arXiv:2610.19413cs.ROOct 2026

Manuscript Offprint

arXiv:2610.19413

A VR Training System for Robot-Assisted Vascular Intervention with Dual-Mode Control and Model-Based Interaction Response

Yining Shang, Shuxiang Guo, Sheng Cao, Pengcheng Li, Bin Wang, Mingchao Ding

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Yining Shang, Shuxiang Guo, Sheng Cao, Pengcheng Li, Bin Wang, and Mingchao Ding

Abstract—Virtual reality training for robot-assisted vascular intervention requires accurate robotic control and effective feedback on guidewire–vessel interactions. This paper presents a VR training system combining asynchronous dual-mode command mapping with sensorless interaction estimation. The system supports velocity-controlled navigation and position-driven incremental manipulation. Guidewire deformation and a Hunt-Crossley-type contact response are combined and temporally filtered to estimate interaction intensity, while trajectory-based visualization supports post-procedure review. Tests confirmed distinct command-dispatch patterns across three velocity gears and the position mode. During hybrid-scene navigation, the estimated response remained low during straight insertion and increased as the guidewire bent within the vascular model. Trajectory analysis yielded a model-derived cumulative trauma work (CTW) value of 15.166 in normalized units and localized regions of elevated interaction. In a preliminary study with 15 novices, physical-phantom task success increased from 53.3% to 93.3% after VR training, while mean completion time decreased from 132.4 s to 96.7 s. These results demonstrate the system's dual-mode operation, deformation-sensitive feedback, and potential to improve subsequent physical-phantom task performance.

Index Terms—Vascular interventional surgery, virtual reality training, guidewire simulation, virtual interaction estimation, dual-mode control.

I. INTRODUCTION

VASCULAR interventional surgery (VIS) enables minimally invasive treatment of cardiovascular and cerebrovascular diseases. Robot-assisted VIS (RA-VIS) can improve manipulation precision and reduce occupational radiation exposure [1]–[3]. However, navigating flexible guidewires through tortuous vessels remains demanding because operators must coordinate instrument motion while interpreting guidewire–vessel interactions. This creates a need for training systems that reproduce the control characteristics of robotic platforms and provide interpretable interaction feedback.

Previous studies have investigated robotic delivery [4], force- and torque-feedback interfaces [5], virtual-to-real autonomous navigation [6], curvature-aware delivery control [7], and catheter interaction simulation [8]. Simulation and extended-reality environments have also been investigated for endovascular training and performance assessment [9]–[12]. Although these studies have advanced individual aspects of robot-assisted intervention, few have established an integrated physical-leader–virtual-follower framework that connects robotic manipulation, virtual vascular simulation, interaction feedback, and post-procedure spatial review within one VR training workflow. In particular, reproducing the visual workflow alone does not ensure that inputs from a robotic leader-side device are consistently mapped to virtual guidewire motion or that trainees can interpret the effects of their manipulation.

Two technical gaps remain. First, operator input, command dispatch, physics simulation, and visual rendering may operate at different update rates. Feedback-control and haptically enhanced interfaces have been studied for intervention teleoperation [13], [14], but direct event-by-event forwarding in a hardware-connected simulator may produce uneven actuation under asynchronous execution. This issue becomes more complex when continuous velocity commands and discrete position increments coexist within the same procedure.

Second, interaction feedback in vascular intervention training is difficult to obtain without dedicated force sensors. Guidewire deformation and tip–vessel proximity provide complementary cues, but both are affected by numerical noise, sliding motion, and intermittent contact. Existing deformable-body simulation, catheter interaction modeling, and surgical haptic-feedback methods [8], [15]–[17] provide useful foundations, but they do not directly provide a unified sensorless interaction representation for training. Moreover, instantaneous feedback alone does not record where interaction occurred along the trajectory or quantify its accumulated burden for post-training review. Mixed-reality visualization studies [18] motivate such spatially linked feedback.

To address these gaps, this paper presents an integrated physical-leader–virtual-follower VR training framework for robot-assisted vascular intervention. The framework reproduces the manipulation workflow of the robotic platform, provides sensorless interaction feedback during guidewire navigation, and supports post-procedure spatial review. Within this framework, a multi-cue interaction representation method combines distributed guidewire deformation with local guidewire–vessel proximity to estimate relative interaction intensity without a force sensor. The main con-

Y. Shang, S. Guo, and S. Cao are with the Key Laboratory of Convergence Biomedical Engineering System and Healthcare Technology, Ministry of Industry and Information Technology, Beijing Institute of Technology, Beijing 100081, China.

S. Guo and P. Li are with the Department of Electronic and Electrical Engineering, Southern University of Science and Technology, Shenzhen, Guangdong 518055, China.

B. Wang and M. Ding are with the Department of Peripheral Vascular Intervention, Aerospace Center Hospital, School of Life Science, Beijing Institute of Technology, Beijing 100081, China.

Corresponding author: Shuxiang Guo

e-mail: guo.shuxiang@sustech.edu.cn.

Figure

Fig. 1. VR training architecture. The physical leader console drives virtual catheter and guidewire motion in SOFA. The console also supports a physical follower robot, which is outside the illustrated training loop.

tributions of this paper are:

• An integrated physical-leader–virtual-follower VR training framework with asynchronous dual-mode command mapping. The framework connects physical manipulation, virtual instrument simulation, velocity-controlled navigation, position-driven incremental manipulation, and real-time interaction feedback within a unified training workflow.

  • A sensorless multi-cue interaction representation method for vascular intervention VR training. The method combines distributed guidewire deformation with local guidewire–vessel proximity and a nonlinear virtual contact response to provide relative interaction-intensity feedback without requiring a force sensor.

  • A trajectory-linked post-procedure analysis function that records estimated interaction intensity together with guidewire-tip positions and visualizes its spatial distribution along the vascular geometry.

References (18)