Hall Effect-Based Tactile Force Detection Sensor for Robot-Assisted Minimally Invasive Surgery

Charles DeLorey, Michael Brockdorff, Adam Metcalf, Benjamin Calmé, Pietro Valdastri, Dominic Jones

Charles DeLorey* Graduate Student Member, IEEE, Michael Brockdorff, Adam Metcalf Member, IEEE, Benjamin Calmé Member, IEEE, Pietro Valdastri Fellow, IEEE, and Dominic Jones* Member, IEEE

Abstract—Robotic minimally invasive surgery has revolutionized surgical practice. However, as the surgeons are mechanically separated from the surgical tools, loss of tactile feedback occurs. We present a compact force sensor with three, millimeter scale sensing elements (taxels) that can be affixed to the grasping face of a robotic surgical tool. The sensor features a miniaturized design using Hall effect sensors and magnets embedded in a deformable elastomer contact layer. Sensor calibration is achieved with a 6 degree of freedom (DoF) parallel robot. Each taxel can detect applied normal and shear forces with average RMSE of 2.133 kPa. The calibrated sensor was validated on phantom and ex vivo porcine tissue, demonstrating tactile sensing during retraction as well as slip. This work introduces a small-scale Hall effect based tactile sensor, enabling sensing for force-sensitive robotic surgical applications.

Index Terms—Surgical robotics, force sensing, tactile sensing, grasping, perception for grasping and manipulation, soft sensors and actuators.

I. INTRODUCTION

Robot-Assisted Minimally Invasive Surgery (RAMIS) offers the gold-standard in surgical treatment for a range of procedures and provides improved visualization and dexterity with teleoperated instruments. In traditional open surgery and manual laparoscopy, surgeons heavily rely on tactile sensations through their fingertips to assist in safe manipulation, feeling the tissue contact either directly or through the tool linkage [1]. Modern RAMIS systems, such as Intuitive Surgical's da Vinci platform, physically decouple the surgeon from instrument–tissue interaction via remote teleoperation. This improves precision but removes natural tactile perception at the end-effector, leaving surgeons to infer the forces visually [2]. Trauma from incorrect force application comes through two routes: excessive grasping can crush tissue leading to ischemia, while insufficient force risks slip and abrasion. Additionally, grip force in RAMIS has been shown to vary significantly between instrument as well as platform versions [3], with different tissue types requiring different target levels of force.

Recently, the da Vinci 5 (Intuitive Surgical) integrated on-instrument force sensing and visuohaptic feedback [4]. However, these capabilities only offer a bulk kinesthetic force feedback, with no information as to the contact forces the tissue experiences. Despite extensive research in surgical

Figure

(a)

10 mm

(b)

(c)

Fig. 1: Force sensor in (a) freeform tissue grasping, (b) the assembled sensor on instrument and (c) analogous bioinspired fingertip pinch inspiration.

force sensing [5], reliable local force measurement directly at the grasper–tissue interface remains unresolved. Existing approaches either estimate forces indirectly, or provide bulk interaction forces but are unable to sense compressive force on the tissue.

Force sensing can be introduced at several points along the RAMIS instrument chain: 1. motor-torque estimation at proximal actuators (indirect, model-dependent), 2. bulk sensing in the shaft (global reactions; e.g., da Vinci 5) [6], 3. sensors at the trocar (extra-corporeal, decoupled from local contact), 4. wrist-mounted sensors (requiring kinematic back-projection), and 5. the contact face of the jaws. Of these, on-jaw sensing provides the most direct access to local contact mechanics. Such on-jaw sensing is especially critical for observing contact pressures, distributed shear, and tissue-dependent friction behavior in highly variable surgical conditions.

Prior work has explored several transduction mechanisms for instrumented surgical graspers. Piezoresistive MEMS offer high bandwidth at the cost of drift and the need for careful compensation and encapsulation [7]. Capacitive tactile films provide thin, conformable coverage with normal and shear sensing and low power, but are susceptible to fluids and

This work was supported by the Engineering and Physical Sciences Research Council (EPSRC) under Grant UKRI180 (TARAS), and grant EP/Y010523/1 (UK-RAS Network Project: "OphthaTact").

The authors are with the STORM Lab, School of Electronic and Electrical Engineering, University of Leeds, Leeds, UK. Email: {elchd, d.p.jones}@leeds.ac.uk humidity that shift the dielectric and demand shielding and packaging attention [8], [9]. Optical waveguide or imaging-based tactile devices sized for MIS tips achieve high spatial resolution and multi-axis estimation with remote optoelectronics, though maintaining optical paths and routing within tight curved jaws raises integration complexity [10]. Fiber Bragg grating sensing enables electromagnetic interference-immune, multiplexed measurements for flexible endoscopic tools, yet requires extremely expensive interrogators and bonding strategies that can place sensing away from the true contact interface [11]. Finally, magnetic Hall-effect tactile approaches have demonstrated grasp-aware control with compact packages, electrical simplicity, and high sampling rates, but require compensation to mitigate stray-field and orientation biases [12], [13], [14]. A comparison of these related works is shown in Table I.

Contact interface sensors open the potential for a new approach in surgical grip control: sensing and predicting slip events. Incipient slip is the earliest stage of sliding when only part of the contact patch begins to move while the rest still adheres [17]. Detecting slip in these regions allows the grasping force to be adjusted to prevent a full slip from occurring. An initial approach presented a thermal-based sensor within the surgical jaw to detect movement of the tissue as slip began to occur [18], but only offers a binary state of either slip or no slip. On-grasper optics have been used to detect slip in a surgical grasper [19], however this approach also only measures displacement. Others have utilized distributed force sensing to monitor the progression of slip across a full grasping surface [13], [17]. The approach was successful for rudimentary grip control to prevent the slip [12]; however, it utilized a large-scale model an order of magnitude too large for a real surgical instrument. To our knowledge, localized tactile slip sensing directly on the jaw contact surface has not yet been demonstrated at this scale for RAMIS graspers.

In this letter, we present a replaceable on-jaw Hall-effect magnetic tactile sensor for RAMIS graspers, capable of 3-DoF force and slip observability at the tissue contact interface. Multiple taxels were needed to provide sufficient coverage of the grasping area to detect slip and force across the grasping surface. Given tool-tip dimensions, PCB manufacturing constraints, Hall effect sensor size, low-cost materials, and the need to maintain a safe radius for magnets, a design of three taxels was achieved, resolving a mean pressure of 74.49 kPa. We validate the sensor on silicone phantoms and ex vivo tissue, showing local contact pressures and taxel-wise shear progression during incipient slip under clinically relevant loading, and further demonstrate live force profiles during human manipulation with da Vinci teleoperation.