleton MAE (rad) ↓ | Exoskeleton MSE (rad2) ↓ | Exoskeleton RMSE (rad) ↓ |
|---|---|---|---|
| Stage 3 only | 0.2259 ± 0.1259 | 0.1482 ± 0.1595 | 0.3441 ± 0.1727 |
| Stage 2 + Stage 3 | 0.1835 ± 0.1137 | 0.1148 ± 0.1373 | 0.2914 ± 0.1729 |
| Stage 1 + Stage 2 + Stage 3 | 0.1666 ± 0.1039 | 0.1002 ± 0.1296 | 0.2741 ± 0.1582 |
include mean absolute error (MAE), mean squared error (MSE), and root mean squared error (RMSE), defined as
These metrics are used to evaluate whether the generator can accurately predict future exoskeleton trajectories under different training schemes and multimodal conditional inputs.
- Ablation on the Three-Stage Training Pipeline: To evaluate the contribution of each training stage, three training schemes are compared. The first scheme uses only real-data adaptation. The second scheme adds simulated human–exoskeleton interaction training before real-data adaptation. The third scheme further introduces large-scale human motion pretraining, forming the complete three-stage pipeline. All schemes are evaluated on the same real exoskeleton evaluation set, and the results are reported in Table II.
As shown in Table II, the complete three-stage pipeline achieves the lowest errors among all training schemes. Compared with the complete pipeline, using only real-data adaptation increases the exoskeleton MAE, MSE, and RMSE by 35.59%, 47.90%, and 25.54%, respectively. This indicates that directly adapting the generator with limited real exoskeleton data is insufficient, due to the large inter-user trajectory variations observed even within the same task.
Adding simulated human–exoskeleton interaction training reduces the prediction errors compared with using only real-data adaptation, showing that simulated proprioceptive alignment provides useful exoskeleton-specific motion knowledge. On this basis, further adding large-scale human motion pretraining reduces the exoskeleton MAE, MSE, and RMSE by 9.21%, 12.72%, and 5.93%, respectively, confirming its role in providing a beneficial motion prior for real exoskeleton trajectory generation.
- Ablation on Multimodal Conditional Inputs: To evaluate the contribution of different conditional modalities, the complete three-stage trained generator is tested under different input configurations. During evaluation, one or more modalities are masked while the remaining inputs are kept unchanged. Four settings are compared: full multimodal inputs, without human motion, without task description, and without both human motion and task description. All settings are evaluated on the same real exoskeleton evaluation set using the metrics defined in (50).
The results in Table III indicate that the full multimodal input setting achieves the lowest errors among all input configurations. Compared with this setting, removing human motion leads to the largest performance degradation, increasing the exoskeleton MAE, MSE, and RMSE by 16.47%, 29.74%, and 12.26%, respectively. This suggests that historical human motion provides complementary information beyond exoskeleton proprioception, which helps improve future exoskeleton trajectory prediction. Removing the task description causes only a marginal increase in prediction errors. This can be attributed to the fact that the language input provides static task-level semantics, which mainly helps distinguish task categories rather than describing frame-varying motion states. To examine the role of semantic cues from task descriptions, an additional evaluation is conducted on a task set with ambiguous intentions. In this task set, paired samples share identical historical human motion and exoskeleton proprioception, whereas their task descriptions and corresponding future exoskeleton trajectories are different. Therefore, the future trajectory cannot be uniquely inferred from physical observations alone. During evaluation, the physical inputs are kept unchanged, and the task description is either kept correct or replaced with a mismatched description. Let and denote the prediction errors of the i-th sample obtained with the correct and mismatched task descriptions, respectively. The success rate is defined as
where is the indicator function. This metric measures the percentage of samples in which the correct task description produces a lower prediction error than the mismatched description. Using a mismatched task description substantially degrades the prediction accuracy on the task set with ambiguous intentions. The RMSE increases from rad with the correct description to rad with the mismatched description. The mean error increase is rad, and reaches . These results indicate that the task description is not ignored by the generator. Instead, it provides semantic information that helps disambiguate future exoskeleton trajectories when identical physical observations correspond to different task intentions.
To further interpret how the generator utilizes different modalities, the attention weights assigned to task descriptions, human motion, and exoskeleton proprioception are visualized in Fig. 7. The results show that exoskeleton proprioception receives relatively high attention across most tasks, which is
TABLE III ABLATION RESULTS OF MULTIMODAL CONDITIONAL INPUTS. THE RESULTS ARE REPORTED AS MEAN ± STANDARD DEVIATION.
| Conditional Inputs | Exoskeleton MAE (rad) ↓ | Exoskeleton MSE (rad2) ↓ | Exoskeleton RMSE (rad) ↓ |
|---|---|---|---|
| Full multimodal inputs | 0.1666 ± 0.1039 | 0.1002 ± 0.1296 | 0.2741 ± 0.1582 |
| Without human motion | 0.1940 ± 0.1319 | 0.1300 ± 0.1715 | 0.3077 ± 0.1880 |
| Without task description | 0.1691 ± 0.1059 | 0.1009 ± 0.1265 | 0.2745 ± 0.1599 |
| Without human motion and task description | 0.1799 ± 0.1099 | 0.1122 ± 0.1450 | 0.2921 ± 0.1638 |
expected because it directly reflects the current robot state and human–exoskeleton interaction condition. Human motion also obtains task-dependent attention, indicating that it provides useful complementary information for predicting future exoskeleton trajectories. In contrast, the attention weights of task descriptions are relatively stable and lower than those of the two physical modalities, which is consistent with their role as static task-level semantic cues. These observations are consistent with the quantitative results in Table III, further confirming the complementary roles of multimodal inputs in trajectory generation.

Fig. 7. Attention heatmap of task descriptions, human motion, and exoskeleton proprioception across 13 real exoskeleton tasks.
To further illustrate the contribution of the human motion modality, a representative sample from Task 5 is visualized in Fig. 8. Two phases of the same subject are selected, including the movement onset phase and the object approach phase. In each subplot, the light-blue region denotes the observed input window, while the light-red region denotes the future trajectory segment containing the ground truth and model prediction. It can be observed that the historical exoskeleton proprioceptive states are nearly identical in the two phases, whereas the human motion states exhibit different patterns. Specifically, this subject tends to keep the arm approximately vertical when approaching the object and then flexes the elbow after contacting the object to lift it. Under such conditions, the future exoskeleton trajectory cannot be reliably determined from proprioceptive states alone. Instead, the generator needs to use the variation in human motion to infer the subsequent trajectory. The predicted trajectories in both phases are consistent with the corresponding ground truth, indicating that human motion helps the generator distinguish different future trajectories when the observed exoskeleton states are similar.
(a)
(b)
Fig. 8. Representative visualization of the role of human motion in future exoskeleton trajectory generation in Task 5. The upper plots show the observed exoskeleton joint positions and the future ground-truth and predicted trajectories, while the lower panels show the corresponding RGB frames and extracted human keypoints. The light-blue and light-red regions denote the observed input window and the future prediction window, respectively. Two phases are shown: (a) movement onset phase and (b) object approach phase.
B. Comparative Experiments
The comparative experiments are conducted on seven previously unseen subjects to evaluate the proposed framework against representative baselines from three aspects: trajectory generation, interaction control, and overall assistance performance. No subject has prior experience with the exoskeleton. The seven test subjects are used only for final evaluation and are not involved in training, model selection, or tuning. First, the proposed multimodal trajectory generator is compared with the intention predictor proposed in [18], a recent state-of-the-art baseline for upper-limb exoskeleton intention prediction, to assess its ability to generate future exoskeleton trajectories under the evaluated task conditions. Second, to evaluate the effect of uncertainty-guided stiffness regulation, the proposed variable admittance controller is compared with a representative compliance control method that regulates stiffness based on tracking error [46]. Third, the complete assistance framework is compared with the unassisted condition to assess its effectiveness in reducing human effort during physical tasks.

Fig. 9. Experimental setup for comparative experiments. (a) Host computer. (b) USB-CAN adapter. (c) USB-ADC module. (d) Elbow exoskeleton [44]. (e) Power supply. (f) RGB-D camera.
The experimental setup is shown in Fig. 9. The setup consists of a host computer, a USB-CAN adapter, a USB-ADC module, the elbow exoskeleton, a power supply, and an RGB-D camera. The host computer is used to run the trajectory generator and the control algorithm. Motor commands are transmitted to the elbow exoskeleton through the USB-CAN adapter, while interaction signals are acquired through the USB-ADC module. The RGB-D camera is used to estimate human motion keypoints. The detailed implementation parameters, including the temporal windows, generator hyperparameters, sensor processing procedures, admittance parameters, and real-time latency measurements, are provided in Appendix A. All experimental procedures have been reviewed and approved by the Ethics Committee of the authors' institution. To preserve double-anonymous review, the committee name and approval number are withheld in this manuscript and will be disclosed upon acceptance.
- Comparison of Multimodal Trajectory Generation: The trajectory generation capability of the proposed multimodal generator is evaluated by comparison with the intention predictor proposed in [18]. This baseline is selected as a recent generative learning method for upper limb exoskeleton intention prediction. To ensure a subject-disjoint evaluation and prevent data leakage, all real exoskeleton demonstrations used for the real data adaptation of the proposed generator and for training the baseline are collected exclusively from the three training users. Similar to the proposed generator, the baseline provides a trajectory mean and variance through repeated sampling. The mean trajectory is used as the reference signal for admittance control, while the variance is passed to the same stiffness regulation interface.
The tested tasks include 12 short-horizon tasks and one long-horizon task. Fig. 10 presents representative illustrations of the 12 short-horizon tasks together with the time-normalized elbow joint trajectories across seven subjects. Fig. 11 presents the long-horizon six-box carrying task, in which six sequential carrying phases are performed within one continuous trial. Although several tasks exhibit similar mean motion profiles, such as elbow flexion followed by extension, their variance distributions differ along the normalized time axis. This indicates that similar average trajectories may correspond to different execution patterns. Within the same task, different users may perform the motion with different amplitudes, timing, and bilateral coordination strategies. Across different tasks, the same user may also change the elbow motion pattern according to the object interaction phase. For example, the timing of the maximum trajectory variance suggests different possible sources of variability in Task 1 and Task 2. In Task 1, the variance peak appears before object interaction, which may be related to early elbow flexion before grasping or lifting the object. In Task 2, the variance peak occurs closer to the object approach phase, which may reflect elbow flexion when the hand approaches the object.
Based on these trajectory characteristics, RMSE is used to quantify the tracking error between the reference trajectories generated by each method and the measured exoskeleton trajectories using the same variable admittance controller and parameter settings. Therefore, the resulting RMSE primarily reflects the quality of the generated reference trajectories rather than differences in controller design. Across the seven subjects, the proposed multimodal generator achieves an RMSE of rad, whereas the method in [18] obtains an RMSE of rad. The proposed method reduces the trajectory tracking error by 31.59% compared with [18]. These results indicate that the proposed multimodal generator provides more accurate reference trajectories than [18] under identical control settings.
To further evaluate the assistance effect, the muscle activation levels of the biceps brachii (BB), brachioradialis (BRR), and triceps brachii (TB) are compared between the proposed method and [18]. The sEMG signals from these muscles are recorded using a Trigno wireless sEMG system (Delsys Inc., USA) and then processed to estimate muscle activation levels. For each muscle channel, the raw sEMG signal is filtered with a fourth-order Butterworth band-pass filter between 20 and 450 Hz to attenuate motion artifacts and high-frequency noise. The filtered signal is subsequently full-wave rectified and smoothed using a 10 Hz low-pass filter to obtain the sEMG envelope. Finally, the envelope is expressed as a percentage of the maximum voluntary contraction (MVC) recorded before

Fig. 10. Representative illustrations and joint trajectory distributions of the 12 short-horizon tasks. The curves show the mean trajectories across seven subjects, and the shaded regions show the corresponding trajectory variations.
the experiments to obtain the muscle activation level. To quantify the relative change in muscle activation, the percentage reduction with respect to the method in [18] is defined as
where and denote the mean muscle activation levels in %MVC obtained under [18] and the proposed method, respectively. For each combination of task and muscle, is computed for each subject, and a two-sided paired t-test is performed across the seven subjects. The null hypothesis is , and the alternative hypothesis is , where denotes the mean percentage reduction. The resulting p-values are corrected using the false discovery rate (FDR) method across all task and muscle comparisons. The results across the seven subjects are summarized in Table IV. Detailed statistical results are reported in Appendix B. Overall, the proposed method yields lower muscle activation levels than [18] in most cases. The most evident reduction is observed in BB. The proposed method produces lower BB activation in all reported tasks, with statistically significant reductions in nearly all cases. As the primary agonist muscle for elbow flexion, BB is closely associated with user flexion effort, and its reduced activation suggests that the proposed method provides more effective flexion assistance. For BRR and TB, lower activation
TABLE IV
MUSCLE ACTIVATION LEVELS (%MVC) ACROSS SUBJECTS FOR 13 TASKS UNDER DIFFERENT ASSISTANCE METHODS. THE RESULTS ARE REPORTED AS MEAN ± STANDARD DEVIATION.
| Task | Right BRR | Right BB | Right TB | |||
|---|---|---|---|---|---|---|
| Proposed | Method in [18] | Proposed | Method in [18] | Proposed | Method in [18] | |
| 1 | * | |||||
| 2 | * | |||||
| 3 | * | * | ||||
| 4 | * | * | ||||
| 5 | * | |||||
| 6 | * | * | * | |||
| 7 | * | |||||
| Left BRR | Left BB | Left TB | ||||
| 1 | * | * | ||||
| 2 | * | |||||
| 3 | * | * | ||||
| 4 | * | * | ||||
| 5 | * | * | ||||
| 6 | ||||||
| 7 | * | |||||
| 8 | * | |||||
| 9 | * | * | ||||
| 10 | * | |||||
| 11 | * | * | ||||
| 12 | * | * | ** | |||
| 13 | * | * | ||||
Lower values are highlighted in bold. *p_FDR < 0.05 and **p_FDR < 0.01 indicate statistically significant differences between the proposed method and [18] after FDR correction across all task–muscle comparisons.
levels are also observed in most tasks. This suggests that the reduced BB activation is not accompanied by compensatory increases in the auxiliary flexor or antagonist muscles. Instead, the proposed method reduces the primary elbow flexion effort while maintaining or lowering the activation of other related muscles, demonstrating more effective assistance than [18].
- Comparison of Variable Admittance Control: Since the proposed controller uses trajectory variance to regulate the admittance stiffness, it is necessary to verify whether this variance reflects the reliability of the generated reference trajectory. Therefore, an uncertainty validation experiment is conducted using the data from all 13 tasks and seven unseen subjects. Pearson correlation and Spearman rank correlation are computed between the trajectory variance and the trajectory tracking error measured by RMSE.
Overall, the trajectory variance is positively correlated with the trajectory tracking error, with a Pearson correlation of and a Spearman correlation of across all tasks and subjects. The correlation results at the task and subject levels are presented in Table V. Positive correlations are observed for every task and every subject. At the task and subject levels, the Pearson correlations range from 0.4390 to 0.7288 and from 0.5351 to 0.6128, respectively, while the Spearman correlations range from 0.5727 to 0.7737 and from 0.5498 to 0.6813, respectively. These results show that trajectory variance consistently indicates reference reliability across tasks and subjects. The consistency of this relationship is also examined by dividing all samples into low, medium, and high variance groups. The tracking error increases monotonically from in the low variance group to in the medium variance group.
TABLE V
TASK AND SUBJECT LEVEL CORRELATIONS BETWEEN TRAJECTORY VARIANCE AND TRACKING ERROR.
| Task | Task level | Subject level | |||
|---|---|---|---|---|---|
| Pearson | Spearman | Subject | Pearson | Spearman | |
| 1 | 0.6379 | 0.7346 | 1 | 0.6128 | 0.6801 |
| 2 | 0.6892 | 0.7737 | 2 | 0.5685 | 0.6427 |
| 3 | 0.6811 | 0.7297 | 3 | 0.5690 | 0.6813 |
| 4 | 0.7236 | 0.7400 | 4 | 0.5597 | 0.5498 |
| 5 | 0.7288 | 0.7737 | 5 | 0.6125 | 0.6636 |
| 6 | 0.6597 | 0.7232 | 6 | 0.5618 | 0.6365 |
| 7 | 0.6804 | 0.7175 | 7 | 0.5351 | 0.5603 |
| 8 | 0.4787 | 0.6357 | - | - | - |
| 9 | 0.4877 | 0.7266 | - | - | - |
| 10 | 0.4390 | 0.5727 | - | - | - |
| 11 | 0.5138 | 0.7110 | - | - | - |
| 12 | 0.4670 | 0.6499 | - | - | - |
| 13 | 0.6264 | 0.7577 | - | - | - |
group and in the high variance group. The high variance group shows approximately 5.82 times larger tracking error than the low variance group, which further supports the use of trajectory variance as a reliability indicator for online stiffness regulation.
The effect of uncertainty-guided stiffness regulation is evaluated by comparing the proposed variable admittance controller with the compliance control method in [46]. In both controller conditions, the reference trajectory is provided by the same multimodal generator, so that the comparison isolates the influence of the stiffness regulation strategy in the admittance controller. For the method in [46], the stiffness is

Fig. 11. Representative illustration and joint trajectory distribution of the long-horizon six-box carrying task. The task consists of six sequential box-carrying phases, and the lower panel shows the normalized joint trajectory distribution across seven subjects.
adjusted according to the tracking error as
where is the learning gain.
The comparison is conducted on the dumbbell weightlifting task under three load conditions: 0 kg, 2 kg, and 5 kg dumbbell plates. Fig. 12 presents the results across seven subjects under the three load conditions. Each column includes the task snapshot, elbow joint position, regulated stiffness, and human–exoskeleton interaction force along normalized time. The variance of the joint trajectory increases during the rising phase of the lifting motion, where the regulated stiffness decreases accordingly. This behavior indicates that the controller increases compliance when the generated trajectory exhibits higher uncertainty. During the descending phase or near the end of the motion, the trajectory variance decreases and the regulated stiffness increases, allowing the exoskeleton to provide stronger reference tracking when the motion becomes more consistent across subjects. The interaction force also increases with the external load. This trend can be interpreted from the torque control law in (26). The commanded motor torque consists of an admittance control term and a gravity compensation term. Under heavier loading, the gravity compensation term accounts for a larger proportion of the motor command. Since gravity compensation is not ideal in real experiments, the associated compensation error also increases with the load. This residual error may contribute to the larger interaction force and stronger force fluctuation observed under heavier load conditions.
To quantitatively compare the interaction force under the two control strategies, the RMS interaction force is used as the evaluation metric:

Fig. 12. Variable admittance control results for the dumbbell weightlifting task under 0 kg, 2 kg, and 5 kg loads. Curves and shaded regions denote the mean and standard deviation across seven subjects, respectively.
where denotes the number of samples in each interaction force sequence. Compared with [46], the proposed variable admittance controller yields lower RMS interaction force under all three load conditions. Specifically, the RMS interaction force decreases from N to N under the 0 kg condition, from N to N under the 2 kg condition, and from N to N under the 5 kg condition. To further evaluate whether the reduced interaction force is accompanied by lower muscular effort, the muscle activation levels of BRR, BB, and TB are compared between [46] and the proposed controller. The same sEMG processing procedure described above is used to obtain the muscle
TABLE VI MUSCLE ACTIVATION LEVELS (%MVC) ACROSS USERS UNDER [46] AND THE PROPOSED CONTROLLER. THE RESULTS ARE REPORTED AS MEAN ± STANDARD DEVIATION.
| Load | Left BRR | Left BB | Left TB | |||
|---|---|---|---|---|---|---|
| Method in [46] | Proposed | Method in [46] | Proposed | Method in [46] | Proposed | |
| 0 kg | 14.81 ± 6.18 | 10.30 ± 5.74* | 14.93 ± 7.78 | 10.96 ± 3.93* | 11.76 ± 10.17 | 9.00 ± 8.42* |
| 2 kg | 30.49 ± 15.18 | 21.29 ± 9.31* | 24.89 ± 15.92 | 20.23 ± 10.74+ | 14.98 ± 16.17 | 14.31 ± 16.36 |
| 5 kg | 37.94 ± 15.41 | 32.87 ± 14.46 | 37.42 ± 15.19 | 32.58 ± 13.45 | 19.55 ± 13.17 | 16.92 ± 13.23 |
Lower values are highlighted in bold. and indicate statistically significant and marginal differences between [46] and the proposed controller, respectively, after FDR correction across all load and muscle comparisons.
activation levels in %MVC. For each combination of load and muscle, the paired comparison across the seven subjects is performed using the same statistical analysis and FDR correction procedure. Detailed statistical results are reported in Appendix C. As shown in Table VI, the proposed controller yields lower muscle activation levels than [46] for all measured muscles under the three load conditions. For the unloaded condition, significant reductions are observed in BRR, BB, and TB, indicating that uncertainty-aware stiffness regulation reduces muscular effort when the external load is small. With the 2 kg load, BRR remains significantly reduced, while BB exhibits a marginal reduction. With the 5 kg load, the proposed controller produces lower mean activation levels for all three muscles. These results are consistent with the interaction force analysis, suggesting that the proposed controller reduces physical interaction and muscular effort more effectively than [46], particularly under low and moderate load conditions.
- Comparison of Overall Assistance Performance: The overall assistance performance of the proposed framework is evaluated during a repetitive 5 kg box lifting and placing task by comparing natural human motion with motion under exoskeleton assistance. The motion pattern of this task is similar to that of Task 6, except that subjects are required to maintain an upright trunk posture so that the movement mainly relies on upper limb effort. Seven subjects participate in the experiment. The order of the natural and exoskeleton-assisted conditions is randomized across subjects. Each subject completes only one condition per day, and two consecutive test conditions are separated by at least one day to reduce potential fatigue, learning, and order effects. Each subject performs four 2 min sessions, with a 40 s rest period between two consecutive sessions. Respiratory metabolism is measured using a PACER VO2 Analyzer (Neumafit Co. Ltd., South Korea). The respiratory measurements are converted into energy expenditure normalized by body mass using the Brockway equation:
where and denote the volumetric rates of oxygen consumption and carbon dioxide production, respectively, and m denotes the subject's body mass.
Figure 13 shows the energy expenditure normalized by body mass during the repetitive 5 kg box lifting and placing task. In the first session, energy expenditure increases rapidly from the initial level and approaches a quasi-steady state after approxi- mately 2 min. In the following sessions, the curves do not show a continuous upward drift and remain within a relatively stable range. Energy expenditure increases during each 2 min task period and decreases during each 40 s rest period. This trend indicates that the initial transient metabolic response mainly occurs in the first session, while the last three sessions better reflect the stable metabolic response during repetitive upper-limb work. Motion with exoskeleton assistance shows lower energy expenditure than natural human motion throughout the experiment. Over the full trial, the average energy expenditure decreases from W·kg during natural human motion to W·kg with exoskeleton assistance, corresponding to an average reduction of 9.0%. The total energy expenditure also decreases from kJ to kJ. The reduction ratio of energy expenditure is defined as

Fig. 13. Energy expenditure comparison during repetitive 5 kg box lifting and placing. Curves and shaded regions denote the mean and standard deviation across seven subjects, respectively.
where and denote the energy expenditure during natural human motion and exoskeleton-assisted motion, respectively. The same statistical analysis is performed across subjects. The reduction over the full trial is statistically significant (p = 0.0041).
Based on this observation, the first session is excluded from the session comparison in Fig. 14 to reduce the influence of the initial transient metabolic response. In the last three sessions,

Fig. 14. Energy expenditure comparison in the last three sessions. Bars and error bars denote the mean and standard deviation across seven subjects, respectively. Statistical significance is marked as * for and ** for .
motion with exoskeleton assistance consistently yields lower energy expenditure than natural human motion, with reductions of 9.9%, 7.0%, and 9.4%, respectively. The reductions are statistically significant in all three sessions (p = 0.0081, p = 0.0030, and p = 0.0272). These results further confirm that the proposed framework significantly reduces respiratory metabolic cost during repetitive upper limb box lifting and placing.
V. CONCLUSION
This article proposes an uncertainty-aware exoskeleton assistance framework that couples multimodal trajectory generation with variable admittance control. The proposed generator primarily uses whole-body human motion and exoskeleton proprioception to generate future exoskeleton reference trajectories, while task descriptions provide auxiliary semantic cues. The trajectory variance estimated from multiple generated samples further regulates the admittance stiffness online, allowing the exoskeleton to balance reference tracking and compliant interaction. Experimental results demonstrate that the proposed framework achieves significant reductions in muscle activation levels compared with the baselines, while also decreasing users' metabolic energy expenditure during repetitive box lifting and placing. Future work will further validate the proposed framework on multi-joint exoskeletons under more complex real-world assistance conditions.
APPENDIX A IMPLEMENTATION DETAILS
To facilitate reproducibility, the main implementation parameters of the proposed framework are summarized in Table VII. These parameters include the temporal windows used by the trajectory generator, the hyperparameters of the flow-matching inference process, the network and training settings, the sensor processing procedures, the admittance control parameters, and the measured real-time latency.
TABLE VII IMPLEMENTATION PARAMETERS OF THE PROPOSED FRAMEWORK
| Category | Parameter | Value |
|---|---|---|
| Data split | Training/test subjects | 3/7, disjoint |
| RGB-D sensing | RGB-D camera | ZED 2i |
| Multimodal input | Input frequency | 20 Hz |
| Input window | Human motion history | 20 frames |
| Input window | Proprioception history | 20 frames |
| Prediction | Future horizon | 40 frames |
| Prediction | Number of generated samples N | 100 |
| Flow matching | Integration steps K | 10 |
| Network | Latent dimension | 64 |
| Network | Decoder layers | 4 |
| Network | Attention heads | 4 |
| Network | Feedforward dimension | 1024 |
| LoRA | Rank r | 64 |
| LoRA | Scaling factor | 256 |
| Training | Optimizer | AdamW |
| Training | Stage 1 training steps | 250k |
| Training | Stage 1 learning rate | |
| Training | Stage 1 batch size | 64 |
| Training | Stage 2 training steps | 5k |
| Training | Stage 2 learning rate | |
| Training | Stage 2 batch size | 4 |
| Training | Stage 3 training steps | 45k |
| Training | Stage 3 learning rate | |
| Training | Stage 3 batch size | 39 |
| Real-time system | Inner-loop control frequency | 100 Hz |
| Real-time system | Outer-loop update frequency | 2 Hz |
| Real-time system | Average inference time for N samples | 325 ms |
| Real-time system | Average communication delay | 3 ms |
| Force sensing | ADC sampling rate | 50 Hz |
| Force sensing | Force smoothing window | 0.4 s |
| Admittance | 1 | |
| Admittance | 40 | |
| Admittance | 4 | |
| Admittance | 400 | |
| Admittance | 0.5 | |
| Admittance | 0.98 | |
| Admittance | 0.9 | |
| Admittance | 0.82 |
APPENDIX B
DETAILED STATISTICAL RESULTS OF TRAJECTORY GENERATOR COMPARISON
The trajectory generator comparison evaluates whether the proposed trajectory generator reduces muscle activation compared with [18] across different tasks. The detailed statistical results for all task–muscle comparisons are reported in Appendix Table VIII to support the significance analysis in Table IV.
APPENDIX C
DETAILED STATISTICAL RESULTS OF CONTROLLER COMPARISON
The controller comparison evaluates whether the proposed uncertainty-aware variable admittance controller reduces muscle activation compared with [46] under different external loads. The detailed statistical results are reported in Appendix Table IX to support the significance analysis in the Table VI.
REFERENCES
[1] J. L. Pons, Wearable Robots: Biomechatronic Exoskeletons. Hoboken, NJ, USA: Wiley, 2008.
[2] Y. Zimmermann, M. Sommerhalder, P. Wolf, R. Riener, and M. Hutter, "ANYexo 2.0: A fully actuated upper-limb exoskeleton for manipulation and joint-oriented training in all stages of rehabilitation," IEEE Transactions on Robotics, vol. 39, no. 3, pp. 2131–2150, 2023.
TABLE VIII DETAILED STATISTICAL RESULTS FOR MUSCLE ACTIVATION COMPARISON BETWEEN THE PROPOSED METHOD AND [18].
| Task | Muscle | Reduction (%) | 95% CI (%) | ||
|---|---|---|---|---|---|
| 1 | Right BRR | [-16, 56] | 0.509 | 0.2268/0.2474 | |
| 1 | Right BB | [12, 57] | 1.421 | 0.0094/0.0447* | |
| 1 | Right TB | [-9, 32] | 0.514 | 0.2229/0.2474 | |
| 1 | Left BRR | [-3, 55] | 0.839 | 0.0681/0.0908 | |
| 1 | Left BB | [15, 50] | 1.710 | 0.0040/0.0413* | |
| 1 | Left TB | [6, 42] | 1.235 | 0.0171/0.0447* | |
| 2 | Right BRR | [3, 56] | 1.022 | 0.0354/0.0579 | |
| 2 | Right BB | [23, 63] | 2.025 | 0.0017/0.0346* | |
| 2 | Right TB | [-9, 21] | 0.355 | 0.3836/0.3968 | |
| 2 | Left BRR | [-11, 45] | 0.568 | 0.1835/0.2078 | |
| 2 | Left BB | [9, 61] | 1.228 | 0.0175/0.0447* | |
| 2 | Left TB | [0, 43] | 0.928 | 0.0494/0.0705 | |
| 3 | Right BRR | [6, 53] | 1.176 | 0.0208/0.0447* | |
| 3 | Right BB | [13, 46] | 1.645 | 0.0048/0.0413* | |
| 3 | Right TB | [-13, 40] | 0.482 | 0.2492/0.2669 | |
| 3 | Left BRR | [15, 38] | 2.158 | 0.0012/0.0346* | |
| 3 | Left BB | [6, 51] | 1.172 | 0.0211/0.0447* | |
| 3 | Left TB | [0, 46] | 0.941 | 0.0471/0.0690 | |
| 4 | Right BRR | [8, 62] | 1.208 | 0.0187/0.0447* | |
| 4 | Right BB | [13, 62] | 1.432 | 0.0091/0.0447* | |
| 4 | Right TB | [2, 55] | 1.012 | 0.0366/0.0579 | |
| 4 | Left BRR | [3, 48] | 1.031 | 0.0342/0.0579 | |
| 4 | Left BB | [9, 52] | 1.293 | 0.0141/0.0447* | |
| 4 | Left TB | [5, 39] | 1.173 | 0.0210/0.0447* | |
| 5 | Right BRR | [-10, 58] | 0.652 | 0.1353/0.1592 | |
| 5 | Right BB | [6, 58] | 1.141 | 0.0234/0.0460* | |
| 5 | Right TB | [-1, 56] | 0.903 | 0.0541/0.0738 | |
| 5 | Left BRR | [5, 45] | 1.131 | 0.0242/0.0460* | |
| 5 | Left BB | [17, 57] | 1.699 | 0.0041/0.0413* | |
| 5 | Left TB | [-9, 46] | 0.625 | 0.1496/0.1726 | |
| 6 | Right BRR | [5, 42] | 1.177 | 0.0207/0.0447* | |
| 6 | Right BB | [9, 39] | 1.473 | 0.0080/0.0447* | |
| 6 | Right TB | [4, 27] | 1.240 | 0.0168/0.0447* | |
| 6 | Left BRR | [0, 25] | 0.942 | 0.0470/0.0690 | |
| 6 | Left BB | [2, 44] | 1.012 | 0.0367/0.0579 | |
| 6 | Left TB | [-3, 29] | 0.739 | 0.0984/0.1256 | |
| 7 | Right BRR | [-4, 29] | 0.722 | 0.1045/0.1306 | |
| 7 | Right BB | [7, 37] | 1.348 | 0.0118/0.0447* | |
| 7 | Right TB | [-8, 10] | 0.094 | 0.8112/0.8112 | |
| 7 | Left BRR | [-32, 22] | -0.180 | 0.6513/0.6623 | |
| 7 | Left BB | [4, 31] | 1.227 | 0.0175/0.0447* | |
| 7 | Left TB | [-7, 21] | 0.467 | 0.2632/0.2770 | |
| 8 | Left BRR | [-5, 42] | 0.712 | 0.1084/0.1327 | |
| 8 | Left BB | [6, 36] | 1.285 | 0.0145/0.0447* | |
| 8 | Left TB | [-3, 44] | 0.823 | 0.0724/0.0944 | |
| 9 | Left BRR | [5, 66] | 1.091 | 0.0278/0.0506 | |
| 9 | Left BB | [21, 74] | 1.681 | 0.0043/0.0413* | |
| 9 | Left TB | [14, 72] | 1.378 | 0.0108/0.0447* | |
| 10 | Left BRR | [0, 45] | 0.912 | 0.0523/0.0729 | |
| 10 | Left BB | [9, 40] | 1.452 | 0.0086/0.0447* | |
| 10 | Left TB | [-7, 48] | 0.681 | 0.1215/0.1458 | |
| 11 | Left BRR | [5, 54] | 1.127 | 0.0246/0.0460* | |
| 11 | Left BB | [13, 62] | 1.425 | 0.0093/0.0447* | |
| 11 | Left TB | [4, 60] | 1.042 | 0.0329/0.0579 | |
| 12 | Left BRR | [7, 43] | 1.304 | 0.0136/0.0447* | |
| 12 | Left BB | [8, 57] | 1.242 | 0.0167/0.0447* | |
| 12 | Left TB | [22, 38] | 3.436 | 0.0001/0.0060** | |
| 13 | Left BRR | [2, 59] | 0.986 | 0.0402/0.0618 | |
| 13 | Left BB | [9, 48] | 1.344 | 0.0120/0.0447* | |
| 13 | Left TB | [6, 52] | 1.165 | 0.0216/0.0447* |
and indicate statistical significance after FDR correction across all task-muscle comparisons.
TABLE IX DETAILED STATISTICAL RESULTS FOR THE CONTROLLER COMPARISON UNDER DIFFERENT EXTERNAL LOADS.
| Load | Muscle | Reduction (%) | 95% CI (%) | ||
|---|---|---|---|---|---|
| 0 kg | Left BRR | [17, 47] | 1.942 | 0.0021/0.0110* | |
| 0 kg | Left BB | [5, 36] | 1.193 | 0.0197/0.0443* | |
| 0 kg | Left TB | [9, 40] | 1.438 | 0.0089/0.0267* | |
| 2 kg | Left BRR | [14, 39] | 1.892 | 0.0024/0.0110* | |
| 2 kg | Left BB | [2, 28] | 1.063 | 0.0307/0.0553 | |
| 2 kg | Left TB | [-5, 22] | 0.554 | 0.1929/0.1929 | |
| 5 kg | Left BRR | [-3, 28] | 0.734 | 0.1003/0.1290 | |
| 5 kg | Left BB | [-5, 27] | 0.617 | 0.1539/0.1731 | |
| 5 kg | Left TB | [-2, 35] | 0.814 | 0.0749/0.1123 |
indicates statistical significance after FDR correction across all load-muscle comparisons in the controller comparison experiment.
[3] H. Kazerooni, J.-L. Racine, L. Huang, and R. Steger, “On the control of the Berkeley lower extremity exoskeleton (BLEEX),” in Proceedings of the 22nd IEEE International Conference on Robotics and Automation, Barcelona, Spain, 2005, pp. 4353–4360.
[4] G. Rinaldi et al., “Towards a healthier workplace: How Flexos, an active and bilateral shoulder exoskeleton, provides support in weight-lifting and carrying tasks,” IEEE Transactions on Robotics, vol. 42, pp. 1389–1404, 2026.
[5] T. Proietti, V. Crocher, A. Roby-Brami, and N. Jarrassé, “Upper-limb robotic exoskeletons for neurorehabilitation: A review on control strategies,” IEEE Reviews in Biomedical Engineering, vol. 9, pp. 4–14, 2016.
[6] F. Nazari, N. Mohajer, D. Nahavandi, A. Khosravi, and S. Nahavandi, “Applied exoskeleton technology: A comprehensive review of physical and cognitive human–robot interaction,” IEEE Transactions on Cognitive and Developmental Systems, vol. 15, no. 3, pp. 1102–1122, 2023.
[7] K. L. Scherpereel, M. C. Gombolay, M. K. Shepherd, C. A. Carrasquillo, O. T. Inan, and A. J. Young, “Deep domain adaptation eliminates costly data required for task-agnostic wearable robotic control,” Science Robotics, vol. 10, no. 108, art. no. eads8652, 2025.
[8] G. Durandau, W. F. Rampeltshammer, H. van der Kooij, and M. Sartori, “Neuromechanical model-based adaptive control of bilateral ankle exoskeletons: Biological joint torque and electromyogram reduction across walking conditions,” IEEE Transactions on Robotics, vol. 38, no. 3, pp. 1380–1394, 2022.
[9] F. Lanotte, Z. McKinney, L. Grazi, B. Chen, S. Crea, and N. Vitiello, “Adaptive control method for dynamic synchronization of wearable robotic assistance to discrete movements: Validation for use case of lifting tasks,” IEEE Transactions on Robotics, vol. 37, no. 6, pp. 2193–2209, 2021.
[10] J. Kou, Y. Wang, Z. Chen, Y. Shi, and Q. Guo, “Gait planning and multimodal human-exoskeleton cooperative control based on central pattern generator,” IEEE/ASME Transactions on Mechatronics, vol. 30, no. 4, pp. 2598–2608, 2025.
[11] T. Lenzi, S. M. M. De Rossi, N. Vitiello, and M. C. Carrozza, “Intention-based EMG control for powered exoskeletons,” IEEE Transactions on Biomedical Engineering, vol. 59, no. 8, pp. 2180–2190, 2012.
[12] K. Gui, H. Liu, and D. Zhang, “A practical and adaptive method to achieve EMG-based torque estimation for a robotic exoskeleton,” IEEE/ASME Transactions on Mechatronics, vol. 24, no. 2, pp. 483–494, 2019.
[13] L. J. Han, H. C. Li, Y. Zhang, Y. X. Zou, and L. Cheng, “Physics-informed deep transfer learning for sEMG-based multiple joint angle and torque estimation,” IEEE Transactions on Instrumentation and Measurement, vol. 74, art. no. 2531213, 2025.
[14] D. D. Molinaro, K. L. Scherpereel, E. B. Schonhaut, G. Evangelopoulos, M. K. Shepherd, and A. J. Young, “Task-agnostic exoskeleton control via biological joint moment estimation,” Nature, vol. 635, pp. 337–344, 2024.
[15] P. Huang, Z. Li, M. Zhou, and Z. Kan, “Divergent component of motion planning and adaptive repetitive control for wearable walking exoskeletons,” IEEE Transactions on Cybernetics, vol. 54, no. 4, pp. 2244–2256, 2024.
[16] C. Lauretti et al., “Learning by demonstration for motion planning of upper-limb exoskeletons,” Frontiers in Neurorobotics, vol. 12, art. no. 5, 2018.
[17] P. Huang, Z. Li, M. Zhou, G. Li, Y. Song, and R. Cui, “Hybrid long short-term motor optimization and control of a walking exoskeleton,” IEEE Transactions on Robotics, vol. 41, pp. 3820–3840, 2025.
[18] Y. Chen et al., “Upper-limb rehabilitation with a dual-mode individualized exoskeleton robot: A generative-model-based solution,” The International Journal of Robotics Research, art. no. 02783649251333479, 2025.
[19] N. V. Divekar, G. C. Thomas, A. R. Yerva, H. B. Frame, and R. D. Gregg, “A versatile knee exoskeleton mitigates quadriceps fatigue in lifting, lowering, and carrying tasks,” Science Robotics, vol. 9, no. 94, art. no. eadr8282, 2024.
[20] X. Zhang, H. Zhang, J. Hu, J. Deng, and Y. Wang, “Motion forecasting network (MoFCNet): IMU-based human motion forecasting for hip assistive exoskeleton,” IEEE Robotics and Automation Letters, vol. 8, no. 9, pp. 5783–5790, 2023.
[21] Q. Ye et al., “Assistive trajectory planning for lower limb exoskeletons: Strategies from laboratory-optimized gait to environmentally-adaptive locomotion through multimodal parameter awareness,” IEEE Reviews in Biomedical Engineering, vol. 19, pp. 41–64, 2026.
[22] M. A. Díaz et al., “Human-in-the-loop optimization of wearable robotic devices to improve human–robot interaction: A systematic review,” IEEE Transactions on Cybernetics, vol. 53, no. 12, pp. 7483–7496, 2023.
[23] Y. P. Zhang, G. Z. Cao, L. L. Li, and D. F. Diao, “Interactive control of lower limb exoskeleton robots: A review,” IEEE Sensors Journal, vol. 24, no. 5, pp. 5759–5784, 2024.
[24] N. Hogan, “Impedance control: An approach to manipulation,” in Proceedings of the American Control Conference, San Diego, CA, USA, 1984, pp. 304–313.
[25] A. Q. L. Keemink, H. van der Kooij, and A. H. A. Stienen, “Admittance control for physical human–robot interaction,” The International Journal of Robotics Research, vol. 37, no. 11, pp. 1421–1444, 2018.
[26] M. Sharifi, V. Azimi, V. K. Mushahwar, and M. Tavakoli, “Impedance learning-based adaptive control for human–robot interaction,” IEEE Transactions on Control Systems Technology, vol. 30, no. 4, pp. 1345–1358, 2022.
[27] P. Huang, Z. Li, M. Zhou, X. Li, and M. Cheng, “Fuzzy enhanced adaptive admittance control of a wearable walking exoskeleton with step trajectory shaping,” IEEE Transactions on Fuzzy Systems, vol. 30, no. 6, pp. 1541–1552, 2022.
[28] R. Perez-San Lazaro, R. Q. Fuentes-Aguilar, and I. Chairez, “Human-robot cooperative adaptive reinforcement constraint control for a lower limb rehabilitation exoskeleton based on user’s movement intention,” IEEE Transactions on Medical Robotics and Bionics, vol. 7, no. 2, pp. 607–620, 2025.
[29] Q. Wu, Z. Wang, Y. Chen, and H. Wu, “Barrier Lyapunov function-based fuzzy adaptive admittance control of an upper limb exoskeleton using RBFNN compensation,” IEEE/ASME Transactions on Mechatronics, vol. 30, no. 1, pp. 3–14, 2025.
[30] A. Srisuchinnawong, C. Akkawutvanich, and P. Manoonpong, “Adaptive modular neural control for online gait synchronization and adaptation of an assistive lower-limb exoskeleton,” IEEE Transactions on Neural Networks and Learning Systems, vol. 35, no. 9, pp. 12449–12458, 2024.
[31] X. Xia, L. Cheng, M. Ma, H. Zhang, L. Han, and H. Li, “Neurologically inspired transparent interaction for wearable exoskeletons,” IEEE/ASME Transactions on Mechatronics, vol. 30, no. 6, pp. 4814–4824, 2025.
[32] W. Huo et al., “Impedance modulation control of a lower-limb exoskeleton to assist sit-to-stand movements,” IEEE Transactions on Robotics, vol. 38, no. 2, pp. 1230–1249, 2022.
[33] Q. Zhang, K. Lambeth, Z. Sun, A. Dodson, X. Bao, and N. Sharma, “Evaluation of a fused sonomyography and electromyography-based control on a cable-driven ankle exoskeleton,” IEEE Transactions on Robotics, vol. 39, no. 3, pp. 2183–2202, 2023.
[34] W. Chen, B. Zhang, X. Tan, Y. Zhao, L. Liu, and X. Zhao, “Hip-knee-ankle rehabilitation exoskeleton with compliant actuators: From human–robot interaction control to clinical evaluation,” IEEE Transactions on Robotics, vol. 41, pp. 269–288, 2025.
[35] Y. Lipman, R. T. Q. Chen, H. Ben-Hamu, M. Nickel, and M. Le, "Flow matching for generative modeling," in Proceedings of the 11th International Conference on Learning Representations, Kigali, Rwanda, 2023.
[36] Stereolabs, “Body Tracking Overview,” [Online]. Available: https://www.stereolabs.com/docs/body-tracking
[37] M. Loper, N. Mahmood, J. Romero, G. Pons-Moll, and M. J. Black, "SMPL: A skinned multi-person linear model," in Seminal Graphics Papers: Pushing the Boundaries, Volume 2, New York, NY, USA: ACM, 2023, pp. 851–866.
[38] V. Sanh, L. Debut, J. Chaumond, and T. Wolf, “DistilBERT, a distilled version of BERT: Smaller, faster, cheaper and lighter,” 2019, arXiv:1910.01108.
[39] C. Guo et al., “Generating diverse and natural 3D human motions from text,” in Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition, New Orleans, LA, USA, 2022, pp. 5142–5151.
[40] G. Tevet et al., “CLoSD: Closing the loop between simulation and diffusion for multi-task character control,” in Proceedings of the International Conference on Learning Representations, Singapore, 2025.
[41] M. Ma et al., “Ensuring interaction safety in multitask exoskeleton control: A simulation-trained variable impedance framework,” 2026, arXiv:2606.06370.
[42] H. Wang, V. Caggiano, G. Durandau, M. Sartori, and V. Kumar, "MyoSim: Fast and physiologically realistic MuJoCo models for musculoskeletal and exoskeletal studies," in Proceedings of the 38th
International Conference on Robotics and Automation, Philadelphia, PA, USA, 2022, pp. 8104–8111.
[43] E. J. Hu et al., “LoRA: Low-rank adaptation of large language models,” in Proceedings of the International Conference on Learning Representations, Online, 2022.
[44] X. Xia, L. Han, H. Li, Y. Zhang, Z. Liu, and L. Cheng, “A compliant elbow exoskeleton with an SEA at interaction port,” in Proceedings of the 30th International Conference on Neural Information Processing, Changsha, China, 2024, pp. 146–157.
[45] K. Kronander and A. Billard, “Stability considerations for variable impedance control,” IEEE Transactions on Robotics, vol. 32, no. 5, pp. 1298–1305, 2016.
[46] X. Xing, E. Burdet, W. Si, C. Yang, and Y. Li, “Impedance learning for human-guided robots in contact with unknown environments,” IEEE Transactions on Robotics, vol. 39, no. 5, pp. 3705–3721, Oct. 2023.