Bridging the Markerless Tracking Validation Gap: Intraoperative X-ray Workflow Integration

Connor Daly, Salvatore Russo, Jinendra Ekanayake, Daniel S. Elson, Ferdinando Rodriguez y Baena

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Connor Daly 1(⊗)^{1(\otimes)}, Salvatore Russo 2^{2}, Jinendra Ekanayake 3^{3}, Daniel S. Elson 1^{1}, and Ferdinando Rodriguez y Baena 1^{1}

1^{1} The Hamlyn Centre for Robotic Surgery, Imperial College London, London, UK, cd1723@ic.ac.uk

2^{2} Department of Neurosurgery, Imperial College Healthcare NHS Trust, London, UK

3^{3} Stanford University School of Medicine, Stanford University, California, USA

Abstract. Markerless intraoperative tracking of spinal anatomy via RGB-D sensors is a promising alternative to invasive bone-anchored fiducials, yet no existing clinical dataset provides independent subsurface ground truth for validating these methods. We address this gap with a clinically viable pipeline that derives per-vertebra 6-DoF pose from routine intraoperative C-arm fluoroscopy, co-registered to the RGB-D camera frame through a multimodal fiducial array detectable in both imaging modalities. Critically, the pipeline requires no additional radiation, instrumentation, or deviation from standard surgical workflows, relying solely on the fluoroscopic images already acquired during routine level checks. A calibrated transform chain links the vertebral pose recovered via multi-view 2D–3D CT-to-fluoroscopy registration to the RGB-D coordinate frame, enabling direct comparison of any markerless registration estimate against a subsurface reference. Ex vivo validation against optically tracked ovine vertebrae confirms end-to-end accuracy of 2.06±0.292.06 \pm 0.29 mm and 3.35±0.93∘3.35 \pm 0.93^\circ, sufficient to serve as independent ground truth for methods with substantially larger errors. We demonstrate the pipeline's utility by evaluating two markerless registration strategies on clinical data spanning 11 vertebrae across 1,500 frames from three surgical sequences: the first such evaluation against an independent subsurface reference in a clinical setting. The proposed framework provides reusable infrastructure for benchmarking emerging markerless tracking methods on real surgical data.

Keywords: RGB-D Tissue Tracking · Orthopedic Surgery · 2D–3D Registration

1 Introduction and Motivation

Markerless intraoperative tracking of orthopedic anatomy via RGB-D sensors is a promising emerging direction for surgical guidance [11, 15]. Current surgical navigation systems depend on invasive bone-anchored fiducials to register (and subsequently track) preoperative imaging with the surgical scene [26, 27], introducing added operative time and complication risk [8]. Markerless approaches instead

C. Daly et al.

extract anatomically meaningful correspondences directly from intraoperative point clouds via learned segmentation, enabling registration and tracking without physical instrumentation [11, 15]. Daly et al. [5] advanced this direction with the first clinical markerless tracking dataset for spine surgery. However, their ground truth is inherently approximate: alignment relies on surgeon-indicated surface landmarks refined against intraoperative topology, neither of which references the pose of the subsurface anatomy. For clinical adoption, markerless systems must be validated against a gold-standard measurement offering quantifiable translational and rotational error metrics [1], a requirement no existing clinical dataset satisfies. Cadaveric studies have obtained accurate vertebral pose through rigidly fixed optical tracking arrays [17, 16]. Liebmann et al. [16], for example, fixed 3D-printed constructs to vertebral bodies via K-wires, achieving sub-2 mm accuracy but requiring anterior exposure incompatible with standard posterior approaches, additional tracking hardware within the operative field, and an intraoperative CT. These requirements are not permissible within a clinical observational study, leaving a methodological gap: there is currently no clinically viable means of obtaining precise vertebral pose ground truth for evaluating markerless registration in real surgical settings.

To address this, we present a pipeline that leverages routine intraoperative C-arm imaging to derive ground-truth vertebral pose within the RGB-D camera frame. Our contributions include:

  1. A multimodal data capture and labelling pipeline that co-registers intraoperative RGB-D recordings with ground-truth vertebral poses derived from routine intraoperative fluoroscopy, requiring no additional radiation, no instrumentation within the surgical field, or deviation from standard surgical workflows.

  2. The first quantitative evaluation of markerless spine registration methods on clinical data against an independent subsurface reference, reporting per-vertebra translational and rotational error across three surgical sequences.

  3. A clinically grounded evaluation framework, validated ex vivo to 2.06±0.292.06 \pm 0.29 mm and 3.35±0.93∘3.35 \pm 0.93^\circ, which can serve as reusable infrastructure for benchmarking markerless registration methods against an independent subsurface reference. Upon publication we will release CAD files for the intraoperative marker array and calibration phantom, along with the evaluation code and annotated data.

This work reports initial results from a large-scale ongoing clinical data collection study; the pipeline and evaluation framework are designed to scale as further cases are captured, enabling progressively larger-scale benchmarking of markerless methods.

References (24)