When Magnitude Suffices: Reconstructing Time-Domain-Gated Microwave Signals Without Phase on a Thoracic Phantom

Maciej Cichoń, Patrycja Gołaś, Maksymilian Bartosik, Angelika Walczak, Agata Bielecka-Dąbrowa, Maciej Ślot

Abstract

Time-domain gating separates the arrivals that form a microwave S-parameter measurement. It requires the measurement of complex values, so this application almost always uses a vector network analyzer (VNA). We show that, on one calibrated measurement configuration, the gated magnitude the VNA would give can be reconstructed from a magnitude-only sweep of a spectrum analyzer with a tracking generator (SA). We select the gate center for separability of the measured states, correct each instrument once by a per-bin subtraction of a reference state, and train a fully connected network on sweeps from both instruments. The test object is a torso phantom with lung-shaped compartments whose filling takes eleven dielectric states, each held out in turn. On S11S_{11} the reconstruction from SA input reaches a shape error of 0.08 dB20.08 \, \text{dB}^2 with a state ordering of 0.81. This matches the reconstruction from VNA input. On S21S_{21} it reaches 0.82 dB20.82 \, \text{dB}^2 and keeps the ordered groups of the states. An explicit phase prediction performs much worse, because gating needs the phase to about 0.2 rad RMS. Across eleven measurement configurations the dependence of phase on magnitude is strong for reflection and weak for transmission. This explains the difference between S11S_{11} and S21S_{21}. Where the magnitude carries enough of the phase, a scalar instrument gives the gated response without any phase measurement after installation. On the open UM-BMID breast data set, with whole phantoms held out, the shape error of the network is 30 percent lower than that of a per-bin linear map.

Keywords: magnitude-only measurement; scalar network analysis; time-domain gating; spectrum analyzer; microwave sensing; thoracic phantom; neural network regression

1 Introduction

Microwave measurements in the low-GHz band respond to the dielectric state of tissue. Soft tissues differ enough in permittivity [1] that fluid collecting in the thorax changes the S-parameters of an antenna pair held at the body surface. The radiation is non-ionizing, and the measurement needs no shielded room and no specialized imaging staff. Systems built on this principle range from multi-antenna torso scanners [2, 3] to deep-tissue tumor detection [4, 5, 6] and to the simplified body-mounted transmission geometry that has reached clinical tests for pulmonary congestion [7, 8, 9]. Microwave assessment of lung water was demonstrated as early as 1979 [10], wideband systems for non-invasive monitoring of pulmonary edema followed [11], and the field is surveyed in reviews of electromagnetic torso scanning [12] and of microwave medical sensing and imaging [13].

With the antennas outside the body, the receiver sees several arrivals at once: the antenna's own mismatch, returns from the cabling, energy that travels around the body, and the part that has passed through the region we want to interrogate. They arrive at different times, but at each frequency the

*Corresponding author: maciej.cichon@fis.uni.lodz.pl instrument reports their sum, so in the frequency domain they cannot be told apart. Time-domain gating is the standard way to separate them: the complex S-parameter is transformed into the time domain; a window keeps the interval that holds the arrival of interest and the rest is discarded. Gating is therefore what makes the measurement interpretable, and it is also what sets the instrument. Time-domain gating requires the complex S-parameter spectrum, including both magnitude and phase. In conventional measurements, this information is provided by a vector network analyzer (VNA). The requirement for a VNA therefore arises from the signal-processing procedure rather than from the underlying physical interaction itself. This raises a practical question: how firm is this requirement?

The answer depends on how much of the phase the magnitude already carries. The two are measured separately, but they are not free of each other. The impulse response of one of these measurements is short. A few arrivals, each with a delay and an amplitude. Those few numbers generate the whole spectrum at every frequency. In the limiting case of a minimum-phase system, the phase can be derived from the magnitude response through the Hilbert-transform relation [14]. A measurement that contains a delayed arrival is not minimum phase, so the magnitude does not fix the phase completely. How close it comes depends on the antennas, the geometry, and which arrivals compete. It is a property of one measurement configuration, not a constant, and it can be measured before any reconstruction is attempted. Where the magnitude comes close enough, the gated output can be recovered from a magnitude-only sweep, and the phase never has to be measured again after installation.

What is gained by not measuring phase is more than the price difference between the two instruments. Phase is the demanding half of a network measurement. It needs a coherent reference channel, a vector calibration to fix a reference plane, cables whose electrical length does not change when they are moved or warmed, and a sweep slow enough for the source to settle at every point. None of this is needed for magnitude. Each of them is also a way for an instrument in the field to drift out of specification without the operator seeing anything wrong in the data. A spectrum analyzer (SA) equipped with a tracking generator can provide a frequency-dependent magnitude response without directly measuring phase and without requiring a coherent reference channel. Due to this fact it does not need through-calibration and has no reference plane to lose. It reports only ∣S(f)∣|S(f)|, so a raw SA sweep cannot be gated. In this work we show how to remove this limitation in the context of lung fluid detection in lossy phantom measurements.

Scalar network analyzers and spectrum analyzers with a tracking generator report ∣S(f)∣|S(f)| without a reference channel. The classical method from such data to a complex result is the minimum-phase relation [14]. In this relation the phase follows from the magnitude through the Hilbert transform when the system has no excess delay. A measurement that contains delayed arrivals is not minimum phase. Section 6 reports what the relation gives in that case.

Analytical methods recover error-corrected two-port S-parameters from uncalibrated measurements under stated conditions [15]. Reference [16] propagates the uncertainty of a gated S-parameter through the inverse transform and the gate. That result gives the uncertainty against which a reconstructed gated response must be compared. Reference [17] maps received signal strength to the permittivity of liquids with a learned model. In that work every state is present in training. In this work each state is held out in turn. The target is the gated response. Section 7 reports the dependence of phase on magnitude that we measured on eleven configurations.