A Belief-State World Model for Catheter Navigation under Sparse Fluoroscopy: A Planar Proof of Concept

Damini Rijhwani

Damini Rijhwani

Automation Core Inc.

Cambridge, MA, USA

damini@automationcoreinc.com

ORCID: 0009-0000-2253-6111

Abstract—Endovascular catheter navigation relies on continuous fluoroscopy, exposing patients and clinical staff to ionizing radiation throughout the procedure. We investigate whether a physics-based world model can sustain the navigation task between deliberately sparse X-ray acquisitions, and whether the model can signal when its internal state estimate is no longer reliable. We formulate sparse-fluoroscopy navigation as a partially observable Markov decision process where a Cosserat rod simulator supplies transition dynamics, sparse noisy projections provide observations, and a particle filter maintains a belief over the device state, reduced in this implementation to a tip state with a geometric contact proxy. We evaluate this formulation in a deliberately simplified setting: a synthetic planar vessel phantom with a quasi-static rod model and simulated projections, without clinical or animal data. In this setting, the belief-state model tracks the simulated tip with a root mean square error of 1.30 mm while acquiring observations at one fifteenth of the continuous rate (0.97 mm at every frame, 1.51 mm at one thirtieth), reported 90 percent credible intervals achieve 0.92 empirical coverage, and the belief-derived contact risk estimate discriminates unsafe contact events with an AUROC of 0.68. These results constitute a proof of concept on a simplified simulation rather than a demonstration of clinical readiness; their purpose is to establish that calibrated belief, rather than point-estimate accuracy alone, is the essential property a sparse-imaging world model must deliver.

Index Terms—world models, belief state, POMDP, Cosserat rod, endovascular navigation, fluoroscopy, uncertainty calibration, surgical digital twins

I. INTRODUCTION

Endovascular interventions are guided almost entirely by fluoroscopy. Clinicians advance a guidewire or catheter while viewing a continuous X-ray stream, running the imaging system throughout much of the procedure. The resulting radiation burden presents a recognized occupational hazard for interventional teams and a dose concern for patients; professional guidelines advise operators to keep exposure as low as reasonably achievable and treat every second of beam-on time as a cost [1]. Current dose-reduction practices center on pulsed acquisition, collimation, and operator discipline. The imaging schedule itself, however, remains essentially continuous because operators lack alternative sources of information regarding device location.

Robotics provides a candidate for that missing information source. A world model is an internal simulation of the environment that an agent rolls forward to predict action consequences, querying real observations only to correct internal drift [2], [3]. The components for a physical catheter world model already exist across separate literatures. Cosserat rod theory provides a mechanically grounded model of slender flexible instruments under tendon actuation and external loading [4], and has been applied to intraoperative contact-force estimation [5]. Open simulation frameworks support interactive rod and contact mechanics for medical scenarios [6]. Deep networks segment catheters in fluoroscopy frames in real time [7], enabling acquired frames to serve as direct measurements. Cardiovascular digital-twin research emphasizes exactly this form of patient-specific predictive model [8]. In principle, procedures could rely primarily on internal simulation, querying X-rays sparsely at low frame rates or on demand to reset accumulated error.

What is missing is not forward prediction, but calibrated belief. Between observations, the internal model operates open-loop, with error accumulating from contact, friction, and actuation noise. A system providing an operator with a confident but inaccurate tip position, or an incorrect wall contact force, is worse than no assistance. The fundamental question for a sparse-imaging world model is therefore whether it recognizes when its estimate becomes unreliable: whether its reported uncertainty is statistically honest, and whether safety metrics derived from the belief, such as contact-force threshold exceedance, remain informative as observation frequency decreases.

This paper takes an initial, focused step toward addressing that question. Our contributions are:

  • a POMDP formulation of catheter navigation under sparse fluoroscopy, where the transition model is a Cosserat rod simulator, the observation model is a sparse noisy projection, and safety is evaluated as a contact-force exceedance probability under the belief;

  • a planar proof-of-concept implementation pairing a quasi-static rod simulator with a particle-filter belief over device state, reduced here to a tip state with a geometric contact

proxy;

  • a calibration-focused evaluation protocol reporting tip tracking error (1.30 mm at one fifteenth of continuous frame rate), empirical coverage of reported credible intervals (0.92), and unsafe contact event discrimination (AUROC 0.68) across reduced observation rates.

We state the experimental scope directly. All evaluations run on a synthetic planar vessel phantom with a simplified quasi-static rod model; there are no clinical or animal data, no real fluoroscopic streams, and no human operator in the loop. These results do not constitute a claim of clinical readiness. Our goal is to establish the formulation and evaluation protocol at a focused scale where calibration can be assessed rigorously, providing a clear foundation for future work on complex anatomy and imaging.