Plausible but Unimplantable: A Trustworthiness Audit of Shape Completion Foundation Models for Cranial Defect Reconstruction

Marek Wodzinski, Kamil Kwarciak, Daria Hemmerling, Mateusz Daniol

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original. This text is shared under the same licence.

Marek Wodzinski 1,2^{1,2}[0000-0002-8076-6246], Kamil Kwarciak 1^{1}[0000-0002-1392-4291], Daria Hemmerling 1^{1}[0000-0002-2193-7690], and Mateusz Daniol 1^{1}[0000-0003-2363-7912]

1^{1} AGH University of Krakow, Krakow, Poland

2^{2} Sano Centre for Computational Medicine, Krakow, Poland

wodzinski@agh.edu.pl

Abstract. A patient who has lost part of the skull vault needs an implant that fits the defect exactly, and such implants are still drawn by hand in CAD by specialists, which makes each case slow and expensive. The task can be considered as a shape completion problem, so pretrained completion models, and now the first shape completion foundation models, look like a way to remove that manual step. We audited three publicly released pretrained shape completion models, two of them foundation models, applied zero-shot to 239 cranial defects, some carved from public skull datasets and some real post-craniectomy cases. Each output was compared with simply returning the patient's own remaining bone, and tested for whether the predicted material could be placed without intersecting the bone that is still there. On the 210 carved defects none of the three improved on doing nothing, and on the 29 real cases only one did so beyond chance. Yet not one output, from any model in any configuration, was implantable: none managed both to stay clear of the bone the patient still has and to place more material than doing nothing, where a task-specific volumetric model, trained for this task and selected on these cases, was implantable on all but one. The output that most convincingly resembles a skull is no more placeable than the rest, and at these point densities implant thickness cannot be read off the emitted points at all, which is a property of the output representation rather than of any one model. On this evidence the shape completion foundation models we could obtain cannot be trusted and, used as released, are not usable in clinical practice. Substantial research towards trustworthy foundation models for shape completion is still needed.

Keywords: Cranial implant design · Shape completion · Foundation models · Trustworthy machine learning · Point clouds

∗^{*} Accepted at the Workshop on Trustworthy Foundation Models in Medical Imaging (TrustFMI), ACCV 2026. This is the authors' accepted manuscript; the Version of Record will appear in the ACCV 2026 workshop proceedings.

M. Wodzinski et al.

1 Introduction

After a decompressive craniectomy, a head injury or a tumour resection that takes bone, a patient is left with an opening in the skull vault, closed with an implant shaped to that patient's anatomy [10,12]. The implant should protect the brain and restore the original shape of the skull, and whether it fits is settled in the operating room [9,11].

Such implants are designed by hand, in CAD, by specialists, which makes each case slow and expensive and limits how many patients can be offered one. Automation has been pursued for a decade, mostly as volumetric completion of the missing bone, and networks trained for that task now do well on public benchmarks [9,10,12], most recently point-cloud diffusion and flow matching [5,7,8]. Every one of them has to be trained on data for the task, and that is the cost a foundation model promises to remove.

Foundation models are becoming increasingly popular in medical imaging [17, 28], and the first of them aimed at shape completion are now appearing [14, 25]. A skull with a defect is a partial shape and the implant completes it, so implant design is close to what these models advertise [13], and a checkpoint used as it comes would remove both the training data and the manual step.

Completion models are judged by how closely the completed shape resembles a reference, and that resemblance is not what a surgeon needs to know: it cannot say whether the model added anything the patient's own bone did not already supply, nor whether what it predicted could physically be placed in that patient [15, 16, 24].

Contribution: We audit three publicly released pretrained shape completion models, applied zero-shot to cranial defects, and judge them by what the clinic needs rather than by what a benchmark reports. First, we score every output against an explicit do-nothing baseline, the patient's own retained bone. Second, we introduce a physical acceptance criterion, under which the predicted material must sit in the defect without intersecting the bone that remains, and we report how many patients could receive each output. Third, we test whether the language conditioning these models advertise carries anatomical information, reading each against its own noise floor. Together they let a completion model be judged on the thing that decides its use: whether what it produced could be put in a patient.