Abstract—Robotic vascular intervention can reduce radiation exposure to physicians and improve manipulation stability, but effective guidance remains challenging because conventional fluoroscopic images provide limited spatial information. This paper developed a vascular intervention interactive system that connects a real robotic master controller with the SOFA(Simulation Open Framework Architecture) physical simulation platform and reused successful operation trajectories as visual guidance in physical phantom experiments. Experiments were conducted to validate the master-to-SOFA control loop and to compare three guidance conditions: no guidance, centerline guidance, and successful-trajectory guidance. The control validation results showed reliable command delivery under velocity-mode operation. In the physical phantom experiment, centerline guidance showed no obvious improvement over no guidance. In contrast, successful trajectory guidance provided a clear performance benefit. Compared with the no-guidance condition, it reduced task completion time, collision number, rollback number, and rotation number by approximately 66.8%, 34.3%, 55.6%, and 66.7%, respectively. These results indicate that successful trajectories obtained under instrument-vessel interaction can provide practical guidance cues, offering a simple and effective visual assistance strategy for robotic vascular intervention experiments.

Index Terms—Vascular intervention robot; master-follower control; trajectory record; visual guidance.

I. INTRODUCTION

INTRAVASCULAR intervention has become an effective treatment for cardiovascular and cerebrovascular diseases, owing to its advantages of minimal trauma, reduced postoperative complications, and rapid recovery [1]–[4]. During such procedures, catheters and guidewires are advanced through narrow and tortuous vessels to reach target lesions under fluoroscopic or digital subtraction angiography (DSA) guidance.

Research supported by the Shenzhen "Medical Engineering Integration" Special Support Program Project under Grant F-2024-Z99-502365, Shenzhen, China; in part by the High level of special funds from Southern University of Science and Technology under Grant G03034K003, Shenzhen, China. (Corresponding author: Shuxiang Guo.)

S. Guo is with the Department of Electronic and Electrical Engineering, Southern University of Science and Technology, Shenzhen, Guangdong 518055, China, and also with the School of Life Science and the Key Laboratory of Convergence Medical Engineering System and Healthcare Technology, Ministry of Industry and Information Technology, Beijing Institute of Technology, Beijing 100081, China. (e-mail: guo.shuxiang@sustech.edu.cn).

M. Zou and S. Cao are with the Beijing Institute of Technology, Beijing 100081, China.

P. Li is with the Department of Electronic and Electrical Engineering, Southern University of Science and Technology, Shenzhen, Guangdong 518055, China.

J. Guo is with Shenzhen Institute of Advanced Biomedical Robot Company Limited, Shenzhen 518000, Guangdong Province, China.

B. Wang and M. Ding are with Aerospace Center Hospital and the Key Laboratory of Convergence Medical Engineering System and Healthcare Technology, Beijing Institute of Technology, Beijing 100081, China.

However, conventional manual intervention highly depends on the operator's experience, tactile sensitivity, and hand-eye coordination. Maintaining stable and precise manipulation during long procedures remains challenging. In addition, interventional physicians are exposed to ionizing radiation for prolonged periods, which poses non-negligible occupational health risks [5], [6]. To address these limitations, vascular interventional robotic systems have been developed. By allowing operators to manipulate instruments from a remote console behind radiation shielding, robotic systems can reduce radiation exposure, improve motion stability, and enhance manipulation precision [7].

Although robotic systems can improve the stability and repeatability of catheter and guidewire manipulation, physicians still need to interpret fluoroscopic or digital subtraction angiography images and continuously determine the spatial relationship between the instruments and the surrounding vasculature. Reinok et al [8]. proposed an automated passive tracking method for MR-guided endovascular interventions that detects discrete passive markers on an MR-visible guidewire in real time. The detection results are integrated into a custom software interface, offering a cost-effective, radiation-free, and near-real-time guidance assistance method for MR-guided interventions. Zhang et al [9]. proposed HD-CAR, a 3D-2D registration network for abdominal aortic vessels that aligns preoperative CTA-based 3D vascular models with intraoperative DSA images. This method effectively compensates for missing vascular structures in single-frame DSA images, providing more precise imaging guidance and assistance for endovascular aneurysm repair procedures. However, these methods do not provide a structured opportunity for less experienced operators to explore and practice surgical measures in advance. The experience gained from successful virtual surgeries has not been transferred to subsequent actual surgeries. This limitation is important for young or less experienced physicians. Before performing an actual surgery, they may need to understand not only the anatomical route, but also how the guidewire and catheter should be advanced, rotated, retracted, and coordinated when passing through difficult vascular regions.

A successful surgical trajectory is more than a geometric curve. It implicitly records the physical response of the instrument during guidewire-vessel interaction, including bending, contact-induced deviation, passive compliance, and feasible advancement direction under operational constraints. If such successful trajectories can be collected and reused, they may provide operators with prior knowledge that is unavailable from vessel geometry alone. This successful trajectory con-

Master-to-SOFA Communication Middleware

SOFA Vascular

Intervention Simulation

Figure

Fig. 1. Overview of the interactive intervention system.

training prior knowledge can serve as a practical guidance reference, reflecting not only where the guidewire should go, but also how a successful route progresses through the vascular structure under physical constraints.

This study aims to enable doctors to conduct virtual training before actual surgery and transform successful operational experience into interpretable visual references. The main contributions are summarized as follows. First, a vascular intervention interactive system is developed by connecting a real robotic master controller with the SOFA physical simulation platform, enabling real-time guidewire/catheter manipulation and complete trajectory recording. Second, a successful trajectory acquisition and reuse method is proposed, in which successful trajectories containing instrument-vessel interaction information are used as visual guidance, mapping the recorded 3D successful trajectory onto real-time images of a physical vascular phantom. Finally, comparative experiments are conducted under the same vascular model, including no-guidance operation, centerline-guidance operation, and successful-trajectory-guidance operation, to evaluate the effectiveness of the proposed guidance strategy.