Selectively Actuated Fiberbots for High-precision Minimally Invasive Surgery

Jinshi Zhao, Libaihe Tian, Alex Ranne, Qindong Zheng, Yue Li, Xiaotong Guo, Ziqi Yang, Ali Anil Demircali, Mohamed E. M. K. Abdelaziz, Bruno Gil Rosa, Ferdinando Rodriguez y Baena, Burak Temelkuran

Jinshi Zhao1∗^{1*}, Libaihe Tian1∗^{1*}, Alex Ranne2∗^{2*}, Qindong Zheng3^{3}, Yue Li2^{2}, Xiaotong Guo4^{4}, Ziqi Yang1^{1}, Ali Anil Demircali1^{1}, Mohamed E. M. K. Abdelaziz2^{2}, Bruno Gil Rosa2^{2}, Ferdinando Rodriguez y Baena2^{2}, and Burak Temelkuran1^{1}

Abstract—Long, slender instruments are essential for minimally invasive surgery (MIS), but achieving precise distal manipulation becomes increasingly difficult when actuation extends along the entire instrument. Here, we present an electrothermally actuated fiber-based robot designed for MIS procedures, in which Joule heating is concentrated in the distal active segment. The fiber is fabricated using a thermal drawing process and uses embedded metallic wires for thermal actuation. This robotic fiber, 1.9 mm in diameter and 1.2 m in length, can precisely manipulate surgical laser fibers with positional precision below 100 μ\mum. The compact size of the fiber robot allows it to pass through the working channel of a flexible endoscope, while keeping both devices mechanically uncoupled to allow independent motion. A phantom demonstration combining the robotic fiber with a da Vinci instrument conceptually demonstrated its compatibility with existing endoscopic systems for clinical applications. Our robotic fiber enables precise distal motion and holds the potential to improve clinical outcomes by enhancing precision in MIS procedures.

Index Terms—Soft Robot Materials and Design, Medical Robots and Systems, Dexterous Manipulation, Flexible Robotics, Soft Robot Applications

Figure

Fig. 1. Schematic diagram of the selectively actuated fiberbot.

I. INTRODUCTION

Minimally invasive surgery (MIS) has reduced the need for large surgical incisions by enabling diagnostic and therapeutic procedures through small access ports or natural anatomical pathways. Laparoscopic, endoscopic, and catheter-based procedures have consequently expanded the range of lesions that can be reached with reduced damage to surrounding tissues. Robotic technologies can further enhance these procedures by improving instrument dexterity, motion precision, and operator control within confined anatomical environments [1], [2].

Many minimally invasive procedures require long and slender instruments to pass through access ports, flexible endoscopes, or tortuous anatomical pathways before reaching the target site. Although this geometry facilitates access to deep anatomical regions, it makes precise distal manipulation challenging. Conventional instruments are commonly controlled from their proximal ends, and their distal response can be affected by shaft deformation, friction, backlash, and interactions with the surrounding anatomy or the working channel of an endoscope. These effects become more pronounced as the instrument length and pathway complexity increase, limiting the ability to position or scan a small distal end-effector accurately. Miniaturized continuum and soft robotic instruments offer a promising means of introducing active steering and fine manipulation at the surgical site while retaining the flexibility required for minimally invasive access [3].

In this paper, we present the next generation of electrothermally actuated fiber robots (Fiberbots). These robots take the form of a long (1.2 m in length) and slender (1.9 mm in diameter) catheter, with an electrothermally actuated tip capable of achieving sub-millimeter precision motion. A selective actuation mechanism was developed by localizing the high-resistance heating element to the distal active section. The fiberbot avoids dissipating substantial electrical power along mechanically passive portions of the shaft. Fiberbots are designed to fit through the working channels of off-the-shelf endoscopes, enabling seamless integration into an operating theater. In Sect. II, we begin by investigating common existing designs for laparoscopic robots used in MIS surgery, while highlighting their limitations that need to be addressed. Then in Sect. III, we present a new fiberbot design, highlighting its novelty in selective actuation. Using

∗^{*} These authors contributed equally to this work.

1^{1}Jinshi Zhao, Libaihe Tian, Ziqi Yang, Ali Anil Demircali, and Burak Temelkuran are with the Department of Metabolism, Digestion and Reproduction, Imperial College London, SW7 2AZ, UK (Corresponding author: Burak Temelkuran, email: b.temelkuran@imperial.ac.uk)

2^{2}Alex Ranne, Yue Li, Mohamed E. M. K. Abdelaziz, Bruno Gil Rosa, and Ferdinando Rodriguez y Baena are with the Hamlyn Centre for Robotic Surgery, Imperial College London, SW7 2AZ, UK

3^{3}Qindong Zheng is with the Department of Bioengineering, Imperial College London, SW7 2AZ, UK

4^{4}Xiaotong Guo is with the Department of Biomedical Engineering, University of Glasgow, G12 8QQ, UK thermal fiber drawing technology, a bespoke polymer multilumen fiber was fabricated. This technique provides the capabilities to rapidly prototype devices with complex cross-sectional designs, while mitigating the difficulties faced by conventional extrusion or molding techniques. This section is followed by a summary of circuit assembly. Finally, in Sect. IV, we performed motion characterization to evaluate the fiberbot's precision, repeatability, and reachable workspace. We also conducted temperature characterization to examine the temperature profile under different current levels and validate the efficiency and clinical viability of the selective actuation mechanism. To demonstrate system-level feasibility, we integrated the fiberbot with a da Vinci surgical instrument and used it to perform a moving task in an endoluminal phantom. The proposed selective-actuation architecture addresses a fundamental scaling limitation of electrothermally actuated fiberbots [4]. By localizing the high-resistance heating element to the distal active section, the fiberbot avoids dissipating substantial electrical power along mechanically passive portions of the shaft, thereby improving its suitability for high-precision minimally invasive surgery.

References (24)