arXiv:2610.08464cs.CROct 2026

Manuscript Offprint

arXiv:2610.08464

Federated Bayesian Surveillance of Mechanical Thrombectomy Adverse Events: A Population Risk Layer for Surgical Digital Twins

Damini Rijhwani

Damini Rijhwani

Automation Core Inc., Cambridge, MA, USA

damini@automationcoreinc.com ORCID: 0009-0000-2253-6111

Abstract—Learned surgical simulators and world models can roll out plausible procedural futures, but they carry no grounded estimate of how often interventional devices actually harm patients. We propose treating population-scale adverse-event surveillance as a distinct belief layer of the surgical digital twin, and we evaluate a federated Bayesian protocol for learning it under formal privacy guarantees. Each site holds per-class Gamma-Poisson posteriors over adverse-event rates and exchanges only Rényi-differentially-private natural-parameter updates. We benchmark on the complete FDA MAUDE cohort for thrombus-retrieval catheters (product code NRY): 8,617 reports, of which 6,491 are classified by transparent keyword rules into five thrombectomy complication classes and partitioned across K=8 manufacturer sites. At a matched privacy budget of ( ε≈2.09\varepsilon\approx2.09, δ=10−5\delta=10^{-5}), the conjugate protocol attains a held-out Poisson score of -5.78 per test event versus -26.58 for FedAvg with differential privacy. The non-private federated model also outperforms centralized pooling (+3.19 vs +2.93), evidence that manufacturer-specific complication profiles are real and that federation preserves them. Because MAUDE lacks procedure denominators, outputs are relative rate orderings rather than absolute risks, and we report all privacy-utility operating points.

Index Terms—surgical digital twin, adverse-event surveillance, federated learning, differential privacy, mechanical thrombectomy, Bayesian inference

I. INTRODUCTION

Surgical robotics is increasingly adopting world models, using learned simulators to predict action outcomes before execution. World foundation models now evaluate surgical robot policies in silico [1], digital twins are proposed to ground foundation models in operating rooms [2], [3], and surgical digital twin research is developing rapidly [4]. Endovascular intervention, especially mechanical thrombectomy, is a primary target of this effort.

These systems share a clear blind spot. A learned simulator can predict that a retriever will deform a vessel, but cannot estimate how often retrievers of that class perforate vessels in clinical practice, since that information resides in fragmented, sensitive post-market records rather than surgical video. Real-world failure statistics for thrombectomy devices exist in scattered form: registry studies of the FDA Manufacturer and User Facility Device Experience (MAUDE) database document device-specific complication patterns in peripheral and pulmonary procedures [5], [6], with reproducibility remaining an active concern [7]. Individual hospitals rarely see enough events to estimate rates reliably, and procedural records cannot be centralized.

We propose that surgical digital twins require a distinct population risk layer: calibrated beliefs over adverse-event rates by device class and complication type, learned across institutions without transferring raw records, to condition per-procedure simulations. This paper introduces this layer and evaluates it on real public data:

  1. A federated Bayesian surveillance protocol. Sites maintain per-class Gamma-Poisson posteriors and exchange only Rényi-differentially-private (RDP) natural-parameter updates of dimension 2Z per round. Conjugacy reduces each update to a two-scalar sufficient statistic per class instead of a gradient.

  2. A thrombectomy benchmark from complete public data. We provide the recipe for a manufacturer-federated partition of the complete MAUDE cohort for thrombus-retrieval catheters: 8,617 reports, with 6,491 classified into five complication classes across major thrombectomy manufacturers.

  3. Evidence that federation preserves manufacturer heterogeneity. The federated per-site model outperforms centralized pooling even in the absence of privacy noise, indicating that complication profiles differ across device makers, a structure that pooled models erase.

Scope. This paper evaluates population-level rate estimation on public surveillance data. It does not evaluate per-procedure simulation, device control, or clinical decision-making, and its outputs are not evidence of comparative device safety.