Information-based Unified Design Framework for NMR Magnetic Tracking System

Seçkin Eren Yetim, Mehmet Efe Tiryaki

Seçkin Eren Yetim 1^{1} and Mehmet Efe Tiryaki 1,2^{1,2}

Abstract—Nuclear magnetic resonance (NMR) magnetic tracking offers a promising alternative to magnetic resonance imaging (MRI)-based localization for medical robotic guidance. However, scalar field measurements and field-aligned marker moments constrain position and orientation tracking, requiring careful configuration design. We present a unified information-based framework that uses local covariance analysis to evaluate receiver geometry, marker separation, and external source configuration in on/off-board NMR tracking. Numerical evaluation shows that planar-array precision depends jointly on receiver count and geometry, with triangular lattice arrays achieving the lowest best-case position uncertainty among the evaluated layouts for 9–20 receivers. Two spatially separated, field-aligned markers enable device-axis tracking, with predicted angular uncertainty below 2∘2^{\circ} at the evaluated workspace center for suitable marker separations. We further establish point-dipole source–receiver duality, unifying the analysis of on/off-board NMR tracking. These findings provide model-based guidance for tracking configuration design.

Index Terms—Medical robots and systems, magnetic tracking, sensor design, Fisher information, observability, high-field MRI.

I. INTRODUCTION

NMR sensing has been demonstrated for patient head-motion tracking and wireless localization of medical tools in MRI scanners [1]–[3]. This approach could provide an alternative to MRI-based localization for minimally invasive robotic interventions, including steerable guidewires and robotic catheters [4]–[6]. By separating device localization from anatomical image acquisition, external NMR sensing of magnetic markers offers the potential for faster 3-D tracking while reducing the RF energy deposition and acoustic noise associated with repeated tracking scans [4], [7]. Such feedback could support robotic guidance between anatomical image updates.

NMR-based tracking can place the field sensor either outside or on the tracked object. In off-board tracking, external NMR probes measure field perturbations generated by device-mounted magnetic markers [3]. In on-board tracking, object-mounted NMR probes measure externally generated magnetic fields, as demonstrated for head-motion tracking using dedicated gradient tones or native imaging gradients [1], [2]. Both architectures measure only the scalar field perturbation along the scanner's main field to first order, so tracking precision depends on the source-sensor configuration. In off-board tracking, the strong MRI field saturates the magnetic markers and aligns their moments with the main field regardless of device orientation, so a single marker cannot reveal the device's orientation. These constraints motivate a common design framework for selecting receiver arrangements, marker geometries, and external source configurations to achieve the required position and device-axis precision. We present a unified information-based framework for configuration design under the scalar-readout and field-alignment constraints of NMR magnetic tracking (Fig. 1). Through analytical modeling and numerical covariance analysis, we:

• demonstrate that planar NMR tracking precision depends jointly on receiver count and array geometry;

• show how two spatially separated, field-aligned markers provide device-axis information for NMR tracking; and

• establish point-dipole source–receiver duality, unifying the analysis of on/off-board NMR tracking.