Determination of the Human-Robot Interaction Workspace for Robot Assisted Head Stabilization Control

Xinmin Liu, Alexander L. Van Slyke, Destinee Holloman, Ahmad Sakaamini, Rodney D. Wiersma

Xinmin Liu, PhD 1^{1}, Alexander L. Van Slyke, PhD 2^{2}, Destinee Holloman 3^{3}, Ahmad Sakaamini, PhD 4^{4}, Rodney D. Wiersma, PhD 1^{1}

1^{1} Department of Radiation Oncology, University of California at Los Angeles, CA

2^{2} Department of Radiation Oncology, Yale New Haven Hospital, Trumbull, CT

3^{3} University of Colorado, Anschutz Medical Campus, Aurora, CO

4^{4} Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA

Abstract—Accurate head immobilization is critical for medical procedures such as neuroimaging and intracranial radiation therapy, where even minor motion can compromise precision. Traditional methods, such as rigid stereotactic frames and thermoplastic face masks, are invasive, uncomfortable for patients, and require a time-consuming setup. To overcome these limitations, a novel approach employs a 6-degree-of-freedom (6DoF) robotic head holder that is integrated with real-time optical motion tracking, enabling dynamic compensation for head movements without physical restraint. However, implementing this system requires a detailed biomechanical understanding of skull-cervical spine interactions to define safe and effective motion limits. In this study, we conduct a human volunteer investigation using a 6DoF Stewart platform robot to characterize biokinematic workspace of the head-neck system and establish foundational data for human-robot interaction in precision head stabilization.

Index Terms—stereotactic surgery, medical robotics, neck kinematics, robot human interaction

1. INTRODUCTION

There has been significant interest in achieving high-precision positioning of the head in medical applications 1−3^{1-3}. In diagnostic brain imaging, head motion can lead to movement-related artifacts such as blurring and ghosting, which can hinder clinical diagnosis 4^{4}. This is especially true for imaging modalities such as functional magnetic resonance imaging (MRI) and positron emission tomography (PET), where the collection of data required for 3D image reconstruction requires patients to lie still for long periods of time 5,6^{5,6}. This issue is especially challenging when scanning children or adults who exhibit involuntary motion 7−9^{7-9}. Prospective motion correction (PMC) is a popular technique that updates the image acquisition in real-time to keep the field of view constant relative to the moving object in brain MRI. However, PMC has some limitations. It cannot correct higher-order effects such as changing magnetic field inhomogeneities and gradient imperfections. When using multiple receiver coil arrays for better signal quality, motion can change the coil sensitivities relative to the moving object. Motion during image acquisition can cause blurring and spatial distortion because data is collected from different positions within the nonlinear gradient fields. Even with accurate real-time pose correction, motion can still cause magnetic field changes due to inhomogeneities and object-induced field changes. These local magnetic field changes are often the main source of remaining artifacts 10−14^{10-14}. Therefore, it is desirable to use a robotic platform to stabilize the head position.

Although computed tomography (CT) can typically acquire images faster than MRI or PET, in certain cases, head motion artifacts have been reported 15,16^{15,16}. In therapeutic applications, accurate head positioning is essential for procedures involving the treatment of the brain using conventional surgery or radiation therapy 17−19^{17-19}. Stereotactic radiosurgery (SRS) is a treatment modality employed for managing brain disorders by utilizing highly precise radiation dose placement. SRS has been used to treat benign or metastatic brain tumors 20^{20}, arteriovenous malformations (AVMs) 21,22^{21,22}, trigeminal neuralgia 23,24^{23,24}, and acoustic neuromas 25^{25}. For SRS, accurate head positioning is needed to avoid damaging critical neural and vascular structures within the brain.

The most common approach to stabilize the head position is the use of stereotactic frames. In this approach, a metal ring is attached to the patient's skull using screws, and then the frame is bolted to the treatment table, see top row in Fig. 1 26,27^{26,27}. The invasive nature and discomfort associated with the frame have been identified as a central cause of poor patient compliance.

and poor clinical efficacy. Another approach involves thermoplastic face masks, as shown in the bottom row of Fig. 1. However, such masks have decreased accuracy due to patient-induced mask flex and can suffer from material changes over time, such as shrinkage 28^{28}. Even with the mask in place, a patient's head can still drift laterally due to rotational movement around the fulcrum at the back of the skull. In some cases, this has resulted in systematic shifts of up to 2 mm from the initial target setup position 29^{29}. A recent study reported significant errors, which were attributed to improper patient setup stemming from factors such as limitations in mask manufacturing quality, loosening or shrinkage of the mask during treatment, or physical changes in the patient's head due to facial swelling or weight loss 30^{30}. Moreover, many patients find the thermoplastic mask invasive and psychologically confining, as it severely restricts natural head motion. These factors can contribute to unsafe conditions, potentially leading to treatment interruptions for patient realignment to ensure proper target coverage 27^{27}. The clinical significance of localization inaccuracies in frameless stereotactic radiosurgery (SRS) remains a subject of debate. For critical applications—such as treating acoustic neuromas, pituitary adenomas, trigeminal neuralgia, or lesions adjacent to vital structures—frameless SRS methods have not yet achieved the same level of confidence as frame-based approaches. To mitigate this concern, image-guided radiotherapy (IGRT) is commonly used to monitor patient positioning, complemented by 6D robotic couch adjustments to reduce setup errors in frameless SRS 31−33^{31-33}. However, IGRT is typically performed intermittently and does not compensate for real-time patient motion deviations during radiation delivery. In general, the accuracies of masks are not suitable for deep tumors located near critical structures, such as in trigeminal neuralgia cases.

One recent method that allows highly accurate human head positioning without the use of head frames or thermoplastic masks is through the use of a 6-degree-of-freedom (6DoF) robotic head holder 35,36^{35,36}. By performing real-time 6DoF patient head motion tracking through a suitable camera system that is used as an input into a control system, it was demonstrated that a Stewart platform (hexapod) type of robot can be used to continuously correct the head target position both translationally and rotationally during treatment (Fig. 2). The system efficacy was evaluated on healthy volunteers and indicated that head motion remained under a 0.5mm/0.5deg threshold > 99% of the time, compared to less than 10% of the time without the use of robotic stabilization demonstrating that the robotic SRS has potential to match or exceed the accuracies of frame

Figure

Fig. 1. SRS immobilization. Top row: Traditional frame-based SRS setup where a metal ring is fixed to the patient's skull using four screws 26^{26}. Bottom row: Frameless SRS immobilization. A triple mask configuration to minimize head movement: (1) bottom mask, (2) intermediate mask with chin strap, and (3) top masks 34^{34}.

based procedures.

Currently, robotic head position stabilization systems do not consider the biokinematics of the head and neck, but rather treat the head as a rigid body object that can freely move in space. This control method can therefore lead to robot head correction trajectories that do not follow natural head-neck motion and can be uncomfortable for patients. The use of an accurate

Figure

Fig. 2. Robotic SRS prototype head motion stabilization for volunteers using IR marker tracking head motions. Camera inputs are automatically transformed and 6DoF errors are determined. A cancellation trajectory is calculated, and motion to accomplish cranial stabilization is transmitted to the robot's linear actuators.

biomechanical model of the head and neck could allow for improved patient comfort and more effective robotic head motion correction. However, such kinematic modeling of the human head and neck is a complex subject that requires a multidisciplinary approach, combining knowledge of anatomy, biomechanics, and engineering. The study 37^{37} aimed to examine the three-dimensional kinematics of the cervical spine during maximal passive head rotation, with a focus on coupled motion, using low-dose CT scans of twenty healthy volunteers in both neutral and maximum axial rotation positions. The study employed the finite helical-axis method to describe the 3D motion between the discrete positions of each separated vertebra. In another work, investigating the spine’s role during walking, a 3-dimensional multi-segment kinematic spine model 38^{38} was developed for noninvasive analysis of spinal motion during walking. Validation of the kinematic model was carried out by comparing the calculated angles to the known angles from a mechanical model. The study involved collecting data on the motion of the spine from volunteers walking at their self-selected speeds. Kinematic modeling can also be useful in the development of devices aimed at addressing issues in the head and neck region. A novel study 39^{39} was reported where a dynamic neck brace was utilized to analyze the head movements of patients with amyotrophic lateral sclerosis (ALS) while also recording the surface electromyography (EMG) of their neck muscles. The ALS participants were instructed to perform three head-neck movements in a single plane: flexion-extension in the sagittal plane, lateral bending in the coronal plane, and axial rotation in the transverse plane. These movements were repeated five times at self-selected speeds. There are also many studies on occupational neck disorders as a result of the workforce shifting from industry to sedentary office work, and on the mechanical behavior of the cervical vertebra under normal conditions and under conditions that result in minor or major injuries 40−42^{40-42}.

In this work, we present results from a healthy volunteer study investigating head and neck reaction to a 6DoF robot that is aimed for potential use in SRS radiation therapy. Optical camera monitoring was used to track the real-time response of the human head to various translational and rotational input trajectories induced by a compact 6DoF robot located behind the head. Analysis of motion data allowed for determination of general kinematic parameters as needed for the development of a head-and-neck biomechanical model. Knowledge of the head's dynamic motion constraints is an important considerations for the interaction between humans and robots, not only in the medical space but potentially in other research areas such as the development of head motion control through robotic exoskeletons.

II.. METHODS AND MATERIALS

A. Robotic Head Motion Control System

The robot utilizes a parallel kinematic approach and is equipped with six linear actuators to enable full 6DoF head control, including translations in x, y, z, and rotations in pitch, yaw, and roll. The six extendable legs are positioned in a precise geometrical configuration, providing several advantages over serial robotic systems. The motion system features a platform with a radius of 75 mm and a stationary height of 183 mm, and provides a motion workspace of ±15\pm15 mm along the XYZ axes and ±15∘\pm15^{\circ} in pitch, roll, and yaw 43^{43}. These advantages include the distribution of weight load across all legs, resulting in high compactness, and the division of overall positional error among all linear actuators.

Real-time 6DoF head motion tracking is accomplished using an infrared (IR) stereoscopic camera (Polaris, NDI, Canada) with IR markers attached to the volunteer's head 44^{44}. Real-time 6DoF monitoring of head motion was performed at a 12 Hz sampling frequency. The head motion, as tracked by the four reflective IR markers, was subsequently transmitted automatically to an in-house developed LabVIEW (National Instruments) code, which controls the robotic device shown in the volunteer setup in Figure 2. This control code communicates between the IR camera system and the robotic device simultaneously, transmitting the six linear actuator trajectories using an 8-axis stepper motor controller (PCI-7358, National Instruments) 43^{43}. Movement of the robot produces a corresponding response motion in the volunteer's head by moving the cranial support structure (Fig. 3). The tracked motion is then input into a control computer and analyzed using a fast iterative closest path control algorithm that is then sent to the robot to perform head positioning.

In this study, the head support consisted of four cushioned pads angled to help ensure contact with the volunteer. These pads were each connected to the robot's mobile top platform and were free to swivel around the AP axis to provide improved universal contact between volunteers while maintaining sufficient levels of comfort. The plastic posts were designed to be easy to install and remove, making them convenient for use with different head sizes of volunteers. The slope of the plastic posts was 45∘45^{\circ}. Fig. 3 shows a schematic diagram of the head support system.

Figure

Fig. 3. A configuration of robotic head support with four plastic posts for facilitating head rotational control with the robot. A simple pillow support with only contact in the occipital region would not generate enough torque to rotate the head.

Figure

Figure

Figure

Fig. 4. Different coordinate frames used for radiation delivering and position measuring for a patient in a LINAC system. Left: A LINAC machine shown in a treatment room (image source: https://radformation.com). Middle: the coordinate frames related to the IR camera that tracks the patient's position are displayed. Right: the coordinate frame of a human lying on their back is defined.

B. Coordinate frames of LINAC, Target, Camera and IR tools

A medical linear accelerator (LINAC) is a device used to deliver external radiation beams for patient treatment. The LINAC coordinate system, with its origin at the isocenter, is defined to facilitate precise targeting. A mock tumor target's 6D translational and rotational position within this system can be tracked using a camera system equipped with an IR marker tool (see Fig. 4). Understanding the relationship between the four coordinate frames is crucial for enabling robotic systems to generate the required translational and rotational movements of a target point in the human head. The following subsections outline key kinematic computations and control algorithms used in the robotic system. The following notation to specify the 6D position of the IR tool in the LINAC frame: (Rli,ℓli)(R_l^i, \ell_l^i). Here RliR_l^i is the rotation matrix of the IR tool in LINAC frame (where the subscript ll represents LINAC frame, and the superscript ii represents the IR tool), and ℓli\ell_l^i is the IR tool translational position (xyz) in LINAC frame.

  1. Calibrating the orientation of IR Camera to LINAC with translational motion: One way of calibrating IR Camera and LINAC is rigidly attach an IR tool to the robotic platform, control the platform to make a pure translational motion in LINAC frame, and measure IR tool positions, see the top plot of Fig. 5.

Let the treatment table make a translational motion, and let the IR Camera record its positions at each point. Suppose these mm points in the LINAC frame are L∈R3×mL \in \mathbb{R}^{3 \times m}, and the corresponding IR Camera measurements are P∈R3×mP \in \mathbb{R}^{3 \times m}. For LL compute a mean vector lˉ∈R3×1\bar{l} \in \mathbb{R}^{3 \times 1}, then horizontally concatenate lˉ\bar{l} to generate a mean matrix Lˉ∈R3×m\bar{L} \in \mathbb{R}^{3 \times m}. Similarly, generate the mean vector pˉ\bar{p} and the mean matrix Pˉ\bar{P} for the matrix PP.

C=L−Lˉ,D=P−Pˉ,C=L-\bar{L},\quad D=P-\bar{P}, E=CD′,[U,S,V]=svd(E)E=CD^{\prime},\quad[U,S,V]=svd(E) Rlp=U[10001000det⁡(UV′)]V′R_{l}^{p}=U\begin{bmatrix}{{{1}}}&{{{0}}}&{{{0}}} \\{{{0}}}&{{{1}}}&{{{0}}} \\{{{0}}}&{{{0}}}&{{{\det(UV^{\prime})}}}\end{bmatrix}V^{\prime}

Figure

Fig. 5. Calibrating the IR Camera frame to the LINAC frame. Top: calibrating the orientation of two frames with translational motion. Bottom: determining translational offset of two frames with rotational motion.

A virtual coordinate frame can be defined in which the orientation of the xyz axes matches that of the LINAC frame, but the origin is positioned at the virtual point (origin) of the IR tool when the treatment table is at the isocenter. That is, Rvp=RlpR_v^p = R_l^p. The vector vvvp\mathbf{v}_v^{vp} in the virtual frame is ℓvp=lˉ−Rlppˉ\ell_v^p = \bar{l} - R_l^p \bar{p}.

For IR tool position specified by (Rpi,ℓpi)(R_{p}^{i}, \ell_{p}^{i}) in IR Camera frame, it can be specified by (Rvi,ℓvi)(R_{v}^{i}, \ell_{v}^{i}) in the virtual frame as following,

Rvi=RvpRpi=RlpRpi,R_{v}^{i}=R_{v}^{p}R_{p}^{i}=R_{l}^{p}R_{p}^{i}, ℓvi=Rvp(ℓpi−ℓpv)=Rlp(ℓpi−ℓpv)=Rlpℓpi+ℓvp\ell_{v}^{i}=R_{v}^{p}(\ell_{p}^{i}-\ell_{p}^{v})=R_{l}^{p}(\ell_{p}^{i}-\ell_{p}^{v})=R_{l}^{p}\ell_{p}^{i}+\ell_{v}^{p}

Align the target to LINAC frame, and consider this is target zero position in LINAC frame. Record the initial (Rˉvi,ℓˉvi)(\bar{R}_{v}^{i}, \bar{\ell}_{v}^{i}). The target position in Virtual frame (Rvt,ℓvt)(R_{v}^{t}, \ell_{v}^{t}) can be calculated by Rvt=Rvi(Rˉvi)′R_{v}^{t} = R_{v}^{i}(\bar{R}_{v}^{i})', ℓvt=ℓvi−Rvtℓˉvi\ell_{v}^{t} = \ell_{v}^{i} - R_{v}^{t} \bar{\ell}_{v}^{i}.

  1. Computing the coordinates of the virtual point of the IR tool within the LINAC frame: Determining the coordinates of the virtual point (denote as ℓlv\ell_l^v) of the IR tool in the LINAC frame is an important intermediate step for computing the target position relative to the robot, see the bottom plot of Fig. 5. To obtain these coordinates, we perform pure rotational motions around the origin of the LINAC frame, and record corresponding (Rvt,ℓvt)(R_v^t, \ell_v^t). For each rotation Rvtℓlv−ℓlv=ℓvtR_v^t \ell_l^v - \ell_l^v = \ell_v^t. Suppose nn pure rotations are performed. Vertically concatenate all (Rvt−I3)(R_v^t - I_3) and obtain a matrix A∈R3n×3A \in \mathbb{R}^{3n \times 3}, and similarly vertically concatenate all ℓvt\ell_v^t and obtain a b∈R3n×1b \in \mathbb{R}^{3n \times 1}. Then ℓlv=A\b\ell_l^v = A \backslash b. For IR tool position specified by (Rpi,ℓpi)(R_p^i, \ell_p^i) in IR Camera frame, it can be specified by (Rli,ℓli)(R_l^i, \ell_l^i) in LINAC frame as following,
Rli=Rvi,ℓli=ℓvi+ℓlv(1)R_{l}^{i}=R_{v}^{i},\quad\ell_{l}^{i}=\ell_{v}^{i}+\ell_{l}^{v} \tag{1}
  1. Calibrate IR Camera to LINAC: With the algorithms in the subsections II.B.1. and II.B.2., IR camera coordinate frame can be calibrated to LINAC frame. For IR tool position specified by (Rpi,ℓpi)(R_p^i, \ell_p^i) in IR Camera frame, it can be specified by (Rli,ℓli)(R_l^i, \ell_l^i) in LINAC as following,
Rli=RlpRpi,ℓli=Rlpℓpi+ℓvp+ℓlv(2)R_{l}^{i}=R_{l}^{p}R_{p}^{i},\quad\ell_{l}^{i}=R_{l}^{p}\ell_{p}^{i}+\ell_{v}^{p}+\ell_{l}^{v} \tag{2}

Since the orientation of the Virtual frame and LINAC is the same, one has ℓlp=ℓvp+ℓlv\ell_l^p = \ell_v^p + \ell_l^v.

Figure

Fig. 6. Target tracking by using IR tool. Top: During volunteer setup, the mock target on the volunteer's head is aligned with the LINAC frame. Bottom: The displacement of the target is monitored in the LINAC frame using the IR tool's 6D position tracking.

  1. Tracking target : IR tool was rigidly attached to a volunteer head for position tracking. During setup, as shown in the top plot of Fig. 6, one can first align the orientation and origin of a mock target within a volunteer's head to the LINAC frame, then record the IR tool position at that moment, and denote as (Rˉli,ℓˉli)(\bar{R}_{l}^{i}, \bar{\ell}_{l}^{i}). During target tracking as in the bottom plot of Fig. 6,
Rli=RltRti,ℓli=vllt+vlti=ℓlt+RltℓtiR_{l}^{i}=R_{l}^{t}R_{t}^{i},\quad\ell_{l}^{i}=\mathbf{v}_{l}^{l t}+\mathbf{v}_{l}^{t i}=\ell_{l}^{t}+R_{l}^{t}\ell_{t}^{i}

where (Rti,ℓti)(R_t^i, \ell_t^i) represent the IR 6D position in the target frame. Since target and IR tool are rigid connected, (Rti,ℓti)=(Rˉli,ℓˉli)(R_t^i, \ell_t^i) = (\bar{R}_l^i, \bar{\ell}_l^i). Then the target position in

LINAC frame, (Rlt,ℓlt)(R_l^t, \ell_l^t), can be calculated by Rlt=Rli(Rˉli)′R_l^t = R_l^i(\bar{R}_l^i)', ℓlt=ℓli−Rltℓˉli\ell_l^t = \ell_l^i - R_l^t\bar{\ell}_l^i.