Velocity Optimization via Thermal Imaging Enhances Autonomous Tumor Resection

Naveed D. Riaziat, Jiawei Ge, Ethan Kilmer, Michael W. Mikula, Paul Maria Scheikl, Michael Kam, Leila J. Mady, Axel Krieger, Jeremy D. Brown

Abstract

Objective: To demonstrate and evaluate THERMO, a thermal-velocity optimization algorithm for autonomous electrosurgery, in ex vivo porcine tissue using the UR10e-based ASTR system. Methods: Thermography for Electrosurgical Rate Modulation via Optimization (ThERMO), a closed-loop algorithm that uses real-time thermal imaging data to minimize thermal injury while ensuring accurate cut margins, was tested on ex vivo porcine tissue. Histological analysis was performed to detect thermal injury extent at a cellular level, and 3D photogrammetric measurement was used to verify margin accuracy. ThERMO's performance was compared against two previously used fixed velocity methods. Results: ThERMO outperformed both fixed velocity comparison methods without requiring data-driven training or iterative tuning, reducing thermal injury by up to 52%, improving cut accuracy by up to 59%, and maintaining

task completion speed. Conclusions: Closed-loop thermal feedback via ThERMO improves autonomous robotic electrosurgery by simultaneously reducing thermal injury and improving margin accuracy, reinforcing its value over fixed velocity approaches for tumor resection applications. Significance: This work advances autonomous robotic tumor resection by demonstrating that real-time closed-loop thermal feedback control can substantially reduce collateral tissue damage and improve surgical margin accuracy.

1 Introduction

Oral cavity cancer (OCC) is cancer of the mouth, tongue, lips, and cheeks. Approximately 377,000 new OCC cases are diagnosed each year globally, leading to 178,000 deaths. Tobacco use, alcohol consumption, and human papillomavirus (HPV) are risk factors for developing OCC. Over 90% of OCC cases are squamous cell carcinoma, or cancer of the cells that line the outer layer of the tongue, cheeks, gums, and throat [1]. For treatment of pre-metastatic oral squamous cell carcinoma (OSCC), surgical intervention is often the most effective treatment [2].

The standard surgical approach is the partial glossectomy procedure, where the portion of the tongue consisting of the OSCC tumor is removed. Cancer recurrence rates are minimized if the surgical margins (the region between the tumor and the cut boundary) reach 5 mm [3]. The term negative margin indicates that no cancer cells were found in the margin, while positive margin indicates the opposite. Though positive margins increase

*Laboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, Maryland (email: nriazia1@alumni.jh.edu).

†^{\dagger}Department of Pathology, Johns Hopkins Hospital, Baltimore, Maryland

‡^{\ddagger}Department of Otolaryngology–Head and Neck Surgery, The Johns Hopkins University School of Medicine, Baltimore, Maryland

Funding for this project was provided by the Link Foundation Modeling, Simulation, and Training Fellowship, ARPA-H grants AY1AX000023 and D24AC00415, and NIH Grant P30CA006973.

This work has been submitted to the IEEE for possible publication. Copyright may be transferred without notice, after which this version may no longer be accessible.

Figure

Zivid 2+

Thermal Camera

Porcine Tongue

Figure 1: The ASTR system is an autonomous robotic surgical system for conducting autonomous electrosurgical tumor resection on ex vivo tissue. It consists of a UR10e arm (Universal Robots, Odense, Denmark), a Zivid 2+ RGB-D camera (Zivid, Oslo, Norway), and an electrosurgical system. A T3 Pro thermal camera (InfiRay, Yantai, China) is attached to the tool flange to provide thermal feedback for ThERMO. The porcine tongue sample is placed on the grounding pad on the table and the UR10e autonomously performs tumor resection using information from the Zivid and T3 cameras.

recurrence risk, excess negative margins indicate excess healthy tissue removal and worsen post-operative recovery and quality of life [4, 5]. Therefore, it is critical to ensure accurate surgical margins during tumor resection.

Electrosurgery is a common technique for removing tumor tissue, wherein a surgical tool conducts electricity to generate heat in tissue, thereby causing it to denature. The heat-induced tissue denaturation reduces mechanical cutting effort while controlling bleeding. Monopolar electrosurgery directly heats the tissue by passing current from the tool tip through the body to an alternate-site grounding pad. Excess application of electrosurgery, however, may cause trauma to the lingual nerve in the tongue and permanently impair tongue sensory function. Insufficient application, by contrast, reduces cut accuracy by causing tissue snagging, which negatively impacts surgical margins and outcomes. As such, electrosurgery requires surgeons to diligently monitor tissue conditions to prevent adverse outcomes.

Monitoring and safely controlling electrosurgery becomes increasingly difficult in the tight constraints of the oral cavity. Consequently, robotics has provided advances in ergonomics and visualization during oral surgeries [6, 7]. Systems like the da Vinci Single Port (Intuitive Surgical Inc., Sunnyvale, California, USA) have made trans-oral robotic surgery (TORS) possible by reducing tool size while maintaining high-quality stereoscopic vision. The application of robotics to head and neck surgery offers the possibility of procedural enhancements with sensing, feedback, and automation.

Though current commercial soft tissue surgical robots have little to no autonomous capability, increased electrosurgical autonomy could greatly reduce surgeon workload and improve surgical outcomes [8]. In the few early demonstrations of autonomous tumor resection has been an active research field, a discrepancy exists between the surgeon's technique and the robot's. While experienced surgeons adjust their cutting speed based on visual feedback of cut accuracy and thermal injury, current autonomous electrosurgical robots do not.