arXiv:2610.12780cs.ROOct 2026

Manuscript Offprint

arXiv:2610.12780

Multimodal Image-based Robotic Assistance and Guidance Environment (MIRAGE): A Novel Unified Framework for Surgical Navigation and Visualization

Ehsan Nasiri, Yi Zheng, Michael A. Kokko, Xiaoyao Fan, William R. Warner, Keith D. Paulsen, Ryan J. Halter

ABSTRACT

This paper presents the computational architecture and quantitative preclinical evaluation of MIRAGE, a multimodal image-guided robotic surgery framework for real-time visualization and navigation. It integrates stereovision scene reconstruction, augmented-reality (AR) video overlay, quantified heads-up display (HUD) navigation, and intraoperative ultrasound within the da Vinci console's TilePro display (Intuitive Surgical). Additional modules include optical/electromagnetic tracker registration, stereo calibration, fiducial rigid registration with RMSE feedback, and 3D point-cloud reconstruction. Video I/O and rendering run in a C++ backend, while a Python GUI coordinates calibration, registration, reconstruction, visualization, and user interaction through an in-process zero-copy binding layer. The framework formalizes spatial transforms, module states, and synchronization of heterogeneous video and tracking streams. Preoperative calibration uses synchronized checkerboard image pairs and optimization, with estimated intrinsics, extrinsics, and rectification reused intraoperatively. MIRAGE was validated in phantom experiments across da Vinci S,Si, Xi, and SP platforms. Across clinically relevant working distances, mean 3D reprojection error was 0.6mm, with best 0.1mm RMSE at 100mm on Xi. Results support MIRAGE as a preclinical technical-validation platform for accurate, responsive, and extensible multimodal surgical guidance.

Keywords: Medical Robots and Systems; Surgical Robotics: Laparoscopy; Visual-Based Navigation; Computer Vision for Medical Robotics

*Corresponding author: Ehsan Nasiri ehsan.nasiri@dartmouth.edu

0^{0}This work was supported in part by the Advanced Research Projects Agency for Health (ARPA-H) under award number 1AY2AX000043. The views and conclusions contained in this document are those of the authors and should not be interpreted as representing the official policies, either expressed or implied, of the United States Government.

1 INTRODUCTION

Robotic surgery has been shown to improve patient outcomes, with lower morbidity rates compared to traditional open surgery [1–6]. Image-guided surgery, also known as surgical navigation, is another form of computer-assisted intervention designed to give surgeons real-time intraoperative guidance. Its purpose is to enable precise and efficient access to target regions, support surgical planning, and minimize the risk of damage to surrounding critical structures [7–9]. Imaging modalities integrated with surgical robots include fluorescence imaging, augmented-reality overlays, and navigational displays [10,11]. In urologic surgery for example, image-guided robotic systems employing near-infrared fluorescence imaging and augmented reality overlays have enhanced visualization of tumors and critical anatomy, improving surgical decision-making and functional outcomes [12,13]. Similarly, in transoral robotic (TORS) procedures, the introduction of intraoperative imaging and navigational augmented reality helps surgeons overcome confined operative spaces and the lack of haptic feedback, thereby increasing the accuracy of tumor resection and safety near vital structures [14]. These advances, underscore a broader trend toward more precise, image-enhanced robotic interventions that improve oncologic and functional results for patients [10].

Recent development in image-guided and robotic surgery have explored a wide spectrum of visualization strategies, ranging from integral videography (IV) autostereoscopic displays for 3D overlays without head-mounted devices [15] to goggle-based and monitor-based augmented reality (AR) systems that project preoperative models into the surgical field [16,17]. In urology, numerous groups have applied computer vision and machine-learning techniques to automate 3D model registration during robotic partial nephrectomy [18–20], while systematic reviews highlight the broader use of extended reality (AR/VR) tools for planning, navigation, and education [21,22]. Similar approaches have been extended to TORS to assist with tumor localization [23,24] and to robotic prostatectomy, where AR improved anatomical orientation but remained hindered by organ motion [17]. Across these studies, the main benefits include enhanced depth perception, improved anatomical visualization, and reduced surgeon workload, whereas persistent drawbacks involve soft-tissue deformation, registration inaccuracy, high computational demands, hardware costs, and limited large-scale clinical validation.

Many image-guided robotic surgery systems are still evaluated as isolated demonstrations rather than integrated, time-synchronized navigation frameworks. This paper asks whether heterogeneous imaging, instrument tracking, modeling, reconstruction, and overlay streams can be synchronized, fused, and rendered within a unified robotic surgery framework while preserving measurable navigation accuracy and real-time responsiveness across multiple robotic platforms.

Prior multimodal guidance systems in robotic surgery (e.g., [25]) have typically addressed only subsets of the sensing/visualization stack and left key integration gaps. In robotic prostatectomy, for example, Ukimura et al. [26] reconstructed TRUS-derived, patient-specific 3D models and displayed them on the da Vinci console via TilePro, reporting favorable margin outcomes but relying on preoperative data and static alignment. Samei et al. [27] inte-grated preoperative MRI with intraoperative 3D TRUS and live 2D TRUS in a “partial AR” scene shown beneath the endoscope feed on the surgeon’s console, improving situational awareness yet still constrained by registration quality and without optical/EM tracking fusion. Fusing optical and EM tracking instrument provides complementary robustness and tighter, continuous pose estimation of the endoscope and TRUS probe in a common frame, improving overlay accuracy under real operative dynamics. In [28] authors registered ultrasound/photoacoustic images to the da Vinci stereo camera via an arc-to-line optimizer and achieved sub-centimeter target registration error (TRE) in simulation/bench experiments, but the workflow remains preclinical and depends on photoacoustics hardware. Meza et al. [29] fused freehand ultrasound with structured-light surface capture (optically tracked) to one frame with sub-mm inter-modality error, demonstrating fusion accuracy but outside the robotic console and without endoscopic integration, so the surgeon is not able to see/fuse the modalities natively in the binocular console view. Classic stereo-vision pipelines (e.g., Stoyanov [30]) deliver real-time 3D tissue reconstruction from robotic endoscopy but are vision-only and lack fusion with other modalities (e.g., without any alternative imaging/sensing modules). In contrast, we introduce a unified, integrated framework that fuses preoperative stereo calibration, dual optical–EM tracking within a heads-up display(HUD)-TilePro navigation-visualization interface, live 3D stereo reconstruction (point clouds), 3D AR overlay, and live ultrasound streaming. This integration supports dynamic intraoperative registration within a console workflow. Component calibration accuracy and end-to-end target accuracy are distinct endpoints, since the latter also depends on registration, tracking, motion, and temporal association.

The novelty of this work is not the individual use of stereo calibration, rigid registration, learned disparity estimation, or AR rendering, which are established in the literature. Instead, the contribution is a unified, time-aware, and quantitatively evaluated architecture that coordinates these components within a single console-attached robotic surgery workflow. By defining cross-modal states, timestamp association, transform propagation, buffer ownership, synchronization rules, and evaluation metrics, the framework enables reproducible testing of multimodal surgical guidance under realistic system-level constraints.

Recent consensus guidance recommends evaluating surgical robotics through staged translation rather than technical feasibility alone, spanning device performance, usability, patient acceptability, health-system impact, and long-term monitoring (IDEAL) [31]. This is directly relevant to multimodal image-based robotic assistance and guidance environment (MIRAGE), where value depends not only on accuracy and frame rate, but also on workflow integration, user trust, failure mitigation, and reproducible evaluation across platforms and procedures.

Prior systems that utilize C++-Python software frameworks largely follow three major patterns: (1) Code-generated monoliths (e.g., large simulators) auto-generate both the C++ API and the pybind11 [32] glue to form a single expansive Python module; this scales to evolving engines but introduces a generator/continuous integration (CI) toolchain and is not tailored to low-latency, heterogeneous device streams [33]. (2) Custom-designed interface over a single high-performance library deliver near-C++ speed, but typically bind one math-control stack rather than multiple device-module subsystems [34]. (3) Thin Python fronts over real-time C++ processes (e.g., shared-memory or bursting-mode toolboxes) keep hard real-time loops out-of-process in C++ and expose a minimal Python API. This helps determinism but adds orchestration and deployment complexity beyond the process boundary [35].

In the proposed multi-threaded multimodal system, most high-level modalities are implemented in Python, while video acquisition, rendering, calibration support, and selected hardware-facing modules are implemented in C++ and bound in-process via a custom-developed binding interface into a single Python orchestrator and GUI. The contribution is not the use of pybind11 alone, but the surrounding integration method: explicit buffer ownership, timestamped frame exchange, mutex-protected shared state, asynchronous worker management, and a unified state machine that coordinates C++ and Python modules. This design binds multiple heterogeneous C++ subsystems alongside Python-native modalities under one event loop, prioritizes in-process frame access over out-of-process shared memory, and exposes C++ frame buffers to Python through the buffer protocol when copy-free access is possible. Long-running C++ operations are isolated from the GUI event loop through worker threads and status callbacks. The system maintains a consistent timebase and timestamping policy to support Python-side data fusion and quantitative synchronization analysis.

Figure

Fig. 1: The overview of the framework pipeline.

The binding layer connects C++ acquisition and rendering to the Python interface through timestamped image and pose exchange, shared configuration, and worker-status callbacks.

The formulation brings the spatial transforms, timestamp association, and module interfaces into a common description of the integrated workflow. Registration and calibration use established methods [30, 36–38]; the integration analysis concerns their coupled spatial, temporal, and configuration dependencies.

Compared with general-purpose robotic middleware and application frameworks such as the robot operating system (ROS/ROS2), the da Vinci research kit (dVRK) or NVIDIA Isaac, MIRAGE is deliberately implemented as a single workstation, console-attached framework focused on image guidance rather than robot actuation. In our configuration, the robot is driven entirely by its native clinical console, while MIRAGE concentrates on integrating stereo video, tracking, and multimodal visualization-navigation on the same machine. Introducing an additional middleware layer would add serialization, inter-process communication, and dependency-management overhead without clear benefit for this tightly coupled setup. The single-process design prioritizes direct frame access and a common application state on the console workstation. Related infrastructure includes OpenIGTLink for data exchange [39], Plus for tracked-image acquisition and calibration [40], 3D Slicer for extensible medical-image computing [41], and ROS 2 for communication and component composition [42]. Table 1 compares these roles with the application scope of MIRAGE. These complementary roles motivate comparisons with equivalent tasks, hardware, data-transfer policies, and timing boundaries.

Table 1: Roles and integration scope of MIRAGE and related software frameworks. The comparison describes documented capabilities, not measured performance rankings.

FrameworkPrimary roleScope relevant to multimodal guidance
OpenIGTLink [39]Communication protocolData exchange and interoperability across image-guided therapy components.
Plus [40]US-guided intervention toolkitTracked ultrasound acquisition, calibration, reconstruction, and streaming.
3D Slicer [41]Medical-image computing platformProvides visualization, segmentation, registration, and an extensible environment for image-analysis applications.
ROS2 [42]Robotics middlewareProvides communication and component composition for robotic applications; transport and quality-of-service settings depend on the application configuration.
MIRAGEMultimodal surgical image guidance frameworkUnifies multimodal calibration, stereo vision, EM/optical tracking, tracked ultrasound, AR/HUD guidance, and 3D stereo reconstruction within a unified multimodal image-guidance framework.

The contributions of this research paper are summarized as follows:

[1.] A novel time-synchronized multimodal robotic surgery architecture that formalizes the coupling among stereo video, optical/electromagnetic tracking, AR overlay, stereo reconstruction, ultrasound, HUD navigation, and surgeon-console visualization within one state-driven framework.

[2.] A novel cross-language real-time integration method that defines explicit function mappings, buffer ownership, synchronization invariants, thread-safety rules, and state-transition models between a C++ video/rendering backend and Python-based navigation, reconstruction, registration, and GUI modules.

[3.] A temporal association and interpolation strategy for heterogeneous surgical data streams operating at different update rates, including video frames, tracker poses, rendering loops, and GUI events.

[4.] A quantitative evaluation protocol for multimodal image-guided robotic surgery systems, including calibration/reprojection error, fiducial registration error, overlay reprojection error, stereo reconstruction error, end-to-end response time, synchronization error, and platform-specific repeatability.

[5.] A comparative system-level analysis against representative prior image-guided robotic surgery approaches, identifying which capabilities are integrated within MIRAGE and which remain limitations.

[6.] Preclinical validation with phantom experiments across the da Vinci S, Si, Xi, and SP platforms, demonstrating the feasibility of real-time multimodal guidance and defining the next evaluation steps needed for IDEAL-aligned translational assessment.

References (47)