Differentiable Jitter Correction using Deep Learning-based Image Quality Metric for Phase-Contrast Micro-CT

Junan Chen, Yiting Jia, Joscha Maier, Dominik John, Sami Wirtensohn, Imke Greving, Silja Flenner, Matthias Wieczorek, Julia Herzen

Abstract

X-ray phase-contrast micro computed tomography is an advanced imaging modality capable of revealing the structural and material properties of soft tissue at micrometer-scale resolution. However, the high spatial resolution that makes this technique advantageous also imposes stringent requirements on the calibration of the imaging setup, its mechanical stability, and the sample's stability. Even minor geometric errors have a substantially greater effect on reconstruction quality than in conventional CT. Performing reconstruction without accounting for these instabilities introduces artifacts, such as blurring and streaking, which substantially degrade image quality. This paper proposes a fully differentiable iterative correction method using a deep learning-based image quality metric that estimates and compensates per-projection rigid jitter directly from the acquired projection data, without a pre-scan motion-free reference. The approach builds on a gradient-based auto-focus strategy adapted to parallel-beam geometry. A set of candidate objective functions is benchmarked in a controlled study, and the sensitivity of the visual information fidelity (VIF) metric to the jitter artifact is verified with the target phase-contrast data. To operate without a clean reference, a compact 3D convolutional neural network is trained to predict the VIF score from a single corrupted volume. A spatially selective total variation penalty applied exclusively to the image background is introduced to penalize spurious high-frequency structures that otherwise emerge during optimization. Experiments on biological specimens acquired at different synchrotron beamlines are conducted. Evaluation uses jitter motion applied to simulated and experimentally acquired projection data. The result confirms that the integrated pipeline reliably recovers fine structural detail lost due to jitter, with generalization demonstrated across morphologically distinct samples.

Keywords: Deep Learning, Motion Compensation, Motion Metric, Phase-contrast CT, Micro-CT

1 Introduction

Among the modalities available for high-resolution 3D imaging, X-ray phase-contrast micro computed tomography (μCT) stands out as an exceptionally sensitive tool for visualizing soft tissue at micrometer-scale resolution. In contrast to conventional CT, which exploits differences in X-ray absorption, phase-contrast μCT leverages refraction and interference phenomena to achieve markedly superior contrast for weakly absorbing biological structures, enabling medical and biological investigations at scales previously inaccessible to standard clinical systems [1]. This increased image quality comes, however, at the cost of considerably tighter operational constraints. Because phase-contrast μCT reveals structural information at spatial frequencies much higher than conventional CT, even geometric inconsistencies at the micrometer scale, which are negligible in clinical scanning, can propagate into the reconstructed volume as significant artifacts. Two sources of such inconsistency are particularly relevant in practice: sample drift or jitter, driven by the inherently long acquisition time associated with high-resolution scanning, and device jitter, whose impact grows as voxel dimensions shrink to the micrometer range. The instabilities in a phase-contrast μCT setup are stochastic in nature and persist throughout the scan, meaning that even with synchronization-based hardware measures in place, residual misalignments can still degrade image quality [2, 3]. When left unaddressed, these inter-projection misalignments manifest in the reconstruction as blurring of fine microstructural features or streak-like intensity variations that obscure the very structural information that the modality is designed to reveal [4], hence affecting the outcome of downstream tasks such as cross-correlation with or mapping to a second modality [5, 6].

Methods for CT geometry calibration and motion compensation are widely studied in clinical applications, spanning a broad spectrum of approaches: from methods leveraging external motion tracking signals (e.g., cardiac CT gated with electrocardiography (ECG) [7]), to prior-knowledge-based approach that eliminates the need for gating by directly predicting displacement vector fields (DVFs) using a U-net-based model [8], to image-quality-driven methods such as auto-focus motion compensation, which optimizes projection geometry with respect to a predefined motion metric [9]. Auto-focus motion compensation treats correction as an iterative optimization over the acquisition geometry. An image quality criterion guides the search for parameters that best restore reconstruction quality. Sisniega et al. applied auto-focus correction to extremity cone-beam CT using a sharpness metric and the derivative-free CMA-ES optimizer, which iteratively reconstructs the volume during the optimization for the motion model [9]. Thies et al. embedded the full reconstruction pipeline into a differentiable graph, enabling gradient-based optimization and achieving convergence nearly an order of magnitude faster than CMA-ES [10]. Phase-contrast μ\muCT routinely involves morphologically diverse specimens subject to unknown, per-projection jitter motion. Auto-focus methods require only the acquired data and an image-quality criterion, and the motion model can be specified independently of the metric. We therefore consider auto-focus methods and their variants to offer the greatest potential in this setting. However, hand-crafted auto-focus metrics such as sharpness and total variation (TV) can easily reach their limitation when attempting to correlate with the image degradation caused by various levels of motion. Huang et al. replaced the hand-crafted quality criterion in the auto-focus motion compensation with a learned surrogate trained to predict a visual information fidelity (VIF) [12] score without a clean reference; their framework was demonstrated on human head CT [13]. However, the motion metrics proposed in this context were originally developed for conventional CT. To our knowledge, a learned, reference-free image-quality criterion has not previously been used to drive jitter correction in X-ray phase-contrast μ\muCT. Furthermore, existing metrics are typically tailored to specific anatomical structures or regions, such as the human abdomen or head, limiting their generalizability to arbitrary biological samples in phase-contrast μ\muCT.

In this work, we propose a method that uses differentiable gradient-based optimization with a learned, reference-free quality metric for jitter correction in parallel-beam phase-contrast μ\muCT, in which the projection geometry and specimen diversity differ substantially from those in clinical CT. We derive the gradient matrix of the reconstruction operator for parallel-beam geometry. We train a compact 3D convolutional neural network (CNN) tailored to jitter-induced motion artifacts to predict VIF from a single degraded volume, removing the need for a paired clean reference at inference time. To resolve hallucination failure modes in which certain motion trajectories generate spurious high-frequency patterns that artificially inflate the objective function, we append a TV penalty restricted to the image background for suppressing non-physical textures without smoothing genuine internal sample structure.

References (29)